Beyond the frame rate: measuring high-frequency fluctuations with light-intensity modulation

Wesley P. Wong and Ken Halvorsen  »View Author Affiliations


Optics Letters, Vol. 34, Issue 3, pp. 277-279 (2009)
http://dx.doi.org/10.1364/OL.34.000277


View Full Text Article

Acrobat PDF (140 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Power-spectral-density measurements of any sampled signal are typically restricted by both acquisition rate and frequency response limitations of instruments, which can be particularly prohibitive for video-based measurements. We have developed a new method called intensity modulation spectral analysis that circumvents these limitations, dramatically extending the effective detection bandwidth. We demonstrate this by video tracking an optically trapped microsphere while oscillating an LED illumination source. This approach allows us to quantify fluctuations of the microsphere at frequencies over 10 times higher than the Nyquist frequency, mimicking a significantly higher frame rate.

© 2009 Optical Society of America

Measuring the power spectral density (PSD) is a useful way to characterize fluctuations and noise for a diverse range of physical processes. Optical measurements of the PSD have been used to study single-molecule dynamics [1

1. L. Oddershede, J. K. Dreyer, S. Grego, S. Brown, and K. Berg-Sørensen, Biophys. J. 83, 3152 (2002). [CrossRef] [PubMed]

], bacterial chemotaxis and motion [2

2. C. V. Gabel and H. C. Berg, Proc. Natl. Acad. Sci. USA 100, 8748 (2003). [CrossRef] [PubMed]

, 3

3. E. Korobkova, T. Emonet, J. M. G. Vilar, T. S. Shimizu, and P. Cluzel, Nature 428, 574 (2004). [CrossRef] [PubMed]

], quantum dot blinking [4

4. M. Pelton, D. G. Grier, and P. Guyot-Sionnest, Appl. Phys. Lett. 85, 819 (2004). [CrossRef]

], and microrheology [5

5. F. Gittes, B. Schnurr, P. D. Olmsted, F. C. MacKintosh, and C. F. Schmidt, Phys. Rev. Lett. 79, 3286 (1997). [CrossRef]

, 6

6. M. T. Valentine, P. D. Kaplan, D. Thota, J. C. Crocker, T. Gisler, R. K. Prud’homme, M. Beck, and D. A. Weitz, Phys. Rev. E 64, 061506 (2001). [CrossRef]

] and to calibrate optical traps [7

7. K. Svoboda and S. M. Block, Annu. Rev. Biophys. Biomol. Struct. 23, 247 (1994). [CrossRef] [PubMed]

]. Unfortunately, limited acquisition rate and detector frequency response restrict PSD measurements in frequency space, which is especially detrimental to video applications. Notably, the highest frequency that can be sampled directly, with few exceptions [8

8. F. R. Verdun, T. L. Ricca, and A. G. Marshall, Appl. Spectrosc. 42, 199 (1988). [CrossRef]

, 9

9. M. H. El-Shafey, in Proceedings of the International Conference on Electrical, Electronic, and Computer Engineering (IEEE, 2004), pp. 210–213.

, 10

10. W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]

, 11

11. C. Donciu, C. Temneanu, and R. Ciobanu, in Proceedings of the IEEE International Conference on Virtual Environments, Human–Computer Interfaces, and Measurement Systems (IEEE, 2005), pp. 50–53.

], is half the acquisition rate, or Nyquist frequency.

We have developed a method called intensity modulation spectral analysis (IMSA), which overcomes these limitations in a simple and economical way. By simply oscillating the light intensity of an optical signal prior to detection, the PSD at the oscillation frequency can be determined from the measured variance, even if that frequency is above the acquisition rate. This is similar in spirit to signal processing methods that can spectrally shift a signal (e.g., heterodyne detection, lock-in techniques), or extract high-frequency information folded down via aliasing (e.g., undersampling [11

11. C. Donciu, C. Temneanu, and R. Ciobanu, in Proceedings of the IEEE International Conference on Virtual Environments, Human–Computer Interfaces, and Measurement Systems (IEEE, 2005), pp. 50–53.

, 12

12. R. G. Vaughan, N. L. Scott, and D. R. White, IEEE Trans. SignalProcess. 39, 1973 (1991). [CrossRef]

]). Practically, IMSA can dramatically extend the frequency range of an existing measurement device, allowing, for example, an inexpensive camera to be used in place of a significantly more expensive one. Here we present the framework for IMSA and an experimental demonstration using an optically trapped microsphere, in which modulating the brightness of an LED allows PSD measurement well beyond the Nyquist frequency and camera frame rate.

Physical acquisition systems do not make instantaneous measurements when sampling a signal but rather collect data over finite integration times. Consider a stationary random trajectory X(t). The measured trajectory Xm(t) can be expressed as a convolution of the true trajectory and an impulse response H(t),
Xm(t)=(XH)(t)X(t)H(tt)dt.
(1)
Then the measured power spectrum Pm(ω) differs from P(ω), the true power spectrum of X, according to the relation Pm(ω)=P(ω)H̃(ω)2, and the total measured variance is the integral of Pm(ω) over all frequencies,
var[Xm]=12πP(ω)H̃(ω)2dω,
(2)
where the tilde designates the Fourier transform, X̃(ω)=X(t)exp(iωt)dt. Integrals are taken from to + unless otherwise specified.

In the simplest case, the measured quantity Xm is the unweighted time average of the true value X over the integration time W, i.e., the impulse response is a rectangular function,
H0(t)={1WW2<tW20elsewhere}.
(3)
Correspondingly, the measured power spectrum is the original power spectrum multiplied by
H̃0(ω)2=(sin(ωW2)ωW2)2.
(4)

This simple case of a rectangular impulse response is a good model for video-imaging acquisition systems, where W is the exposure time. This averaging leads to the common problem of video image blur, which not only adds errors in the position of tracked objects but also causes systematic biases when quantifying fluctuations [10

10. W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]

, 13

13. R. Yasuda, H. Miyata, and K. Kinosita Jr, J. Mol. Biol. 263, 227 (1996). [CrossRef] [PubMed]

, 14

14. T. Savin and P. S. Doyle, Biophys. J. 88, 623 (2005). [CrossRef]

, 15

15. T. Savin and P. S. Doyle, Phys. Rev. E 71, 041106 (2005). [CrossRef]

]. As we have previously demonstrated [10

10. W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]

], by measuring the variance for different exposure times W and fitting to Eq. (2), the power spectrum can be characterized above the acquisition rate of the detection system. However, this approach requires that the functional form of the power spectrum is known a priori.

Interestingly, by oscillating the intensity of a source signal (e.g., light in the case of video imaging) and measuring the variance of the resulting signal, the power spectrum can be reconstructed without prior knowledge. This is the fundamental idea behind IMSA.

During detection, the finite integration time W causes the true power spectrum to be multiplied by the low-pass filter H̃0(ω)2 [Eq. (4)]. As W becomes longer, H̃0(ω)2 approaches an unnormalized delta function, i.e., H̃0(ω)2αδ(ω). Multiplying the rectangular impulse response by a complex exponential shifts this delta function, i.e., if H(t)=exp(iωt)H0(t), then H̃(ω)2αδ(ωω). From Eq. (2) we see that the variance approaches αP(ω)2π. Thus, as demonstrated in this simple example, the power spectrum can be sampled by shifting the filter to any frequency of interest ω and measuring the variance.

Practically, we use the real-valued impulse response
H(t)=L(t)H0(t)N,
(5)
where L(t)sin(ωt+ϕ)+B and N is a normalization factor (see Fig. 1 ). We set N=L(t)H0(t)dt, so that the convolution represents a time-averaged signal weighted by the source intensity modulation L(t), and let B0; the parameters (W, B) should be chosen such that N0. Note that N=B whenever ϕ (the phase shift between modulation and sampling) is an integer multiple of 2π, or whenever there are an integer number of oscillations within each exposure window.

By substituting H̃(ω) [the Fourier transform of Eq. (5)] into Eq. (2), and noting that P(ω)=P(ω) for a real-valued wide-sense stationary process, we obtain for the measured variance
var[Xm]=14πN2P(ω)H̃0(ωω)2dω[t1]
+B22πN2P(ω)H̃0(ω)2dω[t2]
+Bsin(ϕ)πN2P(ω)H̃0(ω)H̃0(ωω)dω[t3]
cos(2ϕ)4πN2P(ω)H̃0(ω+ω)H̃0(ωω)dω[t4].
(6)
This key equation can be used to determine the power spectrum from the measured variance, since term t1 approaches P(ω)2WN2 as W gets longer (see [16

16. Note that 14πN2H̃0(ω)2dω=12WN2.

] and Fig. 1), while terms t3 and t4 become small and term t2 can be measured directly. Additionally, proper phase selection can completely eliminate t3 and t4.

To determine the power spectral density of a signal at the frequency of interest ω, the signal should be convolved with L(t)H0(t)N [Eq. (5)] by modulating its intensity, and then its variance should be measured. Writing the measured variance as var[Xm](ω), the PSD is given by the following formulas:
P(ω)=2WN2var[Xm](ω)whenB=0,
(7)
provided ϕ=π4+mπ2, where m is an integer, and
P(ω)=2WB2(var[Xm](ω)var[Xm](0))
whenB0,
(8)
provided that the exposure time W is chosen such that W=n2πω, where n is a natural number, and ϕ is chosen according to (see [17

17. Determination of ϕ to eliminate terms t3 and t4 can be done by calculating H̃(ω)2, setting the sum of the ϕ-dependent terms to zero, and solving for the specific case where W=n2πω.

])
ϕ=(1)nsin1(B2+12B).
(9)

On a practical note, terms t3 and t4 are small when W1ω (i.e., there are many oscillation cycles within each exposure window). If both cross terms t3 and t4 are negligible, the selection of ϕ is irrelevant. In fact, oscillations may not even have to be synchronized to the acquisition device to make a good IMSA measurement—simply multiplying the input signal by L(t) prior to detection can yield acceptable results.

The error in P(ω) is governed primarily by the error in the variance (e.g., for N samples the relative statistical standard error is (2N), barring instrumental error [10

10. W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]

]). The resolution in ω is given by the width of H̃0(ω)2 (see Fig. 1), which is approximately πW in each direction (77% of the area under the curve).

To demonstrate IMSA experimentally, we measured the power spectrum of an optically trapped polystyrene microsphere (2.5μm, Corpuscular) using a machine vision camera (GE680, Prosilica) and an LED (MRMLED, Thorlabs) with intensity modulation capable of up to 300kHz (custom current source, Rowland Institute electronics lab). Details of the particle tracking have been discussed previously for a functionally identical setup, where it was demonstrated that the system is well described by Eq. (3) [10

10. W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]

]. The light intensity was sinusoidally modulated [as in Eq. (5)] with B=0.5 and ϕ determined from Eq. (9), and the integration time of the camera was W=2.5ms. Modulated light at five different frequencies ω was interspersed with dc light on a frame-by-frame basis. Each of the ten variances was calculated, and each PSD data point was calculated according to Eq. (8), where the mean of the two neighboring dc frames was used as var[Xm](0) for each of the five measurements. Data were collected for two beads at different laser powers. The spring constant and the friction factor for each bead were measured using the blur-corrected power-spectrum method [10

10. W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]

]. The IMSA measured values, which have no free-fitting parameters, are in excellent agreement with the expected power spectra (Fig. 2 ).

As we have shown, IMSA provides a unique and practical way for measuring the PSD of a signal that overcomes acquisition rate and frequency response limitations of instruments. While especially beneficial to video-imaging applications (owing to the almost-negligible cost of implementation compared with the expense of fast video cameras), IMSA is a general method applicable to signals in which the time-averaged weighting can be controlled before sampling.

Owing to its potential to benefit a broad range of fields and its flexibility of implementation (see [18

18. For example, measurement sensitivity can be improved by removing the dc offset (i.e., set B=0). For video imaging, this can be accomplished by simulating positive and negative light (e.g., by using two colors or polarizations) or, for low-light applications, by properly synchronizing two camera frames to each illumination cycle. At the sensor level, IMSA can be implemented by using a camera capable of modulating the incoming signal [19, 20, 21], or by having two collection wells for each pixel to simulate positive and negative light.

, 19

19. A. C. Mitchell, J. E. Wall, J. G. Murray, and C. G. Morgan, J. Microsc. 206, 225 (2002). [CrossRef] [PubMed]

, 20

20. P. R. Dmochowski, B. R. Hayes-Gill, M. Clark, J. A. Crowe, M. G. Somekh, and S. P. Morgan, Electron. Lett. 40, 1403 (2004). [CrossRef]

, 21

21. D. Van Nieuwenhove, W. van der Tempel, R. Grootjans, J. Stiens, and M. Kuijk, IEEE Sens. J. 7, 317 (2007). [CrossRef]

]), we expect that IMSA will become a common laboratory method for measuring power spectra. Microrheology measurements [6

6. M. T. Valentine, P. D. Kaplan, D. Thota, J. C. Crocker, T. Gisler, R. K. Prud’homme, M. Beck, and D. A. Weitz, Phys. Rev. E 64, 061506 (2001). [CrossRef]

, 22

22. J. C. Crocker and D. G. Grier, J. Colloid Interface Sci. 179, 298–310 (1996). [CrossRef]

] can take immediate advantage of the up to 10000 fold increase in frequency range over standard video imaging without losing the ability to track multiple targets. Widely used fluorescence-imaging techniques [23

23. E. Toprak and P. R. Selvin, Annu. Rev. Biophys. Biomol. Struct. 36, 349 (2007). [CrossRef] [PubMed]

] stand to benefit as well, with IMSA enabling measurements at frequencies that are currently impossible by any other method owing to light limitations.

The dramatic increase in frequency range enabled by IMSA can be used to push the envelope of high-frequency measurements and to realize significant cost savings in instrumentation. Using an inexpensive LED illuminator, we transformed a standard video camera into a spectrum analyzer with a frequency range of up to 300kHz.

The authors thank M. M. Burns, W. Hill, A. M. Turner, and M. M. Forbes for helpful conversations and C. Stokes for the LED current source. Funding was provided by the Rowland Junior Fellows program.

Fig. 1 Graphical representation of IMSA showing (a) unnormalized impulse responses H(t)=L(t)H0(t) for ϕ=0 and B=0 [solid line; see Eq. (5)] and H(t)=H0(t) [dotted line; see Eq. (3)] and (b) the squared magnitude of their Fourier transforms, which filter P(ω) in terms t1 and t2 of Eq. (6).
Fig. 2 Measurement of the power spectrum using IMSA for two trapped beads with different spring constants (data points). The expected power spectra are superimposed (solid curves), showing good agreement well beyond the Nyquist frequency (vertical dashed line). Error bars represent statistical error.
1.

L. Oddershede, J. K. Dreyer, S. Grego, S. Brown, and K. Berg-Sørensen, Biophys. J. 83, 3152 (2002). [CrossRef] [PubMed]

2.

C. V. Gabel and H. C. Berg, Proc. Natl. Acad. Sci. USA 100, 8748 (2003). [CrossRef] [PubMed]

3.

E. Korobkova, T. Emonet, J. M. G. Vilar, T. S. Shimizu, and P. Cluzel, Nature 428, 574 (2004). [CrossRef] [PubMed]

4.

M. Pelton, D. G. Grier, and P. Guyot-Sionnest, Appl. Phys. Lett. 85, 819 (2004). [CrossRef]

5.

F. Gittes, B. Schnurr, P. D. Olmsted, F. C. MacKintosh, and C. F. Schmidt, Phys. Rev. Lett. 79, 3286 (1997). [CrossRef]

6.

M. T. Valentine, P. D. Kaplan, D. Thota, J. C. Crocker, T. Gisler, R. K. Prud’homme, M. Beck, and D. A. Weitz, Phys. Rev. E 64, 061506 (2001). [CrossRef]

7.

K. Svoboda and S. M. Block, Annu. Rev. Biophys. Biomol. Struct. 23, 247 (1994). [CrossRef] [PubMed]

8.

F. R. Verdun, T. L. Ricca, and A. G. Marshall, Appl. Spectrosc. 42, 199 (1988). [CrossRef]

9.

M. H. El-Shafey, in Proceedings of the International Conference on Electrical, Electronic, and Computer Engineering (IEEE, 2004), pp. 210–213.

10.

W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]

11.

C. Donciu, C. Temneanu, and R. Ciobanu, in Proceedings of the IEEE International Conference on Virtual Environments, Human–Computer Interfaces, and Measurement Systems (IEEE, 2005), pp. 50–53.

12.

R. G. Vaughan, N. L. Scott, and D. R. White, IEEE Trans. SignalProcess. 39, 1973 (1991). [CrossRef]

13.

R. Yasuda, H. Miyata, and K. Kinosita Jr, J. Mol. Biol. 263, 227 (1996). [CrossRef] [PubMed]

14.

T. Savin and P. S. Doyle, Biophys. J. 88, 623 (2005). [CrossRef]

15.

T. Savin and P. S. Doyle, Phys. Rev. E 71, 041106 (2005). [CrossRef]

16.

Note that 14πN2H̃0(ω)2dω=12WN2.

17.

Determination of ϕ to eliminate terms t3 and t4 can be done by calculating H̃(ω)2, setting the sum of the ϕ-dependent terms to zero, and solving for the specific case where W=n2πω.

18.

For example, measurement sensitivity can be improved by removing the dc offset (i.e., set B=0). For video imaging, this can be accomplished by simulating positive and negative light (e.g., by using two colors or polarizations) or, for low-light applications, by properly synchronizing two camera frames to each illumination cycle. At the sensor level, IMSA can be implemented by using a camera capable of modulating the incoming signal [19, 20, 21], or by having two collection wells for each pixel to simulate positive and negative light.

19.

A. C. Mitchell, J. E. Wall, J. G. Murray, and C. G. Morgan, J. Microsc. 206, 225 (2002). [CrossRef] [PubMed]

20.

P. R. Dmochowski, B. R. Hayes-Gill, M. Clark, J. A. Crowe, M. G. Somekh, and S. P. Morgan, Electron. Lett. 40, 1403 (2004). [CrossRef]

21.

D. Van Nieuwenhove, W. van der Tempel, R. Grootjans, J. Stiens, and M. Kuijk, IEEE Sens. J. 7, 317 (2007). [CrossRef]

22.

J. C. Crocker and D. G. Grier, J. Colloid Interface Sci. 179, 298–310 (1996). [CrossRef]

23.

E. Toprak and P. R. Selvin, Annu. Rev. Biophys. Biomol. Struct. 36, 349 (2007). [CrossRef] [PubMed]

OCIS Codes
(120.4820) Instrumentation, measurement, and metrology : Optical systems
(170.4520) Medical optics and biotechnology : Optical confinement and manipulation
(110.4155) Imaging systems : Multiframe image processing

ToC Category:
Instrumentation, Measurement, and Metrology

History
Original Manuscript: October 16, 2008
Revised Manuscript: November 21, 2008
Manuscript Accepted: November 26, 2008
Published: January 26, 2009

Virtual Issues
Vol. 4, Iss. 4 Virtual Journal for Biomedical Optics

Citation
Wesley P. Wong and Ken Halvorsen, "Beyond the frame rate: measuring high-frequency fluctuations with light-intensity modulation," Opt. Lett. 34, 277-279 (2009)
http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-34-3-277


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. L. Oddershede, J. K. Dreyer, S. Grego, S. Brown, and K. Berg-Sørensen, Biophys. J. 83, 3152 (2002). [CrossRef] [PubMed]
  2. C. V. Gabel and H. C. Berg, Proc. Natl. Acad. Sci. USA 100, 8748 (2003). [CrossRef] [PubMed]
  3. E. Korobkova, T. Emonet, J. M. G. Vilar, T. S. Shimizu, and P. Cluzel, Nature 428, 574 (2004). [CrossRef] [PubMed]
  4. M. Pelton, D. G. Grier, and P. Guyot-Sionnest, Appl. Phys. Lett. 85, 819 (2004). [CrossRef]
  5. F. Gittes, B. Schnurr, P. D. Olmsted, F. C. MacKintosh, and C. F. Schmidt, Phys. Rev. Lett. 79, 3286 (1997). [CrossRef]
  6. M. T. Valentine, P. D. Kaplan, D. Thota, J. C. Crocker, T. Gisler, R. K. Prud'homme, M. Beck, and D. A. Weitz, Phys. Rev. E 64, 061506 (2001). [CrossRef]
  7. K. Svoboda and S. M. Block, Annu. Rev. Biophys. Biomol. Struct. 23, 247 (1994). [CrossRef] [PubMed]
  8. F. R. Verdun, T. L. Ricca, and A. G. Marshall, Appl. Spectrosc. 42, 199 (1988). [CrossRef]
  9. M. H. El-Shafey, in Proceedings of the International Conference on Electrical, Electronic, and Computer Engineering (IEEE, 2004), pp. 210-213.
  10. W. P. Wong and K. Halvorsen, Opt. Express 14, 12517 (2006). [CrossRef] [PubMed]
  11. C. Donciu, C. Temneanu, and R. Ciobanu, in Proceedings of the IEEE International Conference on Virtual Environments, Human-Computer Interfaces, and Measurement Systems (IEEE, 2005), pp. 50-53.
  12. R. G. Vaughan, N. L. Scott, and D. R. White, IEEE Trans. Signal Process. 39, 1973 (1991). [CrossRef]
  13. R. Yasuda, H. Miyata, and K. Kinosita, Jr, J. Mol. Biol. 263, 227 (1996). [CrossRef] [PubMed]
  14. T. Savin and P. S. Doyle, Biophys. J. 88, 623 (2005). [CrossRef]
  15. T. Savin and P. S. Doyle, Phys. Rev. E 71, 041106 (2005). [CrossRef]
  16. Note that 1/4πN2∫∣H̃0(ω)∣2dω=1/2WN2.
  17. Determination of ϕ to eliminate terms t3 and t4 can be done by calculating ∣H̃(ω)∣2, setting the sum of the ϕ-dependent terms to zero, and solving for the specific case where W=n2π/ω′.
  18. For example, measurement sensitivity can be improved by removing the dc offset (i.e., set B=0). For video imaging, this can be accomplished by simulating positive and negative light (e.g., by using two colors or polarizations) or, for low-light applications, by properly synchronizing two camera frames to each illumination cycle. At the sensor level, IMSA can be implemented by using a camera capable of modulating the incoming signal , or by having two collection wells for each pixel to simulate positive and negative light.
  19. A. C. Mitchell, J. E. Wall, J. G. Murray, and C. G. Morgan, J. Microsc. 206, 225 (2002). [CrossRef] [PubMed]
  20. P. R. Dmochowski, B. R. Hayes-Gill, M. Clark, J. A. Crowe, M. G. Somekh, and S. P. Morgan, Electron. Lett. 40, 1403 (2004). [CrossRef]
  21. D. Van Nieuwenhove, W. van der Tempel, R. Grootjans, J. Stiens, and M. Kuijk, IEEE Sens. J. 7, 317 (2007). [CrossRef]
  22. J. C. Crocker and D. G. Grier, J. Colloid Interface Sci. 179, 298-310 (1996). [CrossRef]
  23. E. Toprak and P. R. Selvin, Annu. Rev. Biophys. Biomol. Struct. 36, 349 (2007). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.