Biochemical sensing with a polymer-based micromachined Fabry-Perot sensor

Tianhua Zhang, Shantan Talla, Zhongcheng Gong, Sukrut Karandikar, Rebecca Giorno, and Long Que  »View Author Affiliations


Optics Express, Vol. 18, Issue 17, pp. 18394-18400 (2010)
http://dx.doi.org/10.1364/OE.18.018394


View Full Text Article

Acrobat PDF (899 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

A white-light source operated polymer-based micromachined Fabry-Perot biochemical sensor is reported. As a refractive-index sensitive optical sensor, its transducing signal varies upon the changes of the effective refractive index in the Fabry-Perot cavity. This sensor is fabricated from PDMS and glass. More specifically, this sensor is a micromachined Fabry-Perot interferometer (µFPI) and is fabricated by bonding a glass substrate and the soft-lithographically patterned PDMS. Several biochemicals have been detected with the µFPI biochemical sensors. Measurements show that rabbit IgG at a concentration of as low as 5 to 50 ng/ml can be detected even without any performance optimization of the devices.

© 2010 OSA

1. Introduction

The utilization of an optical technique for sensing applications offers some attractive features over other techniques such as high degree of sensitivity, capability of multiplexing and immunity to environment noise. With the advancement of micro and nanofabrication technologies, many optical components or systems such as lenses, gratings, mirrors, micro ring resonators and interferometers have been miniaturized [1

1. Z. D. Popovic, R. A. Sprague, and G. A. Connell, “Technique for monolithic fabrication of microlens arrays,” Appl. Opt. 27(7), 1281–1284 (1988). [CrossRef] [PubMed]

5

5. C. Wang, B. Wherrett, and T. Harvey, “Fabrication and characterization of a 4×4 array of asymmetric Fabry-Perot reflection modulators,” Electron. Lett. 30(15), 1219–1220 (1994). [CrossRef]

]. As a result, the field of microoptics and nanooptics has emerged, triggering extensive research to scale down the optics-based sensing platforms for the past decades. One example is the miniaturization of Fabry-Perot interferometers. Micromachined Fabry-Perot interferometers (µFPIs) or optical fiber based miniature FPIs have been designed and implemented for chemical sensing, gas sensing, biosensing, ultrasonic sensing and optical modulation [5

5. C. Wang, B. Wherrett, and T. Harvey, “Fabrication and characterization of a 4×4 array of asymmetric Fabry-Perot reflection modulators,” Electron. Lett. 30(15), 1219–1220 (1994). [CrossRef]

16

16. J. Liu, Y. Sun, and X. Fan, “Highly versatile fiber-based optical Fabry-Pérot gas sensor,” Opt. Express 17(4), 2731–2738 (2009). [CrossRef] [PubMed]

]. For chemical and gas sensing, the output signals are transmitted intensity of light through FPI, which vary with different chemicals or gases inside the FPI cavity [7

7. J. Han, “Fabry-Perot cavity chemical sensors by silicon micromachining techniques,” Appl. Phys. Lett. 74(3), 445–447 (1999). [CrossRef]

9

9. M. Blomberg, O. Rusanen, and K. Keranen, “A silicon microsystem-miniaturized infrared spectrometer,” Proc. 9th Int. Conf. On Solid-state Sensors, Actuators and Microsystems, 1257–1258 (1997)

]. For ultrasonic sensing, the micromachined capacitive acoustic transducer consists of a FPI cavity with embedded optical diffraction gratings. Optimization of the detection sensitivity can be achieved by deflecting the membrane of the FPI [6

6. N. Hall and F. Degertekin, “Integrated optical interferometric detection method for micromachined capacitive acoustic transducers,” Appl. Phys. Lett. 80(20), 3859–3861 (2002). [CrossRef]

]. For optical modulation, a FPI contains a crosslinked electro-optic polymer inside its cavity, offering the potential for high time-bandwidth modulation [5

5. C. Wang, B. Wherrett, and T. Harvey, “Fabrication and characterization of a 4×4 array of asymmetric Fabry-Perot reflection modulators,” Electron. Lett. 30(15), 1219–1220 (1994). [CrossRef]

]. µFPI has also been utilized for studying the nanoscale fluidic dynamics [14

14. K. M. van Delft, J. C. Eijkel, D. Mijatovic, T. S. Druzhinina, H. Rathgen, N. R. Tas, A. van den Berg, and F. Mugele, “Micromachined Fabry-Pérot interferometer with embedded nanochannels for nanoscale fluid dynamics,” Nano Lett. 7(2), 345–350 (2007). [CrossRef] [PubMed]

], indicating its potential for nanoscience and nanotechnology research. Use of MEMS actuation technology, tunable µFPIs have been designed and implemented successfully, which have significantly expanded and enriched the applications of FPIs. Specifically, tunable µFPIs have been developed for wavelength division multiplexing (WDM) in optical communications [8

8. L. Que, “Two-dimensional tunable filter array for a matrix of integrated fiber optic input-output light channels”, US Patent 6,449,410 (2002)

] and Raman spectroscopy [11

11. L. Que, A. Zribi, A. Banerjee, and D. Hays, “Raman system on a chip,” US Patent 7,505,128 (2009)

], oxygen detection in blood samples [12

12. T. Dohi, K. Matsumoto, and I. Shimoyama, “The optical blood test device with the micro Fabry-Perot interferometer”, Proceedings of IEEE MEMS, 403–406 (2004)

] and spectral endoscope optical imaging [13

13. T. Dohi, K. Matsumoto, and I. Shimoyama, “The micro Fabry-Perot interferometer for the spectral endoscope”, Proceedings of IEEE MEMS, 830–833 (2005)

]. For WDM applications, a tunable µFPI is used as a tunable filter, selecting different wavelengths by changing the FPI cavity length. For oxygen detection in blood samples, the tunable µFPI device is used for a broad wavelength range from visible to near infrared light, offering sufficient characteristics to analyze the spectrum of the blood. For spectral endoscope optical imaging, the tunable device is used for acquisition of spectral images of a target at each pixel. The structural materials for most of the reported µFPIs, which are mainly operated using a laser source, are silicon, polysilicon, silicon nitride, silicon oxide thin film or other semiconductor materials [5

5. C. Wang, B. Wherrett, and T. Harvey, “Fabrication and characterization of a 4×4 array of asymmetric Fabry-Perot reflection modulators,” Electron. Lett. 30(15), 1219–1220 (1994). [CrossRef]

14

14. K. M. van Delft, J. C. Eijkel, D. Mijatovic, T. S. Druzhinina, H. Rathgen, N. R. Tas, A. van den Berg, and F. Mugele, “Micromachined Fabry-Pérot interferometer with embedded nanochannels for nanoscale fluid dynamics,” Nano Lett. 7(2), 345–350 (2007). [CrossRef] [PubMed]

]. In this article, we report a polymer-based µFPI biochemical sensor operated with a white-light source.

Fig. 1 (A) Schematic of a PDMS-based µFPI biochemical sensor, showing immobilized probes (Protein A) on the glass surface and the binding between Immunoglobulin G (IgG) antibodies and probes; (B) Operational principle of the µFPI: the reflected light from the µFPI as the output transducing optical signals

2. Operation principle of the device

Figure 1 shows the schematic and operational principle of a white-light source operated µFPI biochemical sensor. This sensor is micromachined from polydimethylsiloxane (PDMS) and glass. The µFPI cavity serves as the sensing region with a gap size of 50 µm. Different from a traditional macroscale FPI or other reported µFPIs, neither of the FPI plates has been coated with any thin films for the enhancement of their reflectivity. The operation principle of this sensor is described as the following: Upon entering the FPI cavity, the operation light undergoes multiple internal reflections between the two FPI plates and interferences inside the cavity. As a result, modulated output transducing signals such as reflected interference fringes are generated as show in Fig. 1(B) and Fig. 2(A). More specifically, Protein A molecules serve as the probes and are immobilized on one FPI plate surface (Fig. 1(A)). When a variety of Immunoglobulin G (IgG) antibodies are flown through the cavity, they bind to the Protein A, causing changes of the effective index of refraction, resulting in the interference fringes shift (Δλ) (Fig. 2(A)), which serves as the optical transducing signals.

Fig. 2 (A) A sketch showing the inference fringe shift (Δλ) due to the effective index of refraction changes in the FPI cavity with the binding between Protein A and porcine IgG; (B) Testing setup: a custom-designed optical fiber probe (Ocean Optics, Inc) consists of a tight bundle of 7 optical fibers in a stainless steel ferrule. The center fiber is to collect the reflected light while the outer 6 fibers deliver illumination light to the biochemical sensor

Figure 2(B) gives the sketch of the testing setup for the µFPI biochemical sensor. A custom-designed optical fiber probe (Ocean Optics, Inc.) delivers white light from a tungsten halogen source (Ocean Optics, Inc.) perpendicularly to a biochemical sensor. This probe also receives the reflected signals from the sensor, which is eventually coupled to a HR4000 Spectrometer (Ocean Optics, Inc.).

3. Device design, modeling and fabrication

Figure 3
Fig. 3 (A) Sketch of the optical system including the optical fibers and the µFPI device: the indexes of refraction of PDMS, medium in FPI cavity and glass are n1, n2 and n3; (B) Based on Eq. (1) and (2), the calculated matches the measured transducing optical signals from the biochemical sensor
shows the sketch of the optical system consisting of an optical fiber and a µFPI for the modeling. The white light delivered from the optical fiber to the sensor can be modeled as a Gaussian beam [17

17. M. Born, and E. Wolf, Principal of Optics (John Wiley & Sons, Inc., 2000)

19

19. T. Zhang, Z. Gong, and L. Que, “A white-light source operated polymer-based micromachined Fabry-Perot chemo/biosensor,” Proc. of IEEE Intl. Conf. on NEMS, 177–180 (2009)

]. Due to the fiber core finite size of the optical fiber probe, the loss of the optical coupling between the fiber and µFPI sensor needs to be considered.

Assume the optical power loss coefficient of µFPI cavity is LFPI and the coupling loss between the fiber probe and µFPI is Lfiber-FPI, then the reflected intensity Ir from this system, coupled to the spectrometer, is written as:
Ir=Iiexp(2(λλ0)2ω2)×LfiberFPI×f(RairPDMS,Rairglass)
(1)
Where
f(RairPDMS,Rairglass)=r22+r32LFPI22r2r3LFPIcos(2kn2d)1r22r32LFPI22r2r3LFPIcos(2kn2d)
(2)
andr2=r1=n2n1n2+n1, r3=n2n3n2+n3

Where Iiexp(−2(λ-λ0)22) is incident light intensity; λ0 is the center wavelength of the light source, ω is the beam radius at the beam waist; f(Rair-PDMS, Rair-glass) is the reflectivity from the µFPI; Rair-PDMS, Rair-glass are the reflectivity at the interface between the air and PDMS, the air and glass, respectively and d is the gap size of FPI cavity; n1, n2 and n3 are the indices of refraction of PDMS, medium in FPI cavity and glass, respectively.

The fabrication process flow of the devices utilizing an inexpensive and rapid soft lithography process is shown in Fig. 4 (A). Briefly, using conventional optical lithography a 50 µm thick SU8 mold of the device is fabricated on a silicon substrate. PDMS is then casted on the mold, followed by 1.5 hour curing at the temperature of 65 °C. The patterned PDMS layer and the glass substrate are bonded together after oxygen plasma treatment, followed by the formation of the input and output wells in the PDMS layer. Optical micrographs of SU8 molds, fabricated PDMS devices and assembled devices with tubing are given in Fig. 4(B-F). Note that the arrayed devices can be readily and rapidly batch-fabricated, offering a cost-effective and disposable biosensing technical platform with high throughput.

As an example for the validation of the optical model described above, the measurement of the output signals from a µFPI device with air inside its cavity has been performed. The calculated signals with air in the FPI cavity match the measured signals very well as shown in Fig. 3 (B). The intensity profile offset of the calculated results from the measured results is mainly due to the fact that the light source is not an ideal Gaussian beam.

4. Experimental results and discussions

A protein-protein binding assay has been used to demonstrate the operation and biomolecular detection capability of the polymer-based µFPI at room temperature. The detailed experimental materials and procedure are described in the following.

The assay reagents used in the experiments include Protein A (Pierce Biotechnology), buffer solution phosphate buffered saline (PBS) (Sigma-Aldrich), Sea Block (Pierce Biotechnology), and porcine, rabbit, or sheep immunoglobulin G (IgG) (Sigma-Aldrich).

Specifically, Protein A is prepared in PBS solution at a concentration of 500µg/ml, which can be diluted accordingly for different concentrations. The blocking buffer, Sea Block, used to minimize the non-specific binding is mixed with PBS solution with a ratio of 1:4. Similarly, the porcine, rabbit, sheep IgG anti-sera are diluted with PBS to achieve final concentrations of 500 µg/ml, 500 ng/ml, 50 ng/ml, and 5 ng/ml, for each species respectively.

As a reference, the first measurement is taken with air inside the FPI cavity. The PBS solution is flowed into the cavity with a flow rate of 0.5µl/minute the slowest possible flow rate to avoid any deformation of the PDMS plate of the FPI. In addition, care is taken to eliminate any air bubbles inside the cavity throughout the experiment. The measurement is
Fig. 4 (A) The fabrication process flow: (a) form the 50µm tall SU8 mold on the silicon wafer; (b) transfer the mold to PDMS; (c) bond patterned PDMS to glass wafer after plasma treatment, and then form the input /output wells; (B) Optical micrograph of SU8 mold for a PDMS-FPI with integrated microfluidic channel to deliver chemicals to the FPI cavity; (C) Optical micrograph of 2 × 2 PDMS-FPIs; (D-E) Photos of one SU8 mold and one assembled PDMS-FPI biochemical sensor; (F) A photo of arrayed PDMS-FPI biochemical sensors
taken with PBS solution, which is used to calibrate the cavity gap based on the measured index of refraction of PBS being 1.336 [20

20. S. Zangooi, R. Bjorklund, and H. Arwin, “Protein adsorption in thermally oxidized porous silicon layers,” Thin Solid Films 313–314(1-2), 825–830 (1998). [CrossRef]

]. Thereafter, the Protein A solution (500µg/ml) is slowly passed into the device with flow rate of 0.5µl/minute and allowed to stay inside the FPI cavity for 30 minutes at room temperature. This will allow the Protein A to bind to the glass surface and serve as probes. The unbound Protein A is washed away using PBS solution. The washing routine is done for at least 3 cycles of 3 minutes each. The measurement is taken at each interval and checked for measurement repeatability. The Sea Block solution is passed into the device with the same flow rate to block the unmodified or unbound sites in the FPI cavity. The Sea Block solution is allowed to stay inside FPI cavity for 15 minutes at room temperature. After the specified time the Sea block solution is pumped out and the wash routine is performed again to take some measurements. Similarly the rabbit IgG solution made with PBS solution is passed into the device and allowed it settle for 15 to 30 minutes in the FPI cavity at room temperature. The measurements are taken at regular intervals after washing the excess unbound rabbit IgG using PBS solution. Similarly the same procedure is repeated with porcine and sheep IgGs for different concentrations and measurements are carried out. Note that all the measurements are performed after the fluidic channel has become dried.

As examples, the measured transducing signals for PBS, immobilized Protein A, Protein A bound with porcine IgG and Protein A bound with rabbit IgG are shown in Fig. 5(A-B). The shift in interference fringes is clearly observed. In Fig. 5(C), the transducing signals of Protein A and its binding with porcine, sheep and rabbit IgGs are plotted together. Again the fringe shift is clearly visible, indicating the difference of the effective of refraction indices of these three different types of IgGs. In Table 1

Table 1. Fringe peak shift relative to air for Protein A and different IgG samples

table-icon
View This Table
| View All Tables
, the average fringe peak shifts relative to the air inside the FPI cavity for different cases are summarized. The average shift for each case is obtained by averaging the shift of the fringe peaks between the 600 nm –720 nm spectrum window. The fringe peaks are identified based on the fitting curves using the proposed mathematical model, which also facilitates the determination of the peak shift larger than free spectral range (FSR), which is around 3–4 nm for this µFPI device. Measurement error is ± 0.01 nm due to the resolution of the HR4000 Spectrometer is 0.02 nm. Note that the concentrations of Protein A, porcine, rabbit and sheep IgGs are all 500 µg/ml. Use of the index of refraction of PBS 1.336, the FPI cavity length is calculated to be 48 µm, which is 2 µm offset from the nominal length of the design. This reduction of the cavity length is probably due to the thickness change of the SU8 mold during the microfabrication process. Based on the FPI cavity calibration with PBS solution, using similar calculation method, the effective refractive indices of Protein A and Protein A bound with rabbit IgG at concentration of 50 ng/ml are 1.521 and 1.550, respectively. Note that the effective refractive indices for Protein A bound with different types of IgG or the same IgG but at varied concentrations are different.

In Table 2

Table 2. Fringe peak shift relative to air for rabbit IgG samples with varied concentrations

table-icon
View This Table
| View All Tables
, the fringe shift for rabbit IgG with varied concentrations is summarized. The higher the concentration of the rabbit IgG, the more binding between Protein A and rabbit IgG occurs, resulting in higher effective index of refraction changes thus bigger effective optical thickness (EOT) changes inside the FPI cavity. Hence a greater shift of the fringe peaks is achieved. In other words, as the concentration of rabbit IgG decreases, the shift also decreases. It is observed that when the concentration of rabbit IgG is reduced to 5 ng/ml, the fringe shift relative to the air inside the FP cavity is 0.68 nm, the same as that (0.69 nm) of Protein A without binding to rabbit IgG, suggesting the possible detection limit (DL) to be between 50 ng/ml and 5 ng/ml for this FPI biochemical sensor before optimization.

Fig. 5 (A-B) Measured signals for PBS, immobilized Protein A (500 µg/ml), Protein A bound with porcine IgG (500 µg/ml) and protein A bound with rabbit IgG (500 µg/ml); (C) Measured transducing signals for immobilized Protein A (500 µg/ml), Protein A bound with porcine IgG (500 µg/ml), rabbit IgG (500 µg/ml)and sheep IgG (500 µg/ml) respectively, showing clear shift of the interference fringes

5. Conclusion

In summary, a polymer-based micromachined FPI biochemical sensor has been developed and its biomolecular detection capability has been demonstrated successfully. The simple, cost-effective and disposable nature of this type of sensor is attractive for point-of-care biodetection applications. The experiments show that detection limit (DL) of this sensor is between 50 and 5 ng/ml for the rabbit IgG. It is highly possible to enhance the sensitivity and DL significantly by introducing some nanostructures inside the FPI cavity [21

21. T. Zhang, Z. Gong, R. Giorno, and L. Que, “Signal sensitivity and intensity enhancement for a polymer-based Fabry-Perot interferometer with embedded nanostructures in its cavity,” Proceeding of 15th International Conference on Solid-State Sensors, Actuators & Microsystems (Transducers'09), 2310–2313 (2009)

,22

22. J. Zhao, X. Zhang, C. R. Yonzon, A. J. Haes, and R. P. Van Duyne, “Localized surface plasmon resonance biosensors,” Nanomedicine (Lond) 1(2), 219–228 (2006). [CrossRef]

].

Acknowledgement

This research is supported in part by grants of NSF CAREER Award 2009 and NSF 2008 PFUND-110. The authors thank the technical support from the technical staffs at Institute for Micromanufacturing at Louisiana Tech University

References and links

1.

Z. D. Popovic, R. A. Sprague, and G. A. Connell, “Technique for monolithic fabrication of microlens arrays,” Appl. Opt. 27(7), 1281–1284 (1988). [CrossRef] [PubMed]

2.

C. H. Lee, H. Yoshida, Y. Miura, A. Fujii, and M. Ozaki, ““Fabrication of micro-grating structures by direct laser writing based on two photon process and their liquid crystal alignment abilities,” IEICE Transactions on Electronics,” E 91-C(10), 1581–1586 (2008).

3.

L. Hornbeck, “Deformable-mirror spatial light modulators,” Proc. SPIE, Spatial Light Modulators and Applications III, v.1150, 86–102 (1990)

4.

Q. Xu and M. Lipson, “All-optical logic based on silicon micro-ring resonators,” Opt. Express 15(3), 924–929 (2007). [CrossRef] [PubMed]

5.

C. Wang, B. Wherrett, and T. Harvey, “Fabrication and characterization of a 4×4 array of asymmetric Fabry-Perot reflection modulators,” Electron. Lett. 30(15), 1219–1220 (1994). [CrossRef]

6.

N. Hall and F. Degertekin, “Integrated optical interferometric detection method for micromachined capacitive acoustic transducers,” Appl. Phys. Lett. 80(20), 3859–3861 (2002). [CrossRef]

7.

J. Han, “Fabry-Perot cavity chemical sensors by silicon micromachining techniques,” Appl. Phys. Lett. 74(3), 445–447 (1999). [CrossRef]

8.

L. Que, “Two-dimensional tunable filter array for a matrix of integrated fiber optic input-output light channels”, US Patent 6,449,410 (2002)

9.

M. Blomberg, O. Rusanen, and K. Keranen, “A silicon microsystem-miniaturized infrared spectrometer,” Proc. 9th Int. Conf. On Solid-state Sensors, Actuators and Microsystems, 1257–1258 (1997)

10.

N. Neumann, M. Ebermann, K. Hiller, and S. Kurth, “Tunable infrared detector with integrated micromachined Fabry-Perot filter,” Proceedings of the SPIE, v. 6466, 646606–12 (2007)

11.

L. Que, A. Zribi, A. Banerjee, and D. Hays, “Raman system on a chip,” US Patent 7,505,128 (2009)

12.

T. Dohi, K. Matsumoto, and I. Shimoyama, “The optical blood test device with the micro Fabry-Perot interferometer”, Proceedings of IEEE MEMS, 403–406 (2004)

13.

T. Dohi, K. Matsumoto, and I. Shimoyama, “The micro Fabry-Perot interferometer for the spectral endoscope”, Proceedings of IEEE MEMS, 830–833 (2005)

14.

K. M. van Delft, J. C. Eijkel, D. Mijatovic, T. S. Druzhinina, H. Rathgen, N. R. Tas, A. van den Berg, and F. Mugele, “Micromachined Fabry-Pérot interferometer with embedded nanochannels for nanoscale fluid dynamics,” Nano Lett. 7(2), 345–350 (2007). [CrossRef] [PubMed]

15.

Y. Zhang, H. Shibru, K. L. Cooper, and A. Wang, “Miniature fiber-optic multicavity Fabry-Perot interferometric biosensor,” Opt. Lett. 30(9), 1021–1023 (2005). [CrossRef] [PubMed]

16.

J. Liu, Y. Sun, and X. Fan, “Highly versatile fiber-based optical Fabry-Pérot gas sensor,” Opt. Express 17(4), 2731–2738 (2009). [CrossRef] [PubMed]

17.

M. Born, and E. Wolf, Principal of Optics (John Wiley & Sons, Inc., 2000)

18.

S. Jiang, B. Zeng, Y. Liang, and B. Li, “Optical fiber sensor for tensile and compressive strain measurement by white-light Faby-Perot interferometry,” Opt. Eng. 46(3), 034402 (2007). [CrossRef]

19.

T. Zhang, Z. Gong, and L. Que, “A white-light source operated polymer-based micromachined Fabry-Perot chemo/biosensor,” Proc. of IEEE Intl. Conf. on NEMS, 177–180 (2009)

20.

S. Zangooi, R. Bjorklund, and H. Arwin, “Protein adsorption in thermally oxidized porous silicon layers,” Thin Solid Films 313–314(1-2), 825–830 (1998). [CrossRef]

21.

T. Zhang, Z. Gong, R. Giorno, and L. Que, “Signal sensitivity and intensity enhancement for a polymer-based Fabry-Perot interferometer with embedded nanostructures in its cavity,” Proceeding of 15th International Conference on Solid-State Sensors, Actuators & Microsystems (Transducers'09), 2310–2313 (2009)

22.

J. Zhao, X. Zhang, C. R. Yonzon, A. J. Haes, and R. P. Van Duyne, “Localized surface plasmon resonance biosensors,” Nanomedicine (Lond) 1(2), 219–228 (2006). [CrossRef]

OCIS Codes
(050.2230) Diffraction and gratings : Fabry-Perot
(130.6010) Integrated optics : Sensors
(230.3990) Optical devices : Micro-optical devices

ToC Category:
Sensors

History
Original Manuscript: May 18, 2010
Revised Manuscript: June 28, 2010
Manuscript Accepted: August 2, 2010
Published: August 12, 2010

Virtual Issues
Vol. 5, Iss. 13 Virtual Journal for Biomedical Optics

Citation
Tianhua Zhang, Shantan Talla, Zhongcheng Gong, Sukrut Karandikar, Rebecca Giorno, and Long Que, "Biochemical sensing with a polymer-based micromachined Fabry-Perot sensor," Opt. Express 18, 18394-18400 (2010)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-18-17-18394


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. Z. D. Popovic, R. A. Sprague, and G. A. Connell, “Technique for monolithic fabrication of microlens arrays,” Appl. Opt. 27(7), 1281–1284 (1988). [CrossRef] [PubMed]
  2. C. H. Lee, H. Yoshida, Y. Miura, A. Fujii, and M. Ozaki, ““Fabrication of micro-grating structures by direct laser writing based on two photon process and their liquid crystal alignment abilities,” IEICE Transactions on Electronics,” E 91-C(10), 1581–1586 (2008).
  3. L. Hornbeck, “Deformable-mirror spatial light modulators,” Proc. SPIE, Spatial Light Modulators and Applications III, v.1150, 86–102 (1990)
  4. Q. Xu and M. Lipson, “All-optical logic based on silicon micro-ring resonators,” Opt. Express 15(3), 924–929 (2007). [CrossRef] [PubMed]
  5. C. Wang, B. Wherrett, and T. Harvey, “Fabrication and characterization of a 4×4 array of asymmetric Fabry-Perot reflection modulators,” Electron. Lett. 30(15), 1219–1220 (1994). [CrossRef]
  6. N. Hall and F. Degertekin, “Integrated optical interferometric detection method for micromachined capacitive acoustic transducers,” Appl. Phys. Lett. 80(20), 3859–3861 (2002). [CrossRef]
  7. J. Han, “Fabry-Perot cavity chemical sensors by silicon micromachining techniques,” Appl. Phys. Lett. 74(3), 445–447 (1999). [CrossRef]
  8. L. Que, “Two-dimensional tunable filter array for a matrix of integrated fiber optic input-output light channels”, US Patent 6,449,410 (2002)
  9. M. Blomberg, O. Rusanen, and K. Keranen, “A silicon microsystem-miniaturized infrared spectrometer,” Proc. 9th Int. Conf. On Solid-state Sensors, Actuators and Microsystems, 1257–1258 (1997)
  10. N. Neumann, M. Ebermann, K. Hiller, and S. Kurth, “Tunable infrared detector with integrated micromachined Fabry-Perot filter,” Proceedings of the SPIE, v. 6466, 646606–12 (2007)
  11. L. Que, A. Zribi, A. Banerjee, and D. Hays, “Raman system on a chip,” US Patent 7,505,128 (2009)
  12. T. Dohi, K. Matsumoto, and I. Shimoyama, “The optical blood test device with the micro Fabry-Perot interferometer”, Proceedings of IEEE MEMS, 403–406 (2004)
  13. T. Dohi, K. Matsumoto, and I. Shimoyama, “The micro Fabry-Perot interferometer for the spectral endoscope”, Proceedings of IEEE MEMS, 830–833 (2005)
  14. K. M. van Delft, J. C. Eijkel, D. Mijatovic, T. S. Druzhinina, H. Rathgen, N. R. Tas, A. van den Berg, and F. Mugele, “Micromachined Fabry-Pérot interferometer with embedded nanochannels for nanoscale fluid dynamics,” Nano Lett. 7(2), 345–350 (2007). [CrossRef] [PubMed]
  15. Y. Zhang, H. Shibru, K. L. Cooper, and A. Wang, “Miniature fiber-optic multicavity Fabry-Perot interferometric biosensor,” Opt. Lett. 30(9), 1021–1023 (2005). [CrossRef] [PubMed]
  16. J. Liu, Y. Sun, and X. Fan, “Highly versatile fiber-based optical Fabry-Pérot gas sensor,” Opt. Express 17(4), 2731–2738 (2009). [CrossRef] [PubMed]
  17. M. Born, and E. Wolf, Principal of Optics (John Wiley & Sons, Inc., 2000)
  18. S. Jiang, B. Zeng, Y. Liang, and B. Li, “Optical fiber sensor for tensile and compressive strain measurement by white-light Faby-Perot interferometry,” Opt. Eng. 46(3), 034402 (2007). [CrossRef]
  19. T. Zhang, Z. Gong, and L. Que, “A white-light source operated polymer-based micromachined Fabry-Perot chemo/biosensor,” Proc. of IEEE Intl. Conf. on NEMS, 177–180 (2009)
  20. S. Zangooi, R. Bjorklund, and H. Arwin, “Protein adsorption in thermally oxidized porous silicon layers,” Thin Solid Films 313–314(1-2), 825–830 (1998). [CrossRef]
  21. T. Zhang, Z. Gong, R. Giorno, and L. Que, “Signal sensitivity and intensity enhancement for a polymer-based Fabry-Perot interferometer with embedded nanostructures in its cavity,” Proceeding of 15th International Conference on Solid-State Sensors, Actuators & Microsystems (Transducers'09), 2310–2313 (2009)
  22. J. Zhao, X. Zhang, C. R. Yonzon, A. J. Haes, and R. P. Van Duyne, “Localized surface plasmon resonance biosensors,” Nanomedicine (Lond) 1(2), 219–228 (2006). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1 Fig. 2 Fig. 3
 
Fig. 4 Fig. 5
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited