Three-dimensional Čerenkov tomography of energy deposition from ionizing radiation beams

Adam K. Glaser, William H. A. Voigt, Scott C. Davis, Rongxiao Zhang, David J. Gladstone, and Brian W. Pogue  »View Author Affiliations


Optics Letters, Vol. 38, Issue 5, pp. 634-636 (2013)
http://dx.doi.org/10.1364/OL.38.000634


View Full Text Article

Acrobat PDF (379 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Since its discovery during the 1930s the Čerenkov effect (light emission from charged particles traveling faster than the local speed of light in a dielectric medium) has been paramount in the development of high-energy physics research. The ability of the emitted light to describe a charged particle’s trajectory, energy, velocity, and mass has allowed scientists to study subatomic particles, detect neutrinos, and explore the properties of interstellar matter. However, to our knowledge, all applications of the process to date have focused on the identification of particles themselves, rather than their effect upon the surroundings through which they travel. Here we explore a novel application of the Čerenkov effect for the recovery of the spatial distribution of ionizing radiation energy deposition in a medium and apply it to the issue of dose determination in medical physics. By capturing multiple projection images of the Čerenkov light induced by a medical linear accelerator x-ray photon beam, we demonstrate the successful three-dimensional tomographic reconstruction of the imparted dose distribution.

© 2013 Optical Society of America

Alternative modalities to overcome the prohibitive spatial profiling capabilities of ionization chambers have been proposed, including plastic or liquid scintillation and gel dosimetry, in which the number of emitted scintillation photons or a chemical change in a polymer gel is assumed to scale with imparted dose [5

5. V. Collomb-Patton, P. Boher, T. Leroux, J.-M. Fontbonne, A. Vela, and A. Batalla, Med. Phys. 36, 317 (2009). [CrossRef]

7

7. M. J. Maryanski, G. S. Ibbott, P. Eastman, R. J. Schulz, and J. C. Gore, Med. Phys. 23, 699 (1996).

]. Although the techniques offer several advantages, each requires a material other than water to record the deposited dose, an unfavorable requirement that introduces dosimetric inaccuracies due to differences in material properties. Therefore, there is immediate interest in a simple, fast, high-resolution, and noninvasive modality capable of reconstructing full three-dimensional (3D) dose distributions originating from the native water volume itself.

Here we present a method to recover volumetric dose distributions in pure water by tomographically capturing optical projection images of the induced Čerenkov radiation from a megavoltage x-ray photon LINAC beam using an intensified charge-coupled device (ICCD). Figure 1(a) presents a schematic of the experimental optical dosimetry system. X-ray photons generated by electron bombardment of a target within a medical LINAC form a pulsed radiation beam (5 μs in duration at 180 Hz) and travel downward into a perpendicularly placed water-filled glass tank, where either the primary collimator (capable of providing rectangular shapes) or the multileaf collimator (capable of producing irregular shapes) dictates the cross section of the propagating radiation beam. The two beam configurations used in the current study, field A and field B, were chosen to explore the ability of the modality to reconstruct axisymmetric and asymmetric distributions. The current produced at the target during electron irradiation is converted into a trigger voltage suitable for opening the shutter, which subsequently closes at the end of the pulse. The process repeats iteratively as signal accumulates on the camera (PI-MAX3, 1024i, Princeton Instruments, Acton, Massachusetts).

Fig. 1. (a) Experimental schematic of the proposed optical dosimetry system. The telecentric lens captures optical rays emerging from the transparent water tank that are nearly parallel to the optical axis and focuses them onto the ICCD. The two field shapes corresponding to the primary and multileaf collimator are shown in green. (b) Atomic diagram of the radiation transport interactions governing electron energy loss in water. (c) Electron energy loss per unit path length, resulting in local energy deposition and emitted Čerenkov photons as a function of electron energy.

As each radiation pulse enters the tank, secondary electrons are liberated from water molecules by photon–electron interactions, which then dissipate their energy through electromagnetic interactions with neighboring water molecules [see Fig. 1(b)]. During propagation, a small fraction of each electron’s energy is emitted as optical photons as a result of the Čerenkov effect, which is focused by a peripherally placed telecentric lens (0.06X Gold Series, Edmund Optics, Barrington, New Jersey). Because of the approximate proportionality between the electron energy loss due to ionizing events leading to energy deposition and the energy released in the form of Čerenkov photons [see Fig. 1(c)], the imaged light serves as a surrogate to indirectly determine the imparted dose distribution [i.e., the number of Čerenkov photons captured is directly correlated to the electron energy deposition at any spatial location in the irradiated medium; see [8

8. A. K. Glaser, S. C. Davis, D. M. McClatchy, R. Zhang, B. W. Pogue, and D. J. Gladstone, Med. Phys. 40, 012101 (2013). [CrossRef]

] for background theory and definitions of equations used to produce Fig. 1(c)].

To demonstrate tomographic reconstruction of 3D dose distributions, images of the induced Čerenkov radiation (10cm×10cm field of view) from a 6 MeV x-ray photon beam for both field sizes shown in Fig. 1(a) are recorded by rotating the LINAC collimators through an angle (0°θ180° in 2° increments) to provide the telecentric lens with a number of angled projections (91 total). Each projection is acquired by imaging individual pulses over 18 s and processed for noise induced by indirect irradiation of the ICCD using a 5×5pixel median filter, resulting in a 1 mm resolution (0.2 mm native image resolution), and total scan time less than 30 min. Once all projections are captured, a sinogram is constructed at each depth, z, in the irradiated water volume corresponding to a single row of pixels in each captured projection. Representative sinograms for fields A and B are shown in Figs. 2(a) and 2(d), respectively, at a depth of z=1.5cm. The resulting sinograms are then used to reconstruct two-dimensional (2D) cross sections of the induced light volume using a parallel-beam back-projection algorithm and cosine filter to prevent amplification of high-spatial-frequency noise in the reconstruction. The recovered cross sections for the sinograms in Figs. 2(a) and 2(d) are, respectively, shown in Figs. 2(b) and 2(e). Finally, the full 3D distributions are created by parsing together the 2D reconstructions from each depth.

Fig. 2. (a), (d) Sinograms and (b), (e) reconstructed cross sections for fields A and B, respectively. (c), (f) Full 3D reconstructions of fields A and B. The intensity decreases with depth as a result of exponential attenuation of the radiation beam. The full image data sets used to construct (a) and (d) (Media 3 and Media 4), as well as multimedia files for (c) and (f) (Media 1 and Media 2), are included in the supplementary material.

To examine the accuracy of the proposed system, a well-characterized and known dose profile along the central depth axis (i.e., x=0cm, y=0cm) of field A from the Varian Eclipse treatment-planning system was compared with the experimentally reconstructed distribution based on the captured Čerenkov light. The results are plotted in Fig. 3, where the recovered light profile decays exponentially and is indicative of the known dose at all depths (within ±3%) beyond the dose maximum, before which the difference is higher as a result of the large dose gradient. An alternative accuracy metric in the buildup region is the distance to agreement (i.e., shortest distance to equivalence between the two curves), which is less than 1 mm for the curves presented in Fig. 3.

Fig. 3. Comparison of the reconstructed central-axis light profile for field A with the known dose profile.

In contrast to initial reports investigating the capture of a single 2D projection, the results in the present study extend the concept to optical tomography, a significant advancement providing full 3D distributions [8

8. A. K. Glaser, S. C. Davis, D. M. McClatchy, R. Zhang, B. W. Pogue, and D. J. Gladstone, Med. Phys. 40, 012101 (2013). [CrossRef]

,9

9. A. K. Glaser, S. C. Davis, W. H. A. Voigt, R. Zhang, B. W. Pogue, and D. J. Gladstone, Phys. Med. Biol. 58, 601 (2013). [CrossRef]

]. We introduce a critical innovation by using a telecentric lens to establish a constant magnification at all imaging distances and thereby avoid parallax (i.e., objects closer appear larger, and vice versa), a technique used previously for bubble-chamber photography [10

10. K. H. Carnell, N. C. Gortmans, W. T. Welford, and C. Pataky, Opt. Acta 15, 187 (1968). [CrossRef]

]. By eliminating the perspective error associated with all conventional imaging lenses, true orthographic projections of the imparted 3D light volume suitable for optical tomography are captured. The use of a telecentric lens also provides a novel solution to the main challenge in quantitatively imaging the anisotropic light volume, an issue that was previously resolved using a Monte Carlo derived correction factor, or a fluorophore to convert the anisotropic Čerenkov light to isotropic fluorescence [8

8. A. K. Glaser, S. C. Davis, D. M. McClatchy, R. Zhang, B. W. Pogue, and D. J. Gladstone, Med. Phys. 40, 012101 (2013). [CrossRef]

,9

9. A. K. Glaser, S. C. Davis, W. H. A. Voigt, R. Zhang, B. W. Pogue, and D. J. Gladstone, Phys. Med. Biol. 58, 601 (2013). [CrossRef]

]. However, by only accepting rays parallel to the optical axis of the detection system, the telecentric lens samples the same solid angle of the anisotropic phase function of Čerenkov emission from each spatial location within the image, inherently providing accurate results (see [8

8. A. K. Glaser, S. C. Davis, D. M. McClatchy, R. Zhang, B. W. Pogue, and D. J. Gladstone, Med. Phys. 40, 012101 (2013). [CrossRef]

]).

In conclusion, we report full 3D optical tomography of ionizing radiation beams using only the induced Čerenkov radiation in a pure water volume, at a high speed and with a resolution of 1.0 mm per pixel. With the ability to noninvasively interrogate energy deposition of energetic charged particles, the technique could be applied to additional fields utilizing high-energy ionizing radiation.

This work was financially supported by NIH grants R01 CA120368 and R01 CA109558 and Department of Defense award W81XWH-09-1-0661 (SCD).

References

1.

W. C. Röntgen, Nature 53, 274 (1896).

2.

M. Weissbluth, C. J. Karzmark, R. E. Steele, and A. H. Selby, Radiology 72, 242 (1959).

3.

P. R. Almond, P. J. Biggs, B. M. Coursey, W. F. Hanson, M. S. Huq, R. Nath, and D. W. O. Rogers, Med. Phys. 26, 1847 (1999). [CrossRef]

4.

E. B. Podgorsak, Radiation Oncology Physics: A Handbook for Teachers and Students (IAEA, 2005).

5.

V. Collomb-Patton, P. Boher, T. Leroux, J.-M. Fontbonne, A. Vela, and A. Batalla, Med. Phys. 36, 317 (2009). [CrossRef]

6.

F. Ponisch, L. Archambault, T. M. Briere, N. Sahoo, R. Mohan, S. Beddar, and M. T. Gillin, Med. Phys. 36, 1478 (2009). [CrossRef]

7.

M. J. Maryanski, G. S. Ibbott, P. Eastman, R. J. Schulz, and J. C. Gore, Med. Phys. 23, 699 (1996).

8.

A. K. Glaser, S. C. Davis, D. M. McClatchy, R. Zhang, B. W. Pogue, and D. J. Gladstone, Med. Phys. 40, 012101 (2013). [CrossRef]

9.

A. K. Glaser, S. C. Davis, W. H. A. Voigt, R. Zhang, B. W. Pogue, and D. J. Gladstone, Phys. Med. Biol. 58, 601 (2013). [CrossRef]

10.

K. H. Carnell, N. C. Gortmans, W. T. Welford, and C. Pataky, Opt. Acta 15, 187 (1968). [CrossRef]

OCIS Codes
(170.0110) Medical optics and biotechnology : Imaging systems
(170.1610) Medical optics and biotechnology : Clinical applications
(170.3890) Medical optics and biotechnology : Medical optics instrumentation
(170.6960) Medical optics and biotechnology : Tomography

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: December 21, 2012
Revised Manuscript: January 18, 2013
Manuscript Accepted: January 18, 2013
Published: February 21, 2013

Virtual Issues
Vol. 8, Iss. 4 Virtual Journal for Biomedical Optics

Citation
Adam K. Glaser, William H. A. Voigt, Scott C. Davis, Rongxiao Zhang, David J. Gladstone, and Brian W. Pogue, "Three-dimensional Čerenkov tomography of energy deposition from ionizing radiation beams," Opt. Lett. 38, 634-636 (2013)
http://www.opticsinfobase.org/ol/abstract.cfm?URI=ol-38-5-634


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. W. C. Röntgen, Nature 53, 274 (1896).
  2. M. Weissbluth, C. J. Karzmark, R. E. Steele, and A. H. Selby, Radiology 72, 242 (1959).
  3. P. R. Almond, P. J. Biggs, B. M. Coursey, W. F. Hanson, M. S. Huq, R. Nath, and D. W. O. Rogers, Med. Phys. 26, 1847 (1999). [CrossRef]
  4. E. B. Podgorsak, Radiation Oncology Physics: A Handbook for Teachers and Students (IAEA, 2005).
  5. V. Collomb-Patton, P. Boher, T. Leroux, J.-M. Fontbonne, A. Vela, and A. Batalla, Med. Phys. 36, 317 (2009). [CrossRef]
  6. F. Ponisch, L. Archambault, T. M. Briere, N. Sahoo, R. Mohan, S. Beddar, and M. T. Gillin, Med. Phys. 36, 1478 (2009). [CrossRef]
  7. M. J. Maryanski, G. S. Ibbott, P. Eastman, R. J. Schulz, and J. C. Gore, Med. Phys. 23, 699 (1996).
  8. A. K. Glaser, S. C. Davis, D. M. McClatchy, R. Zhang, B. W. Pogue, and D. J. Gladstone, Med. Phys. 40, 012101 (2013). [CrossRef]
  9. A. K. Glaser, S. C. Davis, W. H. A. Voigt, R. Zhang, B. W. Pogue, and D. J. Gladstone, Phys. Med. Biol. 58, 601 (2013). [CrossRef]
  10. K. H. Carnell, N. C. Gortmans, W. T. Welford, and C. Pataky, Opt. Acta 15, 187 (1968). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1. Fig. 2. Fig. 3.
 

Supplementary Material


» Media 1: MP4 (1501 KB)     
» Media 2: MP4 (1543 KB)     
» Media 3: MP4 (930 KB)     
» Media 4: MP4 (904 KB)     

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited