1. Introduction
Recently, second harmonic generation (SHG) has proven to be a useful window into the amount
and organization of fibrillar collagen in biological tissues due to its relative specificity and
the fact that it is an intrinsic signal [
1
R. M. Williams, W. R. Zipfel, and W. W. Webb, “Interpreting second-harmonic generation images of collagen I fibrils,” Biophys. J.
88(2), 1377–1386 (2005). [CrossRef]
–
6
S. V. Plotnikov, A. C. Millard, P. J. Campagnola, and W. A. Mohler, “Characterization of the myosin-based source for second-harmonic generation from muscle sarcomeres,” Biophys. J.
90(2), 693–703 (2006). [CrossRef]
]. SHG is a coherent phenomenon, which implies that SHG is
sensitive not only to the amplitude of the illumination field but also to its phase. In addition
to a spatial resolution that is equal to other imaging techniques (such as two photon excited
fluorescence), SHG microscopy can provide information about the sample’s molecular
structure. For example, the ratio of the forward-propagating to backward propagating SHG signal
(the “F/B ratio”) can help us to understand the axial extent of ordering in
collagen fibers [
1
R. M. Williams, W. R. Zipfel, and W. W. Webb, “Interpreting second-harmonic generation images of collagen I fibrils,” Biophys. J.
88(2), 1377–1386 (2005). [CrossRef]
–
5
A. C. Kwan, D. A. Dombeck, and W. W. Webb, “Polarized microtubule arrays in apical dendrites and axons,” Proc. Natl. Acad. Sci. U.S.A.
105(32), 11370–11375 (2008). [CrossRef]
[PubMed]
].
Previously, in vitro measurements of SHG F/B ratios have been used to study collagen fiber
ordering in various tissue samples such as rat tail, ovarian cancer biopsies, mouse models of
breast cancer, and dermis from mouse models of Osteogenesis Imperfecta (OIM) [
1
R. M. Williams, W. R. Zipfel, and W. W. Webb, “Interpreting second-harmonic generation images of collagen I fibrils,” Biophys. J.
88(2), 1377–1386 (2005). [CrossRef]
–
5
A. C. Kwan, D. A. Dombeck, and W. W. Webb, “Polarized microtubule arrays in apical dendrites and axons,” Proc. Natl. Acad. Sci. U.S.A.
105(32), 11370–11375 (2008). [CrossRef]
[PubMed]
,
7
O. Nadiarnykh, R. B. LaComb, M. A. Brewer, and P. J. Campagnola, “Alterations of the extracellular matrix in ovarian cancer studied by second harmonic generation imaging microscopy,” BMC Cancer
10(1), 94 (2010). [CrossRef]
[PubMed]
]. The F/B ratio revealed the length scale of ordering in the
fibers and in the case of OIM and ovarian cancer, was able to discriminate pathological tissue
from healthy tissue [
2
R. Lacomb, O. Nadiarnykh, and P. J. Campagnola, “Quantitative second harmonic generation imaging of the diseased state osteogenesis imperfecta: experiment and simulation,” Biophys. J.
94(11), 4504–4514 (2008). [CrossRef]
[PubMed]
,
7
O. Nadiarnykh, R. B. LaComb, M. A. Brewer, and P. J. Campagnola, “Alterations of the extracellular matrix in ovarian cancer studied by second harmonic generation imaging microscopy,” BMC Cancer
10(1), 94 (2010). [CrossRef]
[PubMed]
]. In these measurements a second objective lens was needed to collect forward
propagating SHG signal. Hence, the tissue sample had to be dissected from the animal and
sectioned to 100um slices to allow signal to reach the second detection lens. For clinical
application, such as in endoscopy, it is impossible to put an objective and a PMT detector
underneath the tissue sample to collect the forward propagating SHG. The excision and sectioning
required to use a second detector for forward propagating SHG also prevents dynamic measurements
of collagen ordering over time. Therefore, it becomes highly desirable to develop a new optical
system to measure collagen SHG F/B ratio in vivo, using only epi-detection and on intact thick
tissue samples.
2. Experimental methods
2.1 Optical setup
The optical setup we are using to determine the SHG F/B ratio in vivo using only epidetection
is shown in
Fig. 1
below. SHG signal was generated by a Spectra Physics MaiTai Ti:Sapphire laser providing
100fs pulses at 80 MHz and 810 nm. The excitation beam was directed into an Olympus Fluoview
F300 scanhead connected to an Olympus BX61WI upright microscope. The focusing objective is an
Olympus UMPLFL20XW water immersion lenses (20 × , 0.5 N.A.), with supporting
measurements performed using an Olympus LUMPlanFL/IR (40x, 0.8 NA) water immersion lens. This
objective was used to focus the excitation beam on the sample and at the same time collect the
direct backward-propagating SHG as well as the forward-propagating SHG signal that was
subsequently backscattered within the tissue. The SHG signal was collected by the objective,
converged by the tube lens, and collimated by the pupil lens. The collimated SHG beam was then
de-scanned and focused again on the pinhole plane by the collector lens. All of these lenses
are intrinsic to the BX61WI microscope or the Fluoview F300 scanhead. The focal lengths of the
pupil lens and the collector lens are 54mm and 185mm respectively, and size of the pinholes on
the F300 variable pinhole turret are 60um, 100um, 150um, 200um and 7000um. The SHG beam was
separated from the excitation beam by a dichroic mirror (Chroma 670 DCSX) inserted into the
F300 scanhead, as well as a band pass filter centered at 405 nm (Chroma HQ405/30m-2P) placed
after the pinhole, and detected by the FV300’s intrinsic photomultiplier tubes. The only
significant equipment modifications are insertion of the aforementioned dichroic mirror and
filter in the appropriate location in the scanhead, and the punching out of one of the pinholes
in the pinhole turret to produce one extremely large pinhole setting (see below).
Fig. 1 Experimental setup for in vivo measurement of tumor collagen SHG F/B ratio.
2.2 SHG Back Scattering and Monte Carlo Simulation
Using the apparatus described above, we generate an SHG image using confocal detection of the
resultant SHG. While much of the forward-propagating SHG signal will travel into the tissue
sample and be lost to our objective lens, a fraction of this signal undergoes multiple
scattering events and passes back through the object plane, traveling towards the objective
lens. As shown in
Fig. 2
, this signal will pass through the object plane at multiple locations. Conversely, at
shallow imaging depths the backwards-propagating SHG signal will emanate from the image plane
only from the two-photon focal volume with minimal subsequent scatter. When the object plane is
imaged onto the confocal pinhole, the spatial distribution of SHG signal on the confocal plane
will consist of a sharp central peak due to the backward propagating SHG plus a diffuse signal
due to the forwards propagating and subsequently backscattered SHG. As discussed below, our new
method consists of repeatedly imaging the sample through a series of different sized pinholes,
whereby the shape of this total SHG distribution can be measured and, with suitable
calibration, the underlying F/B ratio can be determined.
Fig. 2 Back scattering of forward-propagating SHG and propagation of direct backward propagating
SHG
In order to implement this method we must first understand the spatial distribution, in the
object plane, of forward propagated light that subsequently backscatters and reaches that
plane. 405 nm SHG signal is easily scattered while propagating in the tissue sample. In Legare
et al 2007 [
8
F. Légaré, C. Pfeffer, and B. R. Olsen, “The role of backscattering in SHG tissue imaging,” Biophys. J.
93(4), 1312–1320 (2007). [CrossRef]
[PubMed]
], it was determined that ~21% of the
forward-propagating SHG signal was subsequently backscattered and reached the tissue surface,
and that the epi-detected image from a 5mm thick Achilles tendon tissue block was actually a
combination of the direct backward propagating SHG signal and the forward propagating SHG
signal that was subsequently backscattered. In fact for collagen fibers in some tissue samples,
such as in mouse breast tumor models, the SHG F/B ratio is very high (~30) [
4
X. Han, R. M. Burke, M. L. Zettel, P. Tang, and E. B. Brown, “Second harmonic properties of tumor collagen: determining the structural relationship between reactive stroma and healthy stroma,” Opt. Express
16(3), 1846–1859 (2008). [CrossRef]
[PubMed]
], and most SHG signal propagates in the forward direction.
In these cases epi-detected SHG microscopy is greatly enhanced due to the subsequent
backscattering of the forward propagating SHG.
The propagation of light in turbid media can be well modeled by Monte Carlo simulation of the
paths that photons make as they travel through tissue, which are chosen by statistically
sampling the probability distributions for step size and angular deflection per scattering
event. Using a Monte Carlo code based closely upon that of Wang et al [
9
L. Wang and S. L. Jacques, “Hybrid model of Monte Carlo simulation and diffusion theory for light reflectance by turbid media,” J. Opt. Soc. Am. A
10(8), 1746–1752 (1993). [CrossRef]
] (see Appendix), we sent 405nm SHG photons into an infinitely deep
scattering tissue in the forward direction. The SHG photons were emitted from a point source
and we ignored the direct backward propagating SHG photons. We experimentally measured the
angular distribution of the forward propagating SHG emission by imaging the back focal plane of
an objective lens which was collecting the forward propagating SHG from a 100 um section of
tumor tissue, and found that the forward SHG from collagen fibers is emitted in a rather tight
beam confined to +/−15 degrees around the laser axis (data not shown). Hence we used the
same initial angular distribution (+/−15 degrees around the axis normal to the
tissue-air surface) in the Monte Carlo simulation. The scattering particles are modeled as 10um
diameter cells, the refractive indexes inside and outside of the cells are 1.38 and 1.42
respectively, and the scattering and absorption coefficients we used in the simulation are
µa = 0.7 cm−1 and µs = 150 cm−1 [
10
W. F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron.
26(12), 2166–2185 (1990). [CrossRef]
]. The probability distribution of scattering angles after every scattering event
were calculated according to Mie theory with the cell size and the refractive index inside and
outside of the cells, with a 99% chance the scattering angle will fall between +/−6
degrees. We then counted the number of photons which escape from the air-tissue surface in the
backward direction and plotted out the steady state distribution of the backscattered photons
over the radial distance from the emission point.
The simulation results are shown in
Fig. 3
.
Figure 3(a) shows the steady state distribution
of the forward propagating and subsequently backscattered SHG photons which reach the object
plane, over a large range of radial positions i.e. from the center to 3mm away from the
emission point.
Figure 3(b) shows the same distribution
over a short range i.e. from the center to only 50um away from the emission point, which is the
length scale of our largest pinhole setting (where the pinhole-bearing foil has been punched
out). We can see from
Fig. 3(a) that over a large length
scale the forward propagating and subsequently back-scattered SHG photon intensity decays
exponentially with the distance from emission point, which agrees with previous literature
[
11
T. J. Farrell, M. S. Patterson, and B. Wilson, “A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,” Med. Phys.
19(4), 879–888 (1992). [CrossRef]
[PubMed]
]. But from
Fig.
3(b) we can see over a very short range of 50 um, the range covered by our pinhole
settings and where the diffusion approximation does not hold [
11
T. J. Farrell, M. S. Patterson, and B. Wilson, “A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,” Med. Phys.
19(4), 879–888 (1992). [CrossRef]
[PubMed]
], the intensity of forward propagating and subsequently back scattered SHG photons
is constant and does not depend on distance from the focal volume.
Fig. 3 Monte Carlo simulation of forward propagating and subsequently backscattered SHG photons
which reach the object plane.(a) Steady state radial distribution of the backscattered SHG
photons over a large range of radial position (3mm) (b) The same distribution over a short
radial range (50um)
2.3 Fitting model
To measure collagen fiber SHG F/B ratio in vivo with only the epi-detection objective, we
will generate a series of collagen fiber SHG images, in the backward channel, through a series
of confocal pinholes of different sizes which correspond to a diameter in the object plane
ranging from 0.874 um to 102 um (see 3.1 below). As discussed above, the SHG signal we collect
after the pinhole is a mixture of both the direct backward propagating SHG and the forward
propagating and subsequently backscattered SHG. To analyze our experimental data we model the
image in the pinhole plane of the direct backward propagating SHG signal plane as a Gaussian
spot [
12
W. R. Zipfel, R. M. Williams, and W. W. Webb, “Nonlinear magic: multiphoton microscopy in the biosciences,” Nat. Biotechnol.
21(11), 1369–1377 (2003). [CrossRef]
[PubMed]
], while the forward propagating SHG that
subsequently backscatters is modeled as a uniform distribution over these length scales (based
upon the results of our Monte Carlo simulation, as described above). So the total SHG signal
intensity distribution on the object plane can be expressed as:
where ω is the e−2 Gaussian
spot size of the direct backward propagating SHG, F and B are absolute intensities of forward
and backward propagating SHG signals, the parameter C relates the initial forward propagating
signal intensity to the average intensity of the uniform distribution of SHG light that reaches
the object plane, and is a function of scattering and absorption properties of the underlying
tissue. Alternatively, this expression can be written in another way,
where F/B represents the collagen fiber
SHG F/B ratio.
When we generate collagen fiber SHG images through a series of confocal pinholes of different
sizes, each pixel on that image represents an integration of the total SHG signal over the
pinhole area
where R is the size of the pinhole with
respect to Gaussian spot size of the direct backward propagating SHG, i.e. R =
rpinhole/ω. If the pixel intensities at various pinhole sizes are normalized to the
maximum pixel intensity at the largest pinhole size, the relative pixel intensity is a function
of relative pinhole size R:
where Rmax is the largest pinhole size in
our system. With this expression we can plot the relative pixel intensity versus pinhole size
and fit the data to produce (F/B)C. In order to determine F/B we must then eliminate C, the
fraction of signal which originally propagates in the forward direction but is eventually
backscattered by the tissue and reaches the pinhole plane. This can be done by introducing into
the sample a reference of known F/B ratio, whose C value is the same. In these experiments we
used blue fluorescent polystyrene beads (peak λem = 424nm) sprinkled onto the sample
surface. The quantity (F/B)C is then determined for both the calibration beads in the image, as
well as the fibers of interest. Next, the correction factor that is required to convert the
measured values of bead (F/B)C to the known value of F/B is determined, and the previously
unknown value of fiber F/B is determined from the measured (F/B)C using the same correction
factor:
To illustrate the fitting function given by
Eq.
(4) in the case of the typical backscattering fraction C = 0.001 and Rmax = 28 we can
plot out the relative pixel intensity v.s. pinhole size curves at various collagen fiber SHG
F/B ratios in
Fig. 4
:
Fig. 4 Relative SHG intensity vs. relative pinhole size curves at different collagen SHG F/B
ratios, based upon
Eq. (4). The relative SHG
intensities equal 1 at a pinhole size of r = Rmax = 28.
From this plot one can see the sharp curvature due to the Gaussian distribution of direct
back-propagating SHG (most evident in the steep early rise of the curves with small values of
F/B) and the slow and steady rise due to forward-propagating light that is subsequently
backscattered and produces a diffuse signal which does not vary with r (most evident in the
slow rise of the curves with large values of F/B). This reveals the utility in punching out one
of the metal foils containing a pinhole from the pinhole turret of the scanner, as this
produces one pinhole of extremely large size (R = 28 in the case of the Fluoview 300) whose
radius is far from the initial sharp curvature due to the Gaussian distribution of direct
back-propagating SHG, allowing the collection of a great deal of the diffuse forward-propagated
signal.
3. Results
3.1 Determining backward propagating SHG spot size on the pinhole plane
In order to produce the required plot of SHG signal versus pinhole size and fit it with our
model, we must first determine the sizes of the pinholes, relative to the Gaussian spot size of
the direct backward propagating SHG on the pinhole plane. Since we are determining the spot
size in the pinhole plane of the direct backward propagating SHG, we must avoid significant
backscattering of the forward propagating SHG. Therefore rat tail collagen samples were fresh
frozen and sectioned into 10um thin slices, spread out on a cover slip and dried overnight in a
refrigerator for good adhesion between sample sections and cover slip. The cover slip was then
flipped over with the sample side facing down, submerged in PBS and the excitation beam
transmitted through the cover slip. The sample was submerged in saline to minimize the
refractive index change and thus reduce the subsequent backscattering of the forward
propagating SHG. The saline container is a cup with 4 cm diameter and 4 cm depth. It was
painted black to absorb forward propagating SHG that goes through the sample section (see
Fig. 5
).
Fig. 5 Configuration of objective, sample and scatterer holder
To determine the size of the image of the SHG spot in the pinhole plane, we measured the SHG
spot directly with a CCD camera (SPOT RT3, SciTech) at the intermediate image plane depicted on
Fig. 1. The SHG spot in the pinhole plane is just a
magnified version of the spot we captured with the CCD camera with a magnification factor of
the ratio of the focal length of the pupil lens and collection lens.
As shown in
Fig. 6(a)
above, we drew a straight line through the center of the SHG spot captured by CCD
camera, and measured SHG intensity along this line. The SHG intensity vs. horizontal location
data set was then plotted in
Fig. 6(b) as a series of
separate points, and fitted to a Gaussian model producing an average spot size in the
intermediate image plane of 9.34 pixels or 69 um. Taking into account the magnification factor
of the pupil and collector lens, the average pinhole plane spot size of the backward
propagating SHG is therefore 236.8 um. We already knew the absolute diameters of the five
pinholes on the pinhole turret are 60um 100um 150um 200um to 7000um respectively (7000um
corresponds to the location in the pinhole turret where we punched out the pinhole-bearing
metal foil). That corresponds to 0.24 ω, 0.4 ω, 0.6 ω, 0.8 ω and 28
ω, where ω stands for the e−2 SHG spot size in the pinhole plane.
Fig. 6 Determination of SHG spot size. (a) CCD captured image of SHG focal spot after the tube
lens (b) SHG intensity along the straight line that goes through the center of SHG spot. The
red straight line in (a) is the line along which SHG intensities are measured. The red curve
in (b) is the Gaussian fit to the intensities we measured in (a)
3.2 Measurement of SHG intensity varied with pinhole size
To evaluate our new method we chose the rat tail tendon, a sample whose SHG properties have
been well studied [
1
R. M. Williams, W. R. Zipfel, and W. W. Webb, “Interpreting second-harmonic generation images of collagen I fibrils,” Biophys. J.
88(2), 1377–1386 (2005). [CrossRef]
]. A whole rat tail was removed from
a previously sacrificed animal (removal from the animal is not necessary for the technique but
is convenient for handling the sample). To generate a clear SHG image of the tendon in
particular (to match the sample previously studied in the literature) we peeled a thin layer of
outer skin off the rat tail at the location of interest and exposed the tendon beneath it. We
then put the rat tail on a glass slide, with the exposed collagen fiber facing up and we put
another coverslip on top of the collagen fibers to ensure a stable sample. The rat tail and
coverslip were then fixed on the glass slide with plastic tape and the collagen fibers were
imaged though the coverslip. Note that, unlike previous methods which measure F/B with a second
detector for the forward-propagating SHG, the tendons remained within the ~1cm diameter
tail.
We prepared whole rat tail samples from 5 separate animals (see
Fig. 7
). On each rat tail we chose 5 image fields. And for each image field we took 5 back
detected SHG images with the pinhole size varied from 60um 100um 150um 200um to 7000um, plus
one image with no sample in order to quantify the background noise. One of these image sets is
shown below.
Fig. 7 In vivo SHG images of rat tail collagen fibers in an intact rat tail. Images 1-5 are SHG
images of the same ROI when the size of the pinhole varied from 60um, 100um, 150um, 200um,
to 7000um, respectively. The bright spots are blue fluorescent polystyrene beads for
calibration. Images are 600 um across.
3.3 Curve fitting and prediction of collagen SHG F/B ratio
In each set of images, we picked 5 small regions of interest (ROIs) around collagen fibers
and 3 regions of interest around the calibration beads. ROIs were drawn around fibers that
extended over at least ~100 um in the image plane, to ensure that fibers were close to
perpendicular to the optical axis. A length 100 um in an optical section of ~12 um thickness
(e−2 z diameter of the PSF) corresponds to a maximum angle in the object plane of ~7
degrees. Signal intensities in these collagen and beads ROIs were measured with ImageJ,
background was subtracted, and signal intensities were normalized so that the maximum intensity
measured with the largest pinhole was set to 1. The average relative SHG intensity of all 25
collagen ROIs (5 ROIs in each of 5 image stacks) and average relative TPEF intensity of all 15
beads ROIs in one rat tail vs. relative pinhole size plot is shown below. Note the separation
between the collagen fiber curve and the bead curve due to the different F/B ratios of rat tail
collagen and fluorescent calibration beads.
As described above, for one rat tail we measured relative intensity in 25 collagen ROIs and
15 bead ROIs. For each ROI 5 relative intensities were measured with size of the pinhole
varying from the smallest to the largest. Average relative SHG intensity of all 25 collagen
ROIs and average relative TPEF intensity of all 15 bead ROIs, as shown in
Fig. 8
, were considered a set of data for one rat tail. We then fit this set of real measured
data to the model given in
Eq. (4) and calculated
the rat tail collagen fiber SHG F/B ratio in one animal by eliminating C using
Eq. (5). This data collection/averaging/curve
fitting/calculation process was repeated 5 times to calculate rat tail collagen fiber SHG F/B
ratio in 5 animals. The results are listed in
Table 1
below.
Fig. 8 Epidetected total collagen SHG intensity (blue triangles), or total bead TPEF intensity
(blue squares) vs. pinhole size, fit to the model given by
Eq. (4) (red lines). The horizontal axis is pinhole size in units of
backward propagating SHG spot size on the pinhole plane i.e. fraction of ω. The
vertical axis is normalized SHG intensity. Note that the fitting curves identically equal 1
when the pinhole size reaches R = 28, whose data point is not shown, and that the intensity
information for R = 28 provides the normalization value and hence is included in the overall
fit.
Table 1 Results of in vivo measured rat tail collagen SHG F/B ratio
| Animal I | Animal II | Animal III | Animal IV | Animal V | Average |
|---|
| F/B ratio | 2.10 ± 0.19 | 1.81 ± 0.30 | 2.50 ± 0.34 | 1.34 ± 0.13 | 2.26 ± 0.23 | 2.00 ± 0.45 |
For each animal, means ± 95% confidence intervals are presented, with the data from 25
ROIs being pooled to produce a single curve fit for each animal. The average value for the 5
animals is then presented ± the standard deviation.
To verify the validity of our method using thick samples and only epi-detection, we also
directly measured rat tail collagen SHG F/B ratio with both forward and backward detectors and
with the sample sectioned in ~10um thin slices, as previously described [
4
X. Han, R. M. Burke, M. L. Zettel, P. Tang, and E. B. Brown, “Second harmonic properties of tumor collagen: determining the structural relationship between reactive stroma and healthy stroma,” Opt. Express
16(3), 1846–1859 (2008). [CrossRef]
[PubMed]
]. In summary, the SHG image captured in the forward detector was divided by
the SHG image captured in the backward detector. We then averaged SHG F/B ratio over all pixels
within collagen fibers and ignored all pixels outside the collagen fibers. The scattering of
the tissue, though very low in this case, was corrected for by measuring the F/B ratio of the
TPEF signals from 10um diameter calibration fluorescent beads.
The measurement results from 5 animals, using forward and backwards detectors, are listed in
Table 2
.
Table 2 Results of in vitro measured rat tail collagen SHG F/B ratio
| Animal VI | Animal VII | Animal VIII | Animal IX | Animal X | Average |
|---|
| F/B ratio | 1.43 ± 0.80 | 1.22 ± 0.27 | 2.29 ± 0.49 | 1.75 ± 0.50 | 2.11 ± 0.86 | 1.76 ± 0.45 |
For each animal, means ± standard deviations are presented from 5 F/B measurements.
The average value for the 5 animals is then presented ± the standard deviation.
The overall average F/B ratio from 5 animals is 1.76+/−0.45. This result was not
statistically significantly different from our measurement using only the epi-detection
objective lens and the model given by
Eq. (4)
(Student’s t-test p = 0.42) demonstrating that the results from the two methods are in
good agreement. To evaluate the new method at a higher NA we also measured the F/B ratio using
a 0.8 NA lens, producing a value for F/B (1.37+/−0.28) which is again not statistically
significantly different from the direct measurement using forward and backwards detectors (p =
0.17 N = 5).
4. Discussion
In this paper, we present a method to measure the SHG F/B ratio with only an epi-detection
objective lens, allowing measurement on thick tissue samples. In our demonstration we used the
whole rat tail (~1 cm thick), with the rat removed for convenience, but the technique is equally
applicable to a live specimen. By obviating the need for thin sectioning of the sample, this
technique provides the opportunity to do time-dependent studies, as well as the possibility of
use in an endoscopic setting. This is significant because SHG F/B ratios have been shown to be
of interest in discriminating skin with Osteogenesis Imperfecta [
2
R. Lacomb, O. Nadiarnykh, and P. J. Campagnola, “Quantitative second harmonic generation imaging of the diseased state osteogenesis imperfecta: experiment and simulation,” Biophys. J.
94(11), 4504–4514 (2008). [CrossRef]
[PubMed]
] from normal dermis [
2
R. Lacomb, O. Nadiarnykh, and P. J. Campagnola, “Quantitative second harmonic generation imaging of the diseased state osteogenesis imperfecta: experiment and simulation,” Biophys. J.
94(11), 4504–4514 (2008). [CrossRef]
[PubMed]
] and SHG F/B ratio
measurements have been used to help determine the organization of fibrillar collagen in samples
such as breast tumor models [
4
X. Han, R. M. Burke, M. L. Zettel, P. Tang, and E. B. Brown, “Second harmonic properties of tumor collagen: determining the structural relationship between reactive stroma and healthy stroma,” Opt. Express
16(3), 1846–1859 (2008). [CrossRef]
[PubMed]
], in rat tail tendon [
1
R. M. Williams, W. R. Zipfel, and W. W. Webb, “Interpreting second-harmonic generation images of collagen I fibrils,” Biophys. J.
88(2), 1377–1386 (2005). [CrossRef]
], cellulose [
3
O. Nadiarnykh, R. B. Lacomb, P. J. Campagnola, and W. A. Mohler, “Coherent and incoherent SHG in fibrillar cellulose matrices,” Opt. Express
15(6), 3348–3360 (2007). [CrossRef]
[PubMed]
], mouse
Achilles tendon [
8
F. Légaré, C. Pfeffer, and B. R. Olsen, “The role of backscattering in SHG tissue imaging,” Biophys. J.
93(4), 1312–1320 (2007). [CrossRef]
[PubMed]
], muscle fascia [
8
F. Légaré, C. Pfeffer, and B. R. Olsen, “The role of backscattering in SHG tissue imaging,” Biophys. J.
93(4), 1312–1320 (2007). [CrossRef]
[PubMed]
], and ovarian cancer [
7
O. Nadiarnykh, R. B. LaComb, M. A. Brewer, and P. J. Campagnola, “Alterations of the extracellular matrix in ovarian cancer studied by second harmonic generation imaging microscopy,” BMC Cancer
10(1), 94 (2010). [CrossRef]
[PubMed]
] (although see
4.2 Limitations, below). Furthermore, the confocal pinholes used in this technique are available
in most commercial two photon microscopes, allowing the measurement of the SHG F/B ratio in
intact specimens without addition of new equipment (assuming one is already using a two-photon
microscope to study SHG) or extensive equipment modification (except for a single dichroic and
filter inserted in the dichroic holder of the scanbox, and the punching out of one of the
pinhole settings to produce one very large pinhole).
4.1 Sensitivity
In order to determine the radial intensity distribution of the forward propagating and
subsequently backscattered SHG using a Monte Carlo simulation, the tissue scattering parameters
we used in the simulation were µa = 0.7 cm−1 and µs = 150 cm−1,
which are typical values for 405 nm light [
10
W. F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron.
26(12), 2166–2185 (1990). [CrossRef]
]. However,
the scattering properties of tissue vary over a large range among different organs throughout
the body, and it is reasonable to wonder if this fitting model is good for all types of tissues
from different organs. To determine this we varied the scattering parameters over a large range
and repeated the Monte Carlo simulation with the results shown in
Fig. 9
and
Fig. 10
.
Fig. 9 Monte Carlo simulation of forward propagating SHG that is subsequently backscattered and
reaches the object plane, with varying µs. Steady state radial distribution of the
backscattered SHG photons when (a) µs = 100 cm−1 (b) µs = 150
cm−1 (c) µs = 200 cm−1 (d) µs = 250 cm−1 (e) µs =
300 cm−1 (f) µs = 350 cm−1 (g) µs = 400 cm−1 (h)
µs = 450 cm−1 (i) µs = 500 cm−1. In each case µa = 0.7
cm−1
Fig. 10 Monte Carlo simulation of forward propagating SHG that is subsequently backscattered and
reaches the object plane, with varying µa. (a) Steady state radial distribution of
the backscattered SHG photons when (a) µa = 0.05 cm−1 (b) µa = 0.1
cm−1 (c) µa = 1 cm−1 (d) µa = 2 cm−1 (e) µa = 3
cm−1 (f) µa = 4 cm−1 (g) µa = 5 cm−1 (h) µa = 6
cm−1 (i) µa = 7 cm−1. In all three cases µs = 500
cm−1
We can see from the simulation results that when µa decreases or µs increases
more forward propagating SHG photons reach the object plane after subsequent backscattering;
and when µa increases or µs decreases less reach the object plane. However, over
a range of scattering parameters in our Monte Carlo simulations that are comparable to values
found at these wavelengths in the literature [
10
W. F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron.
26(12), 2166–2185 (1990). [CrossRef]
],
varying µa and µs does not alter the distribution of light exiting the tissue
over the 0-50 um length scale, and it remains a constant, independent of radial position (a
slope of zero is within the 95% confidence interval). We therefore conclude that within this
small region of the object plane close to the optical axis, which is the region covered by our
pinhole distribution, the radial intensity distribution of the backscattered SHG is not
affected by a wide variety of tissue scattering parameters, and this fitting model is therefore
applicable for tissue from diverse organs.
4.2 Limitations
This technique has two significant limitations. As shown in
Fig. 4, when the F/B ratio increases, the separation between two SHG intensity vs.
pinhole size curves decreases. Using a reasonable value for C [
8
F. Légaré, C. Pfeffer, and B. R. Olsen, “The role of backscattering in SHG tissue imaging,” Biophys. J.
93(4), 1312–1320 (2007). [CrossRef]
[PubMed]
],
Fig. 4 shows that when the collagen fiber
SHG F/B ratio is significantly more than ~5, the two curves are so close that they are not
likely to be distinguishable assuming typical variation in experimental data (such as shown in
Fig. 8). This implies that the new method might not be
applicable for samples with collagen fiber SHG F/B ratios significantly higher than ~5. To
illustrate this, we applied this method to measure the SHG F/B ratio of collagen fibers in
intact mouse skeletal muscle fascia (with overlying skin removed to allow access). Using the
new technique, we produced a value of the collagen fiber F/B ratio in vivo of
5.05+/−0.81 (N = 5) while after extracting the fascia, mounting it between two
coverslips and directly measuring SHG F/B ratio with two objective lenses we produced a value
of 5.56+/1.03 (N = 5), which was not statistically significantly different (p = 0.43). We also
applied this method to measure the SHG F/B ratio of collagen fibers in 4T1 breast tumor models.
According to our previous work the true collagen fiber SHG F/B ratio in 4T1 mouse breast tumor
is ~30. We measured collagen fiber SHG F/B ratio in vivo in whole tumor samples from 5
different animals (25 ROIs per animal), and the result is 6.64+/−2.98 (data not shown).
As predicted, as the F/B becomes significantly larger than ~5, the result becomes inaccurate
and noisy, with a greater relative standard deviation, because of the normal random noise in
the experimental data combined with the insensitivity of this new technique to large F/B
values. Fortunately, our data and published studies reveal that the F/B ratio in many
biological samples is at or below five, including in the rat tail tendon (F/B~1), skeletal
muscle fascia (F/B~5), cellulose (F/B~4), ovarian cancer (F/B ~3.4), and dermis with (F/B~3.4)
and without (F/B~2.64) Osteogenesis Imperfecta [
1
R. M. Williams, W. R. Zipfel, and W. W. Webb, “Interpreting second-harmonic generation images of collagen I fibrils,” Biophys. J.
88(2), 1377–1386 (2005). [CrossRef]
–
3
O. Nadiarnykh, R. B. Lacomb, P. J. Campagnola, and W. A. Mohler, “Coherent and incoherent SHG in fibrillar cellulose matrices,” Opt. Express
15(6), 3348–3360 (2007). [CrossRef]
[PubMed]
,
7
O. Nadiarnykh, R. B. LaComb, M. A. Brewer, and P. J. Campagnola, “Alterations of the extracellular matrix in ovarian cancer studied by second harmonic generation imaging microscopy,” BMC Cancer
10(1), 94 (2010). [CrossRef]
[PubMed]
,
8
F. Légaré, C. Pfeffer, and B. R. Olsen, “The role of backscattering in SHG tissue imaging,” Biophys. J.
93(4), 1312–1320 (2007). [CrossRef]
[PubMed]
]. Care must be taken when this technique
produces F/B ratios significantly greater than five.
The second limitation arises from the fact that this technique assumes there is no subsequent
scattering of back-propagating SHG signal. Hence it is limited to the surface, or extremely
shallow imaging depths, of the sample and is suitable for quantifying F/B ratios on intact
thick tissue samples, but not in intact thick tissue samples.
5. Conclusion
In this paper, we present a method to measure SHG F/B ratio suitable for intact tissue samples
without sectioning, using just the epidetection objective lens. The method requires minor
modification of the dichroics, filters, and confocal pinholes used in most commercial two photon
microscopes and hence minimizes the purchase of new equipment. This allows F/B ratio
measurements to be done in a dynamic fashion, and offers the possibility of endoscopic
measurements. This technique is sensitive to F/B values up to ~5 and within this range are a
variety of interesting and clinically relevant materials, tissues, and disease states, including
tendon, fascia, cellulose, ovarian cancer, and dermis with and without Osteogenesis Imperfecta.
The fact that OIM is within the range of this technique’s sensitivity offers the
attractive possibility of a non-invasive optical diagnosis of that disease, using a single
objective lens, without biopsy samples being removed from the patient.
Acknowledgements
We thank Dr. Jerome Mertz for helpful conversations. This work is supported by a Department of Defense BCRP Pre-doctoral Traineeship Award (W81XWH-08-1-0323) to Xiaoxing Han, a Department of Defense BCRP Era of Hope Scholar Award (W81XWH05-1-0396), and a Pew Scholar in the Biomedical Sciences Award to Edward Brown III.
References and links
1. |
R. M. Williams, W. R. Zipfel, and W. W. Webb, “Interpreting second-harmonic generation images of collagen I fibrils,” Biophys. J.
88(2), 1377–1386 (2005). [CrossRef]
|
2. |
R. Lacomb, O. Nadiarnykh, and P. J. Campagnola, “Quantitative second harmonic generation imaging of the diseased state osteogenesis imperfecta: experiment and simulation,” Biophys. J.
94(11), 4504–4514 (2008). [CrossRef]
[PubMed]
|
3. |
O. Nadiarnykh, R. B. Lacomb, P. J. Campagnola, and W. A. Mohler, “Coherent and incoherent SHG in fibrillar cellulose matrices,” Opt. Express
15(6), 3348–3360 (2007). [CrossRef]
[PubMed]
|
4. |
X. Han, R. M. Burke, M. L. Zettel, P. Tang, and E. B. Brown, “Second harmonic properties of tumor collagen: determining the structural relationship between reactive stroma and healthy stroma,” Opt. Express
16(3), 1846–1859 (2008). [CrossRef]
[PubMed]
|
5. |
A. C. Kwan, D. A. Dombeck, and W. W. Webb, “Polarized microtubule arrays in apical dendrites and axons,” Proc. Natl. Acad. Sci. U.S.A.
105(32), 11370–11375 (2008). [CrossRef]
[PubMed]
|
6. |
S. V. Plotnikov, A. C. Millard, P. J. Campagnola, and W. A. Mohler, “Characterization of the myosin-based source for second-harmonic generation from muscle sarcomeres,” Biophys. J.
90(2), 693–703 (2006). [CrossRef]
|
7. |
O. Nadiarnykh, R. B. LaComb, M. A. Brewer, and P. J. Campagnola, “Alterations of the extracellular matrix in ovarian cancer studied by second harmonic generation imaging microscopy,” BMC Cancer
10(1), 94 (2010). [CrossRef]
[PubMed]
|
8. |
F. Légaré, C. Pfeffer, and B. R. Olsen, “The role of backscattering in SHG tissue imaging,” Biophys. J.
93(4), 1312–1320 (2007). [CrossRef]
[PubMed]
|
9. |
L. Wang and S. L. Jacques, “Hybrid model of Monte Carlo simulation and diffusion theory for light reflectance by turbid media,” J. Opt. Soc. Am. A
10(8), 1746–1752 (1993). [CrossRef]
|
10. |
W. F. Cheong, S. A. Prahl, and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron.
26(12), 2166–2185 (1990). [CrossRef]
|
11. |
T. J. Farrell, M. S. Patterson, and B. Wilson, “A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,” Med. Phys.
19(4), 879–888 (1992). [CrossRef]
[PubMed]
|
12. |
W. R. Zipfel, R. M. Williams, and W. W. Webb, “Nonlinear magic: multiphoton microscopy in the biosciences,” Nat. Biotechnol.
21(11), 1369–1377 (2003). [CrossRef]
[PubMed]
|