Analyzing absorption and scattering spectra of micro-scale structures with spectroscopic optical coherence tomography

Ji Yi, Jianmin Gong, and Xu Li  »View Author Affiliations


Optics Express, Vol. 17, Issue 15, pp. 13157-13167 (2009)
http://dx.doi.org/10.1364/OE.17.013157


View Full Text Article

Acrobat PDF (426 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We demonstrate the feasibility of characterizing the absorption and scattering spectra of micron-scale structures in a turbid medium using a spectroscopic optical coherence tomography (SOCT) system with a bandwidth of 430–650nm. SOCT measurements are taken from phantoms composed of fluorescent microspheres. The absorption and scattering spectra are recovered with proper selections of spatial window width in the post processing step. Furthermore, we present an analysis using numerical OCT simulation based on full-wave solutions of the Maxwell’s Equation to elucidate the origination of the multiple peaks in the OCT image for a single microsphere. Finally, we demonstrate the possibility of identifying contrast agents concentrated in micron-sized scale in an SOCT image. Two different types of microspheres in gel phantom are discriminated in 3D volume based on their distinguished absorbent feature.

© 2009 Optical Society of America

1. Introduction

It has been extensively demonstrated that the spectroscopic features of light measured from biological tissues contain valuable information on cellular morphology and biological functions. Recently, Spectroscopic Optical Coherence Tomography (SOCT)[1

1. U. Morgner, W. Drexler, F. X. Kärtner, X. D. Li, C. Pitris, E. P. Ippen, and J. G. Fujimoto, “Spectroscopic optical coherence tomography,” Opt. Lett. 25, 111–113 (2000). [CrossRef]

] has been developed as an important enhancement of traditional OCT, providing not only a map of reflectivity intensity but also spatially resolved spectroscopic information. As in all spectroscopy-based optical techniques for tissue characterization, the spectral contrast exploited by SOCT originates from the properties of either endogenous tissue or exogenous contrast agents. Given that SOCT is based on coherence detection approach, the mechanisms investigated by SOCT are mostly focused on chemical-dependent absorption and structure-dependent elastic scattering. Endogenous chemicals such as hemoglobin[2

2. D. J. Faber, E.G. Mik, M. C. G. Aalders, and T. G. van Leeuwen, “Light absorption of (oxy-)hemoglobin assessed by spectroscopic optical coherence tomography,” Opt. Lett. 28, 1436–1438 (2003). [CrossRef] [PubMed]

] and melanin[3

3. S. A. Boppart, M. E. Brezinski, C Pitris, and J. G. Fujimoto, “Optical Coherence Tomography for Neurosurgical Imaging of Human Intracortical Melanoma,” Neurosurgery 43, 834–841 (1998). [CrossRef] [PubMed]

] as well as exogenous contrast-enhancing dyes[4

4. Z. Yaqoob, E. Mc Dowell, J. Wu, X. Heng, J. Fingler, and C. Yang, “Molecular contrast optical coherence tomography: a pump-probe scheme using indocyanine green as a contrast agent,” J. Biomed. Opt. 11, 054017 (2006). [CrossRef] [PubMed]

7

7. C. Xu, J. Ye, D. L. Marks, and S. A. Boppart, “Near-infrared dyes as contrast-enhancing agents for spectroscopic optical coherence tomography,” Opt. Lett. 29, 1647–1649(2004). [CrossRef] [PubMed]

] determine the wavelength-dependent absorption profiles, while the size, shape and refractive index of cellular organelles[8, 9] as well as nanoparticle-based exogenous contrast agents[10

10. T. S. Troutman, J. K. Barton, and M. Romanowski, “Optical coherence tomography with plasmon resonant nanorods of gold,” Opt. Lett. 32, 1438–1440 (2007). [CrossRef] [PubMed]

14

14. A. L. Oldenburg, M.N. Hansen, D. A. Zweifel, A. Wei, and S. A. Boppart, “Plasmon-resonant gold nanorods as low backscattering albedo contrast agents for optical coherence tomography,” Opt. Express 14, 6724–6738 (2006). [CrossRef] [PubMed]

] determine the scattering spectral signatures.

Due to the complex nature of biological tissue, it has always been a challenge to definitely associate the measured spectroscopic features with tissue structure and composition in the micro-nano scale. The development in high-resolution OCT system[15

15. W. Drexler, U. Morgner, F. X. Kartner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, “In vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett. 24, 1221–1223 (1999). [CrossRef]

, 16

16. A. Dubois, K. Grieve, G. Moneron, R. Lecaque, L. Vabre, and C. Boccara, “Ultrahigh-Resolution Full-Field Optical Coherence Tomography,” Appl. Opt. 43, 2874–2883 (2004). [CrossRef] [PubMed]

] presents us the unique opportunity to measure spectroscopic signals from single or a small number of tissue scatterers, and has generated great research interests in correlating SCOT measurements with tissue properties in the micro-scale[17

17. C. Xu, D. L. Marks, M N. Do, and S. A. Boppart, “Separation of absorption and scattering profiles in spectroscopic optical coherence tomography using a least-squares algorithm,” Opt. Express 12, 4790–4803 (2004). [CrossRef] [PubMed]

19

19. T. Dennis, S. D. D., A. Dienstfrey, G. Singh, and P. Rice, “Analyzing quantitative light scattering spectra of phantoms measured with optical coherence tomography,” J. Biomed. Opt. 13, 024004 (2008). [CrossRef] [PubMed]

]. The study presented in this paper focuses on numerical simulation and phantom-based experiments of ultra-wide bandwidth SOCT measurements on dispersed scatterers with both scattering and absorption spectral profiles. These phantom studies are analogous to scenarios of imaging single cells or cellular organelles with endogenous or exogenous absorbing chemicals within turbid biological tissue, such as red-blood cells with hemoglobin or organelles enhanced with exogenous dyes. We demonstrate that it is feasible to locally extract the architecture-dependent scattering spectrum and the fluorophore-determined absorption spectrum from fluorescent microspheres via post-processing of the SOCT measurements. Furthermore, we illustrate the possibility of imaging and identifying particles labeled with dyes in a same phantom with ultra-wide bandwidth SOCT measurements.

2. Methods

2.1 Experimental setup

Our experimental setup adopts a parallel[20

20. R. N. Graf, W. J. Browmn, and A. Wax, “Parallel frequency-domain optical coherence tomography scatter-mode imaging of the hamster cheek pouch using a thermal light source,” Opt. Express 33, 1285–1287 (2008).

] common-path Fourier-domain OCT(FDOCT) with broadband thermal light source as shown in Fig. 1(a). Light from a 100W Xenon lamp (Oriel Apex, Newport) is focused by an achromatic lens L1 on an aperture with 0.8cm in diameter, which is then directly imaged onto the sample with another lens L2 and a 0.1-NA objective lens, creating a full-field illumination area with about 400µm in diameter. The effective numerical aperture of L1 and L2 is estimated as 1.8. The sample is attached to the bottom surface of a 3mm-thick quartz plate. The interference signal that is formed from light reflected by the bottom surface of the plate and light backscattered from the sample is detected by a spectrograph (Acton SP2150i) coupled to a CCD camera (Princeton instruments PIMAX-2 HB512). The bandwidth and the resolution of the spectrograph are 430–650nm and 0.4nm respectively, yielding an axial resolution of 0.8µm in water. The lateral resolution of the imaging system is of about 3µm in water as determined by the objective lens.

The post processing involves inverse Fourier transform followed by short-time Fourier transform (briefly, IFT-STFT method) from FDOCT signal. The specific data collection and processing routine are illustrated in Fig. 1(b)–(d). The spectrograph passes 1D signal in image plane through a 10µm-wide entrance slit and disperses the signal into spectra. The CCD camera takes one shot of a 2D matrix with its x-axis corresponding to a spatial position along the 1D image signal and its y-axis as the spectrum in wavelength, as shown in Fig. 1(c). Before further processing, these spectral measurements are normalized by light source spectrum and resampled for equal k-domain interval. Then IFT is performed to obtain the spatial intensity profile in the depth (z) direction. Thus a B-scan image is formed in the x-z plane of the sample as shown in Fig. 1(d). In order to obtain spectral profile from a specific depth, we performed STFT with a Gaussian-shaped window located on the corresponding z position (Fig. 1(b)). Three sub-Figs. in Fig. 1(b) demonstrate the sequence of the processing by plotting the k-domain A-scan signal, the z-domain intensity distribution after IFT, and the recovered spectra after STFT. The red Gaussian window selects our interest spatial location. Note that the z-domain data are complex numbers before and after the STFT and only the amplitude is plotted as resultant extracted spectrum. The window width in the STFT determines the trade-off between the spatial and spectral resolution, where applying a narrower Gaussian window in the spatial domain results in a wider spectral point spread function (PSF) at each sampling.

Fig. 1. Schematic sketch of the system and data processing procedure. (a) Source image are focused by Lens (L1) on the aperture (AP). Lens (L2) and NA=0.1 objective lens (OL) focuses AP filtered light on sample. The beamsplitter (BS) redirects light collected by OL to spectroscope coupled by spectrograph (SG) and CCD camera. (b–d) procedure of post-processing and image formation.

2.2 Simulation

In order to provide better system characterization and interpretation to the resulted OCT image, we have developed a numerical tool to simulate the formation of OCT images with wide-field imaging configuration. Light scattered from the scatterers in the sample is calculated via solving Maxwell’s Equations. For the case of spherical scatterers illuminated by plane waves, Mie theory provides rigorous calculation on both the amplitude and phase of scattered E-field E i in the far-field. We note that the sample illumination field in our current experimental setup can be considered as a spatially incoherent homogeneous field in paraxial region, since the illumination area on sample is a direct projection of the aperture, which spatially filters the image of light source and passes only a homogeneous portion. In our setup, the full-field illumination area is about 400µm in diameter. Although the plane-wave illumination condition is not strictly satisfied here, we speculate that the scattered signal can be approximated by that generated with a plane wave illumination as long as an individual scatterer is located inside the depth of focus, which is about 55µm in our experiment setup. This speculation is supported by the comparison between the numerical and experimental results as discussed in the following section. Here we employ a Mie-theory based computational code developed by Xu et al [21

21. Y. L. Xu “Electromagnetic scattering by an aggregate of spheresParallel frequency-domain optical coherence tomography scatter-mode imaging of the hamster cheek pouch using a thermal light source,” Appl. Opt. 34, 4573–4588 (1995). [CrossRef] [PubMed]

] to calculated E i(θ,ϕ) of a sphere with specific size and complex refractive index n, where θ and ϕ are the zenith and azimuth scattering angle.

Next, we calculate the E-field distribution at the pupil of the objective lens. This E-field is contributed by two components: the scattered field by the sample and the reflected field from the reference plane. The field distribution calculation is based on the geometric arrangement illustrated in Fig. 2. The origin of the coordinate system is placed at the center of the sphere and the incident wave propagates in the +z direction. The objective lens is placed at z=-lc, where lc is the distance between the sphere and the pupil of objective lens. The reference plane is placed at z=-lref. Since it is a far-field solution, the scattered light is regarded as spherical wave spreading from the microsphere as in Fig. 2(a). According to the definition of far-field field quantities, the scattered E-field component at the pupil of the objective lens can be calculated as:

Es(rp)=1krpEi(θ,ϕ)exp(ikrp);
(1)

where k is the wave number, rp is the position vector of a point on the pupil of lens and |rp| is calculated as rp=lc1+tan2(θ).. The reference E-field component E ref(p) can be approximated as a Gaussian beam:

Eref(p)=Econstexp(p2w2(lc))exp(iklc+ikp22R(lc))exp(ik·2lref),
(2)

which is reflected by the reference arm. Here, p is the radial vector in the pupil plane transverse to optical axis as shown in Fig. 2(b), w(lc) is the beam 1/e radius and R(lc) is the curvature radius of the wavefront at the pupil. The distance from reference plane to the focus (origin) is much smaller than the focal length so that we approximate the reflected Gaussian beam backward follows illumination profile with only a phase delay. Given that the initial phase of E i(θ,ϕ) is zero, we add the phase delay in the last exponential term corresponding to optical pathlength difference between reference and the sphere. Econst is complex constant dependent only on the distance between the reference arm and the pupil, which remains unchanged during experiment. The Gaussian profile is designed with beam waist w 0=3µm so that the lateral resolution is kept consistent with experimental setup. The spatial incoherent illumination will just generate stable interference fringe by light reflected from one particular Gaussian beam, since it is completely phase irrelevant with every other Gaussian beams. The total field at the objective pupil can thus be calculated as E(rp)=E ref(p)+E s(rp). In the above formulation, we note that both rp and p indicate the position of a spot within the objective pupil and are functions of θ and ϕ. They are written in different vector definitions for convenience in the formulation.

Fig. 2. Schematic sketch to illustrate the image formation algorithm. (a) Scattering field from microsphere located at origin. E s is the scattered field from the sample when reaching the pupil of objective lens. (b) Reflected reference field. The reference plane is placed between the objective lens and the scatterers. The reference E-field E ref (dots line) is approximated as a reflected Gaussian beam, backward following the illumination Gaussian profile (solid line) with a phase delay.

EI(rq)=02ππΔθ2π+Δθ2E(rp)exp(ikrprqM)dθ.
(3)

Here, rq is the position vector of a point in the image plane. The integration range of the zenith angle Δθ is determined by the numerical aperture (NA) of the objective lens as NA=sin(Δθ/2) and M represents the magnification of whole imaging system, which is set to be 1 through this study. This formulation is similar to the image formation algorithm proposed in [22

22. J. L. Hollmann, A. K. Dunn, and C. A. DiMarzio, “Computational microscopy in embryo imaging,” Opt. Express 29, 2267–2269 (2004).

] for synthesizing regular wide-field microscopy images. Both algorithms use the paraxial approximation, and are thus only valid for imaging systems with low or moderate NA. However, the two algorithms are different in their phase compensation terms within the integral. In our case, we assume that the lens re-adjusts the phase distribution of E(rp) in such a manner that for a point source at any location in the focal plane, its corresponding image in the image plane is formed with complete constructive interference. In other words, our formulation also assumes that the imaging system is free of any type of aberrations.

The above calculation is repeated for a range of wavelength covering the same bandwidth of our experimental system and thus the FDOCT data set E I(rp,λ) is obtained numerically with a specific scattering sphere. A B-scan wavelength-dependent intensity signal can be obtained which is further resampled into k-domain and IFT is taken to ultimately synthesize the image. The same IFT-STFT method is performed to extract and analyze the spectrum. In order to incorporate the absorbent feature of fluorescent sphere, we use the fluorophore absorption spectrum provided by manufacturer to calculate the wavelength-dependent imaginary part of the refractive index n’(λ) in simulation. The absorption spectrum was first normalized to a range between 0 and 1, and further scaled by a constant τ, which is the imaginary part of refractive index at absorption spectrum peak. By comparing the simulation and experimental backscattering spectrum, we empirically set τ equal to 0.02. The linear absorption coefficient µa(λ) can be calculated as µa(λ)=2kn’(λ) and then the molar concentration of fluorophore c also can be calculated as µa(λ)=(λ) if with known molar absorption coefficient ε(λ). We estimated µa=5.6×103 cm-1 at the absorption spectrum peak 450nm for the fluorophore trapped in the 6µm microspheres.

3. Result

Our first experimental phantom is made of fluorescent microspheres embedded in transparent gels. A solution composed of agarose powder (Sigma-aldrich) and water was microwaved until boiling and the solution appears transparent after the powder is completely dissolved. Microspheres solution is added in and stirred when the gel solution cools down at about 45–50 degree C, then the gel phantom is solidified until further cooling to room temperature. Fluorescent microspheres (yellowgreen 6µm nominal diameter, standard deviation 0.146µm, Polyscences Inc.) were experimentally imaged with our FDOCT setup. Since OCT is based on coherent scattering, the fluorescence emission signal will not be present in our interference signal. Therefore, these microspheres can be regarded as scattering structures with an absorbing spectral profile depending on the fluorophore. The particle concentration is controlled to be low enough so that the microspheres are well dispersed and their scattering spectrum can be analyzed individually. The example image of one single microsphere is plotted in Fig. 3(b). In the corresponding simulation, a 6µm sphere is placed 50µm below the reference plane and the OCT image was synthesized for a 2-D longitudinal cross-section intensity image of the 3-D space as I I(x,z:y=0) with the sphere in focus as shown in Fig. 3(a). In these two Figs, the single microsphere is presented as multiple peaks in both experimental and simulated OCT images. The patterns of the multiple peaks are nearly identical, supporting the accuracy of our simulation methodology.

We then perform STFT with different Gaussian windows to recover both fluorophore-dependent absorption and architecture-dependent scattering feature from OCT images. First, we conduct STFT with a window located at the strongest intensity peak, where z=55µm and 50.5µm in Figs. 3(a) and 3(b) respectively. The window size is selected to yield 1.2µm axial resolution, corresponding to a spectral resolution of about 35nm, and the extracted spectra are plotted in Fig. 3(c). Both experimental and simulated data show an absorption band (420–490nm) in recovered spectral profiles, matching with the standard absorption spectrum provided by manufacturer as plotted in the same Fig. Subsequently, a wider window is applied at z=62µm in Fig. 3(a) and 57µm in Fig. 3(b) to cover the second and the third peaks during STFT. The axial and spectral resolutions of the second window are about 6µm and 8nm respectively. Figure 3(d) shows the corresponding extracted oscillating spectra from both experimental and simulated data, and they fit well with backscattering spectrum calculated from a sphere of the same size by Mie theory. This extracted spectrum is dependent on the size of the sphere. Some discrepancy was observed around 500–530nm between Mie backscattering and extracted spectra and this may be caused by the ignorance of the first peak when we applied the larger window.

Fig. 3. Simulated (a) and experimental (b) OCT image of 6µm yellow-green fluorescent microspheres and the extracted spectra by two different spatial Gaussian windows (c–d).

Fig. 4. Simulated en face microscopic (a–c) and OCT images (e–f) of 6µm microsphere for an objective lens with NA=0.1, NA=0.2 and NA=0.3 respectively.

Fig. 5. (a). (Media 1) 3D OCT image of the scattering phantom containing yellowgreen and polystyrene microspheres; volume=125*85*100µm in x-y-z. See the animation of scanning and 360 degree view in Movie1.mpg (3.5M). (b) En face projection images on x-y plane of original 3D OCT, (d) 3D-YG and (e) 3D-POLY data. (c) En face fluorescence image at the same position taken prior to OCT scanning. Bar=20µm.

Figure 6 shows a close look at one x-z B scan OCT image and demonstrated the effectiveness of our algorithm. Figure 6(a) is the original B scan OCT image taken from the yellow line in Fig. 5(b) and we developed a false-color coded map in Fig. 6(b) in HSV color space to identify the positions of two different kinds of microspheres. We encoded Hue with a resulting matrix of subtracting yellowgreen coefficient map from polystyrene one and encoded Saturation by the result of summing two coefficient maps. And the Value is encoded by the original OCT intensity map as Fig. 6(a). The color-coded image shows that there is one yellowgreen sphere at about x=140µm and three polystyrene ones at around x=105µm, 125µm and 150µm. The discrimination of yellowgreen microsphere can be verified by the fluorescence image in Fig. 5(c). Thanks to our color-coded Figs, we gathered more information based on their spectral behavior and managed to differentiate spheres with their characteristic absorbing feature.

Fig. 6. (a). Intensity based OCT image of the scattering phantom containing fluorescent and non-absorbing microsphere; (b). False-color coded SOCT image for discrimination; Gray scale colorbar is designated for the dB intensity and Hue colorbar encodes the result of subtracting YG coefficient map from polystyrene one.

4. Discussion and conclusion

Biological tissue scatterers are usually heterogeneous, nonspherical, and organized in a complex manner. This work represents a starting point for understanding SOCT data measured from localized scattering and absorbing structures. It demonstrates that in order to interpret SOCT measurements in details for even relatively simple scatterers, a thorough understanding of the full-wave optical response of the scatterer is required. Numerical techniques that solve the Maxwell’s equation for arbitrary geometry are useful for investing and understanding OCT images of a more complex structure.

Several algorithms have been developed to perform time-frequency analysis of SOCT data [1

1. U. Morgner, W. Drexler, F. X. Kärtner, X. D. Li, C. Pitris, E. P. Ippen, and J. G. Fujimoto, “Spectroscopic optical coherence tomography,” Opt. Lett. 25, 111–113 (2000). [CrossRef]

, 7

7. C. Xu, J. Ye, D. L. Marks, and S. A. Boppart, “Near-infrared dyes as contrast-enhancing agents for spectroscopic optical coherence tomography,” Opt. Lett. 29, 1647–1649(2004). [CrossRef] [PubMed]

, 9

9. R. N. Graf and A. Wax, “Nuclear morphology measurements using Fourier domain low coherence interferometry,” Opt. Express 13, 4693–4698 (2005). [CrossRef]

, 25

25. R. Leitgeb, M. Wojtkowski, A. Kowalczyk, C. K. Hitzenberger, M. Sticker, and A. F. Fercher, “Spectral measurement of absorption by spectroscopic frequency-domain optical coherence tomography,” Opt. Lett. 25, 820–822 (2000). [CrossRef]

28

28. R. N. Graf and A. Wax, “Temporal coherence and time-frequency distributions in spectroscopic optical coherence tomography,” J. Opt. Soc. Am. A , 24, 2186–2195 (2007). [CrossRef]

]. This work adopts a straight-forward IFT-STFT method to extract spectra from a specified spatial window. Our investigation shows that in order to extract absorption spectral profile from localized structures, it is important to use a small window size in the IFT-STFT algorithm. Spectra selected by a bigger-sized window are very likely corrupted by complex scattering spectra, originating not only from the structural properties of the scatterers themselves, but also from coherent multiple scattering effects which create speckle in image and high frequency disturbance on recovered spectra.

In conclusion, we have demonstrated the feasibility of characterizing the absorption and scattering spectra of micro-scale structures in a turbid medium using a SOCT system. We have also presented an analysis using numerical OCT simulations based on full-wave solutions of the Maxwell’s Equations to elucidate the origin of the multiple peaks in the OCT image for a single microsphere. Finally we have demonstrated the possibility of identifying contrast agents concentrated in micron-sized scale in a 3D SOCT image.

Acknowledgements

The authors would like to thank Dr. Xingde Li for his advices on improving our system performance.

References and links

1.

U. Morgner, W. Drexler, F. X. Kärtner, X. D. Li, C. Pitris, E. P. Ippen, and J. G. Fujimoto, “Spectroscopic optical coherence tomography,” Opt. Lett. 25, 111–113 (2000). [CrossRef]

2.

D. J. Faber, E.G. Mik, M. C. G. Aalders, and T. G. van Leeuwen, “Light absorption of (oxy-)hemoglobin assessed by spectroscopic optical coherence tomography,” Opt. Lett. 28, 1436–1438 (2003). [CrossRef] [PubMed]

3.

S. A. Boppart, M. E. Brezinski, C Pitris, and J. G. Fujimoto, “Optical Coherence Tomography for Neurosurgical Imaging of Human Intracortical Melanoma,” Neurosurgery 43, 834–841 (1998). [CrossRef] [PubMed]

4.

Z. Yaqoob, E. Mc Dowell, J. Wu, X. Heng, J. Fingler, and C. Yang, “Molecular contrast optical coherence tomography: a pump-probe scheme using indocyanine green as a contrast agent,” J. Biomed. Opt. 11, 054017 (2006). [CrossRef] [PubMed]

5.

K. D. Rao, M. A. Choma, S. Yazdanfar, A. M. Rollins, and J. A. Izatt, “Molecular contrast in optical coherence tomography by use of a pump-probe technique,” Opt. Lett. 28, 340–342 (2003). [CrossRef] [PubMed]

6.

C. Yang, L. E. L. McGuckin, J. D. Simon, M. A. Choma, B. E. Applegate, and J. A. Izatt, “Spectral triangulation molecular contrast optical coherence tomography with indocyanine green as the contrast agent,” Opt. Lett. 29, 2016–2018 (2004). [CrossRef] [PubMed]

7.

C. Xu, J. Ye, D. L. Marks, and S. A. Boppart, “Near-infrared dyes as contrast-enhancing agents for spectroscopic optical coherence tomography,” Opt. Lett. 29, 1647–1649(2004). [CrossRef] [PubMed]

8.

D. C. Adler, T. H. Ko, P. R. Herz, and J. G. Fujimoto, “Optical coherence tomography contrast enhancement using spectroscopic analysis with spectral autocorrelation,” Opt. Express 12, 5487–5501 (2004). [CrossRef] [PubMed]

9.

R. N. Graf and A. Wax, “Nuclear morphology measurements using Fourier domain low coherence interferometry,” Opt. Express 13, 4693–4698 (2005). [CrossRef]

10.

T. S. Troutman, J. K. Barton, and M. Romanowski, “Optical coherence tomography with plasmon resonant nanorods of gold,” Opt. Lett. 32, 1438–1440 (2007). [CrossRef] [PubMed]

11.

J. Chen, F. Saeki, B. J. Wiley, Hu Cang, M. J. Cobb, Z.-Y. Li, L. Au, H. Zhang, M. B. Kimmey, X. Li, and Y. Xia, “Gold Nanocages: Bioconjugation and Their Potential Use as Optical Imaging Contrast Agents,” Nano Lett. 5, 473–477 (2005). [CrossRef] [PubMed]

12.

J. K. Barton, N. J. Halas, J. L. West, and R. A. Drezek, “Nanoshells as an Optical Coherence Tomography Contrast Agent,” Proc. SPIE, 5316, (2004). [CrossRef]

13.

A. M. Gobin, M. H. Lee, N. J. Halas, W. D. James, R. A. Drezek, and J. L. West, “Near-Infrared Resonant Nanoshells for Combined Optical Imaging and Photothermal Cancer Therapy,” Nano Lett. 7, 1929–1934 (2007). [CrossRef] [PubMed]

14.

A. L. Oldenburg, M.N. Hansen, D. A. Zweifel, A. Wei, and S. A. Boppart, “Plasmon-resonant gold nanorods as low backscattering albedo contrast agents for optical coherence tomography,” Opt. Express 14, 6724–6738 (2006). [CrossRef] [PubMed]

15.

W. Drexler, U. Morgner, F. X. Kartner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, “In vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett. 24, 1221–1223 (1999). [CrossRef]

16.

A. Dubois, K. Grieve, G. Moneron, R. Lecaque, L. Vabre, and C. Boccara, “Ultrahigh-Resolution Full-Field Optical Coherence Tomography,” Appl. Opt. 43, 2874–2883 (2004). [CrossRef] [PubMed]

17.

C. Xu, D. L. Marks, M N. Do, and S. A. Boppart, “Separation of absorption and scattering profiles in spectroscopic optical coherence tomography using a least-squares algorithm,” Opt. Express 12, 4790–4803 (2004). [CrossRef] [PubMed]

18.

C. Xu, P. S. Carney, and S. A. Boppart, “Wavelength-dependent scattering in spectroscopic optical coherence tomography,” Opt. Express 13, 5450–5462(2005). [CrossRef] [PubMed]

19.

T. Dennis, S. D. D., A. Dienstfrey, G. Singh, and P. Rice, “Analyzing quantitative light scattering spectra of phantoms measured with optical coherence tomography,” J. Biomed. Opt. 13, 024004 (2008). [CrossRef] [PubMed]

20.

R. N. Graf, W. J. Browmn, and A. Wax, “Parallel frequency-domain optical coherence tomography scatter-mode imaging of the hamster cheek pouch using a thermal light source,” Opt. Express 33, 1285–1287 (2008).

21.

Y. L. Xu “Electromagnetic scattering by an aggregate of spheresParallel frequency-domain optical coherence tomography scatter-mode imaging of the hamster cheek pouch using a thermal light source,” Appl. Opt. 34, 4573–4588 (1995). [CrossRef] [PubMed]

22.

J. L. Hollmann, A. K. Dunn, and C. A. DiMarzio, “Computational microscopy in embryo imaging,” Opt. Express 29, 2267–2269 (2004).

23.

Y. Liu, X. Li, Y. L. Kim, and V. Backman, “Elastic backscattering spectroscopic microscopy,” Opt. Lett. 30, 2445–2447 (2005). [CrossRef] [PubMed]

24.

R. Cubeddu, A. Pifferi, P. Taroni, A. Torricelli, and G. Valentini, “A solid tissue phantom for photon migration studies,” Phys. Med. Biol. 42, 1971–1979 (1997). [CrossRef] [PubMed]

25.

R. Leitgeb, M. Wojtkowski, A. Kowalczyk, C. K. Hitzenberger, M. Sticker, and A. F. Fercher, “Spectral measurement of absorption by spectroscopic frequency-domain optical coherence tomography,” Opt. Lett. 25, 820–822 (2000). [CrossRef]

26.

C. Xu, F. Kamalabadi, and S. A. Boppart, “Comparative performance analysis of time-frequency distributions for spectroscopic optical coherence tomography,” Appl. Opt. 44, 1813–1822 (2005). [CrossRef] [PubMed]

27.

A. Wax, C. Yang, and J. A. Izatt, “Fourier-domain low-coherence interferometry for light-scattering spectroscopy,” Opt. Lett. 28, 1230–1232 (2003). [CrossRef] [PubMed]

28.

R. N. Graf and A. Wax, “Temporal coherence and time-frequency distributions in spectroscopic optical coherence tomography,” J. Opt. Soc. Am. A , 24, 2186–2195 (2007). [CrossRef]

OCIS Codes
(170.4500) Medical optics and biotechnology : Optical coherence tomography
(290.1350) Scattering : Backscattering
(300.1030) Spectroscopy : Absorption

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: April 13, 2009
Revised Manuscript: June 22, 2009
Manuscript Accepted: July 10, 2009
Published: July 17, 2009

Virtual Issues
Vol. 4, Iss. 9 Virtual Journal for Biomedical Optics

Citation
Ji Yi, Jianmin Gong, and Xu Li, "Analyzing absorption and scattering spectra of micro-scale structures with spectroscopic optical coherence tomography," Opt. Express 17, 13157-13167 (2009)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-17-15-13157


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. U. Morgner, W. Drexler, F. X. Kärtner, X. D. Li, C. Pitris, E. P. Ippen, and J. G. Fujimoto, "Spectroscopic optical coherence tomography," Opt. Lett. 25,111-113 (2000). [CrossRef]
  2. D. J. Faber, E.G. Mik, M. C. G. Aalders, and T. G. van Leeuwen, "Light absorption of (oxy-)hemoglobin assessed by spectroscopic optical coherence tomography," Opt. Lett. 28, 1436-1438 (2003). [CrossRef] [PubMed]
  3. S. A. Boppart, M. E. Brezinski, C Pitris, and J. G. Fujimoto, "Optical Coherence Tomography for Neurosurgical Imaging of Human Intracortical Melanoma," Neurosurgery 43, 834-841 (1998). [CrossRef] [PubMed]
  4. Z. Yaqoob, E. Mc Dowell, J. Wu, X. Heng, J. Fingler, and C. Yang, "Molecular contrast optical coherence tomography: a pump-probe scheme using indocyanine green as a contrast agent," J. Biomed. Opt. 11, 054017 (2006). [CrossRef] [PubMed]
  5. K. D. Rao, M. A. Choma, S. Yazdanfar, A. M. Rollins, and J. A. Izatt, "Molecular contrast in optical coherence tomography by use of a pump-probe technique," Opt. Lett. 28, 340-342 (2003). [CrossRef] [PubMed]
  6. C. Yang, L. E. L. McGuckin., J. D. Simon, M. A. Choma, B. E. Applegate, and J. A. Izatt, "Spectral triangulation molecular contrast optical coherence tomography with indocyanine green as the contrast agent," Opt. Lett. 29, 2016-2018 (2004). [CrossRef] [PubMed]
  7. C. Xu, J. Ye, D. L. Marks, and S. A. Boppart, "Near-infrared dyes as contrast-enhancing agents for spectroscopic optical coherence tomography," Opt. Lett. 29, 1647-1649 (2004). [CrossRef] [PubMed]
  8. D. C. Adler, T. H. Ko, P. R. Herz, and J. G. Fujimoto, "Optical coherence tomography contrast enhancement using spectroscopic analysis with spectral autocorrelation," Opt. Express 12, 5487-5501 (2004). [CrossRef] [PubMed]
  9. R. N. Graf and A. Wax, "Nuclear morphology measurements using Fourier domain low coherence interferometry," Opt. Express 13, 4693-4698 (2005). [CrossRef]
  10. T. S. Troutman, J. K. Barton, and M. Romanowski, "Optical coherence tomography with plasmon resonant nanorods of gold," Opt. Lett. 32, 1438-1440 (2007). [CrossRef] [PubMed]
  11. J. Chen, F. Saeki, B. J. Wiley, Hu Cang, M. J. Cobb, Z.-Y. Li, L. Au, H. Zhang, M. B. Kimmey, X. Li, and Y. Xia, "Gold Nanocages: Bioconjugation and Their Potential Use as Optical Imaging Contrast Agents," Nano Lett. 5, 473-477 (2005). [CrossRef] [PubMed]
  12. J. K. Barton, N. J. Halas, J. L. West, and R. A. Drezek, "Nanoshells as an Optical Coherence Tomography Contrast Agent," Proc. SPIE, 5316, (2004). [CrossRef]
  13. A. M. Gobin, M. H. Lee, N. J. Halas, W. D. James, R. A. Drezek, and J. L. West, "Near-Infrared Resonant Nanoshells for Combined Optical Imaging and Photothermal Cancer Therapy," Nano Lett. 7, 1929-1934 (2007). [CrossRef] [PubMed]
  14. A. L. Oldenburg, M.N. Hansen, D. A. Zweifel, A. Wei, and S. A. Boppart, "Plasmon-resonant gold nanorods as low backscattering albedo contrast agents for optical coherence tomography," Opt. Express 14, 6724-6738 (2006). [CrossRef] [PubMed]
  15. W. Drexler, U. Morgner, F. X. Kartner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, "In vivo ultrahigh-resolution optical coherence tomography," Opt. Lett. 24, 1221-1223 (1999). [CrossRef]
  16. A. Dubois, K. Grieve, G. Moneron, R. Lecaque, L. Vabre, and C. Boccara, "Ultrahigh-Resolution Full-Field Optical Coherence Tomography," Appl. Opt. 43, 2874-2883 (2004). [CrossRef] [PubMed]
  17. C. Xu, D. L. Marks, M N. Do, and S. A. Boppart, "Separation of absorption and scattering profiles in spectroscopic optical coherence tomography using a least-squares algorithm," Opt. Express 12, 4790-4803 (2004). [CrossRef] [PubMed]
  18. C. Xu, P. S. Carney, and S. A. Boppart, "Wavelength-dependent scattering in spectroscopic optical coherence tomography," Opt. Express 13, 5450-5462 (2005). [CrossRef] [PubMed]
  19. T. Dennis, S. D. D., A. Dienstfrey, G. Singh and P. Rice, "Analyzing quantitative light scattering spectra of phantoms measured with optical coherence tomography," J. Biomed. Opt. 13, 024004 (2008). [CrossRef] [PubMed]
  20. R. N. Graf, W. J. Browmn, and A. Wax, "Parallel frequency-domain optical coherence tomography scatter-mode imaging of the hamster cheek pouch using a thermal light source," Opt. Express 33, 1285-1287 (2008).
  21. Y. L. Xu "Electromagnetic scattering by an aggregate of spheresParallel frequency-domain optical coherence tomography scatter-mode imaging of the hamster cheek pouch using a thermal light source," Appl. Opt. 34, 4573-4588 (1995). [CrossRef] [PubMed]
  22. J. L. Hollmann, A. K. Dunn, and C. A. DiMarzio, "Computational microscopy in embryo imaging," Opt. Express 29, 2267-2269 (2004).
  23. Y. Liu, X. Li, Y. L. Kim, and V. Backman, "Elastic backscattering spectroscopic microscopy," Opt. Lett. 30, 2445-2447 (2005). [CrossRef] [PubMed]
  24. R. Cubeddu, A. Pifferi, P. Taroni, A. Torricelli, and G. Valentini, "A solid tissue phantom for photon migration studies," Phys. Med. Biol. 42, 1971-1979 (1997). [CrossRef] [PubMed]
  25. R. Leitgeb, M. Wojtkowski, A. Kowalczyk, C. K. Hitzenberger, M. Sticker, and A. F. Fercher, "Spectral measurement of absorption by spectroscopic frequency-domain optical coherence tomography," Opt. Lett. 25, 820-822 (2000). [CrossRef]
  26. C. Xu, F. Kamalabadi, and S. A. Boppart, "Comparative performance analysis of time-frequency distributions for spectroscopic optical coherence tomography," Appl. Opt. 44, 1813-1822 (2005). [CrossRef] [PubMed]
  27. A. Wax, C. Yang, and J. A. Izatt, "Fourier-domain low-coherence interferometry for light-scattering spectroscopy," Opt. Lett. 28, 1230-1232 (2003). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Supplementary Material


» Media 1: MPG (3554 KB)     

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited