Abstract
In this paper, the imaging performance of an Offner concentric imaging spectrometer is analyzed when the spectrometer entrance slit is disposed arbitrarily on the plane that is parallel to the grating grooves and contains the common center of curvature. Astigmatism-corrected designs are obtained for off-plane incidence on the grating if one point on the slit is located on the Rowland circle of the primary mirror. In this case, the combined system of primary mirror plus diffraction grating provides two astigmatic line images oriented parallel and orthogonal to the plane of diffraction, with the former located on the same plane as the slit. Consequently, these images can be brought to a single focus on this plane by the tertiary mirror if its radius of curvature is chosen properly. In addition, coma aberration is simultaneously removed. These results can be applied to the design of two-mirror or three-mirror spectrometers, generalizing the concept of the best imaging circle and providing solutions to get anastigmatic imaging for two object points and two wavelengths.
© 2011 Optical Society of America