The laser-based microjet injection system uses the hydrodynamic impact of a narrow liquid
jet onto skin. The immediate delivery would enable minimized prescription of topical
drugs intended to work on the outer layers of the skin, avoiding any skin irritation or
allergic reaction and preventing uncontrolled evaporation of active ingredient and
unpleasant odor associated with noninvasive procedures. Several types of injection
mechanism have been considered, including spring compression, expansion of piezoelectric
transducer, linear Lorentz-force-driven piston actuator, and expansion of
laser-initiated waves in water [
1A. Arora, M. R. Prausnitz, and S. Mitragotri, Int. J. Pharm.
364, 227 (2008). [CrossRef]
T. Han and J. J. Yoh, J. Appl. Phys.
107, 103 (2010). [CrossRef]
–
3A. Taberner, N. C. Hogan, and I. W. Hunter, “Needle-free jet injection using
real-time controlled linear Lorenz-force actuators,”
Med. Eng. Phys. (to be published). [CrossRef]
]. Such
mechanisms would eliminate abundant needle wastes, and they are favored for highly
needle-phobic patients [
4S. Mitragotri, Nat. Rev. Drug Discov.
5, 543 (2006). [CrossRef]
].
A narrow, high-pressure jet of 100 to
velocity is required to accelerate the drug to
penetrate the animal skin with
of yield strength [
4S. Mitragotri, Nat. Rev. Drug Discov.
5, 543 (2006). [CrossRef]
]. Reducing the jet diameter to a 100 μm size has shown the
advantage in drug delivery of minimizing damage to the tissue [
1A. Arora, M. R. Prausnitz, and S. Mitragotri, Int. J. Pharm.
364, 227 (2008). [CrossRef]
]. The present scheme of injection via Er:YAG laser beam
at 250 μs pulse duration
generates pressure by the
displacement of liquid via
laser-induced vapor bubbles and the elastic pumping of the drug through a nozzle by a
membrane separating the driving liquid from the drug.
Success in jet injection requires sufficient impulse of the jet to penetrate the target
tissue. In the case of an Nd:YAG laser at 7 ns pulse duration, high irradiance of a
-switched beam produced instantaneous expansion of
bubbles and generated multiple shockwaves [
2T. Han and J. J. Yoh, J. Appl. Phys.
107, 103 (2010). [CrossRef]
]. The
consequent jets due to both shock and bubble expansion reached
while the short length of the jet is insufficient for
ensuring a sizable dose of drug targeting a treatment site.
Here we adopted a quasi-long-pulsed Er:YAG beam at 250 μs to
source an injector. The high absorbance of the beam to water (2940 nm) also enables
a stronger generation of a jet at lower laser energies. We present the enhanced
controllability and dosage of delivered drug into a guinea pig’s skin through
fluorescent staining on both postmortem abdominal and living dorsal skins. Furthermore,
to verify test results, the Plesset equation of the vapor bubble theory has been adapted
to confirm the measured jet velocity resulting from the laser-initiated microjet.
When the low-irradiance laser energy reaches the driving liquid in the
upper chamber of the injector [
2T. Han and J. J. Yoh, J. Appl. Phys.
107, 103 (2010). [CrossRef]
], the temperature
rise in a focal point in water admits a sudden vaporization and generation of a vapor
bubble [
5C. E. Brennen, Cavitation and Bubble Dynamics
(Oxford University,
1995).
]. The process of constant-pressure phase
transformation is marked by a superheated liquid formation of vapor voids. The Er:YAG
laser has 250 μs pulse duration and wavelength of 2940 nm at which water
best absorbs the beamed energy. Once vapor is accumulated in the bubble, any additional
radiation passes through the expanding bubble without being directly absorbed by the
vapor. Elongation of the bubble discontinues as the initiation of the bubble stabilizes
beyond 250 μs and reaches a maximum radius.
We adopt explosive bubble expansion induced by a laser irradiation as an
actuator for ejecting a coherent microjet. The present injector consists of a
micronozzle for storing liquid drug, a chamber for driving fluid separated by a
heat-resistant flexible membrane between drug and water, and air-tight confinement glass
at the beam incident end with O-ring-type sealing. A highly water-absorbant
Er:YAG beam irradiates the water for vapor
bubble generation within the driving liquid chamber. Ideally sealed in the chamber,
growth of the bubble would cause a sizable pressure impulse on the elastic membrane. The
elastic response of the membrane ejects liquid drug out from a 150 μm nozzle
at a velocity needed to penetrate skin.
In Fig.
1(a), the ejected
jet shown in an air background reaches
, which has larger injection volume than the speed
previously attainable by a nanosecond pulsed Nd:YAG laser system [
2T. Han and J. J. Yoh, J. Appl. Phys.
107, 103 (2010). [CrossRef]
]. The laser system offers various jet properties with a change in
laser energy. The flow through a narrow nozzle experiences turbulent and frictional
losses. Figure
1(b) indicates that friction
at the nozzle exit causes velocity reduction in the region below 800 mJ and the
turbulence shortens the jet length in the region beyond 800 mJ [
6A. H. Lefebvre, Atomization and Sprays
(Hemisphere,
1989).
]. The instability causes spray and a decrease in
energy of the jets, a process known as atomization. In all injections of the present
scheme, the standoff distance between the nozzle and the skin is less than 3 mm to
avoid any instability due to these jet properties.
Fig. 1. Ejected microjet in air. (a) Images of
Er:YAG microjet at 408 mJ,
250 μs pulse duration showing jet velocity of
. (b) Jet velocity shown for varying laser
energy
The injection liquid was prepared by dissolving fluorescein
isothiocyanate (FITC, ) in dimethyl sulfoxide (DMSO) solution for verifying
microjet penetration performance. The treated skin from the FITC test was
analyzed by a fluorescence microscope
(Nikon Eclipse Ti-U), and hematoxylin and eosin (H&E) staining was used to monitor
alteration of tissue after injection.
The stained skin samples were frozen and chopped by cryotome (Leica). Three pieces of a
sample were made with the injection point being the center. Each cross section was
vertical to the injected area and made into a slide to analyze with a microscope. The
embedded sites of fluorescent trace observed with the microscope would confirm drug
penetration and the range of spread underneath the outer skin layer.
A male Hartley guinea pig was laid on a table to perform jet injection on the back for a
dorsal skin test. In the abdominal case, the relevant skin was first sectioned into a
target and cleaned to remove fat and subdermal tissues.
The prepared skin sample was affixed on a foam board with pins. The laser irradiated the
driving fluid, and the jet was ejected from the nozzle vertically to a target material.
The guinea pig used was 250 grams of weight. The skin was epilated with wax for
24 h before the experiment and immediately used without freezing.
For the skin mockup for instant visualization of penetration, a gelatin gel was used,
which offers controllable mechanical properties depending on its weight percent. Gelatin
of 60 Bloom was dissolved in water at 5 percent. The Bloom number indicates the
toughness of gels, and 60 meets the typical animal skin toughness.
Figure
2 is the staining
result on a sectioned sample of guinea pig abdominal skin. The fluorescence is dispersed
in all directions around an impact point. The microjet initiated with a 19 J laser
beam delivers the drug over the epidermis and dermis within
from the skin (Fig.
3).
Fig. 2. FITC staining of guinea pig abdominal skin treated with .
Fig. 3. FITC staining of guinea pig dorsal skin treated with
(a) and (b) .
Figure
3 shows the results of microjet
injection on dorsal skin as administered on a living guinea pig laid on a table. The
drug is evenly dispersed on the skin tissue similar to the abdominal case. The dorsal
skin is thicker (
) than the abdominal skin such that deeper wetting of
the relevant layers is effectively treated with FITC. The H&E staining shown in the
upper-right windows after microjet injection displays the architecture of the treated
sites, and it shows that the drug is delivered with no alteration of skin morphology
adjacent to the injection site. The
microjet initiated with a 1.57 J laser beam, however, achieves a targeted local
delivery rather than dispersion [Fig.
3(b)].
Unlike the abdominal case of a sectioned sample affixed to an acrylic plate, the
underlying structure of dorsal skin supports jet propagation, allowing a deeper
penetration. Even though the injection spot had been ruined following the path of a jet,
these microstructural changes are expected to be recovered by the barrier recovery
process [
7J. A. Segre, J. Clin. Invest.
116, 1150 (2006). [CrossRef]
].
The jet produced with 1.19 J of laser pulse showed smaller volume than the case of
1.57 J. The jet energy is mostly dissipated at the upper layer of skin. With higher
energy, however, increased jet energy reaches deeper layers of skin. The jet energy is
converted to deformation of the skin barrier or propagation of the stress wave depending
on the skin properties. The jet achieves farther delivery of drugs when more energy is
converted to deformation energy than to
stress wave.
To evaluate jet injection, gel models are used for visualization and mimicking of the
skin elasticity. Jet injection consists of three events: jet impingement, flow into
skin, and dispersion under skin [Fig.
4(a)].
Jet impact creates a hole on the gel with an estimated impact pressure or water-hammer
pressure from Eq. (
1), where the
impact pressure depends on the sound velocity
as well as the jet velocity
. Then at a lower jet pressure proportional to the
square of the velocity, the ejected dose is delivered into the gel making a path of jet
stream. A thin cylindrical jet of 150 μm generated from an injector causes
virtually zero splashback at the contact surface as seen in
Fig.
4(a), allowing smooth penetration. In gel models, the
dense structure with no porosity forces the
jet to agglomerate and causes bounce of the drug in the gel:
This is the pressure needed to overcome
the ultimate strength of a target for the surface erosion. For a typical skin strength
of
, Eq. (
1) suggests a minimum jet velocity of
for water density
and sound speed of
in water.
Fig. 4. (a) Microjet injection shown with no splashback, 150 μm diameter,
and gel penetration of drug at 408 mJ. (b) Penetration depth and width
for varied laser energy.
The penetrated depth and injected volume are evaluated with varying laser
energy on a gel model [Fig.
4(b)].
The penetrated depth increases as the laser
energy is intensified up to 800 mJ. At 600 mJ, the injection efficiency is
reduced due to the recovery response of the gel. Above 800 mJ, increased jet
velocity shortens the jet length according to the jet breakup. The spray characteristic
and instability of the weakening jet beyond a high critical Weber number may be
responsible for this observation.
The jet velocity may be analytically determined. We consider a
408 mJ (low-energy) case for illustration
purposes. We used the empirical data for
temporal development of the bubble radius and the Rayleigh–Plesset approximation
[
8M. S. Plesset and A. Prosperetti, Annu. Rev. Fluid Mech.
9, 145 (1977). [CrossRef]
] to estimate the pressure and temperature
gradients on the driving liquid wall for jet ejection. For vapor bubbles, the thermal
effects play a dominant role and the effect of liquid inertia can be neglected. The
Plesset model considers the evaporation and the heat conduction. The temporal evolution
of bubble radius
can be approximated by the following equation [
5C. E. Brennen, Cavitation and Bubble Dynamics
(Oxford University,
1995).
]
where
is defined as
Here
is the latent heat,
is the equilibrium vapor density corresponding to the
boiling temperature
,
is the temperature on the bubble wall, and
and
are the water thermal diffusivity and conductivity,
respectively.
Solid circles in Fig.
5(a) illustrate the
empirical
Our empirical data obtained from the images of
5(b) implies that
. The dependence of Eq. (
2) for this value of
is illustrated with the solid curve. A good agreement
between asymptotical behavior and the empirical data is observed. Equations (
1) and (
2) suggest that the “superheat”
should be nearly constant, and by substituting values
of the dimensionless radius,
,
,
and
for water, we estimate
.
Fig. 5. Laser-induced vapor bubble: (a) radius of expanding bubble wall (data:
symbol; theory: curve) and (b) images of 408 mJ beam-initiated bubbles
in water.
Equation (
3) can be rewritten in terms
of the pressure difference [
2T. Han and J. J. Yoh, J. Appl. Phys.
107, 103 (2010). [CrossRef]
],
where
is the pressure exerted on the bubble wall by vapor,
is the liquid pressure far from the bubble wall, and
the constant
corresponds to the water heated to 100 °C.
Substituting our value of
in Eq. (
4), we estimate
and thus
.
Suppose that the chamber is closed and the volume of the liquid is comparable with the
actual volume of the bubble in the driving chamber of the injector. Then
in Eq. (
4) should decrease, taking into account the compression of the liquid inside
of the chamber. This is done by the Tait equation of state,
where
,
, the actual chamber volume
, and the volume of the bubble
.
From Eq. (
5), the
pressure on the chamber wall increases with the radius of the bubble. The bubble grows
until the pressure in the liquid compensates the pressure on the bubble wall. Assuming
that the bubble has the same initial value of
, the maximum pressure on the chamber wall should be
reasonably close to
. The rise
of the pressure inside of the chamber pushes the liquid out of the chamber. The
resulting jet velocity
can be estimated with the help of the Bernoulli
equation,
where
is approximately 225 kPa at the exit nozzle. For
the given chamber and laser parameters, the jet velocity (without the elastic membrane)
is
, which is
close to the experimentally obtained jet speed at
of Fig.
2(b).
In this Letter, the performance of an Er:YAG
laser-initiated microjet is evaluated for
transdermal drug delivery. Approximately 500 nl per pulse of drug is delivered
beyond the skin barrier in the form of a microjet as the injected drugs are effectively
dispersed over the epidermis. We ensured controllability of the laser-initiated
microjets via the longer pulse
(250 μs) at lower energy (
), which is an improvement from the Nd:YAG-based
injection scheme delivering 200 nl per pulse [
2T. Han and J. J. Yoh, J. Appl. Phys.
107, 103 (2010). [CrossRef]
].