Video-rate near infrared tomography to image pulsatile absorption properties in thick tissue

Zhiqiu Li, Venkataramanan Krishnaswamy, Scott C. Davis, Subhadra Srinivasan, Keith D. Paulsen, and Brian W. Pogue  »View Author Affiliations


Optics Express, Vol. 17, Issue 14, pp. 12043-12056 (2009)
http://dx.doi.org/10.1364/OE.17.012043


View Full Text Article

Acrobat PDF (779 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

A high frame-rate near-infrared (NIR) tomography system was created to allow transmission imaging of thick tissues with spectral encoding for parallel source implementation. The design was created to maximize tissue penetration through up to 10 cm of tissue, allowing eventual use in human imaging. Eight temperature-controlled laser diodes (LD) are used in parallel with 1.5 nm shifts in their lasing wavelengths. Simultaneous detection is achieved with eight high-resolution, CCD-based spectrometers that were synchronized to detect the intensities and decode their source locations from the spectrum. Static and dynamic imaging is demonstrated through a 64 mm tissue-equivalent phantom, with acquisition rates up to 20 frames per second. Imaging of pulsatile absorption changes through a 72 mm phantom was demonstrated with a 0.5 Hz varying object having only 1% effect upon the transmitted signal. This subtle signal change was used to show that while reconstructing the signal changes in a tissue may not be possible, image-guided recovery of the pulsatile change in broad regions of tissue was possible. The ability to image thick tissue and the capacity to image periodic changes in absorption makes this design well suited for tracking thick tissue hemodynamics in vivo during MR or CT imaging.

© 2009 OSA

1. Introduction

High speed near-infrared spectral tomography of tissue can be used to measure functional properties of tissues such as metabolism, hemodynamics and contrast agents to provide useful diagnostic information. Imaging at a speed much faster than the tissue activity rate of interest is important to provide quantitative information with high signal to noise ratio (SNR), and potentially to allow signal processing methods to be used which can amplify low SNR. But practically building a fast scanning tomography system for biomedical application remains a challenging task, as the diffuse pattern of light propagation in biological tissue demands that each source-detector (S-D) pair be encoded properly through delivery and decoded afterwards. Accumulating enough photons at the detectors in a limited exposure time is difficult, especially when the tissue is thicker, e.g. more than 5–6 cm for imaging most critical organs or tissue volumes. Methods for mechanical switching of S-D pairs have been routinely used, with imaging speeds reported up to 3 Hz [1

1. C. H. Schmitz, M. Löcker, J. M. Lasker, A. H. Hielscher, and R. L. Barbour, “Instrumentation for fast functional optical tomography,” Rev. Sci. Instrum. 73(2), 429 (2002). [CrossRef]

]. Simultaneous source delivery with frequency encoding [2

2. S. Fantini, M. A. Franceschini, G. Gaida, E. Gratton, H. Jess, W. W. Mantulin, K. T. Moesta, P. M. Schlag, and M. Kaschke, “Frequency-domain optical mammography: edge effect corrections,” Med. Phys. 23(1), 149–157 (1996). [CrossRef]

] can be done considerably faster, but is limited by the overwhelming effect of closer strong signals dominating over the farther weaker signals that propagate through a few cm of tissue. This approach, while tremendously successful in sub-surface imaging [3

3. A. M. Siegel, J. J. A. Marota, and D. A. Boas, “Design and evaluation of a continuous-wave diffuse optical tomography system,” Opt. Express 4(8), 287–298 (1999).

], cannot be realized in deep tissue tomography, because of the severe dynamic range variation seen in signals that propagate different distances in tissue.

Our earlier work reported a configuration with spectrally encoded sources around a circular coil, where the spectral encoding was introduced as a way to eliminate signal interference at the detector [4

4. D. Piao, H. Dehghani, S. Jiang, S. Srinivasan, and B. W. Pogue, “Instrumentation for video-rate near-infrared diffuse optical tomography,” Rev. Sci. Instrum. 76(12), 124301 (2005). [CrossRef]

]. The sources were identical LDs with different temperature-controlled wavelength shifts, and using a single imaging spectrometer detection at 30 frames per second was achieved with this setup. The design demonstrated the concept, with limited dynamic range capability because of the single CCD detector and the circular geometry of the coil. It worked for small circular-shaped tissues with a radius of less than 3 cm, but a revised design was needed for larger tissue volumes. In this study, a parallel detector design is implemented which demonstrates the ability to image through much thicker tissues and therefore has applications in breast tissue or brain tissue tomography.

Fig. 1. A diagram of the video-rate near infrared tomography system is shown. Eight spectrally encoded LD systems were integrated onto one cart [5]. Eight fiber-coupled spectrometers with high-resolution CCDs were integrated onto another cart and set to respond to an external TTL trigger signal. The signal was generated by a data acquisition board and then split into eight channels by a customized splitter circuit, and delivered to the EXT SYNC port on each CCD simultaneously. Imaging data are transferred back to the computer for post processing. A flexible fiber mount was customized to hold the phantom or tissue in between.

2. System Configuration

The video-rate tomography system [5

5. Z. Li, V. Krishnaswamy, K. D. Paulsen, and B. W. Pogue, “Video-rate near infrared tomography for imaging thick tissue with dynamically varying absorption properties,” Proc. SPIE 7174, 71742G (2009). [CrossRef]

], depicted in Fig. 1, consists of several sub systems. Eight spectrally encoded laser diodes (LDs) each with a temperature control module and a current control module were integrated on to one cart. Eight fiber-coupled spectrometers, each equipped with a high-resolution charge-coupled device (CCD), were integrated on to another cart and set to respond to an external TTL trigger signal. The signal was generated by a data acquisition (DAQ) board from National Instrument (NI) and programmed through LabVIEW software. A customized splitter circuit was built to convey the signal into eight channels and finally to the EXT SYNC port on each CCD simultaneously. Imaging data are transferred back to the console for post processing through USB connection with the CCDs [5

5. Z. Li, V. Krishnaswamy, K. D. Paulsen, and B. W. Pogue, “Video-rate near infrared tomography for imaging thick tissue with dynamically varying absorption properties,” Proc. SPIE 7174, 71742G (2009). [CrossRef]

]. A customized flexible fiber mount and phantom/tissue interface was built to hold the phantom in the parallel plate slab geometry with the fibers in contact with it. Each component is described in detail below.

2.1 Spectrally encoded sources

In conventional NIR tomography systems, light is delivered to one plane at the same wavelength. Therefore sources are launched in sequence to avoid interference at the detectors. So the time needed for one image is proportional to the number of sources in use, unless temporal frequency encoding methods are used. The new video-rate tomography system uses 8 identical single mode continuous wave (CW) LDs (Hitachi HL-7851G) in the 785 nm band. They were spectrally encoded to have approximately 1.5 nm interval in between by shifting the temperature of each LD differently. This interval is wide enough for the 1200 l/mm grating on the CCD to separate each source signal clearly, but still very narrow so that the tissue optical properties remains almost the same. The difference of the absorption coefficient in this range for whole blood is within 5%. Therefore, all source LDs can be launched simultaneously to improve imaging speed without interference with each other. The stability of the LDs under wavelength shift has been verified by sampling over long periods of time [5

5. Z. Li, V. Krishnaswamy, K. D. Paulsen, and B. W. Pogue, “Video-rate near infrared tomography for imaging thick tissue with dynamically varying absorption properties,” Proc. SPIE 7174, 71742G (2009). [CrossRef]

].

2.2 Optical Detectors

The optical detection system consisted of 8 Princeton/Acton Insight: 400F Integrated Spectroscopy Systems (Acton, MA) residing in one custom designed wheeled carts (8020, Columbia City, IN) [6

6. S. C. Davis, B. W. Pogue, R. Springett, C. Leussler, P. Mazurkewitz, S. B. Tuttle, S. L. Gibbs-Strauss, S. S. Jiang, H. Dehghani, and K. D. Paulsen, “Magnetic resonance-coupled fluorescence tomography scanner for molecular imaging of tissue,” Rev. Sci. Instrum. 79(6), 064302 (2008). [CrossRef]

]. Insight 400F consists of a 0.3 m length, F3.9 imaging spectrograph and a low noise, front illuminated charge coupled device (CCD) (Pixis 400F) cooled to −70 °C. The 16- bit CCD provided a large dynamic range of 4.8 orders and was vertically binned to maximize detector area/wavelength. A 1200 l/mm grating in each spectrometer provided a sufficiently fine spectral resolution, near 0.045 nm, which is much smaller than the 1.5 nm laser bandwidth interval, and a wide enough spectral range of 60 nm to allow simultaneous imaging of all wavelengths.

2.3 Tissue/Phantom Interface

2.4 Exposure Synchronization

3. Experiments and results

3.1 Static Blood Phantom Imaging Experiment

A set of heterogeneous phantom experiments was performed to assess the linearity and accuracy of the system. The slab-shaped resin phantom used for this purpose was 134 mm long and 64.5 mm thick. The background optical properties of the phantom, µa=0.004 mm−1 and µs=1.4 mm−1, were measured separately using a well characterized frequency domain clinical diffuse optical tomography system at 785 nm [7

7. S. Jiang, B. W. Pogue, S. Srinivasan, S. Soho, S. P. Poplack, T. D. Tosteson, and K. D. Paulsen, “Assessment of the menstrual cycle upon total hemoglobin, water concentration and oxygen saturation in the female breast,” Proc. SPIE 4955, 342–348 (2003). [CrossRef]

]. A cylindrical hole with a radius of 10 mm was drilled vertically at about 16 mm away from the center of the phantom surface. A batch of liquid phantoms with linearly varying volume fractions of swine blood, ranging from 0.0% to 4.0% with an interval of 0.5%, were made in 1.0% stock Intralipid solution. Spectrographs were set to operate at 20 frames per second. Approximately 10 mW of optical power was delivered into the phantom by each of the 8 source fibers. The hole was filled up with one phantom solution at a time and NIR tomography images were acquired for about 1 second.

As this is a static imaging experiment, a reconstructed image of one representative frame for each phantom solution was displayed in this paper. The calibration and reconstruction algorithm used was similar to that described by [8

8. H. Dehghani, B. W. Pogue, S. P. Poplack, and K. D. Paulsen, “Multiwavelength three-dimensional near-infrared tomography of the breast: initial simulation, phantom, and clinical results,” Appl. Opt. 42(1), 135–145 (2003). [CrossRef]

]. Briefly, the reconstruction is based on a modified Newton’s method for minimization of the difference between measured and model data; the model used here was a finite element discretization of the diffusion equation. The measured data from the tissue/phantom was first calibrated to account for small offsets due to source–detector fiber transmission, alignment characteristics, and errors in discretization or model–data mismatch. This was accomplished using a homogeneous fitting algorithm [9

9. T. O. McBride, “Spectroscopic Reconstructed Near Infrared Tomographic Imaging for Breast Cancer Diagnosis,” (Dartmouth College, 2001). http://nir.thayer.dartmouth.edu/Downloads.html

] which utilized Newton-Raphson method for fitting of two parameters: the slope of the phase (log(θ)) with respect to distance from the source location, and the slope of the logarithm of intensity times distance with respect to distance (log(rI)). Since this video-rate tomography system used CW LDs and hence had no ability to measure pathlength of transmission, the spatial distribution of the reduced scattering coefficient µs‘ was not attempted to be recovered. Thus an assumption about the spatial distribution of µs was needed to reconstruct the spatial distribution of the absorption coefficient µa. As the concentration of scatterers was constant in all phantoms, a homogeneous distribution of µs‘=1.4 mm−1, which was the same as the background and close to the scattering coefficient of 1.0% Intralipid solution, was used in the reconstruction calculation. The applicability of this assumption will be analyzed later in the Discussions section.

3.1.1 Direct Reconstruction of µa

Direct reconstruction here means to reconstruct for real value of µa which reflects the actual local absorption property across the 2D imaging plane. The data set used for reconstruction was obtained from heterogeneous phantom measurement and calibrated with another data set obtained from homogeneous phantom measurement. The calibration formula is:

Φcalibrated(hetero)=Φmeasured(hetero)(Φmeasured(homo)Φcalculated(homo))Φoffset(net)
(1)

where each Φ is the data set consisting of logarithm of the amplitude and phase shift, and the calibrated data set is the one used for reconstruction. Details about this calibration method can be found in [9

9. T. O. McBride, “Spectroscopic Reconstructed Near Infrared Tomographic Imaging for Breast Cancer Diagnosis,” (Dartmouth College, 2001). http://nir.thayer.dartmouth.edu/Downloads.html

].

For each swine blood concentration, about 20 frames of raw data were acquired continuously. After putting these data sets into the reconstruction program, it was found that the standard deviation of the µa within the ROI was less than 1% of the mean value of µa, so the system and the algorithm were proven to be very stable and only the result of one frame was listed in Fig. 2 for each swine blood concentration. The reconstructed images of µa value on the imaging plane with linearly increasing concentration of swine blood in the inclusion were listed in Figs. 2(a)2(d). The central position of the reconstructed inclusion was recovered accurately compared to its actual position in the phantom. The recovered shape appeared stretched in the vertical direction and compressed in the horizontal direction. This was believed to be due to the slab geometry of the source-detector arrangement, which enhanced imaging and reconstruction sensitivity in the vertical direction comparing to the horizontal direction. The true value of µa in each solution was calculated from oxygen saturation previously measured with a commercial oximeter (Hb 201 +, HemoCue). The average reconstructed µa within the region-of-interest (ROI) inside the inclusion was calculated for each case and a linear response was observed as shown in Fig. 2(f). The absolute value of recovered µa was found to be slightly larger than the true µa, which may be caused by the inaccurate assumption made to µs of the ROI. The artifacts close to the top boundary on these images were mainly due to the coupling error of source and detector fibers, which can be repeated through simulation. Besides, it was also found that the reconstructed µa of ROI differed more from its true value when the µa of ROI was close to or lower than the background. This should be due to the reflection index mismatch on the boundary of the inclusion. Fortunately the µa of the tumor region is usually larger than that of the background, so that is not a big problem for clinical use. As expected, the recovered mean background µa outside the ROI remained constant for all phantoms.

Fig. 2. Direct reconstruction results of µa on the imaging plane of a heterogeneous phantom with different blood concentrations in the inclusion. Figures 2(a)2(d) showed phantoms with swine blood concentration of 1%, 2%, 3% and 4%. The reconstructed and true values of µa in the inclusion of different swine blood concentrations were listed in Fig. 2(e), and plotted in Fig. 2(f) together with the reconstructed average value of background µa.

3.1.2 Difference Reconstruction of µa

Difference reconstruction here means to reconstruct only for the change of µa in the inclusion with data of two frames. The formule is:

Φdifference(hetero)i=Φhomo_fit(hetero)i+(Φcalibrated(hetero)iΦcalibrated(hetero)l)
(2)

Diffuse tomography reconstruction was performed after the experiments were completed on the processed data. The time needed to reconstruct one frame mostly depends on the mesh size. With the current settings, it takes 10 seconds to reconstruct one frame with a standard IBM laptop, and so 200 frames took about 33 minutes total processing time.

The difference reconstruction results of µa with different swine blood concentrations in the inclusion are listed in Figs. 3(a)3(d). For the same reason as the direct reconstruction case, only the result of one frame out of about 20 frames was listed in Fig. 3 for each swine blood concentration. The general position of the reconstructed inclusion was the same as its actual position in the phantom, and the artifact around upper boundary was greatly removed as expected. However, instead of being at the center of the inclusion, the brightest region shifted up a little to be closer to the upper boundary of the phantom. It was verified through simulation that this shift didn’t exist with noise-free data, but became more and more apparently with increasing noise level and towards the boundary, which had higher imaging and reconstruction sensitivity. Although the noise was largely canceled in Eq. (2), it was difficult to estimate the effect of remaining noise on the reconstructed image. The reconstructed mean µa of the ROI fitted with the true value reasonably well, especially when the µa of the ROI was larger than that of the background. This phenomenon was identical to that of direct reconstruction. No change in background µa was observed.

Fig. 3. Difference reconstruction results of µa on the imaging plane of a heterogeneous phantom with different blood concentrations in the inclusion. Figures 3(a)3(d) showed phantoms with swine blood concentration of 1%, 2%, 3% and 4%. The reconstructed and true values of difference µa in the inclusion of different swine blood concentrations were listed in Fig. 3(e), and plotted in Fig. 3(f) together with the mean value of background µa.

3.2 Dynamic Blood Phantom Imaging Experiment

A blood diffusion experiment was carried out to test the responsivity of the system to dynamic changes in absorption contrast. The phantom and system setup were the same as the static phantom experiment described in 3.1. Initially, the phantom inclusion was filled with 1% intralipid solution. After imaging the phantom at 20 frames per second for about 30 seconds, swine blood was added into the inclusion slowly through a syringe, the adding process took about 50 seconds, and the phantom was continuously imaged for about 40 seconds after the completion of the addition.

Results of both direct reconstruction and difference reconstruction were shown in Fig. 4. In both figures, the time of the first rising point and the time µa reached the maximum fitted well with the experiment procedure. The µa value in Fig. 4(a) of direct reconstruction couldn’t be verified, because the solution in the inclusion was not stirred during imaging to get rid of the impact introduced by the stirring tool. Thus the blood was not evenly distributed in the inclusion, and it was meaningless to calculate the true µa value based on the volume of swine blood added into the inclusion. The calibrated data set of the first frame was used in the difference reconstruction as the “baseline” to be subtracted by calibrated data sets of following frames. In the first 30 seconds before adding blood, there is no change of the absorber in the inclusion, so after subtraction only the homogenous fitting part was left unchanged. The first 30 seconds section of the curve µa_ROI met with that very well. And as more blood being added in, the variation of µa in the inclusion increased and reached the peak at around 80 seconds and then went to a stable status, as curve µa_ROI shown in Fig. 4(b). It was noted that the largest difference of µa in Fig. 4(b) was smaller than that in Fig. 4(a), mainly because the µa value of direct reconstruction was slightly larger than the true value (as shown in Fig. 2(f)) and the µa value of difference reconstruction was slightly smaller than the true value (as shown in Fig. 3(f)). Overall, the results shown in Fig. 4 demonstrated the reliable functioning of this video-rate tomography system to catch dynamic varying absorption content at imaging speed of 20 frames per second.

Fig. 4. Results of dynamic blood diffusion phantom experiment. Figure 4(a) showed the direct reconstructed µa of both ROI and background versus time. Figure 4(b) shows the difference reconstructed µa value of both ROI and background versus time. Curve µa_ROI is the mean value of µa in the inclusion, while curve µa_BKG is the mean value of µa of the background. 10 mW LDs were used as source beam. 100 ms exposure time and 100 µm slit width were set up at all spectrographs. Images were acquired at 20 frames per second.

3.3 Pulsatile Phantom Experiment

The clinical goal of this video-rate NIR tomography system is to image tissue hemodynamics. Thus a pulsatile phantom experiment was carried out to test the ability of this system to image an abnormity in real-time whose absorption property varied at a preset frequency with moderate effect upon the transmitted signal. The experimental setup is illustrated in Fig. 5. A solution made of 1% intralipid and low concentration of Indian ink was filled into a 72 mm thick slab-shaped high-density polypropylene bottle (32 oz HDPE Rectangular Bottle, Nalgene) to provide a homogeneous background with low µa. The plastic bottle was colorless with a cloudy appearance that prevented straight light channeling laterally, so it was a fairly good container for this experiment. Several sides of the bottle were sealed with black tape to minimize the impact of boundary mismatch. Another tank held solution made of the same 1% intralipid and 3 times higher concentration of Indian ink than that in the Nalgene bottle. So the absorption contrast of two solutions was 3:1. The high µa solution was continuously pumped from the tank into an incoming latex tubing with an internal diameter (ID) of ¼ inch. This tubing was connected to one end of a T-shape connector. The second end of the T-shape connector was connected to an outgoing clear vinyl tube with the same ID and directed the high absorbing solution back to the tank. A white balloon was tied up to the third end of the connector and sealed with water-proof glue. The connector was fixed on a three-jaw chuck with the balloon submerged into the low µa solution in the Nalgene bottle. The incoming tubing was periodically compressed by the rotational part of a pump (503U/RL, Watson-Marlow L.L.C.) at a preset frequency. The pump was driven by a frequency-adjustable 0~10 Volt square wave voltage generator. Therefore the high µa solution was pumped through the T-shape connector in pulsatile mode at the same frequency, and the balloon expanded and contracted at the driving frequency.

Fig. 5. System setup of the pulsatile phantom experiment. High µa solution was continuously pumped through a balloon at 0.5 Hz. The balloon was submerged in a 72 mm thick slab container filled with low µa solution. The absorption contrast of the solution inside the balloon against outside was 3 to 1. Eight LDs were launched as sources, and 6 spectrometer systems were set as detectors to acquire data at 10 frames per second. Slit widths on all spectrographs were set to 100 µm.
Fig. 6. Signals of different S-D pairs in frequency domain. Figure 6(a) is detected signals of 3 different optical paths in frequency domain after being normalized to their individual mean intensity. Only 0 to 1 Hz of these spectra were displayed. Figure 6(b) illustrates the 3 different optical paths through the phantom with straight lines. D1 to D6 were 6 detection spots connected to spectrographs through optical fibers. S1 to S8 were 8 laser beam input spots connected to LDs through optical fibers.

This pumping system was designed to mimic the cyclical change in absorption seen in vaso-activity within tissue. The video-rate tomography system was used to image and recover this pulsatile change of the absorbing content, and quantify the magnitude of the effect upon the transmitted signal. The experiment was designed such that the elasticity of the balloon would only allow response below pump frequencies of 1 Hz. So a 0.5 Hz drive was chosen as the experimental pumping frequency for this study. Most pumped solution went through the T-shape connector directly, and only a small amount of the solution was to be pushed into the balloon corresponding to the change of pressure difference of the incoming and outgoing tubes. The volume of high µa solution in the balloon and thus the size of the balloon would vary according to the pumping frequency, but only on a very subtle scale which was barely apparent and not significantly measurable. So the variation of µa in the balloon was expected to be very small, and was quantified by the effect it had on the transmission of the light.

The detected signals for 3 different optical paths in frequency domain (FD) are displayed in Fig. 6. After subtracting the mean value of each signal to remove DC component, a Fourier transform was performed on each signal and then the signal was normalized to the intensity of DC component separately, such that the change is relative to 1.0. The relative intensity of every frequency component on the FD spectrum reflected the portion of signal caused by the varying portion of absorbing property versus that by the homogeneous background. It is obvious from Fig. 6(a) that path 1, which was the shortest and went through the balloon, received strongest 0.5 Hz pulsatile information, increasing from 0.005 up to 0.015, or 0.5% to 1.5%. Meanwhile, path 3 received weakest 0.5 Hz pulsatile signal since it was the longest, although it also went through the balloon. The largest effect of the pulsatile flow was set at approximately 1% change in the DC signal, as can be seen in the peak of Figs. 6(a) and 6(b). This low magnitude in pulsatile flow is approximately what is seen in pulse oximetry, and so was chosen as a good example of typical physiological signal level.

After imaging the pulsatile phantom at 10 frames per second for 20 seconds, the raw data set was calibrated and put into the reconstruction program. Mean µa of the ROI using direct reconstruction algorithm versus time is shown in Fig. 7(a). There was a blurry periodical trend in curve µa, but the noise made it not easy to distinguish. A Fourier transform was applied to the µa curve, and the result was shown in Fig. 7(b). Since the experimental imaging speed was 10 frames per second, the highest distinguishable frequency after Fourier transform was 5 Hz. There was a peak right at 0.5 Hz, but several other peaks with similar intensity at higher frequencies made it not convincing to conclude the system had picked out the desired pulse. The result of difference reconstruction was not listed here, because the reconstructed µa difference was very small and could not be clearly discerned from reconstruction artifacts around.

Fig. 7. Reconstructed µa of the ROI in the time-domain and frequency-domain. In Fig. 7(a) the mean recovered µa value in the ROI is plotted versus time with a direct diffuse reconstruction method. In Fig. 7(b) the Fourier transform of the mean µa in Fig. 7(a) is shown. In Fig. 7(c) the mean µa recovered in the ROI versus time is shown using region-guided direct reconstruction methods. Figure 7(d) is the Fourier transformed amplitude data from Fig. 7(c), showing the dominance of the 0.5 Hz signal in this frequency spectrum.

A region-based direct reconstruction method was also applied to this data set. Classically the imaging plane can be segmented into several homogeneous regions based on the structural information obtained in advance with other imaging modalities, e.g. MRI and/or x-ray CT. The iterative update is then simplified from the situation where individual nodes are reconstructed to the problem where only the parameters for each homogenous region are estimated in the inversion [10

10. B. Brooksby, H. Dehghani, B. W. Pogue, and K. D. Paulsen, “Near infrared (NIR) tomography breast image reconstruction with a priori structural information from MRI: algorithm development for reconstructing heterogeneities,” IEEE J. STQE 9, 199–209 (2003).

]. In this case, instead of inverting a matrix that is thousands of rows, it has just 2 rows, one for each material. The advantage of this method is that the contrast is enhanced between the inclusion and the background. The two regions defined in this experiment were from direct measurements of the phantom size and shape and location of the balloon. Since the balloon size change was too small to measure between the pulsatile motion, this was assumed to be constant. In Fig. 7(c) the recovered µa of the ROI versus time showed a clear periodical trend, and after Fourier transformation a sharp peak at 0.5 Hz was recovered in Fig. 7(d), which is much higher than the background. This peak verifies that this region-based method does allow recovery of the pulsatile varying absorption content in the balloon, likely due to noise suppression through the combined effect of using all the measurements together. As such the inversion matrix problem is highly over-conditioned, and so is a well posed problem.

4. Discussion

This work has shown the feasibility of using this spectrally-encoded tomography system to do fast speed imaging through thick tissues, of the size required in clinical breast imaging [7

7. S. Jiang, B. W. Pogue, S. Srinivasan, S. Soho, S. P. Poplack, T. D. Tosteson, and K. D. Paulsen, “Assessment of the menstrual cycle upon total hemoglobin, water concentration and oxygen saturation in the female breast,” Proc. SPIE 4955, 342–348 (2003). [CrossRef]

10

10. B. Brooksby, H. Dehghani, B. W. Pogue, and K. D. Paulsen, “Near infrared (NIR) tomography breast image reconstruction with a priori structural information from MRI: algorithm development for reconstructing heterogeneities,” IEEE J. STQE 9, 199–209 (2003).

]. The goal of the system was to achieve high speed imaging, of blood pulsation dynamics, and show the validation that the system could be used to recover subtle signals in realistic tissue phantoms. The system was built around CCD detection which allows only CW detection, and does not have the capability to image frequency domain data. This choice was intentionally made because of the high cost that PMT or APD detectors would incur in developing high-frame rate systems with a similar dynamic range. The tradeoff is the lack of the ability to accurately recover exact absorption and scattering coefficients independently from one another, because there is no clear ability to estimate photon traveled pathlength in tissue. However the application of the system has focused on vascular pulsatile flow imaging, and in this case the scatter coefficient remains largely static. Thus the errors in assumed scattering coefficient will mostly contribute to an overall offset of recovered absorption coefficient. Over the years, changes in absorption have been repeatedly been shown to be quantifiable with CW systems, such that this design was chosen for this application. The errors in absolute absorption from the unknown value of the scattering coefficient are a problem, but the error incurred is low if we have a reasonable value of scattering coefficient to use in the model fitting. If the estimated scattering coefficient is within 10% of the true value, then our error in absorption is less than this level. Still, changes in absorption are always recoverable at some level, even if the scatter signal is inaccurate by 20%, so the system does have highly robust response for dynamic imaging applications.

The use of spectrally-encoded sources in the tomography system has the implicit assumption that they are treated as interacting with the same tissue optical properties, even though they are at slightly different wavelengths. Strictly speaking, their wavelengths are different, but if treated as a multi-wavelength tomography system, and if we reconstructed the wavelengths separately, we would be able to potentially recover more information about the spectral features of the tissue. But this approach would have obviously fewer measurements at each individual wavelength, and would make a multispectral reconstruction possible, but highly ill-posed at any given wavelength. When assuming most of the absorption is due to whole blood, it can be seen that its absorption coefficient varies only by about 5% across the used wavelength range. Thus, by only incurring 5% error, we assume the attenuation is the same at all sources, and thereby solve the image reconstruction problem with a better posed inversion problem, and avoid the need to get into spectral reconstruction. However the idea of doing multispectral reconstruction at video rate is something to consider in future research in this area.

The dynamic blood diffusion phantom experiments illustrated the ability of the system to image heterogeneities through 64 mm of tissue-like media at high speed. The direct reconstruction method worked well, and so did difference reconstruction methods, since the absorption content in the inclusion changed much during the process. In the pulsatile phantom experiments, the system was used to illustrate the problem of accurately detecting very subtle periodic variations, on the order of 1% intensity signal change. The volume change in the phantom was approximately 0.5%, which resulted in the observed transmission signal change by 1%. The absorption properties of the inclusion, could not be usefully recovered with diffuse tomography imaging, as might be expected for a system where the signal change is only 1%, and the noise level of the system is near 1%. Thus, in a realistic tissue imaging situation, where SNR is close to unity, the application of diffuse tomography alone will be limited without improvements in signal amplification somehow. However, image-guided recovery of the absorption, using a region-fitting based method, did recover useful signal changes, and the amplitude of the periodic signal was found with a factor of 4 between the amplitude peak and the background noise at other frequencies (Fig. 7(d)). Image-based recovery of broad regions turns the recovery problem from an ill-posed under-conditioned situation, to a well-posed over-conditioned inversion problem. Thus in the future, this type of imaging will likely be utilized in the context of MRI-guided hemodynamic imaging.

Results of both direct and difference reconstruction methods were listed and compared. The approach of using difference-data based reconstruction generates images with fewer artifacts, especially at boundaries, but the ROI profile biases a little towards higher sensitivity region, and this is more of a problem towards the closest boundary. Besides, difference reconstruction doesn’t perform well when the difference of nearby images is subtle. So for applications like hemodynamics monitoring, direct reconstruction, maybe image-guided by other modalities, is more suitable.

The slab geometry was adopted here to decrease the differences of optical pathlengths of source-detector pairs, since dynamic range is such a problem in diffuse tomography. To have a very high imaging rate, all sources and detectors have to be kept on during the whole imaging process, and so the system needs to have a reasonably small dynamic range between individual source signals to make the signals useful. The slab geometry has a much smaller dynamic range than the circular geometry, and solves some of this problem. For tissue that is compressed into the slab shape of say 6–7 cm thickness, the 8 sources are placed at one side with 8 mm interval, and 8 detectors are oppositely placed in the similar way. Then the maximum path length difference is near 2 cm, which is much smaller than the circular geometry design, where differences in pathlengths can be on the same order as the diameter of the tissue. Thus, the slab approach solves this seemingly trivial but yet quite important issue of maintaining the dynamic ranges within a few orders of magnitude from each other between signals.

Real tissue imaging applications of this system are ongoing, using MRI imaging to visualize organs, and then rapid imaging with the spectrally encoded system to recover hemoglobin changes as well as exogenous dye uptake in the organs being tracked. The results of this study show that temporal changes in signal intensity at the level of 1% are still reliable enough to provide useful data for this image-guided tomographic approach. These results then allow estimation of how concentration changes of hemoglobin or dye per unit volume would translate into detectable levels of transmitted signal. The system testing in rodent tissue imaging is ongoing.

Acknowledgements

Funding was provided by research grant R33CA100984, PO1CA80139 and R01EB007966. The authors would like to thank Prof. Ryan Halter for help with the pulsatile phantom.

References and links

1.

C. H. Schmitz, M. Löcker, J. M. Lasker, A. H. Hielscher, and R. L. Barbour, “Instrumentation for fast functional optical tomography,” Rev. Sci. Instrum. 73(2), 429 (2002). [CrossRef]

2.

S. Fantini, M. A. Franceschini, G. Gaida, E. Gratton, H. Jess, W. W. Mantulin, K. T. Moesta, P. M. Schlag, and M. Kaschke, “Frequency-domain optical mammography: edge effect corrections,” Med. Phys. 23(1), 149–157 (1996). [CrossRef]

3.

A. M. Siegel, J. J. A. Marota, and D. A. Boas, “Design and evaluation of a continuous-wave diffuse optical tomography system,” Opt. Express 4(8), 287–298 (1999).

4.

D. Piao, H. Dehghani, S. Jiang, S. Srinivasan, and B. W. Pogue, “Instrumentation for video-rate near-infrared diffuse optical tomography,” Rev. Sci. Instrum. 76(12), 124301 (2005). [CrossRef]

5.

Z. Li, V. Krishnaswamy, K. D. Paulsen, and B. W. Pogue, “Video-rate near infrared tomography for imaging thick tissue with dynamically varying absorption properties,” Proc. SPIE 7174, 71742G (2009). [CrossRef]

6.

S. C. Davis, B. W. Pogue, R. Springett, C. Leussler, P. Mazurkewitz, S. B. Tuttle, S. L. Gibbs-Strauss, S. S. Jiang, H. Dehghani, and K. D. Paulsen, “Magnetic resonance-coupled fluorescence tomography scanner for molecular imaging of tissue,” Rev. Sci. Instrum. 79(6), 064302 (2008). [CrossRef]

7.

S. Jiang, B. W. Pogue, S. Srinivasan, S. Soho, S. P. Poplack, T. D. Tosteson, and K. D. Paulsen, “Assessment of the menstrual cycle upon total hemoglobin, water concentration and oxygen saturation in the female breast,” Proc. SPIE 4955, 342–348 (2003). [CrossRef]

8.

H. Dehghani, B. W. Pogue, S. P. Poplack, and K. D. Paulsen, “Multiwavelength three-dimensional near-infrared tomography of the breast: initial simulation, phantom, and clinical results,” Appl. Opt. 42(1), 135–145 (2003). [CrossRef]

9.

T. O. McBride, “Spectroscopic Reconstructed Near Infrared Tomographic Imaging for Breast Cancer Diagnosis,” (Dartmouth College, 2001). http://nir.thayer.dartmouth.edu/Downloads.html

10.

B. Brooksby, H. Dehghani, B. W. Pogue, and K. D. Paulsen, “Near infrared (NIR) tomography breast image reconstruction with a priori structural information from MRI: algorithm development for reconstructing heterogeneities,” IEEE J. STQE 9, 199–209 (2003).

OCIS Codes
(170.0110) Medical optics and biotechnology : Imaging systems
(170.3010) Medical optics and biotechnology : Image reconstruction techniques
(170.3880) Medical optics and biotechnology : Medical and biological imaging
(170.6960) Medical optics and biotechnology : Tomography
(170.7160) Medical optics and biotechnology : Ultrafast technology

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: March 18, 2009
Revised Manuscript: May 19, 2009
Manuscript Accepted: May 20, 2009
Published: July 2, 2009

Virtual Issues
Vol. 4, Iss. 9 Virtual Journal for Biomedical Optics

Citation
Zhiqiu Li, Venkataramanan Krishnaswamy, Scott C. Davis, Subhadra Srinivasan, Keith D. Paulsen, and Brian W. Pogue, "Video-rate near infrared tomography to image pulsatile absorption properties in thick tissue," Opt. Express 17, 12043-12056 (2009)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-17-14-12043


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. C. H. Schmitz, M. Löcker, J. M. Lasker, A. H. Hielscher, and R. L. Barbour, “Instrumentation for fast functional optical tomography,” Rev. Sci. Instrum. 73(2), 429 (2002). [CrossRef]
  2. S. Fantini, M. A. Franceschini, G. Gaida, E. Gratton, H. Jess, W. W. Mantulin, K. T. Moesta, P. M. Schlag, and M. Kaschke, “Frequency-domain optical mammography: edge effect corrections,” Med. Phys. 23(1), 149–157 (1996). [CrossRef]
  3. A. M. Siegel, J. J. A. Marota, and D. A. Boas, “Design and evaluation of a continuous-wave diffuse optical tomography system,” Opt. Express 4(8), 287–298 (1999).
  4. D. Piao, H. Dehghani, S. Jiang, S. Srinivasan, and B. W. Pogue, “Instrumentation for video-rate near-infrared diffuse optical tomography,” Rev. Sci. Instrum. 76(12), 124301 (2005). [CrossRef]
  5. Z. Li, V. Krishnaswamy, K. D. Paulsen, and B. W. Pogue, “Video-rate near infrared tomography for imaging thick tissue with dynamically varying absorption properties,” Proc. SPIE 7174, 71742G (2009). [CrossRef]
  6. S. C. Davis, B. W. Pogue, R. Springett, C. Leussler, P. Mazurkewitz, S. B. Tuttle, S. L. Gibbs-Strauss, S. S. Jiang, H. Dehghani, and K. D. Paulsen, “Magnetic resonance-coupled fluorescence tomography scanner for molecular imaging of tissue,” Rev. Sci. Instrum. 79(6), 064302 (2008). [CrossRef]
  7. S. Jiang, B. W. Pogue, S. Srinivasan, S. Soho, S. P. Poplack, T. D. Tosteson, and K. D. Paulsen, “Assessment of the menstrual cycle upon total hemoglobin, water concentration and oxygen saturation in the female breast,” Proc. SPIE 4955, 342–348 (2003). [CrossRef]
  8. H. Dehghani, B. W. Pogue, S. P. Poplack, and K. D. Paulsen, “Multiwavelength three-dimensional near-infrared tomography of the breast: initial simulation, phantom, and clinical results,” Appl. Opt. 42(1), 135–145 (2003). [CrossRef]
  9. T. O. McBride, “Spectroscopic Reconstructed Near Infrared Tomographic Imaging for Breast Cancer Diagnosis,” (Dartmouth College, 2001). http://nir.thayer.dartmouth.edu/Downloads.html
  10. B. Brooksby, H. Dehghani, B. W. Pogue, and K. D. Paulsen, “Near infrared (NIR) tomography breast image reconstruction with a priori structural information from MRI: algorithm development for reconstructing heterogeneities,” IEEE J. STQE 9, 199–209 (2003).

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited