Non-contact detection of oxygen saturation based on visible light imaging device using ambient light

Lingqin Kong, Yuejin Zhao, Liquan Dong, Yiyun Jian, Xiaoli Jin, Bing Li, Yun Feng, Ming Liu, Xiaohua Liu, and Hong Wu  »View Author Affiliations


Optics Express, Vol. 21, Issue 15, pp. 17464-17471 (2013)
http://dx.doi.org/10.1364/OE.21.017464


View Full Text Article

Enhanced HTML    Acrobat PDF (2177 KB) Open Access





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

A method that remotely measures blood oxygen saturation through two cameras under regular lighting is proposed and experimentally demonstrated. Two narrow-band filters with their visible wavelength of 660nm and 520nm are mounted to two cameras respectively, which are then used to capture two photoplethysmographic (PPG) from the subject simultaneously. The data gathered from this system, including both blood oxygen saturation and heart rate, is compared to the output of a traditional figure blood volume pulse (BVP) senor that was employed on the subject at the same time. Result of the comparison showed that the data from the new, non-contact system is consistent and comparable with the BVP senor. Compared to other camera-based measuring method, which requires additional close-up lighting, this new method is achievable under regular lighting condition, therefore more stable and easier to implement. This is the first demonstration of an accurate video-based method for non-contact oxygen saturation measurements by using ambient light with their respective visible wavelength of 660nm and 520nm which is free from interference of the light in other bands.

© 2013 OSA

OCIS Codes
(170.0170) Medical optics and biotechnology : Medical optics and biotechnology
(230.0230) Optical devices : Optical devices
(280.4991) Remote sensing and sensors : Passive remote sensing

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: April 29, 2013
Revised Manuscript: June 15, 2013
Manuscript Accepted: July 1, 2013
Published: July 15, 2013

Virtual Issues
Vol. 8, Iss. 8 Virtual Journal for Biomedical Optics

Citation
Lingqin Kong, Yuejin Zhao, Liquan Dong, Yiyun Jian, Xiaoli Jin, Bing Li, Yun Feng, Ming Liu, Xiaohua Liu, and Hong Wu, "Non-contact detection of oxygen saturation based on visible light imaging device using ambient light," Opt. Express 21, 17464-17471 (2013)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-21-15-17464


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. X. F. Teng and Y. T. Zhang, “The effect of contacting force on photoplethysmographic signals,” Physiol. Meas.25(5), 1323–1335 (2004). [CrossRef] [PubMed]
  2. C. Takano and Y. Ohta, “Heart rate measurement based on a time-lapse image,” Med. Eng. Phys.29(8), 853–857 (2007). [CrossRef] [PubMed]
  3. M. Z. Poh, D. J. McDuff, and R. W. Picard, “Non-contact, automated cardiac pulse measurements using video imaging and blind source separation,” Opt. Express18(10), 10762–10774 (2010). [CrossRef] [PubMed]
  4. W. Verkruysse, L. O. Svaasand, and J. S. Nelson, “Remote plethysmographic imaging using ambient light,” Opt. Express16(26), 21434–21445 (2008). [CrossRef] [PubMed]
  5. F. P. Wieringa, F. Mastik, and A. F. W. van der Steen, “Contactless multiple wavelength photoplethysmographic imaging: a first step toward “SpO(2) camera’ technology,” Ann. Biomed. Eng.33(8), 1034–1041 (2005). [CrossRef] [PubMed]
  6. M. Z. Poh, D. J. McDuff, and R. W. Picard, “Advancements in noncontact, multiparameter physiological measurements using a webcam,” IEEE Trans. Biomed. Eng.58(1), 7–11 (2011). [CrossRef] [PubMed]
  7. C. G. Scully, J. Lee, J. Meyer, A. M. Gorbach, D. Granquist-Fraser, Y. Mendelson, and K. H. Chon, “Physiological parameter monitoring from optical recordings with a mobile phone,” IEEE Trans. Biomed. Eng.59(2), 303–306 (2012). [CrossRef] [PubMed]
  8. J. A. Pollard, “Cardiac arrhythmias and pulse variability: a plethysmographic study,” Anaesthesia25(1), 63–72 (1970). [CrossRef] [PubMed]
  9. H. D. Hummler, A. Engelmann, F. Pohlandt, J. Högel, and A. R. Franz, “Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis,” Intensive Care Med.30(4), 709–713 (2004). [CrossRef] [PubMed]
  10. N. S. Trivedi, A. F. Ghouri, N. K. Shah, E. Lai, and S. J. Barker, “Effects of motion, ambient light, and hypoperfusion on pulse oximeter function,” J. Clin. Anesth.9(3), 179–183 (1997). [CrossRef] [PubMed]
  11. K. Humphreys, T. Ward, and C. Markham, “Noncontact simultaneous dual wavelength photoplethysmography: a further step toward noncontact pulse oximetry,” Rev. Sci. Instrum.78(4), 044304 (2007). [CrossRef] [PubMed]
  12. K. Humphreys, T. Ward, and C. Markham, “A CMOS camera-based pulse oximetry imaging system,” Conf. Proc. IEEE Eng. Med. Biol. Soc.4, 3494–3497 (2005). [PubMed]
  13. J. Zheng, S. Hu, A. S. Echiadis, V. A. Peris, P. Shi, and V. Chouliaras, “A remote approach to measure blood perfusion from the human face,” Proc. SPIE7169, 171–177 (2009). [CrossRef]
  14. W. Verkruysse, L. O. Svaasand, and J. S. Nelson, “Remote plethysmographic imaging using ambient light,” Opt. Express16(26), 21434–21445 (2008). [CrossRef] [PubMed]
  15. H. S. Zhu, Y. J. Zhao, and L. Q. Dong, “Non-contact detection of cardiac rate based on visible light imaging device,” Proc. SPIE8498, 849806, 849806-7 (2012). [CrossRef]
  16. J. M. Bland and D. G. Altman, “Statistical methods for assessing agreement between two methods of clinical measurement,” Lancet327(8476), 307–310 (1986). [CrossRef] [PubMed]
  17. Y. Sun, C. Papin, V. Azorin-Peris, R. Kalawsky, S. Greenwald, and S. Hu, “Use of ambient light in remote photoplethysmographic systems: comparison between a high-performance camera and a low-cost webcam,” J. Biomed. Opt.17(3), 037005 (2012). [CrossRef] [PubMed]
  18. Y. Sun, S. Hu, V. Azorin-Peris, S. Greenwald, J. Chambers, and Y. Zhu, “Motion-compensated noncontact imaging photoplethysmography to monitor cardiorespiratory status during exercise,” J. Biomed. Opt.16(7), 077010 (2011). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited