Adaptive optics photoacoustic microscopy

Minshan Jiang, Xiangyang Zhang, Carmen A. Puliafito, Hao F. Zhang, and Shuliang Jiao  »View Author Affiliations


Optics Express, Vol. 18, Issue 21, pp. 21770-21776 (2010)
http://dx.doi.org/10.1364/OE.18.021770


View Full Text Article

Acrobat PDF (4417 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We have developed an adaptive optics photoacoustic microscope (AO-PAM) for high-resolution imaging of biological tissues, especially the retina. To demonstrate the feasibility of AO-PAM we first designed the AO system to correct the wavefront errors of the illuminating light of PAM. The aberrations of the optical system delivering the illuminating light to the sample in PAM was corrected with a close-loop AO system consisting of a 141-element MEMS-based deformable mirror (DM) and a Shack-Hartmann (SH) wavefront sensor operating at 15 Hz. The photoacoustic signal induced by the illuminating laser beam was detected by a custom-built needle ultrasonic transducer. When the wavefront errors were corrected by the AO system, the lateral resolution of PAM was measured to be better than 2.5 µm using a low NA objective lens. We tested the system on imaging ex vivo ocular samples, e.g., the ciliary body and retinal pigment epithelium (RPE) of a pig eye. The AO-PAM images showed significant quality improvement. For the first time we were able to resolve single RPE cells with PAM.

© 2010 OSA

1. Introduction

The axial resolution of PAOM is determined by the bandwidth and center frequency of the ultrasonic transducer. The lateral resolution of PAOM is determined by the smallest achievable spot size of the illuminating light on the retina. In imaging the eye in vivo, for large pupil diameters, monochromatic aberrations of the eye are the major factors that limit the achievable resolution of retinal images [11

11. P. Artal, A. Guirao, E. Berrio, and D. R. Williams, “Compensation of corneal aberrations by the internal optics in the human eye,” J. Vis. 1(1), 1–8 (2001). [CrossRef]

]. Although the depth resolution of PAOM could be improved by increasing the bandwidth of the ultrasonic detecting system, the lateral resolution of PAOM is still limited by the optics. One possible approach for correcting the ocular aberrations to decrease the spot size of the illuminating light on the retina, and therefore improve the lateral resolution of PAOM, is to use adaptive optics (AO) [12

12. J. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef]

,13

13. E. J. Fernández, I. Iglesias, and P. Artal, “Closed-loop adaptive optics in the human eye,” Opt. Lett. 26(10), 746–748 (2001). [CrossRef]

]. AO has been applied to ophthalmic imaging with great success, which makes it possible to image single photo-receptors [14

14. K. Y. Li and A. Roorda, “Automated identification of cone photoreceptors in adaptive optics retinal images,” J. Opt. Soc. Am. A 24(5), 1358–1363 (2007). [CrossRef]

17

17. T. Y. Chui, H. Song, and S. A. Burns, “Adaptive-optics imaging of human cone photoreceptor distribution,” J. Opt. Soc. Am. A 25(12), 3021–3029 (2008). [CrossRef]

]. AO has also been successfully integrated in OCT [18

18. B. Hermann, E. J. Fernández, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, P. M. Prieto, and P. Artal, “Adaptive-optics ultrahigh-resolution optical coherence tomography,” Opt. Lett. 29(18), 2142–2144 (2004). [CrossRef] [PubMed]

23

23. K. Kurokawa, K. Sasaki, S. Makita, M. Yamanari, B. Cense, and Y. Yasuno, “Simultaneous high-resolution retinal imaging and high-penetration choroidal imaging by one-micrometer adaptive optics optical coherence tomography,” Opt. Express 18(8), 8515–8527 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-8-8515. [CrossRef] [PubMed]

], scanning laser ophthalmoscopy (SLO) [24

24. A. Roorda, F. Romero-Borja, W. Donnelly Iii, H. Queener, T. Hebert, and M. Campbell, “Adaptive optics scanning laser ophthalmoscopy,” Opt. Express 10(9), 405–412 (2002), http://www.opticsinfobase.org/abstract.cfm?URI=OE-10-9-405. [PubMed]

28

28. S. A. Burns, R. Tumbar, A. E. Elsner, D. Ferguson, and D. X. Hammer, “Large-field-of-view, modular, stabilized, adaptive-optics-based scanning laser ophthalmoscope,” J. Opt. Soc. Am. A 24(5), 1313–1326 (2007). [CrossRef]

], and microscopy [29

29. L. Sherman, J. Y. Ye, O. Albert, and T. B. Norris, “Adaptive correction of depth-induced aberrations in multiphoton scanning microscopy using a deformable mirror,” J. Microsc. 206(1), 65–71 (2002). [CrossRef] [PubMed]

33

33. J. W. Cha, J. Ballesta, and P. T. C. So, “Shack-Hartmann wavefront-sensor-based adaptive optics system for multiphoton microscopy,” J. Biomed. Opt. 15(4), 046022 (2010). [CrossRef] [PubMed]

].

In this paper we present a feasibility study on the integration of AO with PAM for ex vivo imaging of ocular tissues, which is a technical preparation for the ultimate AO-PAOM for in vivo retinal imaging. The goal of the current study is to correct the aberrations of the optical system for delivering the illuminating light to the sample.

2. Experimental system

Figure 1
Fig. 1 Schematic of the experimental AO-PAM system. ND: neutral density filter; PD: photodiode; D: diaphragm; L1-L4: achromatic lenses; DM: deformable mirror; SH: Shack-Hartmann wavefront sensor; BS: pellicle beamsplitter (reflection: 8%, transmission 92%); UT: ultrasonic transducer.
shows a schematic of the AO-PAM experimental system. A compact close-loop AO system was integrated into a LSOR-PAM. The AO system consists of a real-time Shack-Hartmann wavefront sensor (SH sensor, 39x31 maximum active lenslets, Thorlabs, USA) operating at 15Hz and a 141-element MEMS-based deformable mirror (DM, Thorlabs, USA).

The illumination light source of the PAM was a frequency-doubled Q-switched Nd:YAG laser (SPOT-10-100-532, Elforlight Ltd, UK) with the following parameters: wavelength = 532nm, pulse energy = 10 µJ/pulse, pulse duration = 2 ns, and maximal pulse repetition rate = 30 kHz. The output laser light was attenuated with a neutral density (ND) filter before coupled into a single mode optical fiber (SMF). The light reflected from the surface of the ND filter is detected by a photodiode (DET10A, Thorlabs, Inc.), which provided trigger for PAM signal acquisition. The output laser light from the SMF was first collimated into a beam diameter of 4.4 mm, passed through a two-f system (L1 and L2, f = 75mm), and was then projected onto the DM (Clear aperture: 4.4mm x 4.4mm). Another two-f system (L3 and L4, f = 100 mm and f = 225 mm, respectively) expanded the beam reflected from the DM into 10mm in diameter and projected the beam onto an X-Y galvanometer scanner. The X-Y galvanometer, DM, and SH were conjugated with each other. The light beam was finally focused on the sample by an objective lens (f = 30 mm, Φ = 25.4 mm, Achromatic Lens, VIS-NIR, Edmund Optics, Inc). The galvanometer scanner was controlled by an analogue-output board, whose sample clock was used to trigger the laser.

To avoid speckles formed by the interference of the reflected light from the sample, a superluminescent diode (SLD, power = 10 mW, bandwidth = 50nm, central wavelength = 840nm, Superlum, Russia) was used as the AO light source. The AO beam was first collimated (beam diameter < 1 mm) and then coupled into the center line of the PAM illuminating light path by a pellicle beam-splitter (BP-108, Uncoated, Throlabs, USA), which reflects 8% of the AO light onto the objective lens while transmitting 92% of the back-scattered light from the sample. The diffusely reflected light from the sample could be considered to have a spherical wavefront emitted from a point light source. This wavefront passed through the DM and was delivered into the SH by another pellicle beam-splitter. Therefore, the AO beam travelled through the same path as the PA laser beam illuminating the samples. The DM consists of a continuous mirror membrane that is deformed by electrostatic actuators ensuring deformation without hysteresis. Each actuator provides a stroke of 3.5µm over a compact area with low influence on neighboring actuators. The control software AOkit (Thorlabs, USA) [34

34. Thorlabs, “Operation Manual of AOkit,” http://www.thorlabs.com/NewGroupPage9.cfm?ObjectGroup_ID=3208

] minimized the wavefront errors by analyzing the signals from the SH and determined the appropriate drive signals for the DM actuators to compensate for the wavefront aberrations. After successfully correcting the wavefront errors for the reflected AO beam, the surface profile of the DM was recorded and used to compensate for the PAM light.

3. Results and discussion

The lateral resolution of the AO-PAM was first evaluated by imaging a United States Air Force resolution target (USAF 1951 Chart, Edmund Optics, Inc.), which reached element 6 of group 7, corresponding to a resolution of 2.19 µm (228.07 cycles/mm). The measured images of Group 6 and 7 of the target are shown in Fig. 3
Fig. 3 (a) – (c) Images of Group 6 and 7 of the USAF 1951 Chart: (a) Light microscopy image, (b) MAP of the PAM image without AO, (c) MAP of the PAM image of the same target with AO. d: measured edge response; e: PSF calculated from the data in (d) without AO; f: PSF calculated from the data in (d) with AO.
: (a) light microscopy image, (b) maximum amplitude projection (MAP) of the PAM image without AO, and (c) MAP of the PAM image with AO. The bars within the 6th element of Group 7 were clearly resolved in Fig. 3(c). We also measured the edge response by imaging the edge of the black square on the same resolution target [Fig. 3(d)]. The PSFs [Fig. 3(c) and 3(d)] were calculated from the measured edge response by differentiation. The calculation showed that the lateral resolution was improved from 5.1 µm without AO to 2.5 µm with AO. From the images we can also see that our system has uniform lateral resolutions in the X and Y directions.

We then tested the system on imaging excised biological samples ex vivo. We first imaged the ciliary body of a pig eye. The eye was acquired from a local slaughter house and was fixed in 10% formalin. The eyeball was cut in half to expose the ciliary body. Figure 4
Fig. 4 MAP of the PAM images of the ciliary body of a pig eye: (a) without AO, (b) with AO, (c) (Media 1) 3D rendering of the PAM data set. The area outlined in each image was enlarged and shown as pointed by the arrows.
shows the experimental results. Figure 4(a) and 4(b) show the MAP of the PAM images without and with AO corrections, respectively. Each image consists of 256 × 256 pixels covering an area of about 1mm × 1mm. Figure 4(c) is a 3D rendering of the PAM data set. The images showed interesting melanin distributions in the ciliary body. Compared with the image without AO correction, the PAM image with AO correction resolved many more details of the sample (see the enlarged areas pointed by the arrows). Due to better light energy concentration within the focus with AO correction, the PA signals are also stronger in Fig. 4(b) than in Fig. 4(a). As a result, the images with AO correction also have a better signal-to-noise ratio (SNR).

We then moved on to image the RPE of the same pig eye. The acquired PAM images without and with AO correction are shown in Fig. 5(a)
Fig. 5 MAP of the PAM images of the RPE of the pig eye: (a) without AO, (b) with AO.
and 5(b), respectively. The MAP images consist of 256 × 256 pixels covering an area of 0.4 mm × 0.4 mm. From the images we can see significant resolution improvement of the image with AO compared with the image without AO. In the image with AO single RPE cells are well resolved while in the image without AO the RPE cells are barely recognizable.

4. Conclusion

We have successfully integrated AO with PAM. We have evaluated the improved lateral resolution by correcting the wavefront aberrations of the illuminating optics of PAM with AO. The experimental results on biological samples have demonstrated not only significant improvement on the lateral resolution of PAM but also improvement on the SNR. For the first time single RPE cells are resolved with AO-PAM based on optical absorption contrast. The current study has laid the foundation for the ultimate AO-PAM for in vivo imaging of the RPE. Although for in vivo imaging the major task of the AO system will be correcting the aberrations of the eye, the system layout will be the same except an ocular lens will be added.

Acknowledgment

The research was supported in part by the National Institutes of Health grant 7R21EB008800-02 and the Juvenile Diabetes Research Foundation Innovative Grant 5-2009-498. Both S. Jiao and H. F. Zhang are corresponding authors for this work.

References and links

1.

K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]

2.

H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]

3.

H. F. Zhang, K. Maslov, and L. V. Wang, “In vivo imaging of subcutaneous structures using functional photoacoustic microscopy,” Nat. Protoc. 2(4), 797–804 (2007). [CrossRef] [PubMed]

4.

H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Imaging acute thermal burns by photoacoustic microscopy,” J. Biomed. Opt. 11(5), 054033 (2006). [CrossRef] [PubMed]

5.

S. Hu and L. V. Wang, “Photoacoustic imaging and characterization of the microvasculature,” J. Biomed. Opt. 15(1), 011101 (2010). [CrossRef] [PubMed]

6.

B. Rao, L. Li, K. Maslov, and L. Wang, “Hybrid-scanning optical-resolution photoacoustic microscopy for in vivo vasculature imaging,” Opt. Lett. 35(10), 1521–1523 (2010). [CrossRef] [PubMed]

7.

S. Jiao, M. Jiang, J. Hu, A. Fawzi, Q. Zhou, K. K. Shung, C. A. Puliafito, and H. F. Zhang, “Photoacoustic ophthalmoscopy for in vivo retinal imaging,” Opt. Express 18(4), 3967–3972 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-4-3967. [CrossRef] [PubMed]

8.

Z. Xie, S. Jiao, H. F. Zhang, and C. A. Puliafito, “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett. 34(12), 1771–1773 (2009). [CrossRef] [PubMed]

9.

S. Jiao, Z. Xie, H. F. Zhang, and C. A. Puliafito, “Simultaneous multimodal imaging with integrated photoacoustic microscopy and optical coherence tomography,” Opt. Lett. 34(19), 2961–2963 (2009). [CrossRef] [PubMed]

10.

H. F. Zhang, J. Wang, Q. Wei, T. Liu, S. Jiao, and C. A. Puliafito, “Collecting back-reflected photons in photoacoustic microscopy,” Opt. Express 18(2), 1278–1282 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-2-1278. [CrossRef] [PubMed]

11.

P. Artal, A. Guirao, E. Berrio, and D. R. Williams, “Compensation of corneal aberrations by the internal optics in the human eye,” J. Vis. 1(1), 1–8 (2001). [CrossRef]

12.

J. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef]

13.

E. J. Fernández, I. Iglesias, and P. Artal, “Closed-loop adaptive optics in the human eye,” Opt. Lett. 26(10), 746–748 (2001). [CrossRef]

14.

K. Y. Li and A. Roorda, “Automated identification of cone photoreceptors in adaptive optics retinal images,” J. Opt. Soc. Am. A 24(5), 1358–1363 (2007). [CrossRef]

15.

M. Pircher, R. J. Zawadzki, J. W. Evans, J. S. Werner, and C. K. Hitzenberger, “Simultaneous imaging of human cone mosaic with adaptive optics enhanced scanning laser ophthalmoscopy and high-speed transversal scanning optical coherence tomography,” Opt. Lett. 33(1), 22–24 (2008). [CrossRef]

16.

J. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef]

17.

T. Y. Chui, H. Song, and S. A. Burns, “Adaptive-optics imaging of human cone photoreceptor distribution,” J. Opt. Soc. Am. A 25(12), 3021–3029 (2008). [CrossRef]

18.

B. Hermann, E. J. Fernández, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, P. M. Prieto, and P. Artal, “Adaptive-optics ultrahigh-resolution optical coherence tomography,” Opt. Lett. 29(18), 2142–2144 (2004). [CrossRef] [PubMed]

19.

R. J. Zawadzki, S. M. Jones, S. S. Olivier, M. Zhao, B. A. Bower, J. A. Izatt, S. Choi, S. Laut, and J. S. Werner, “Adaptive-optics optical coherence tomography for high-resolution and high-speed 3D retinal in vivo imaging,” Opt. Express 13(21), 8532–8546 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85806. [CrossRef] [PubMed]

20.

Y. Zhang, B. Cense, J. Rha, R. S. Jonnal, W. Gao, R. J. Zawadzki, J. S. Werner, S. Jones, S. Olivier, and D. T. Miller, “High-speed volumetric imaging of cone photoreceptors with adaptive optics spectral-domain optical coherence tomography,” Opt. Express 14(10), 4380–4394 (2006), http://www.opticsinfobase.org/abstract.cfm?id=89918&CFID=113107704&CFTOKEN=11132256. [CrossRef] [PubMed]

21.

Y. Zhang, J. Rha, R. Jonnal, and D. Miller, “Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina,” Opt. Express 13(12), 4792–4811 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-12-4792. [CrossRef] [PubMed]

22.

R. J. Zawadzki, S. S. Choi, S. M. Jones, S. S. Oliver, and J. S. Werner, “Adaptive optics-optical coherence tomography: optimizing visualization of microscopic retinal structures in three dimensions,” J. Opt. Soc. Am. A 24(5), 1373–1383 (2007). [CrossRef]

23.

K. Kurokawa, K. Sasaki, S. Makita, M. Yamanari, B. Cense, and Y. Yasuno, “Simultaneous high-resolution retinal imaging and high-penetration choroidal imaging by one-micrometer adaptive optics optical coherence tomography,” Opt. Express 18(8), 8515–8527 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-8-8515. [CrossRef] [PubMed]

24.

A. Roorda, F. Romero-Borja, W. Donnelly Iii, H. Queener, T. Hebert, and M. Campbell, “Adaptive optics scanning laser ophthalmoscopy,” Opt. Express 10(9), 405–412 (2002), http://www.opticsinfobase.org/abstract.cfm?URI=OE-10-9-405. [PubMed]

25.

D. Merino, C. Dainty, A. Bradu, and A. G. Podoleanu, “Adaptive optics enhanced simultaneous en-face optical coherence tomography and scanning laser ophthalmoscopy,” Opt. Express 14(8), 3345–3353 (2006), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-14-8-3345. [CrossRef] [PubMed]

26.

Y. Zhang, S. Poonja, and A. Roorda, “MEMS-based adaptive optics scanning laser ophthalmoscopy,” Opt. Lett. 31(9), 1268–1270 (2006). [CrossRef] [PubMed]

27.

D. X. Hammer, R. D. Ferguson, C. E. Bigelow, N. V. Iftimia, T. E. Ustun, and S. A. Burns, “Adaptive optics scanning laser ophthalmoscope for stabilized retinal imaging,” Opt. Express 14(8), 3354–3367 (2006), http://www.opticsinfobase.org/abstract.cfm?id=89323&CFID=113107704&CFTOKEN=11132256. [CrossRef] [PubMed]

28.

S. A. Burns, R. Tumbar, A. E. Elsner, D. Ferguson, and D. X. Hammer, “Large-field-of-view, modular, stabilized, adaptive-optics-based scanning laser ophthalmoscope,” J. Opt. Soc. Am. A 24(5), 1313–1326 (2007). [CrossRef]

29.

L. Sherman, J. Y. Ye, O. Albert, and T. B. Norris, “Adaptive correction of depth-induced aberrations in multiphoton scanning microscopy using a deformable mirror,” J. Microsc. 206(1), 65–71 (2002). [CrossRef] [PubMed]

30.

M. J. Booth, M. A. Neil, R. Juskaitis, and T. Wilson, “Adaptive aberration correction in a confocal microscope,” Proc. Natl. Acad. Sci. U.S.A. 99(9), 5788–5792 (2002). [CrossRef] [PubMed]

31.

P. Marsh, D. Burns, and J. Girkin, “Practical implementation of adaptive optics in multiphoton microscopy,” Opt. Express 11(10), 1123–1130 (2003), http://www.opticsinfobase.org/abstract.cfm?&id=72382. [CrossRef] [PubMed]

32.

J. M. Girkin, S. Poland, and A. J. Wright, “Adaptive optics for deeper imaging of biological samples,” Curr. Opin. Biotechnol. 20(1), 106–110 (2009). [CrossRef] [PubMed]

33.

J. W. Cha, J. Ballesta, and P. T. C. So, “Shack-Hartmann wavefront-sensor-based adaptive optics system for multiphoton microscopy,” J. Biomed. Opt. 15(4), 046022 (2010). [CrossRef] [PubMed]

34.

Thorlabs, “Operation Manual of AOkit,” http://www.thorlabs.com/NewGroupPage9.cfm?ObjectGroup_ID=3208

35.

Thorlabs, “Operation Manual of Optical Wavefront Sensor,” http://www.thorlabs.com/NewGroupPage9.cfm?ObjectGroup_ID=2946

OCIS Codes
(110.0180) Imaging systems : Microscopy
(110.5120) Imaging systems : Photoacoustic imaging
(110.1085) Imaging systems : Adaptive imaging

ToC Category:
Microscopy

History
Original Manuscript: July 13, 2010
Revised Manuscript: August 23, 2010
Manuscript Accepted: September 19, 2010
Published: September 29, 2010

Virtual Issues
Vol. 5, Iss. 14 Virtual Journal for Biomedical Optics

Citation
Minshan Jiang, Xiangyang Zhang, Carmen A. Puliafito, Hao F. Zhang, and Shuliang Jiao, "Adaptive optics photoacoustic microscopy," Opt. Express 18, 21770-21776 (2010)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-18-21-21770


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. K. Maslov, H. F. Zhang, S. Hu, and L. V. Wang, “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett. 33(9), 929–931 (2008). [CrossRef] [PubMed]
  2. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol. 24(7), 848–851 (2006). [CrossRef] [PubMed]
  3. H. F. Zhang, K. Maslov, and L. V. Wang, “In vivo imaging of subcutaneous structures using functional photoacoustic microscopy,” Nat. Protoc. 2(4), 797–804 (2007). [CrossRef] [PubMed]
  4. H. F. Zhang, K. Maslov, G. Stoica, and L. V. Wang, “Imaging acute thermal burns by photoacoustic microscopy,” J. Biomed. Opt. 11(5), 054033 (2006). [CrossRef] [PubMed]
  5. S. Hu and L. V. Wang, “Photoacoustic imaging and characterization of the microvasculature,” J. Biomed. Opt. 15(1), 011101 (2010). [CrossRef] [PubMed]
  6. B. Rao, L. Li, K. Maslov, and L. Wang, “Hybrid-scanning optical-resolution photoacoustic microscopy for in vivo vasculature imaging,” Opt. Lett. 35(10), 1521–1523 (2010). [CrossRef] [PubMed]
  7. S. Jiao, M. Jiang, J. Hu, A. Fawzi, Q. Zhou, K. K. Shung, C. A. Puliafito, and H. F. Zhang, “Photoacoustic ophthalmoscopy for in vivo retinal imaging,” Opt. Express 18(4), 3967–3972 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-4-3967 . [CrossRef] [PubMed]
  8. Z. Xie, S. Jiao, H. F. Zhang, and C. A. Puliafito, “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett. 34(12), 1771–1773 (2009). [CrossRef] [PubMed]
  9. S. Jiao, Z. Xie, H. F. Zhang, and C. A. Puliafito, “Simultaneous multimodal imaging with integrated photoacoustic microscopy and optical coherence tomography,” Opt. Lett. 34(19), 2961–2963 (2009). [CrossRef] [PubMed]
  10. H. F. Zhang, J. Wang, Q. Wei, T. Liu, S. Jiao, and C. A. Puliafito, “Collecting back-reflected photons in photoacoustic microscopy,” Opt. Express 18(2), 1278–1282 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-2-1278 . [CrossRef] [PubMed]
  11. P. Artal, A. Guirao, E. Berrio, and D. R. Williams, “Compensation of corneal aberrations by the internal optics in the human eye,” J. Vis. 1(1), 1–8 (2001). [CrossRef]
  12. J. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef]
  13. E. J. Fernández, I. Iglesias, and P. Artal, “Closed-loop adaptive optics in the human eye,” Opt. Lett. 26(10), 746–748 (2001). [CrossRef]
  14. K. Y. Li and A. Roorda, “Automated identification of cone photoreceptors in adaptive optics retinal images,” J. Opt. Soc. Am. A 24(5), 1358–1363 (2007). [CrossRef]
  15. M. Pircher, R. J. Zawadzki, J. W. Evans, J. S. Werner, and C. K. Hitzenberger, “Simultaneous imaging of human cone mosaic with adaptive optics enhanced scanning laser ophthalmoscopy and high-speed transversal scanning optical coherence tomography,” Opt. Lett. 33(1), 22–24 (2008). [CrossRef]
  16. J. Liang, D. R. Williams, and D. T. Miller, “Supernormal vision and high-resolution retinal imaging through adaptive optics,” J. Opt. Soc. Am. A 14(11), 2884–2892 (1997). [CrossRef]
  17. T. Y. Chui, H. Song, and S. A. Burns, “Adaptive-optics imaging of human cone photoreceptor distribution,” J. Opt. Soc. Am. A 25(12), 3021–3029 (2008). [CrossRef]
  18. B. Hermann, E. J. Fernández, A. Unterhuber, H. Sattmann, A. F. Fercher, W. Drexler, P. M. Prieto, and P. Artal, “Adaptive-optics ultrahigh-resolution optical coherence tomography,” Opt. Lett. 29(18), 2142–2144 (2004). [CrossRef] [PubMed]
  19. R. J. Zawadzki, S. M. Jones, S. S. Olivier, M. Zhao, B. A. Bower, J. A. Izatt, S. Choi, S. Laut, and J. S. Werner, “Adaptive-optics optical coherence tomography for high-resolution and high-speed 3D retinal in vivo imaging,” Opt. Express 13(21), 8532–8546 (2005), http://www.opticsinfobase.org/abstract.cfm?id=85806 . [CrossRef] [PubMed]
  20. Y. Zhang, B. Cense, J. Rha, R. S. Jonnal, W. Gao, R. J. Zawadzki, J. S. Werner, S. Jones, S. Olivier, and D. T. Miller, “High-speed volumetric imaging of cone photoreceptors with adaptive optics spectral-domain optical coherence tomography,” Opt. Express 14(10), 4380–4394 (2006), http://www.opticsinfobase.org/abstract.cfm?id=89918&CFID=113107704&CFTOKEN=11132256 . [CrossRef] [PubMed]
  21. Y. Zhang, J. Rha, R. Jonnal, and D. Miller, “Adaptive optics parallel spectral domain optical coherence tomography for imaging the living retina,” Opt. Express 13(12), 4792–4811 (2005), http://www.opticsinfobase.org/abstract.cfm?URI=oe-13-12-4792 . [CrossRef] [PubMed]
  22. R. J. Zawadzki, S. S. Choi, S. M. Jones, S. S. Oliver, and J. S. Werner, “Adaptive optics-optical coherence tomography: optimizing visualization of microscopic retinal structures in three dimensions,” J. Opt. Soc. Am. A 24(5), 1373–1383 (2007). [CrossRef]
  23. K. Kurokawa, K. Sasaki, S. Makita, M. Yamanari, B. Cense, and Y. Yasuno, “Simultaneous high-resolution retinal imaging and high-penetration choroidal imaging by one-micrometer adaptive optics optical coherence tomography,” Opt. Express 18(8), 8515–8527 (2010), http://www.opticsinfobase.org/abstract.cfm?URI=oe-18-8-8515 . [CrossRef] [PubMed]
  24. A. Roorda, F. Romero-Borja, W. Donnelly Iii, H. Queener, T. Hebert, and M. Campbell, “Adaptive optics scanning laser ophthalmoscopy,” Opt. Express 10(9), 405–412 (2002), http://www.opticsinfobase.org/abstract.cfm?URI=OE-10-9-405 . [PubMed]
  25. D. Merino, C. Dainty, A. Bradu, and A. G. Podoleanu, “Adaptive optics enhanced simultaneous en-face optical coherence tomography and scanning laser ophthalmoscopy,” Opt. Express 14(8), 3345–3353 (2006), http://www.opticsinfobase.org/oe/abstract.cfm?URI=oe-14-8-3345 . [CrossRef] [PubMed]
  26. Y. Zhang, S. Poonja, and A. Roorda, “MEMS-based adaptive optics scanning laser ophthalmoscopy,” Opt. Lett. 31(9), 1268–1270 (2006). [CrossRef] [PubMed]
  27. D. X. Hammer, R. D. Ferguson, C. E. Bigelow, N. V. Iftimia, T. E. Ustun, and S. A. Burns, “Adaptive optics scanning laser ophthalmoscope for stabilized retinal imaging,” Opt. Express 14(8), 3354–3367 (2006), http://www.opticsinfobase.org/abstract.cfm?id=89323&CFID=113107704&CFTOKEN=11132256 . [CrossRef] [PubMed]
  28. S. A. Burns, R. Tumbar, A. E. Elsner, D. Ferguson, and D. X. Hammer, “Large-field-of-view, modular, stabilized, adaptive-optics-based scanning laser ophthalmoscope,” J. Opt. Soc. Am. A 24(5), 1313–1326 (2007). [CrossRef]
  29. L. Sherman, J. Y. Ye, O. Albert, and T. B. Norris, “Adaptive correction of depth-induced aberrations in multiphoton scanning microscopy using a deformable mirror,” J. Microsc. 206(1), 65–71 (2002). [CrossRef] [PubMed]
  30. M. J. Booth, M. A. Neil, R. Juskaitis, and T. Wilson, “Adaptive aberration correction in a confocal microscope,” Proc. Natl. Acad. Sci. U.S.A. 99(9), 5788–5792 (2002). [CrossRef] [PubMed]
  31. P. Marsh, D. Burns, and J. Girkin, “Practical implementation of adaptive optics in multiphoton microscopy,” Opt. Express 11(10), 1123–1130 (2003), http://www.opticsinfobase.org/abstract.cfm?&id=72382 . [CrossRef] [PubMed]
  32. J. M. Girkin, S. Poland, and A. J. Wright, “Adaptive optics for deeper imaging of biological samples,” Curr. Opin. Biotechnol. 20(1), 106–110 (2009). [CrossRef] [PubMed]
  33. J. W. Cha, J. Ballesta, and P. T. C. So, “Shack-Hartmann wavefront-sensor-based adaptive optics system for multiphoton microscopy,” J. Biomed. Opt. 15(4), 046022 (2010). [CrossRef] [PubMed]
  34. Thorlabs, “Operation Manual of AOkit,” http://www.thorlabs.com/NewGroupPage9.cfm?ObjectGroup_ID=3208
  35. Thorlabs, “Operation Manual of Optical Wavefront Sensor,” http://www.thorlabs.com/NewGroupPage9.cfm?ObjectGroup_ID=2946

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1 Fig. 2 Fig. 3
 
Fig. 4 Fig. 5
 

Supplementary Material


» Media 1: MPG (1842 KB)     

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited