Abstract
The beam quality of the diffracted wave from a bulk acousto-optic frequency shifter has been accurately measured. High accuracy in measuring the output-phase profile was achieved by a unique utilization of heterodyne interferometry. Results for glass and PbMoO4 deflectors show no discernible beam-quality degradation attributible to the acousto-optic interaction to within 0.01 wave rms at 647.1 nm.
© 1978 Optical Society of America
Full Article | PDF ArticleMore Like This
M. G. Gazalet, M. Ravez, F. Haine, C. Bruneel, and E. Bridoux
Appl. Opt. 33(7) 1293-1298 (1994)
Anthony Dieulangard, Jean-Claude Kastelik, Samuel Dupont, and Joseph Gazalet
Appl. Opt. 52(33) 8134-8141 (2013)
Linbo Shao, Neil Sinclair, James Leatham, Yaowen Hu, Mengjie Yu, Terry Turpin, Devon Crowe, and Marko Lončar
Opt. Express 28(16) 23728-23738 (2020)