Automated volumetric evaluation of stereoscopic disc photography

Juan Xu, Hiroshi Ishikawa, Gadi Wollstein, Richard A Bilonick, Larry Kagemann, Jamie E Craig, David A Mackey, Alex W Hewitt, and Joel S Schuman  »View Author Affiliations


Optics Express, Vol. 18, Issue 11, pp. 11347-11359 (2010)
http://dx.doi.org/10.1364/OE.18.011347


View Full Text Article

Acrobat PDF (1520 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

PURPOSE: To develop a fully automated algorithm (AP) to perform a volumetric measure of the optic disc using conventional stereoscopic optic nerve head (ONH) photographs, and to compare algorithm-produced parameters with manual photogrammetry (MP), scanning laser ophthalmoscope (SLO) and optical coherence tomography (OCT) measurements. METHODS: One hundred twenty-two stereoscopic optic disc photographs (61 subjects) were analyzed. Disc area, rim area, cup area, cup/disc area ratio, vertical cup/disc ratio, rim volume and cup volume were automatically computed by the algorithm. Latent variable measurement error models were used to assess measurement reproducibility for the four techniques. RESULTS: AP had better reproducibility for disc area and cup volume and worse reproducibility for cup/disc area ratio and vertical cup/disc ratio, when the measurements were compared to the MP, SLO and OCT methods. CONCLUSION: AP provides a useful technique for an objective quantitative assessment of 3D ONH structures.

© 2010 OSA

1. Introduction

Glaucoma is the second most common cause of blindness in the world. Excavation of the optic nerve head (ONH) is a typical clinical sign of glaucoma. Disruption of the connective tissue support at the optic disc and posterior displacement or thinning of the lamina cribrosa may also play a morphological role in optic cup enlargement [1

1. A. J. Bellezza, C. J. Rintalan, H. W. Thompson, J. C. Downs, R. T. Hart, and C. F. Burgoyne, “Deformation of the lamina cribrosa and anterior scleral canal wall in early experimental glaucoma,” Invest. Ophthalmol. Vis. Sci. 44(2), 623–637 (2003). [CrossRef] [PubMed]

]. Regardless of the mechanism, accurate assessment of the optic disc is paramount in the routine clinical review of patients with ocular hypertension and glaucoma.

The aims of this study were to describe a fully automated method for volumetric measurements of the ONH using a photogrammetric algorithm, to evaluate its performance in comparison with the measurements obtained from CSLO and OCT.

2. Methods

2.1 Study participants

All subjects had a comprehensive ophthalmic evaluation, axial length measurement with the IOL master (Carl Zeiss Meditec, Dublin, CA), reliable visual field test (Humphrey Field Analyzer; Carl Zeiss Meditec, Dublin, CA; with less than 30% fixation losses, false-positive and false-negative responses), CSLO (Heidelberg Retina Tomograph (HRT) 3, Heidelberg Engineering, Heidelberg, Germany) and OCT (Stratus OCT; Carl Zeiss Meditec, Dublin, CA). Subjects’ pupils were dilated with 1% tropicamide and 2.5% phenylephrine (Alcon Laboratories, Fort Worth, TX) and stereoscopic disc photographs were obtained (Nidek 3Dx; Nidek, Gamagori, Japan). All tests were acquired within a six-month period.

2.2 Automated photogrammetry

Nidek 3Dx and a Nikon D1x digital camera unit (Nikon, Tokyo, Japan) was used for optic disc photography. The image resolution was 1312 × 2000 pixels. The stereo-pairs were cropped to 480 × 1020 pixels for image registration and 3D reconstruction. Image analysis and 3D modeling of the ONH were performed using software of our own design, as previously described [3

3. J. Xu, O. Chutatape, C. Zheng, and P. C. Kuan, “Three dimensional optic disc visualisation from stereo images via dual registration and ocular media optical correction,” Br. J. Ophthalmol. 90(2), 181–185 (2006). [CrossRef] [PubMed]

,4

4. J. Xu, H. Ishikawa, G. Wollstein, R. A. Bilonick, K. R. Sung, L. Kagemann, K. A. Townsend, and J. S. Schuman, “Automated assessment of the optic nerve head on stereo disc photographs,” Invest. Ophthalmol. Vis. Sci. 49(6), 2512–2517 (2008). [CrossRef] [PubMed]

]. A process flowchart was illustrated in Fig. 1
Fig. 1 : Overview of the process flow
to give a short introduction of the designed software. A series of technical algorithms were developed to automatically generate ONH volumetric measurements based on conventional stereoscopic photographs. The process included: (1) 3D ONH reconstruction, (2) 2D and 3D magnification correction, (3) Optic disc and cup margin detection, and (4) Optic cup volumetric measurements. Enhancements to the previously described software added new functionality, specifically in step (2), (3) and (4).

2.2.1 3D ONH reconstruction

In the stereo image pair, depth is inversely proportional to the disparity between the two matching points from the left and right images (Fig. 2
Fig. 2 : Parallel stereo image configuration
), where the coordinate difference (vL-vR) of the corresponding points is defined as the disparity [3

3. J. Xu, O. Chutatape, C. Zheng, and P. C. Kuan, “Three dimensional optic disc visualisation from stereo images via dual registration and ocular media optical correction,” Br. J. Ophthalmol. 90(2), 181–185 (2006). [CrossRef] [PubMed]

,4

4. J. Xu, H. Ishikawa, G. Wollstein, R. A. Bilonick, K. R. Sung, L. Kagemann, K. A. Townsend, and J. S. Schuman, “Automated assessment of the optic nerve head on stereo disc photographs,” Invest. Ophthalmol. Vis. Sci. 49(6), 2512–2517 (2008). [CrossRef] [PubMed]

,15

15. K. Deguchi, D. Kawamata, K. Mizutani, H. Hontani, and K. Wakabayashi, ““3D fundus shape reconstruction and display from stereo fundus images,” IEICE Trans. INF. & SYST,” E 83-D(7), (2000).

]. Multiple corresponding points were identified automatically on the left and right images by calculating the combination of the highest correlations and minimal differences of features within the given searching window. The disparities of the corresponding points were converted into depths. The scaled 3D ONH model was then generated from sparse corresponding points and interpolation (Fig. 1A,B).

2.2.2 2D and 3D magnification correction

2.2.3 Optic disc and cup margin detection

The geometrical center of the region with the highest intensity value on the image was located as the disc center. The disc margin was generated using a deformable model technique whereby a smooth boundary of high contrast surrounding the disc center was detected (Fig. 1C). The image features, such as pixel intensity value, gradient, and contour smoothness, were extracted and used in computation of the deformable model [6

6. J. Xu, O. Chutatape, and P. Chew, “Automated optic disk boundary detection by modified active contour model,” IEEE Trans. Biomed. Eng. 54(33 issue. 3), 473–482 (2007). [CrossRef] [PubMed]

]. The cup was located at 150μm posterior to the disc margin for eyes where 0.2mm ≤ cup depth ≤ 1.0mm, or at the 1/5 depth from the disc margin to the deepest point of the cup for eyes with cupping < 0.2mm or > 1.0mm (Fig. 4A
Fig. 4 : Geometric model of optic nerve head (ONH) with cross sectional profile and projected image. (A) Locating cup margin points in deformable model algorithm. (B) Cross sectional profile of 3D ONH model, with definitions of various ONH parameters, i.e., disc area, rim area, cup area, rim volume and cup volume. (C) The definitions of vertical C/D ratio and C/D area ratio.
). Cup margin was extracted using a deformable model technique similar to that described for disc margin detection; however, a different energy function, which amalgamated data relating to cup depth and location, intensity gradient, contour smoothness and shape, was used. Optic cups were assumed to be located in close proximity to the geometrical center of the optic disc to control the flexibility of the deformable model and avoid generating the cup margin of irregular shape.

2.2.4 Optic cup volumetric measurements

Magnification corrections were applied on each stereo disc photograph as described in step (2) to compute the 3D and 2D scales and to generate the actual size 3D disc model. Seven ONH parameters, disc area, rim area, cup area, cup-to-disc (C/D) area ratio and vertical C/D ratio, were automatically generated by the designed software (Fig. 4B,C). Disc area and cup area were defined as the area inside the disc margin and cup margin respectively, while rim area was defined as the area inside the disc margin but outside cup margin. C/D area ratio was the cup area divided by disc area. Vertical C/D ratio was defined as the vertical cup diameter through the geometric center divided by the vertical disc diameter through disc center (Fig. 4C). All the areas and ratios were measured on the 2D stereo disc photograph. A 3D reference plane was generated from the detected cup margin on the 3D disc model, as illustrated by the cross sectional profile in Fig. 4B. The voxel size was computed from the corrected 3D disc model. The cup volume was measured as the summation of the total number of voxels within the disc margin and below the reference plane multiplied by the voxel size, while rim volume was the voxels within the disc margin between the reference plane and the 3D optic disc surface (Fig. 4B).

2.3 Quality assessment of the automatically detected disc margin

The developed software provided a fully automated ONH assessment. Detected disc margin quality was assessed by an independent and automated method using Fourier transform analysis. The disc margin was converted in polar coordinates centered at the geometric center of the detected disc margin and then transformed into Fourier domain. Automated quality assessment was used by analyzing the frequency along the margin in Fourier domain. The assumption is that optic disc is close to oval shape with a smooth boundary. High frequency components correspond to noise on the boundary. Therefore the boundary with an unsmooth and irregular shape should have more high frequency components and is more like to be a failure. The analysis was defined as successful when the low frequency energy was larger than 87.5% (7/8) of the total energy and the high frequency energy was less than 5%. Low frequency was defined as < 7% of the frequency components at the lower end, and high frequency was defined as the higher half of the frequency component. These thresholds were set arbitrarily.

2.4 Manual photogrammetry

Three glaucoma specialists independently and manually defined the disc margin and the cup margin on stereo disc photographs in a randomized and masked fashion. The final location of the disc and cup margins for the analysis was defined by the majority opinion of the observers using the method described by Abramoff et al. [7

7. M. D. Abràmoff, W. L. M. Alward, E. C. Greenlee, L. Shuba, C. Y. Kim, J. H. Fingert, and Y. H. Kwon, “Automated segmentation of the optic disc from stereo color photographs using physiologically plausible features,” Invest. Ophthalmol. Vis. Sci. 48(4), 1665–1673 (2007). [CrossRef] [PubMed]

] Optic disc parameters were generated from the manual disc demarcation using the same processes of 3D modeling and 2D and 3D magnification correction methods as those described above for the automated photogrammetry.

2.5 HRT

HRT scans were performed using a HRT2 device. The files were then transferred to HRT3 software to be processed. Eligible images had pixel standard deviation < 50µm with even illumination, acceptable centration and focus. The optic disc margin was manually delineated at the border of Elschnig’s ring and the disc parameters were generated by the HRT3 software (version 1.5.10.0).

2.6 OCT

Fast ONH scans were taken using Stratus OCT for each eye. Eligible images had signal strength > 7. The disc margin was automatically defined at the termination of the retinal pigment epithelium (RPE) in each of 6 radially sampled cross-sectional retinal images. The OCT software generated a disc margin by interpolating these 12 detected points. Manual correction were required if the disc margin was not correctly placed at the end of RPE. The disc parameters were then calculated by the OCT software (version 4.0.7).

2.7 Statistical analysis

Seven disc parameters including disc area, rim area, cup area, cup/disc area ratio, vertical cup/disc ratio, rim volume and cup volume were generated by the automated photogrammetric algorithm (AP), manual photogrammetry (MP) for each individual observer and the major opinion of the three observers, HRT and OCT. Latent variable measurement error models (LVM) were used to determine the relative biases (the relative systematic error for each technique) and imprecision (random error) for each method. LVM is a statistical method that extracts the shared properties among different parameters that aim to measure the same feature. The model requires that data from both eyes of the same subjects will be used. Both eyes were eliminated from this analysis if only one eye was qualified. A separate LVM was calculated for each of the seven parameters accounting for the correlation between eyes from the same subject. Square root transformation of the measurements was used to reduce positive skewness (lack of Normality) where necessary. In order to assess the difference in relative bias and imprecision between methods the ratio of the estimates was computed. The relative bias indicates whether the measurement from one method is consistently higher (bias ratio > 1) or lower (bias ratio < 1) than another method. Imprecision ratios were adjusted by taking imprecision standard deviation over bias. Higher imprecision SD corresponds to larger random error and lower reproducibility. If the ratio of imprecision SD between method A and B is less than 1, this indicates method A has better reproducibility than method B, and vice versa.

Linear calibration curves were computed (using the relative biases) and combined with estimates of imprecision to evaluate the relationship among the measurements of each method. Automated photogrammetry was used as the reference slope with all other methods calibrated accordingly. Mixed effect models were used to evaluate the differences in cases where there was a constant bias between the measurements. No statistical method is available to assess the difference in the presence of non-constant bias because the difference changes with level - i.e., measurements of MP is bigger than AP for bigger rim area values and smaller than AP for smaller rim area values.

The general statistical modeling software package Mx was used to estimate the LVM parameters [8

8. M.C. Neale, S.M. Boker, G. Xie, H.H. Maes, Mx: Statistical Modeling 6th Ed.(VCU Box 900126, Richmond, VA 23298: Department of Psychiatry.)

]. Other statistical analyses were performed using the R Language and Environment for Statistical Computing (R 2.6.0, R development Core Team, 2007, http://www.r-project.org). P < 0.05 was considered as statistically significant.

3. Results

The mean age of the 61 participants in the study was 56.8 ± 9.6. Twenty-nine eyes (22 subjects) out of the 122 eyes (23.8%) were labeled as no-success by the automated quality assessment of AP. Examples of algorithm success and failure appear in Fig. 5
Fig. 5 Comparison between automatically photogrammetric algorithm and manual disc demarcation. (A) A successful example of disc demarcation. (B) An example of failure in disc demarcation due to prominent peripapillary atrophy.
.

No consistent pattern in terms of mean and standard deviation was noted among the observers when the results obtained from each of the three individual observers were compared, i. e., comparing seven ONH measurements of one observer with measurements of majority opinion of three individual observers, some measurements were larger and others were smaller. For all ONH parameters, the correlation between the latent variables (representing the “true values”) between eyes of the same subject exceeded 0.93 (except for cup volume with a correlation of 0.79) indicating high correlation between eyes. Table 1

Table 1. : Comparison of optic disc measurements (mean ± standard deviation) from automated photogrammetry (AP), manual photogrammetry (MP), Heidelberg retina tomography (HRT) and optical coherence tomography (OCT).

table-icon
View This Table
| View All Tables
summarizes the mean and standard deviation of seven disc parameters measured by the four methods from all qualified eyes (78 eyes, 39 subjects). For most parameters there were non-constant biases so that the bias between methods depended on the level of the parameter. All parameters that had constant bias were statistically significantly lower than AP.

Table 2

Table 2. Imprecision standard deviation of optic disc parameters measured by automated photogrammetry (AP), manual photogrammetry (MP), Heidelberg retina tomography (HRT) and optical coherence tomography (OCT).

table-icon
View This Table
| View All Tables
summarizes the imprecision standard deviation (SD) for each method. Table 3

Table 3. Relative imprecision standard deviation ratio (adjusted by bias, 95% confidence interval) of optic disc parameters measured by automated photogrammetry (AP), manual photogrammetry (MP), Heidelberg retina tomography (HRT) and optical coherence tomography (OCT).

table-icon
View This Table
| View All Tables
summarizes the imprecision ratio (adjusted by bias) among the methods. Disc area measurements obtained by MP had a SD that was 2.0054 times larger than those obtained with AP (Table 3). As the confidence interval did not cross the value of 1, this ratio was statistically significant. MP had the smallest SD among the four methods for most of the disc parameters. The imprecision SDs of the AP were statistically significantly lower for disc area, rim volume and cup volume (except for rim volume comparison between AP and MP that did not reach statistically significant level), and the highest for rim area, cup area, cup/disc area ratio and vertical cup/disc ratio (although cup area comparisons between AP and the other three methods, and rim area comparisons between AP and both HRT and OCT did not reach the statistically significant level), compared with MP, HRT and OCT. This indicates that AP had statistically significantly smaller variability (lower SD) in measuring disc area, rim volume and cup volume. When the comparison was conducted between AP and each individual observer’s measurements, there were inconsistent results with any of the observers or the AP showing alternately the best performance.

The slopes (relative bias ratio) and intercepts of the calibration curves are given in Table 4

Table 4. Linear calibration between optic nerve head measurements of automated photogrammetry (AP), manual photogrammetry (MP), HRT and OCT. AP was set as reference and all measurements were accordingly adjusted.

table-icon
View This Table
| View All Tables
and Fig. 6
Fig. 6 Relative bias comparison between optic nerve head measurements of automated photogrammetry, manual photogrammetry, Heidelberg retina tomography (HRT) and optical coherence tomography (OCT). Automated photogrammetry was used as reference to all other methods
. No constant bias was apparent for most measurements among the four methods, as shown in Fig. 6, with calibration curves intersect each other. For example, disc area measurements obtained by MP increased 6.98% (1-0.9302) slower than disc area measurements that were obtained by AP (Fig. 6A). Disc area, rim area and cup area obtained by AP were constantly (no cross point) and statistically significantly higher than HRT measurements, while disc area and cup volume were statistically significantly higher with AP than OCT measurements. The differences in the calibration curves between AP and MP tended to be smaller compared to the differences with HRT and OCT in all ONH measurements.

4. Discussion

In this study we described an automated ONH photogrammetry technique that provided volumetric parameters (Fig. 1). ONH measurements from the new technique were compared with manual demarcation of the margins from three observers, major opinion of the three observers (MP), HRT and OCT. AP and MP showed high similarity as indicated by the mean and standard deviation values of the seven disc parameters (Table 1) and similar reproducibility to those obtained by the other automated imaging methods (HRT, OCT) (Tables 2, 3). Linear calibration equations were also computed to determine the relationship between measurements obtained by the various methods (Table 4 and Fig. 6).

An automated segmentation algorithm was used to demarcate the optic disc margin based primarily on the components of pixel intensity values, smoothness, and gradient in the image. We found however, that this automated method occasionally placed some disc margins away from the true location. This is likely due to indiscernible differences between Elschnig’s rim and peripapillary tissue in photographs (Fig. 5B).

The extraction or delineation of the optic cup margin is the initial requirement for the automated measurement of disc parameters. 3D depth information has been shown to be a reliable feature by which to segment the cup and as such many definitions for the optic cup have been proposed [9

9. HRT tutorial. Quantitative three-dimensional imaging of the posterior segment with the Heidelberg Retina Tomograph. http://www.heidelbergengineering.com/hrtf/hrt.html

13

13. P. Hrynchak, N. Hutchings, D. Jones, and T. Simpson, “A comparison of cup-to-disc ratio measurement in normal subjects using optical coherence tomography image analysis of the optic nerve head and stereo fundus biomicroscopy,” Ophthalmic Physiol. Opt. 24(6), 543–550 (2004). [CrossRef] [PubMed]

]. The HRT software defines the optic cup to be 50μm below the mean surface height along a six degree annulus at the temporal inferior disc margin [9

9. HRT tutorial. Quantitative three-dimensional imaging of the posterior segment with the Heidelberg Retina Tomograph. http://www.heidelbergengineering.com/hrtf/hrt.html

]. OCT software defines the optic cup below a reference plane located 150μm superior to the RPE/Bruch’s membrane tips. In the automated program we used a hybrid approach where the cup margin was located at a fixed distance from the disc margin for average size cups, and a relative distance for large and small cups [4

4. J. Xu, H. Ishikawa, G. Wollstein, R. A. Bilonick, K. R. Sung, L. Kagemann, K. A. Townsend, and J. S. Schuman, “Automated assessment of the optic nerve head on stereo disc photographs,” Invest. Ophthalmol. Vis. Sci. 49(6), 2512–2517 (2008). [CrossRef] [PubMed]

].

Non-constant bias was noted for most measurements between the methods indicating a difference that changes with the measurement level. Hence, measurements from one method could not simply be substituted by measurements from the other method plus or minus a certain value. In addition, although the linear calibration equation provided a way to adjust the systematic bias from one method to the other, the imprecision of the given method could not be replaced by the other method. Therefore, measurements are not inter-changeable between modalities. AP showed measurements similar to MP and was different from HRT and OCT. The difference, which was noted in cases where the comparison could be computed, might originate from the definitions of disc margin used by the different methods. OCT defines the ONH margin as the termination of the RPE (or Bruch’s membrane) in six radial scans. AP and HRT define the ONH margin as the high contrast edge points in the two-dimensional pseudo fundus photo images. However, disc photography, HRT and OCT used different wavelengths for light sources and different scanning principles, resulting in different anatomical locations as ONH margins [14

14. A. Manassakorn, H. Ishikawa, J. S. Kim, G. Wollstein, R. A. Bilonick, L. Kagemann, M. L. Gabriele, K. R. Sung, T. Mumcuoglu, J. S. Duker, J. G. Fujimoto, and J. S. Schuman, “Comparison of optic disc margin identified by color disc photography and high-speed ultrahigh-resolution optical coherence tomography,” Arch. Ophthalmol. 126(1), 58–64 (2008). [CrossRef] [PubMed]

].

The SD of disc area and cup volume with AP was statistically significantly smaller than with other three methods, indicating that AP of automated ONH margin detection and cup volume assessment provide more reproducible measurements, while cup/disc ratio measurements are more reproducible with the other three methods (Tables 2 and 3). The reason may be that all the other parameters are absolute values, while C/D area ratio and vertical C/D ratio are relative values. The relative value, such as vertical C/D ratio, is more sensitive to the exact locations and shape of the cup. It should be noted that manual intervention was present in all methods except for the AP, which might artificially affect the variability.

The limitation of this study was the substantial failure rate of the AP method (23.8%). This was due to the rigorous quality criteria that were employed for AP. The main causes for the failures were the presence of pathologic features such as peripapillary atrophy, which reduced the visibility of disc margin. It should be noted that all other methods used subjective input for their disc margin definition, which is also a viable option in cases where AP fails. Another limitation of the study was the use of stereo photographs that were acquired with the Nidek 3-DX camera only. However, minor modifications in the algorithm accounting for the specific properties of different cameras can be employed thus expanding the generalisability of this method. A small test sample obtained using another camera showed similar performance as those reported in this manuscript (unpublished preliminary data).

In summary, we present a photogrammetric method that provides a fully automated and objective volumetric quantification of the ONH. The method was comparable to other conventionally accepted methods of automated ONH analysis in term of measurement bias and imprecision although the measurements are not inter-changeable between modalities.

Acknowledgements

This research was supported by NIH grant R01-EY013178, P30-EY008098 (Bethesda, MD), Research to Prevent Blindness (New York, NY), Eye and Ear Foundation (Pittsburgh, PA), Ophthalmic Research Institute of Australia and NHMRC Medical Postgraduate and Practitioner Fellowship Scholarships.

References and links

1.

A. J. Bellezza, C. J. Rintalan, H. W. Thompson, J. C. Downs, R. T. Hart, and C. F. Burgoyne, “Deformation of the lamina cribrosa and anterior scleral canal wall in early experimental glaucoma,” Invest. Ophthalmol. Vis. Sci. 44(2), 623–637 (2003). [CrossRef] [PubMed]

2.

O. C. Holm, B. Becker, C. F. Asseff, and S. M. Podos, “Volume of the optic disk cup,” Am. J. Ophthalmol. 73(6), 876–881 (1972). [PubMed]

3.

J. Xu, O. Chutatape, C. Zheng, and P. C. Kuan, “Three dimensional optic disc visualisation from stereo images via dual registration and ocular media optical correction,” Br. J. Ophthalmol. 90(2), 181–185 (2006). [CrossRef] [PubMed]

4.

J. Xu, H. Ishikawa, G. Wollstein, R. A. Bilonick, K. R. Sung, L. Kagemann, K. A. Townsend, and J. S. Schuman, “Automated assessment of the optic nerve head on stereo disc photographs,” Invest. Ophthalmol. Vis. Sci. 49(6), 2512–2517 (2008). [CrossRef] [PubMed]

5.

D. F. Garway-Heath, A. R. Rudnicka, T. Lowe, P. J. Foster, F. W. Fitzke, and R. A. Hitchings, “Measurement of optic disc size: equivalence of methods to correct for ocular magnification,” Br. J. Ophthalmol. 82(6), 643–649 (1998). [CrossRef] [PubMed]

6.

J. Xu, O. Chutatape, and P. Chew, “Automated optic disk boundary detection by modified active contour model,” IEEE Trans. Biomed. Eng. 54(33 issue. 3), 473–482 (2007). [CrossRef] [PubMed]

7.

M. D. Abràmoff, W. L. M. Alward, E. C. Greenlee, L. Shuba, C. Y. Kim, J. H. Fingert, and Y. H. Kwon, “Automated segmentation of the optic disc from stereo color photographs using physiologically plausible features,” Invest. Ophthalmol. Vis. Sci. 48(4), 1665–1673 (2007). [CrossRef] [PubMed]

8.

M.C. Neale, S.M. Boker, G. Xie, H.H. Maes, Mx: Statistical Modeling 6th Ed.(VCU Box 900126, Richmond, VA 23298: Department of Psychiatry.)

9.

HRT tutorial. Quantitative three-dimensional imaging of the posterior segment with the Heidelberg Retina Tomograph. http://www.heidelbergengineering.com/hrtf/hrt.html

10.

G. J. Jaffe and J. Caprioli, “Optical coherence tomography to detect and manage retinal disease and glaucoma,” Am. J. Ophthalmol. 137(1), 156–169 (2004). [CrossRef] [PubMed]

11.

J. Caprioli, U. Klingbeil, M. Sears, and B. Pope, “Reproducibility of optic disc measurements with computerized analysis of stereoscopic video images,” Arch. Ophthalmol. 104(7), 1035–1039 (1986). [PubMed]

12.

R. Varma and G. L. Spaeth, “The PAR IS 2000: a new system for retinal digital image analysis,” Ophthalmic Surg. 19(3), 183–192 (1988). [PubMed]

13.

P. Hrynchak, N. Hutchings, D. Jones, and T. Simpson, “A comparison of cup-to-disc ratio measurement in normal subjects using optical coherence tomography image analysis of the optic nerve head and stereo fundus biomicroscopy,” Ophthalmic Physiol. Opt. 24(6), 543–550 (2004). [CrossRef] [PubMed]

14.

A. Manassakorn, H. Ishikawa, J. S. Kim, G. Wollstein, R. A. Bilonick, L. Kagemann, M. L. Gabriele, K. R. Sung, T. Mumcuoglu, J. S. Duker, J. G. Fujimoto, and J. S. Schuman, “Comparison of optic disc margin identified by color disc photography and high-speed ultrahigh-resolution optical coherence tomography,” Arch. Ophthalmol. 126(1), 58–64 (2008). [CrossRef] [PubMed]

15.

K. Deguchi, D. Kawamata, K. Mizutani, H. Hontani, and K. Wakabayashi, ““3D fundus shape reconstruction and display from stereo fundus images,” IEICE Trans. INF. & SYST,” E 83-D(7), (2000).

OCIS Codes
(100.0100) Image processing : Image processing
(100.2960) Image processing : Image analysis
(100.6890) Image processing : Three-dimensional image processing

ToC Category:
Image Processing

History
Original Manuscript: July 7, 2009
Revised Manuscript: October 1, 2009
Manuscript Accepted: October 11, 2009
Published: May 14, 2010

Virtual Issues
Vol. 5, Iss. 10 Virtual Journal for Biomedical Optics

Citation
Juan Xu, Hiroshi Ishikawa, Gadi Wollstein, Richard A Bilonick, Larry Kagemann, Jamie E Craig, David A Mackey, Alex W Hewitt, and Joel S Schuman, "Automated volumetric evaluation of stereoscopic disc photography," Opt. Express 18, 11347-11359 (2010)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-18-11-11347


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. A. J. Bellezza, C. J. Rintalan, H. W. Thompson, J. C. Downs, R. T. Hart, and C. F. Burgoyne, “Deformation of the lamina cribrosa and anterior scleral canal wall in early experimental glaucoma,” Invest. Ophthalmol. Vis. Sci. 44(2), 623–637 (2003). [CrossRef] [PubMed]
  2. O. C. Holm, B. Becker, C. F. Asseff, and S. M. Podos, “Volume of the optic disk cup,” Am. J. Ophthalmol. 73(6), 876–881 (1972). [PubMed]
  3. J. Xu, O. Chutatape, C. Zheng, and P. C. Kuan, “Three dimensional optic disc visualisation from stereo images via dual registration and ocular media optical correction,” Br. J. Ophthalmol. 90(2), 181–185 (2006). [CrossRef] [PubMed]
  4. J. Xu, H. Ishikawa, G. Wollstein, R. A. Bilonick, K. R. Sung, L. Kagemann, K. A. Townsend, and J. S. Schuman, “Automated assessment of the optic nerve head on stereo disc photographs,” Invest. Ophthalmol. Vis. Sci. 49(6), 2512–2517 (2008). [CrossRef] [PubMed]
  5. D. F. Garway-Heath, A. R. Rudnicka, T. Lowe, P. J. Foster, F. W. Fitzke, and R. A. Hitchings, “Measurement of optic disc size: equivalence of methods to correct for ocular magnification,” Br. J. Ophthalmol. 82(6), 643–649 (1998). [CrossRef] [PubMed]
  6. J. Xu, O. Chutatape, and P. Chew, “Automated optic disk boundary detection by modified active contour model,” IEEE Trans. Biomed. Eng. 54(33 issue. 3), 473–482 (2007). [CrossRef] [PubMed]
  7. M. D. Abràmoff, W. L. M. Alward, E. C. Greenlee, L. Shuba, C. Y. Kim, J. H. Fingert, and Y. H. Kwon, “Automated segmentation of the optic disc from stereo color photographs using physiologically plausible features,” Invest. Ophthalmol. Vis. Sci. 48(4), 1665–1673 (2007). [CrossRef] [PubMed]
  8. M.C. Neale, S.M. Boker, G. Xie, H.H. Maes, Mx: Statistical Modeling 6th Ed.(VCU Box 900126, Richmond, VA 23298: Department of Psychiatry.)
  9. HRT tutorial. Quantitative three-dimensional imaging of the posterior segment with the Heidelberg Retina Tomograph. http://www.heidelbergengineering.com/hrtf/hrt.html
  10. G. J. Jaffe and J. Caprioli, “Optical coherence tomography to detect and manage retinal disease and glaucoma,” Am. J. Ophthalmol. 137(1), 156–169 (2004). [CrossRef] [PubMed]
  11. J. Caprioli, U. Klingbeil, M. Sears, and B. Pope, “Reproducibility of optic disc measurements with computerized analysis of stereoscopic video images,” Arch. Ophthalmol. 104(7), 1035–1039 (1986). [PubMed]
  12. R. Varma and G. L. Spaeth, “The PAR IS 2000: a new system for retinal digital image analysis,” Ophthalmic Surg. 19(3), 183–192 (1988). [PubMed]
  13. P. Hrynchak, N. Hutchings, D. Jones, and T. Simpson, “A comparison of cup-to-disc ratio measurement in normal subjects using optical coherence tomography image analysis of the optic nerve head and stereo fundus biomicroscopy,” Ophthalmic Physiol. Opt. 24(6), 543–550 (2004). [CrossRef] [PubMed]
  14. A. Manassakorn, H. Ishikawa, J. S. Kim, G. Wollstein, R. A. Bilonick, L. Kagemann, M. L. Gabriele, K. R. Sung, T. Mumcuoglu, J. S. Duker, J. G. Fujimoto, and J. S. Schuman, “Comparison of optic disc margin identified by color disc photography and high-speed ultrahigh-resolution optical coherence tomography,” Arch. Ophthalmol. 126(1), 58–64 (2008). [CrossRef] [PubMed]
  15. K. Deguchi, D. Kawamata, K. Mizutani, H. Hontani, and K. Wakabayashi, ““3D fundus shape reconstruction and display from stereo fundus images,” IEICE Trans. INF. & SYST,” E 83-D(7), (2000).

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.


« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited