Experimental validation of an optimized signal processing method to handle non-linearity in swept-source optical coherence tomography

Sébastien Vergnole, Daniel Lévesque, and Guy Lamouche  »View Author Affiliations


Optics Express, Vol. 18, Issue 10, pp. 10446-10461 (2010)
http://dx.doi.org/10.1364/OE.18.010446


View Full Text Article

Acrobat PDF (1828 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We evaluate various signal processing methods to handle the non-linearity in wavenumber space exhibited by most laser sources for swept-source optical coherence tomography. The following methods are compared for the same set of experimental data: non-uniform discrete Fourier transforms with Vandermonde matrix or with Lomb periodogram, resampling with linear interpolation or spline interpolation prior to fast-Fourier transform (FFT), and resampling with convolution prior to FFT. By selecting an optimized Kaiser-Bessel window to perform the convolution, we show that convolution followed by FFT is the most efficient method. It allows small fractional oversampling factor between 1 and 2, thus a minimal computational time, while retaining an excellent image quality.

© 2010 OSA

1. Introduction

In Fourier-domain optical coherence tomography (FD-OCT) [1

1. A. F. Fercher, C. K. Hitzenberger, G. Kamp, and S. Y. El Zaiat, “Measurement of intraocular distances by backscattering spectral interferometry,” Opt. Commun. 117(1–2), 43–48 (1995). [CrossRef]

, 2

2. G. Häusler and M. W. Linduer, ““Coherence radar” and “spectral radar”-new tools for dermatological diagnosis,” J. Biomed. Opt. 3(1), 21–31 (1998). [CrossRef]

], the signal is collected as a function of the wavelength and the spatial information is recovered by a discrete Fourier transform (DFT). To maximize the image quality, the basic DFT requires to be performed with equally-spaced data in wavenumber space (k-space). Unfortunately, these digitized data are not commonly recorded with such a regular spacing in wavenumber. Various approaches have been developed to overcome this issue in the two branches of FD-OCT. These approaches involve hardware modifications: linear-wavenumber spectrometers have been developed [3–5

3. Z. Hu and A. M. Rollins, “Fourier domain optical coherence tomography with a linear-in-wavenumber spectrometer,” Opt. Lett. 32(24), 3525–3527 (2007). [CrossRef] [PubMed]

] for spectral domain OCT (SD-OCT) and the feasibility of a linear-wavenumber swept source has been demonstrated for swept-source OCT (SS-OCT) [6

6. C. M. Eigenwillig, B. R. Biedermann, G. Palte, and R. Huber, “K-space linear Fourier domain mode locked laser and applications for optical coherence tomography,” Opt. Express 16(12), 8916–8937 (2008), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-16-12-8916. [CrossRef] [PubMed]

]. Another solution in SS-OCT is to use an external k-trigger [7

7. D. C. Adler, Y. Chen, R. Huber, J. Schmitt, J. Connolly, and J. G. Fujimoto, “Three-dimensional endomicroscopy using optical coherence tomography,” Nat. Photonics 1(12), 709–716 (2007). [CrossRef]

, 8

8. M. A. Choma, K. Hsu, and J. A. Izatt, “Swept source optical coherence tomography using an all-fiber 1300-nm ring laser source,” J. Biomed. Opt. 10(4), 044009 (2005). [CrossRef]

]. This feature is available as an option on some commercial swept sources but it increases the total cost of the setup and the complexity of the data acquisition system.

Very recently, non-uniform DFT algorithms have been used [32–35

32. S. S. Sherif, C. Flueraru, Y. Mao, and S. Change, “Swept Source Optical Coherence Tomography with Nonuniform Frequency Domain Sampling,” in Biomedical Optics, OSA Technical Digest (CD) (Optical Society of America, 2008), paper BMD86.

]. The most direct approach is the method which relies on the Vandermonde matrix first proposed in OCT by Sherif et al. [32

32. S. S. Sherif, C. Flueraru, Y. Mao, and S. Change, “Swept Source Optical Coherence Tomography with Nonuniform Frequency Domain Sampling,” in Biomedical Optics, OSA Technical Digest (CD) (Optical Society of America, 2008), paper BMD86.

]. A similar method called Schuster periodogram has been proposed in FD-OCT [33

33. T. H. Chow, S. G. Razul, B. K. Ng, G. Ho, and C. B. A. Yeo, “Enhancement of Fourier domain optical coherence tomorgraphy images using discrete Fourier transform method,” Proc. SPIE 6847, 68472T (2008). [CrossRef]

]. Finally, Hillmann et al. [35

35. D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]

] proposed a method where resampling is done via a convolution of the data with a Kaiser-Bessel window prior to performing a FFT. This method is well known in the magnetic resonance imaging (MRI) community as the gridding method. Reference [35

35. D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]

] suggests that this method is very efficient, although the demonstration is made only on simulated data in absence of noise. A similar convolution approach was proposed by Zhang et al. [36

36. Y. Zhang, X. Li, L. Wei, K. Wang, Z. Ding, and G. Shi, “Time-domain interpolation for Fourier-domain optical coherence tomography,” Opt. Lett. 34(12), 1849–1851 (2009). [CrossRef] [PubMed]

] with a kernel in the k-space equivalent to the zero-filling interpolation described above.

In this paper, we present a comparison of the different methods that can be used in SS-OCT. To provide a true assessment of the performance of the various methods, the comparison is not made with simulated data but with experimental data. Although SS-OCT is used in this work, the results also apply to SD-OCT. We briefly review the selected methods which include non-uniform DFT with Vandermonde matrix or with Lomb periodogram, resampling with linear interpolation or spline interpolation prior to FFT, and resampling with convolution prior to FFT. For the latter, the convolution is performed with a Kaiser-Bessel window as proposed in Ref [35

35. D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]

]. We introduce an improved implementation of the convolution method by using an optimized Kaiser-Bessel window that allows achieving good image quality with minimal computational time. Experimental data are then processed using each method to compare computational times and image quality by evaluating the point spread functions. Finally, we show images of a phantom and of a finger nail obtained with the different methods to better appreciate their performance. All these results do confirm that the most efficient method is the convolution with an optimized Kaiser-Bessel window followed by FFT.

2. Signal processing methods

The methods described below require an accurate calibration vector for the swept-laser source used, so that the wavenumber k=2π/λ is a priori known for all time sampling points, N. These methods are non-uniform discrete Fourier transform (DFT), resampling with interpolation followed by FFT, and resampling with convolution followed by FFT. Practical considerations for the implementation of each algorithm are given.

2.1 Non-uniform DFT

The most direct method to the problem is the use of a non-uniform DFT relying on the Vandermonde matrix [32

32. S. S. Sherif, C. Flueraru, Y. Mao, and S. Change, “Swept Source Optical Coherence Tomography with Nonuniform Frequency Domain Sampling,” in Biomedical Optics, OSA Technical Digest (CD) (Optical Society of America, 2008), paper BMD86.

]. The measured OCT signal in k-space Uk and the OCT profile in spatial space un are linked by a Fourier transform that can be written as:

Uk=n=0N1unγkn,γk=ei2πNk
(1)

In the above expression for the classical DFT, the k values are integers (i.e. non-dimensional wavenumber) and the γk are evenly distributed on a unit circle in the complex plane. A nonuniform DFT can be defined for γk non-evenly spaced, i.e. where k values are not integers. Also, this equation can be written in a matrix form as Uk = Dun, where D is a Vandermonde matrix that can be inverted to relate the SS-OCT signal Uk measured with non-uniform sampling in k-space to the OCT profile as un = D -1 Uk. To avoid any numerical instability in computing D -1, one can use the forward transform instead, with summation over k values in the form:

un=kukγkn=DUk
(2)

where the conjugate-transposed matrix D is:

D=[111γ01γ11γN11γ0N1γ1N1γN1N1]
(3)

Rescaling with the k vector in the range kmin<k<kmax, the γ k in discrete form above are now expressed as:

γk=ei(kkmin)δz
(4)

with the step size in optical path length given by:

δz=2π(kmaxkmin)
(5)

The Lomb periodogram is another DFT for processing OCT signals that circumvents the resampling issue. Basically, such a method attributed to Lomb [39

39. N. R. Lomb, “Least-squares frequency analysis of unequally spaced data,” Astrophys. Space Sci. 39(2), 447–462 (1976). [CrossRef]

] was developed for estimating the power spectrum of a time series. The method uses a least-square fitting of a sinusoid to the data for each frequency and has a well-defined statistical behavior. Starting with a zero-mean time series un = u(tn) of N points with variance σ 2, the Lomb periodogram, function of the angular frequency ω > 0, is defined as [40

40. W. H. Press, S. A. Teukolsky, W. T. Vetterling, and B. P. Flannery, Numerical Recipes in FORTRAN (Cambridge University Press, New York, 1992).

]:

U(ω)2=12σ2{[nuncosω(tnτ)]2ncos2ω(tnτ)+[nunsinω(tnτ)]2nsin2ω(tnτ)},
(6)

where τ = τ(ω) is an offset that makes the periodogram invariant to time translation as given by:

tan2ωτ=nsin2ωtnncos2ωtn.
(7)

Such a method can be used to process SS-OCT data in the k-space since inverse or direct Fourier transform is equivalent. As a periodogram however, only the magnitude or the envelope is obtained, that is as a function of the optical path z (related to the variable ω) from unevenly sampled wavenumbers k (related to the variable t). A direct implementation of the Lomb periodogram requires O(N 2) operations. However, one can implement a fast algorithm of complexity O(N log N) for its evaluation to any desired degree of approximation by spreading each measured point onto L most nearly centered mesh points using Lagrange interpolating polynomials [40

40. W. H. Press, S. A. Teukolsky, W. T. Vetterling, and B. P. Flannery, Numerical Recipes in FORTRAN (Cambridge University Press, New York, 1992).

]. The number of points L is at least 2 and has a direct impact on the computational time. The algorithm also uses FFT, but it is not for calculation of the DFT from the OCT data. Further optimization can be made toward a preprocessing step followed by the calculation only required for each OCT signal. The data windowing, e.g. using a Hamming window, is also allowed to reduce spectral leakage taking into account the measured point locations.

2.2 Interpolation with FFT

δk=(kmaxkmin)=2πNδz,δz=2π(kmaxkmin)
(8)

where α is the oversampling factor in the k-space and N’= is the number of points involved in the FFT operation. The value of α is a tradeoff between the accuracy and computational time, and is commonly set to 2. In our implementation, the FFT algorithm used only requires that N’ is the product of small prime numbers, possibly after padding the vector with a few zeroes. Our implementation relies on the algorithm FFTPack [41

41. P. N. Swarztrauber, Vectorizing the FFTs, in Parallel Computations, G. Rodrigue, ed. (Academic Press, New York, NY, 1982), pp. 51–83.

], but a more sophisticated algorithm exists with a speedup factor of about 2 [42

42. M. Frigo and S. G. Johnson, “FFTW: an adaptive software architecture for the FFT,” in Proceedings of IEEE International Conference on Acoustics, Speech and Signal Processing. (Institute of Electrical and Electronics Engineers, Seattle, 1998), pp. 1381–1384.

]. The windowing on resampled data with fixed step size is also allowed to reduce spectral leakage. For computational costs, the overall scheme has a complexity of O(N’ log N’).

Using the basic linear interpolation, the resampling is made on values of k with the relation:

Uk=Uk+wk(Uk+Uk),wk=kkk+k
(9)

where k - and k + are the neighboring points on each side of k, and the wk are calculated once before processing all OCT signals. The result after the FFT operation usually shows a strong decrease with depth that should not be considered the actual sensitivity roll-off. In fact, the weighted average on each resampled point can be seen as a convolution with a triangle function [35

35. D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]

]. The width of this triangle function is nevertheless not constant due to the nonlinearity of the data in k-space. Therefore, a correction can be applied by assuming that the data is convolved with an average window. Such a correction is applied after the FFT operation by multiplying the result with the vector cn defined as:

cn=(/N)2sin2/N)
(10)

Alternatively, the cubic spline interpolation involves U and its second derivative U” also from nearest neighbor points, as given by [40

40. W. H. Press, S. A. Teukolsky, W. T. Vetterling, and B. P. Flannery, Numerical Recipes in FORTRAN (Cambridge University Press, New York, 1992).

]:

Uk=akUk+bkUk++fkU"k+gkU"k+
(11)

where: ak=1wk,bk=wk,fk=16(ak3ak)(k+k)2andgk=16(bk3bk)(k+k)2.

Its implementation requires the calculation of the second derivative of each OCT signal which makes the resampling more time consuming. Also, such a nearest neighbor calculation on each interpolated point requires a correction after the FFT operation which depends on the nonlinearity of the swept-laser source used, which can be determined empirically. This is made once by comparing the amplitude roll-off as a function of optical path with that of the reference method using the Vandermonde matrix, providing a vector cn as a polynomial fit:

cn=jdj(nδz)j
(12)

where the dj’s are the polynomial coefficients.

Another interpolation scheme has been proposed which is known as the zero-filling method. The standard FFT is first applied to the raw OCT data, the result is then filled with zeroes to a length typically 4N or 8N, and the FFT back into the k-space yields a vector with a finer sampling and proper data interpolation. Although this scheme uses only FFT, it requires larger vectors to handle and makes the method much slower than the above two interpolation methods [15

15. Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K.-P. Chan, M. Itoh, and T. Yatagai, “Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments,” Opt. Express 13(26), 10652–10664 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-26-10652. [CrossRef] [PubMed]

]. Thus, the zero-filling method is not further considered in this paper.

2.3 Convolution with FFT

A method, known as gridding in MRI, can be advantageously used for the data processing of the SS-OCT data [43

43. J. I. Jackson, C. H. Meyer, D. G. Nishimura, and A. Macovski, “Selection of a convolution function for Fourier inversion using gridding computerised tomography application,” IEEE Trans. Med. Imaging 10(3), 473–478 (1991). [CrossRef] [PubMed]

]. In this method, OCT data is resampled by convolution in the k-space with a finite window function as the kernel, followed by the use of conventional FFT. Very recently, such a method was proposed in OCT using a function with cosines, equivalent to the zero-filling interpolation above [36

36. Y. Zhang, X. Li, L. Wei, K. Wang, Z. Ding, and G. Shi, “Time-domain interpolation for Fourier-domain optical coherence tomography,” Opt. Lett. 34(12), 1849–1851 (2009). [CrossRef] [PubMed]

], or using a Kaiser-Bessel function [35

35. D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]

]. The approach uses the same oversampling factor α as for the interpolation method. Additional parameters include a window length of M points and a design parameter β to minimize the amplitude of aliasing side lobes. An improvement to what is found in the current OCT literature is to use the Kaiser-Bessel kernel with a non-integer oversampling factor α between 1 and 2. This implies that smaller vectors are used, insuring small computational time while maintaining high accuracy. The MRI literature prescribes optimal β values [44

44. P. J. Beatty, D. G. Nishimura, and J. M. Pauly, “Rapid gridding reconstruction with a minimal oversampling ratio,” IEEE Trans. Med. Imaging 24(6), 799–808 (2005). [CrossRef] [PubMed]

] and we propose to apply the same optimization in SS-OCT signal processing.

Again with the k vector in the range kmin<k<kmax, the step sizes in wavenumber domain δk and optical path domain δz are the same as previously. The convolution of OCT data is made on center values of k using a Kaiser-Bessel window as:

Uk=j=1MUjCkj
(13)
Ckj=I0(β1(2κ/M)2)/M,κ=kkjδkM2
(14)

where I0(·) is the zero-order modified Bessel function of the first kind, and the kj’s are the neighboring points around the center value k. A careful analysis made between the pass-band and stop-band of the convolution kernel yields a closed-form expression for the optimal design parameter β as [44

44. P. J. Beatty, D. G. Nishimura, and J. M. Pauly, “Rapid gridding reconstruction with a minimal oversampling ratio,” IEEE Trans. Med. Imaging 24(6), 799–808 (2005). [CrossRef] [PubMed]

]:

β=πM2α2(α12)20.8
(15)

For one-dimensional data such as in OCT and values of M typically between 3 and 8, the Ckj are calculated once for processing all SS-OCT signals. Again, N’= is the number of points involved in the FFT operation, and our FFT algorithm only requires that N’ is the product of small prime numbers, possibly after padding the vector with a few zeroes. The windowing on convolved data with a fixed step size is also allowed to reduce spectral leakage. The convolution of the data with a Kaiser-Bessel window requires a correction after the FFT operation, multiplying the result with the vector cn given by [44

44. P. J. Beatty, D. G. Nishimura, and J. M. Pauly, “Rapid gridding reconstruction with a minimal oversampling ratio,” IEEE Trans. Med. Imaging 24(6), 799–808 (2005). [CrossRef] [PubMed]

]:

cn=(nπM/N)2β2sin((nπM/N)2β2)
(16)

where complex arithmetic is allowed. For computational costs, the overall scheme has a complexity of O(N’ log N’).

3. Experimental results

3.1. Experimental setup

The SS-OCT system is an in-house system that operates with a commercial swept source. For this work, we use a Santec source operating at a sweep rate of 30 kHz over 110 nm centered at 1.31 μm wavelength. The SS-OCT system is configured as a Mach-Zehnder interferometer and is packaged as a mobile unit. The acquisition board is an AlazarTech ATS9462. The sampling rate used in our experiment is 50 MS/s meaning that each acquired A-scan is composed of 1666 points. The detector includes a balanced detection from Novacam Technologies using a low-noise preamplifier. With these choices of swept source and sampling frequency, a depth range of 6.5 mm is available for the evaluation of the performance of the various methods. The actual setup is presented in Fig. 1.

Fig. 1. SS-OCT setup. Cp1 and Cp2: couplers, PC: polarization controller, Cir1 and Cir2: circulators and D: balanced detection.

3.2. Comparison of the processing methods

The performance of each signal processing method is evaluated from the point-spread functions (PSFs) obtained with a mirror at various depth locations in the sample arm. The parameters of interest are the amplitude of a PSF, the noise level, and the computational time. The amplitude of the PSF is expected to decrease when the mirror is displaced away from the zero-point delay. This is caused by the limited coherence of the swept-laser source in SS-OCT. This behavior is referred to as roll-off or fall-off. All the methods considered in this paper provide similar roll-off. Thus, the important parameter for comparing the methods is the amplitude of the PSF relative to the noise level. The quantitative determination of this noise level is somewhat arbitrary. It is not constant with depth and it is usually larger in the vicinity of the peak due to side lobes in the envelope function. We thus first compare PSF amplitude to noise level for the various methods from graphical representations of the PSFs. This allows observation of the additional spurious structures resulting from the application of some of the signal processing methods. We then evaluate signal-to-noise ratios for a more quantitative comparison. We finally look at the evolution of both the maximum amplitude of the PSF and the OCT resolution as a function of mirror position.

Averaging A-scans allows a better comparison of the signal processing methods. Figure 2(a) shows PSFs evaluated from a single A-scan for different mirror positions with the Vandermonde matrix method. A Hamming window is applied before processing the data. The noise for a single A-scan is fluctuating around a local mean value. These fluctuations are strong and do not allow a good evaluation of the PSF amplitude relative to the noise level. Our calculations show that the amplitude of the fluctuations is about 6 dB for all PSFs whatever the method used. Consequently, the fluctuations can be reduced by averaging to provide a mean noise level. This mean noise level then allows a better appreciation of the PSF amplitude relative to the noise level. This is illustrated in Fig. 2(b) where 1000 A-scans are averaged for each PSF evaluated at different mirror positions and processed with the Vandermonde matrix method. The noise level of all PSFs is clearly visible throughout the depth. One can always estimate the true noise for a single A-scan by adding a 6 dB fluctuation to the average noise of Fig. 2(b).

Fig. 2. PSFs after Vandermonde processing. (a) 1 A-scan per PSF. (b) averaging of 1000 A-scans per PSF.

The PSFs evaluated with the various methods are presented in Fig. 3 while corresponding computational times are provided in Table 1. Each graph in Fig. 3 presents the PSFs evaluated for mirror positions from 0.1 to 6 mm, the 0 mm position being the zero-delay position of the sample arm. The different rows in Fig. 3 correspond to different signal processing methods. In all methods, a Hamming window is used to reduce leakage. For each row but the first one, the oversampling factor α used in a given method increases from left to right. Table 1 is organized similarly to Fig. 3 and provides the corresponding computational times for 1000 A-scans. Each A-scan is composed of N=1666 points. In the methods were a FFT is used, as mentioned in section 2, we use FFTPack which can be performed with a total number of points that is a product of small prime numbers. Accordingly, FFT are performed on resampled data zero-padded to 1672 points, 3332 points, and 5000 points, for oversampling factors of 1, 2, and 3, respectively. Apart from the Vandermonde matrix method, the computational times of the various methods are provided for optimized algorithms as discussed in Section 2, but running on a single processor. Further optimization could be obtained from multithreading for parallel computing as done in [35

35. D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]

]. Also, reported computational times in recent work of FD-OCT show a wide range of performance, from 3000 ms down to 15 ms for 1000 A-scans, compared to our 200 ms computational time for the case of linear interpolation with an oversampling factor α=3 and FFT [15

15. Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K.-P. Chan, M. Itoh, and T. Yatagai, “Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments,” Opt. Express 13(26), 10652–10664 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-26-10652. [CrossRef] [PubMed]

, 34–36

34. K. Wang, Z. Ding, T. Wu, C. Wang, J. Meng, M. Chen, and L. Xu, “Development of a non-uniform discrete Fourier transform based high speed spectral domain optical coherence tomography system,” Opt. Express 17(14), 12121–12131 (2009), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-17-14-12121. [CrossRef] [PubMed]

]. The factor of about 200 between the different performances is uncertain and cannot be attributable only to the use of a few processors or of a more optimized FFT algorithm [42

42. M. Frigo and S. G. Johnson, “FFTW: an adaptive software architecture for the FFT,” in Proceedings of IEEE International Conference on Acoustics, Speech and Signal Processing. (Institute of Electrical and Electronics Engineers, Seattle, 1998), pp. 1381–1384.

]. Comparing computational times from different publications thus appears to be a hazardous task. Consequently, the computational times of Table 1 should mainly be used to compare the relative efficiency of the various methods. All these computational times would decrease proportionally if multithreading was used.

The PSFs evaluated with non-uniform DFT methods are presented in the first row of Fig. 3. Figure 3(a) is obtained with the Vandermonde matrix method, considered as the reference.

Fig. 3. PSFs for various signal processing methods.

Apart from the variation of the PSF amplitude with the mirror position due to the roll-off, the PSFs are free from artifacts and the noise level for each PSF is low. The noise evolves quite similarly for all mirror positions and decreases with depth. As discussed in Section 2.1, the computational complexity of this method is larger than that of all other methods. Its good results come at a price in computational time. Table 1 indicates that processing 1000 A-scans takes over 3 sec. Our implementation of the Vandermonde matrix approach is not fully optimized, but it is not expected that further optimization would reduce the computational time to that obtained with the other methods. The Lomb periodogram is a faster way for performing non-uniform DFT. Figure 3(b) presents the PSFs obtained with the Lomb2 method: the Lomb periodogram method with 2 Lagrange interpolation points (L=2). The computational time is much improved at 0.5 s for 1000 A-scans, but enlarged replicas appear 3 mm deeper than the main peak for all PSFs. This increases significantly the noise level. Using Lomb3 with 3 Lagrange interpolation points (L=3) as in Fig. 3(c) provides better PSFs with much smaller replicas. The computational time only slightly increases from the Lomb2 case. Consequently, the Lomb3 method is an efficient non-uniform DFT method, but it doesn’t compete in terms of computational time with the methods discussed below.

The PSFs in Figs. 3(g), 3(h), and 3(i) are obtained with a cubic spline-interpolation resampling followed by FFT with oversampling factors α = 1, 2 and 3. We refer to this method as SIFFTα. SIFFTα provides a much better signal-to-noise ratio (SNR) than LIFFTα with the same oversampling factor. There are fewer replicas and the noise level is much lower. Table 1 shows that the computational times are much larger at 230 ms, 328 ms, and 428 ms for the oversampling factors 1, 2, and 3, respectively. SIFFTα is extensively used in the OCT literature, although the level of oversampling is not always clearly specified. Figure 3(h) shows that SIFFT2 provides PSFs that are very close to those obtained with the Vandermonde matrix method in Fig. 3(a). SIFFT2 is nevertheless not optimal due to its rather large computational time.

The last two rows of Fig. 3 present the PSFs obtained with the convolution method using a Kaiser-Bessel window followed by FFT. Cases are identified by the maximum number of points used in the convolution (M=3 and 5) and the oversampling factor (α = 1, 1.2 and 2). They are referred to as KBFFTM,α. The parameter β of the Kaiser-Bessel kernel takes the optimized value from Eq. (15). Figures 3(j) and 3(k) show that this method leads to enlarged replicas away from the main peak of each PSF for KBFFT3,1 and KBFFT3,1.2. These replicas are not present for KBFFT3,2 in Fig. 3(l). The results of KBFFT3,2 are very similar to those of SIFFT2, but with a computational time for 1000 A-scans 1.7 times smaller at 193 ms. Thus KBFFT3,α performs better than LIFFT3 and SIFFT2. Further improvement in computational time can be obtained by using more points in the convolution process. Figures 3(m), 3(n) and 3(o) presents PSFs obtained from KBFFT5,1, KBFFT5,1.2, KBFFT5,2, respectively. The corresponding computational times for KBFFT5,α in Table 1 are essentially the same as those for the KBFFT3,α cases with similar oversampling factors α. An important result is that replica-free PSFs with low noise levels are obtained with a smaller oversampling factor. With an oversampling factor of 1.2, the number of required points for the FFT, using FFTPack, is only 2000 points. This is a net gain when compared to any method requiring an oversampling factor of 2 where 3332 points are required. The results for KBFFT5,1.2 in Fig. 3(n) is very similar to those of SIFFT2 in Fig. 3(h), but with a further reduced computational time more than 2.5 times smaller at 129 ms.

Table 1. Computational timesa for 1000 A-scans (ms).

table-icon
View This Table

Focusing solely on the SNR, Fig. 4 shows the evolution of the SNR as a function of depth for the different processing methods. The SNR is defined in this work as:

SNR=20log10(Aμσ2σ)
(17)

where A is the maximum amplitude of a PSF, μ is the average noise level, and σ is the standard deviation of the noise. The parameters μ et σ are evaluated over a 200 μm region beginning 400 μm to the right of the main peak of the PSF. For a given method and depth, such SNR values are averaged over 100 A-scans. This definition provides a good criterion to assess how a PSF is affected by the noise far from its main peak.

Fig. 4. SNR as a function of depth for different processing methods. For clarity, the results for the various methods are distributed over multiple figures, with the Vandermonde result being present in all the figures for reference.

Finally, Fig. 5 shows the evolution of both the maximum amplitude of the PSF and the OCT resolution as a function of mirror position. The maximum amplitude of the PSF is the same for all the methods. This is why we noted at the beginning of this section that all the methods provided the same roll-off. Consequently, the variations in SNR observed in Fig. 4 between the different methods can be solely attributed to variations in noise level. The OCT resolution is evaluated from the full width at half-maximum (FWHM) of a PSF. Figure 5 shows that all methods provide the same OCT resolution. The OCT resolution is a bit enlarged to more than 16 μm due to the Hamming window applied to the data. Without a Hamming window, the measured OCT resolution is 12 μm but with larger side lobes in the envelope function.

Fig. 5. Maximum amplitude of the PSF (left) and FWHM (right) as a function of depth for different processing methods, after averaging over 100 A-scans.

The results of Figs. 3 to 5 and computational times in Table 1 indicate that KBFFT5,1.2 is the most efficient method. The convolution method in combination with a small fractional oversampling factor provides a very precise result without strongly impacting the computational time. This method produces results comparable to those of the Vandermonde matrix method in terms of PSF amplitude and noise level. It performs much faster than the often used SIFFTα no matter which oversampling factor is used in the latter.

3.3. Phantom and tissue imaging

The performance of the various signal processing methods is further evaluated from real OCT images of a multilayer phantom and of a finger nail. In both cases, a Hamming window is applied before processing the A-scans.

The multilayer phantom is made of 3 silicon layers with different concentrations of alumina powder. The scattering varies from one layer to another, so does the amplitude of the OCT signal. These phantoms are molded in a plastic cup. They are not removed from the cup for measurement and the bottom of the cup provides a 4th layer. The phantom fabrication process is a simplified version of the technology described in Bisaillon et al. [45

45. C.-É. Bisaillon, G. Lamouche, R. Maciejko, M. Dufour, and J.-P. Monchalin, “Deformable and durable phantoms with controlled density of scatterers,” Phys. Med. Biol. 53(237-N), 247 (2008). [CrossRef]

] This artificial structure provides well-defined homogeneous regions and it is an ideal tool to evaluate the efficiency of a signal processing method.

Fig. 6. OCT images of a 4-layer phantom using different signal processing methods. Images are 6.5 cm wide by 6.5 cm deep. Each axis tic mark corresponds to 1 mm. Gray-scale is a log. scale, the dynamic range is 40 dB.

These results are further confirmed in Fig. 7 where OCT images of a finger nail are presented for the various signal processing methods. The arrangement of Fig. 7 is again the same as that of Fig. 3. Ghost artifacts are again present in many images. Best OCT images, similar to that of the Vandermonde matrix method, are obtained for SIFFT2, SIFFT3, KBFFT3,2, KBFFT5,1.2, and KBFFT5,2.

Amongst the methods that provide the best images for the phantom and the finger nail, Table 1 shows that KBFFT5,1.2 is the one with the smallest computational time. Thus KBFFT5,1.2 is the most efficient method providing the best trade-off between image quality and computational time.

Fig. 7. OCT images of a finger nail of a healthy volunteer using different signal processing methods. Images are 14.0 cm wide by 6.5 cm deep. Each axis tic mark corresponds to 1 mm. Gray-scale is a log. scale, the dynamic range is 40 dB.

4. Conclusion

A comparison of various methods to process the data in SS-OCT was performed by evaluating PSFs and by imaging a structured phantom and a finger nail. These methods include non-uniform discrete Fourier transforms with Vandermonde matrix or with Lomb periodogram, resampling with linear interpolation or spline interpolation prior to FFT, and resampling through convolution with a Kaiser-Bessel function followed by FFT. The non-uniform discrete Fourier transform methods provide very good results, but at a hard cost in computational time. The methods involving linear or spline resampling were shown to perform rather poorly in terms of SNR or computational time. We introduced a new approach for SS-OCT signal processing based on an optimized version of the convolution method in which the Kaiser-Bessel window parameter is carefully selected to allow a minimal oversampling factor between 1 and 2. One can thus benefit from an efficient resampling technique while retaining small data structures. This method with a small fractional oversampling factor of 1.2 was shown to provide results similar to the method using spline interpolation with an oversampling factor of 2, but with a computational time 2.5 times smaller. Resampling through convolution using an optimized Kaiser-Bessel function with a small fractional oversampling factor should be considered as one of the methods of choice to obtain high quality OCT images in a small computational time. This conclusion also applies to spectral-domain OCT since the methods to correct for the non-linearity in wavenumber space resulting from spectrometers are similar to those used is SS-OCT.

Acknowledgements

This research was supported by the Genomics and Health Initiative of the National Research Council Canada. The authors gratefully thank Sherif S. Sherif for helpful discussions about the Vandermonde matrix method.

References and links

1.

A. F. Fercher, C. K. Hitzenberger, G. Kamp, and S. Y. El Zaiat, “Measurement of intraocular distances by backscattering spectral interferometry,” Opt. Commun. 117(1–2), 43–48 (1995). [CrossRef]

2.

G. Häusler and M. W. Linduer, ““Coherence radar” and “spectral radar”-new tools for dermatological diagnosis,” J. Biomed. Opt. 3(1), 21–31 (1998). [CrossRef]

3.

Z. Hu and A. M. Rollins, “Fourier domain optical coherence tomography with a linear-in-wavenumber spectrometer,” Opt. Lett. 32(24), 3525–3527 (2007). [CrossRef] [PubMed]

4.

G. V. Gelikonov, V. M. Gelikonov, and P. A. Shilyagin, “Linear wave-number spectrometer for spectral domain optical coherence tomography,” Proc. SPIE 6847, 68470N (2008). [CrossRef]

5.

V. Gelikonov, G. Gelikonov, and P. Shilyagin, “Linear-wavenumber spectrometer for high-speed spectral-domain optical coherence tomography,” Opt. Spectrosc. 106(3), 459–465 (2009). [CrossRef]

6.

C. M. Eigenwillig, B. R. Biedermann, G. Palte, and R. Huber, “K-space linear Fourier domain mode locked laser and applications for optical coherence tomography,” Opt. Express 16(12), 8916–8937 (2008), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-16-12-8916. [CrossRef] [PubMed]

7.

D. C. Adler, Y. Chen, R. Huber, J. Schmitt, J. Connolly, and J. G. Fujimoto, “Three-dimensional endomicroscopy using optical coherence tomography,” Nat. Photonics 1(12), 709–716 (2007). [CrossRef]

8.

M. A. Choma, K. Hsu, and J. A. Izatt, “Swept source optical coherence tomography using an all-fiber 1300-nm ring laser source,” J. Biomed. Opt. 10(4), 044009 (2005). [CrossRef]

9.

B. A. Bower, M. Zhao, R. J. Zawadzki, and J. A. Izatt, “Real-time spectral domain Doppler optical coherence tomography and investigation of human retinal vessel autoregulation,” J. Biomed. Opt. 12(4), 041214–041218 (2007). [CrossRef] [PubMed]

10.

E. Götzinger, M. Pircher, R. A. Leitgeb, and C. K. Hitzenberger, “High speed full range complex spectral domain optical coherence tomography,” Opt. Express 13(2), 583–594 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-2-583. [CrossRef] [PubMed]

11.

M. Szkulmowski, M. Wojtkowski, T. Bajraszewski, I. Gorczynska, P. Targowski, W. Wasilewski, A. Kowalczyk, and C. Radzewicz, “Quality improvement for high resolution in vivo images by spectral domain optical coherence tomography with supercontinuum source,” Opt. Commun. 246(4–6), 569–578 (2005). [CrossRef]

12.

B. Cense, N. A. Nassif, T. C. Chen, M. C. Pierce, S.-H. Yun, B. H. Park, B. Bouma, G. Tearney, and J. de Boer, “Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography,” Opt. Express 12(11), 2435–2447 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-11-2435. [CrossRef] [PubMed]

13.

C. Dorrer, N. Belabas, J.-P. Likforman, and M. Joffre, “Spectral resolution and sampling issues in Fourier-transform spectral interferometry,” J. Opt. Soc. Am. B 17(10), 1795–1802 (2000). [CrossRef]

14.

S. H. Yun, G. J. Tearney, B. E. Bouma, B. H. Park, and J. F. de Boer, “High-speed spectral-domain optical coherence tomography at 1.3 mum wavelength,” Opt. Express 11(26), 3598–3604 (2003), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-26-3598. [CrossRef] [PubMed]

15.

Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K.-P. Chan, M. Itoh, and T. Yatagai, “Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments,” Opt. Express 13(26), 10652–10664 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-26-10652. [CrossRef] [PubMed]

16.

N. A. Nassif, B. Cense, B. H. Park, M. C. Pierce, S. H. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express 12(3), 367–376 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-3-367. [CrossRef] [PubMed]

17.

A. R. Tumlinson, B. Hofer, A. M. Winkler, B. Považay, W. Drexler, and J. K. Barton, “Inherent homogenous media dispersion compensation in frequency domain optical coherence tomography by accurate k-sampling,” Appl. Opt. 47(5), 687–693 (2008). [CrossRef] [PubMed]

18.

T. Endo, Y. Yasuno, S. Makita, M. Itoh, and T. Yatagai, “Profilometry with line-field Fourier-domain interferometry,” Opt. Express 13(3), 695–701 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-3-695. [CrossRef] [PubMed]

19.

B. Baumann, E. Götzinger, M. Pircher, and C. K. Hitzenberger, “Single camera based spectral domain polarization sensitive optical coherence tomography,” Opt. Express 15(3), 1054–1063 (2007), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-15-3-1054. [CrossRef] [PubMed]

20.

C. Fan, Y. Wang, and R. K. Wang, “Spectral domain polarization sensitive optical coherence tomography achieved by single camera detection,” Opt. Express 15(13), 7950–7961 (2007), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-15-13-7950. [CrossRef] [PubMed]

21.

M. A. Choma, M. V. Sarunic, C. Yang, and J. A. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-18-2183. [CrossRef] [PubMed]

22.

R. A. Leitgeb, W. Drexler, A. Unterhuber, B. Hermann, T. Bajraszewski, T. Le, A. Stingl, and A. F. Fercher, “Ultrahigh resolution Fourier domain optical coherence tomography,” Opt. Express 12(10), 2156–2165 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-10-2156. [CrossRef] [PubMed]

23.

M. Wojtkowski, V. J. Srinivasan, T. H. Ko, J. G. Fujimoto, A. Kowalczyk, and J. S. Duker, “Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,” Opt. Express 12(11), 2404–2422 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-11-2404. [CrossRef] [PubMed]

24.

Y. T. Pan, Q. Wu, Z. G. Wang, P. R. Brink, and C. W. Du, “High-resolution imaging characterization of bladder dynamic morphophysiology by time-lapse optical coherence tomography,” Opt. Lett. 30(17), 2263–2265 (2005). [CrossRef] [PubMed]

25.

H. Ren, T. Sun, D. J. MacDonald, M. J. Cobb, and X. Li, “Real-time in vivo blood-flow imaging by moving-scatterer-sensitive spectral-domain optical Doppler tomography,” Opt. Lett. 31(7), 927–929 (2006). [CrossRef] [PubMed]

26.

P. Bu, X. Wang, and O. Sasaki, “Full-range parallel Fourier-domain optical coherence tomography using sinusoidal phase-modulating interferometry,” J. Opt. A, Pure Appl. Opt. 9(4), 422–426 (2007). [CrossRef]

27.

T. S. Ralston, D. L. Marks, P. S. Carney, and S. A. Boppart, “Interferometric synthetic aperture microscopy,” Nat. Phys. 3(2), 129–134 (2007). [CrossRef]

28.

Y. Chen, H. Zhao, and Z. Wang, “Investigation on spectral-domain optical coherence tomography using a tungsten halogen lamp as light source,” Opt. Rev. 16(1), 26–29 (2009). [CrossRef]

29.

A. Liu, R. Wang, K. L. Thornburg, and S. Rugonyi, “Efficient postacquisition synchronization of 4-D nongated cardiac images obtained from optical coherence tomography: application to 4-D reconstruction of the chick embryonic heart,” J. Biomed. Opt. 14(4), 044020–044011 (2009). [CrossRef] [PubMed]

30.

R. Huber, M. Wojtkowski, K. Taira, J. Fujimoto, and K. Hsu, “Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging: design and scaling principles,” Opt. Express 13(9), 3513–3528 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-9-3513. [CrossRef] [PubMed]

31.

B. Hofer, B. Považay, B. Hermann, A. Unterhuber, G. Matz, F. Hlawatsch, and W. Drexler, “Signal post processing in frequency domain OCT and OCM using a filter bank approach,” Proc. SPIE 6443, 64430O (2007). [CrossRef]

32.

S. S. Sherif, C. Flueraru, Y. Mao, and S. Change, “Swept Source Optical Coherence Tomography with Nonuniform Frequency Domain Sampling,” in Biomedical Optics, OSA Technical Digest (CD) (Optical Society of America, 2008), paper BMD86.

33.

T. H. Chow, S. G. Razul, B. K. Ng, G. Ho, and C. B. A. Yeo, “Enhancement of Fourier domain optical coherence tomorgraphy images using discrete Fourier transform method,” Proc. SPIE 6847, 68472T (2008). [CrossRef]

34.

K. Wang, Z. Ding, T. Wu, C. Wang, J. Meng, M. Chen, and L. Xu, “Development of a non-uniform discrete Fourier transform based high speed spectral domain optical coherence tomography system,” Opt. Express 17(14), 12121–12131 (2009), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-17-14-12121. [CrossRef] [PubMed]

35.

D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]

36.

Y. Zhang, X. Li, L. Wei, K. Wang, Z. Ding, and G. Shi, “Time-domain interpolation for Fourier-domain optical coherence tomography,” Opt. Lett. 34(12), 1849–1851 (2009). [CrossRef] [PubMed]

37.

I. Gohberg and V. Olshevsky, “Fast algorithms with preprocessing for matrix-vector multiplication problems,” J. Complexity 10(4), 411–427 (1994). [CrossRef]

38.

V. V. Vityazev, “Time series analysis of unequally spaced data: Intercomparison between the Schuster periodogram and the LS-spectra,” Astron. Astrophys. Trans. 11(2), 139–158 (1996). [CrossRef]

39.

N. R. Lomb, “Least-squares frequency analysis of unequally spaced data,” Astrophys. Space Sci. 39(2), 447–462 (1976). [CrossRef]

40.

W. H. Press, S. A. Teukolsky, W. T. Vetterling, and B. P. Flannery, Numerical Recipes in FORTRAN (Cambridge University Press, New York, 1992).

41.

P. N. Swarztrauber, Vectorizing the FFTs, in Parallel Computations, G. Rodrigue, ed. (Academic Press, New York, NY, 1982), pp. 51–83.

42.

M. Frigo and S. G. Johnson, “FFTW: an adaptive software architecture for the FFT,” in Proceedings of IEEE International Conference on Acoustics, Speech and Signal Processing. (Institute of Electrical and Electronics Engineers, Seattle, 1998), pp. 1381–1384.

43.

J. I. Jackson, C. H. Meyer, D. G. Nishimura, and A. Macovski, “Selection of a convolution function for Fourier inversion using gridding computerised tomography application,” IEEE Trans. Med. Imaging 10(3), 473–478 (1991). [CrossRef] [PubMed]

44.

P. J. Beatty, D. G. Nishimura, and J. M. Pauly, “Rapid gridding reconstruction with a minimal oversampling ratio,” IEEE Trans. Med. Imaging 24(6), 799–808 (2005). [CrossRef] [PubMed]

45.

C.-É. Bisaillon, G. Lamouche, R. Maciejko, M. Dufour, and J.-P. Monchalin, “Deformable and durable phantoms with controlled density of scatterers,” Phys. Med. Biol. 53(237-N), 247 (2008). [CrossRef]

OCIS Codes
(070.4340) Fourier optics and signal processing : Nonlinear optical signal processing
(170.4500) Medical optics and biotechnology : Optical coherence tomography

ToC Category:
Fourier Optics and Signal Processing

History
Original Manuscript: February 1, 2010
Manuscript Accepted: April 18, 2010
Published: May 5, 2010

Virtual Issues
Vol. 5, Iss. 9 Virtual Journal for Biomedical Optics

Citation
Sébastien Vergnole, Daniel Lévesque, and Guy Lamouche, "Experimental validation of an optimized signal processing method to handle non-linearity in swept-source optical coherence tomography," Opt. Express 18, 10446-10461 (2010)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-18-10-10446


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. A. F. Fercher, C. K. Hitzenberger, G. Kamp, and S. Y. El Zaiat, “Measurement of intraocular distances by backscattering spectral interferometry,” Opt. Commun. 117(1-2), 43–48 (1995). [CrossRef]
  2. G. Häusler and M. W. Linduer, ““Coherence radar” and “spectral radar”-new tools for dermatological diagnosis,” J. Biomed. Opt. 3(1), 21–31 (1998). [CrossRef]
  3. Z. Hu and A. M. Rollins, “Fourier domain optical coherence tomography with a linear-in-wavenumber spectrometer,” Opt. Lett. 32(24), 3525–3527 (2007). [CrossRef] [PubMed]
  4. G. V. Gelikonov, V. M. Gelikonov, and P. A. Shilyagin, “Linear wave-number spectrometer for spectral domain optical coherence tomography,” Proc. SPIE 6847, 68470N (2008). [CrossRef]
  5. V. Gelikonov, G. Gelikonov, and P. Shilyagin, “Linear-wavenumber spectrometer for high-speed spectral-domain optical coherence tomography,” Opt. Spectrosc. 106(3), 459–465 (2009). [CrossRef]
  6. C. M. Eigenwillig, B. R. Biedermann, G. Palte, and R. Huber, “K-space linear Fourier domain mode locked laser and applications for optical coherence tomography,” Opt. Express 16(12), 8916–8937 (2008), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-16-12-8916 . [CrossRef] [PubMed]
  7. D. C. Adler, Y. Chen, R. Huber, J. Schmitt, J. Connolly, and J. G. Fujimoto, “Three-dimensional endomicroscopy using optical coherence tomography,” Nat. Photonics 1(12), 709–716 (2007). [CrossRef]
  8. M. A. Choma, K. Hsu, and J. A. Izatt, “Swept source optical coherence tomography using an all-fiber 1300-nm ring laser source,” J. Biomed. Opt. 10(4), 044009 (2005). [CrossRef]
  9. B. A. Bower, M. Zhao, R. J. Zawadzki, and J. A. Izatt, “Real-time spectral domain Doppler optical coherence tomography and investigation of human retinal vessel autoregulation,” J. Biomed. Opt. 12(4), 041214–041218 (2007). [CrossRef] [PubMed]
  10. E. Götzinger, M. Pircher, R. A. Leitgeb, and C. K. Hitzenberger, “High speed full range complex spectral domain optical coherence tomography,” Opt. Express 13(2), 583–594 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-2-583 . [CrossRef] [PubMed]
  11. M. Szkulmowski, M. Wojtkowski, T. Bajraszewski, I. Gorczynska, P. Targowski, W. Wasilewski, A. Kowalczyk, and C. Radzewicz, “Quality improvement for high resolution in vivo images by spectral domain optical coherence tomography with supercontinuum source,” Opt. Commun. 246(4-6), 569–578 (2005). [CrossRef]
  12. B. Cense, N. A. Nassif, T. C. Chen, M. C. Pierce, S.-H. Yun, B. H. Park, B. Bouma, G. Tearney, and J. de Boer, “Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography,” Opt. Express 12(11), 2435–2447 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-11-2435 . [CrossRef] [PubMed]
  13. C. Dorrer, N. Belabas, J.-P. Likforman, and M. Joffre, “Spectral resolution and sampling issues in Fourier-transform spectral interferometry,” J. Opt. Soc. Am. B 17(10), 1795–1802 (2000). [CrossRef]
  14. S. H. Yun, G. J. Tearney, B. E. Bouma, B. H. Park, and J. F. de Boer, “High-speed spectral-domain optical coherence tomography at 1.3 mum wavelength,” Opt. Express 11(26), 3598–3604 (2003), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-26-3598 . [CrossRef] [PubMed]
  15. Y. Yasuno, V. D. Madjarova, S. Makita, M. Akiba, A. Morosawa, C. Chong, T. Sakai, K.-P. Chan, M. Itoh, and T. Yatagai, “Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments,” Opt. Express 13(26), 10652–10664 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-26-10652 . [CrossRef] [PubMed]
  16. N. A. Nassif, B. Cense, B. H. Park, M. C. Pierce, S. H. Yun, B. Bouma, G. Tearney, T. Chen, and J. de Boer, “In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve,” Opt. Express 12(3), 367–376 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-3-367 . [CrossRef] [PubMed]
  17. A. R. Tumlinson, B. Hofer, A. M. Winkler, B. Považay, W. Drexler, and J. K. Barton, “Inherent homogenous media dispersion compensation in frequency domain optical coherence tomography by accurate k-sampling,” Appl. Opt. 47(5), 687–693 (2008). [CrossRef] [PubMed]
  18. T. Endo, Y. Yasuno, S. Makita, M. Itoh, and T. Yatagai, “Profilometry with line-field Fourier-domain interferometry,” Opt. Express 13(3), 695–701 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-3-695 . [CrossRef] [PubMed]
  19. B. Baumann, E. Götzinger, M. Pircher, and C. K. Hitzenberger, “Single camera based spectral domain polarization sensitive optical coherence tomography,” Opt. Express 15(3), 1054–1063 (2007), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-15-3-1054 . [CrossRef] [PubMed]
  20. C. Fan, Y. Wang, and R. K. Wang, “Spectral domain polarization sensitive optical coherence tomography achieved by single camera detection,” Opt. Express 15(13), 7950–7961 (2007), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-15-13-7950 . [CrossRef] [PubMed]
  21. M. A. Choma, M. V. Sarunic, C. Yang, and J. A. Izatt, “Sensitivity advantage of swept source and Fourier domain optical coherence tomography,” Opt. Express 11(18), 2183–2189 (2003), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-11-18-2183 . [CrossRef] [PubMed]
  22. R. A. Leitgeb, W. Drexler, A. Unterhuber, B. Hermann, T. Bajraszewski, T. Le, A. Stingl, and A. F. Fercher, “Ultrahigh resolution Fourier domain optical coherence tomography,” Opt. Express 12(10), 2156–2165 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-10-2156 . [CrossRef] [PubMed]
  23. M. Wojtkowski, V. J. Srinivasan, T. H. Ko, J. G. Fujimoto, A. Kowalczyk, and J. S. Duker, “Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,” Opt. Express 12(11), 2404–2422 (2004), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-12-11-2404 . [CrossRef] [PubMed]
  24. Y. T. Pan, Q. Wu, Z. G. Wang, P. R. Brink, and C. W. Du, “High-resolution imaging characterization of bladder dynamic morphophysiology by time-lapse optical coherence tomography,” Opt. Lett. 30(17), 2263–2265 (2005). [CrossRef] [PubMed]
  25. H. Ren, T. Sun, D. J. MacDonald, M. J. Cobb, and X. Li, “Real-time in vivo blood-flow imaging by moving-scatterer-sensitive spectral-domain optical Doppler tomography,” Opt. Lett. 31(7), 927–929 (2006). [CrossRef] [PubMed]
  26. P. Bu, X. Wang, and O. Sasaki, “Full-range parallel Fourier-domain optical coherence tomography using sinusoidal phase-modulating interferometry,” J. Opt. A, Pure Appl. Opt. 9(4), 422–426 (2007). [CrossRef]
  27. T. S. Ralston, D. L. Marks, P. S. Carney, and S. A. Boppart, “Interferometric synthetic aperture microscopy,” Nat. Phys. 3(2), 129–134 (2007). [CrossRef]
  28. Y. Chen, H. Zhao, and Z. Wang, “Investigation on spectral-domain optical coherence tomography using a tungsten halogen lamp as light source,” Opt. Rev. 16(1), 26–29 (2009). [CrossRef]
  29. A. Liu, R. Wang, K. L. Thornburg, and S. Rugonyi, “Efficient postacquisition synchronization of 4-D nongated cardiac images obtained from optical coherence tomography: application to 4-D reconstruction of the chick embryonic heart,” J. Biomed. Opt. 14(4), 044020–044011 (2009). [CrossRef] [PubMed]
  30. R. Huber, M. Wojtkowski, K. Taira, J. Fujimoto, and K. Hsu, “Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging: design and scaling principles,” Opt. Express 13(9), 3513–3528 (2005), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-13-9-3513 . [CrossRef] [PubMed]
  31. B. Hofer, B. Považay, B. Hermann, A. Unterhuber, G. Matz, F. Hlawatsch, and W. Drexler, “Signal post processing in frequency domain OCT and OCM using a filter bank approach,” Proc. SPIE 6443, 64430O (2007). [CrossRef]
  32. S. S. Sherif, C. Flueraru, Y. Mao, and S. Change, “Swept Source Optical Coherence Tomography with Nonuniform Frequency Domain Sampling,” in Biomedical Optics, OSA Technical Digest (CD) (Optical Society of America, 2008), paper BMD86.
  33. T. H. Chow, S. G. Razul, B. K. Ng, G. Ho, and C. B. A. Yeo, “Enhancement of Fourier domain optical coherence tomorgraphy images using discrete Fourier transform method,” Proc. SPIE 6847, 68472T (2008). [CrossRef]
  34. K. Wang, Z. Ding, T. Wu, C. Wang, J. Meng, M. Chen, and L. Xu, “Development of a non-uniform discrete Fourier transform based high speed spectral domain optical coherence tomography system,” Opt. Express 17(14), 12121–12131 (2009), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-17-14-12121 . [CrossRef] [PubMed]
  35. D. Hillmann, G. Huttmann, and P. Koch, “Using nonequispaced fast Fourier transformation to process optical coherence tomography signals,” Proc. SPIE 7372, 73720R (2009). [CrossRef]
  36. Y. Zhang, X. Li, L. Wei, K. Wang, Z. Ding, and G. Shi, “Time-domain interpolation for Fourier-domain optical coherence tomography,” Opt. Lett. 34(12), 1849–1851 (2009). [CrossRef] [PubMed]
  37. I. Gohberg and V. Olshevsky, “Fast algorithms with preprocessing for matrix-vector multiplication problems,” J. Complexity 10(4), 411–427 (1994). [CrossRef]
  38. V. V. Vityazev, “Time series analysis of unequally spaced data: Intercomparison between the Schuster periodogram and the LS-spectra,” Astron. Astrophys. Trans. 11(2), 139–158 (1996). [CrossRef]
  39. N. R. Lomb, “Least-squares frequency analysis of unequally spaced data,” Astrophys. Space Sci. 39(2), 447–462 (1976). [CrossRef]
  40. W. H. Press, S. A. Teukolsky, W. T. Vetterling, and B. P. Flannery, Numerical Recipes in FORTRAN (Cambridge University Press, New York, 1992).
  41. P. N. Swarztrauber, Vectorizing the FFTs, in Parallel Computations, G. Rodrigue, ed. (Academic Press, New York, NY, 1982), pp. 51–83.
  42. M. Frigo, and S. G. Johnson, “FFTW: an adaptive software architecture for the FFT,” in Proceedings of IEEE International Conference on Acoustics, Speech and Signal Processing. (Institute of Electrical and Electronics Engineers, Seattle, 1998), pp. 1381–1384.
  43. J. I. Jackson, C. H. Meyer, D. G. Nishimura, and A. Macovski, “Selection of a convolution function for Fourier inversion using gridding computerised tomography application,” IEEE Trans. Med. Imaging 10(3), 473–478 (1991). [CrossRef] [PubMed]
  44. P. J. Beatty, D. G. Nishimura, and J. M. Pauly, “Rapid gridding reconstruction with a minimal oversampling ratio,” IEEE Trans. Med. Imaging 24(6), 799–808 (2005). [CrossRef] [PubMed]
  45. C.-É. Bisaillon, G. Lamouche, R. Maciejko, M. Dufour, and J.-P. Monchalin, “Deformable and durable phantoms with controlled density of scatterers,” Phys. Med. Biol. 53(237–N), 247 (2008). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.