Twin-beams digital holography for 3D tracking and quantitative phase-contrast microscopy in microfluidics

Pasquale Memmolo, Andrea Finizio, Melania Paturzo, Lisa Miccio, and Pietro Ferraro  »View Author Affiliations


Optics Express, Vol. 19, Issue 25, pp. 25833-25842 (2011)
http://dx.doi.org/10.1364/OE.19.025833


View Full Text Article

Acrobat PDF (1229 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

We report on a compact twin-beam interferometer that can be adopted as a flexible diagnostic tool in microfluidic platforms with twofold functionality. The novel configuration allows 3D tracking of micro-particles and, at same time, can simultaneously furnish Quantitative Phase-contrast maps of tracked micro-objects by interference microscopy, without changing the configuration. Experimental demonstration is given on for in vitro cells in a microfluidic environment.

© 2011 OSA

1. Introduction

Nowadays the incredible development of bio-microfluidic technology [1

1. D. Psaltis, S. R. Quake, and C. Yang, “Developing optofluidic technology through the fusion of microfluidics and optics,” Nature 442(7101), 381–386 (2006). [CrossRef] [PubMed]

,2

2. J. W. Hong and S. R. Quake, “Integrated nanoliter systems,” Nat. Biotechnol. 21(10), 1179–1183 (2003). [CrossRef] [PubMed]

] highly demands for substantial advancements in multifunctional tools for characterization, monitoring, and manipulation in microfluidic environments. In recent years, the number of proposed and implemented techniques for diagnostic purposes is greatly grown. Single approaches have been demonstrated for imaging, phase-contrast quantitative analysis [3

3. F. Dubois, C. Yourassowsky, O. Monnom, J. C. Legros, O. Debeir, P. Van Ham, R. Kiss, and C. Decaestecker, “Digital holographic microscopy for the three-dimensional dynamic analysis of in vitro cancer cell migration,” J. Biomed. Opt. 11(5), 054032 (2006). [CrossRef] [PubMed]

13

13. Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, and G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express 19(2), 1016–1026 (2011). [CrossRef] [PubMed]

], manipulation and trapping [14

14. L. Friedrich and A. Rohrbach, “Improved interferometric tracking of trapped particles using two frequency-detuned beams,” Opt. Lett. 35(11), 1920–1922 (2010). [CrossRef] [PubMed]

16

16. P. Schiro, C. Dubois, and A. Kwok, “Large capture-range of a single-beam gradient optical trap,” Opt. Express 11(25), 3485–3489 (2003). [CrossRef] [PubMed]

], tracking of micro-objects (i.e. nano-drops, carbon nanotubes, bio-cells, quantum dots, dielectric spheres and for metallic spheres [17

17. A. Rohrbach, H. Kress, and E. H. Stelzer, “Three-dimensional tracking of small spheres in focused laser beams: influence of the detection angular aperture,” Opt. Lett. 28(6), 411–413 (2003). [CrossRef] [PubMed]

], nano- and micro-particles [18

18. M. D. McMahon, A. J. Berglund, P. Carmichael, J. J. McClelland, and J. A. Liddle, “3D particle trajectories observed by orthogonal tracking microscopy,” ACS Nano 3(3), 609–614 (2009). [CrossRef] [PubMed]

]), tracking of both non-fluorescent [19

19. E. Toprak, H. Balci, B. H. Blehm, and P. R. Selvin, “Three-dimensional particle tracking via bifocal imaging,” Nano Lett. 7(7), 2043–2045 (2007). [CrossRef] [PubMed]

] and fluorescent particles [20

20. H. P. Kao and A. S. Verkman, “Tracking of single fluorescent particles in three dimensions: use of cylindrical optics to encode particle position,” Biophys. J. 67(3), 1291–1300 (1994). [CrossRef] [PubMed]

,21

21. S. R. P. Pavani and R. Piestun, “Three dimensional tracking of fluorescent microparticles using a photon-limited double-helix response system,” Opt. Express 16(26), 22048–22057 (2008). [CrossRef] [PubMed]

]. Many techniques accomplish the tracking with high detection accuracy of a single particle [22

22. S. Ram, P. Prabhat, E. S. Ward, and R. J. Ober, “Improved single particle localization accuracy with dual objective multifocal plane microscopy,” Opt. Express 17(8), 6881–6898 (2009). [CrossRef] [PubMed]

24

24. Y. Park, G. Popescu, K. Badizadegan, R. R. Dasari, and M. S. Feld, “Fresnel particle tracing in three dimensions using diffraction phase microscopy,” Opt. Lett. 32(7), 811–813 (2007). [CrossRef] [PubMed]

] and others in which a statistical localization algorithms have been proposed [25

25. M. J. Mlodzianoski, M. F. Juette, G. L. Beane, and J. Bewersdorf, “Experimental characterization of 3D localization techniques for particle-tracking and super-resolution microscopy,” Opt. Express 17(10), 8264–8277 (2009). [CrossRef] [PubMed]

,26

26. M. DaneshPanah, S. Zwick, F. Schaal, M. Warber, B. Javidi, and W. Osten, “3D Holographic Imaging and Trapping for Non-Invasive Cell Identification and Tracking,” J. Disp. Technol. 6(10), 490–499 (2010). [CrossRef]

]. Tracking objects in three-dimensional space using digital multiplexing holography have been presented recently [27

27. J. A. Dominguez-Caballero, N. Loomis, G. Barbastathis, and J. Milgram, Techniques Based on Digital Multiplexing Holography for Three-Dimensional Object Tracking”, in Conference on Lasers and Electro-Optics/Quantum Electronics and Laser Science Conference and Photonic Applications Systems Technologies, OSA Technical Digest (CD) (Optical Society of America, 2007), paper JThD84.

]. Also a three-dimensional tracking of Brownian motion of a particle trapped in optical tweezers was reported [28

28. M.-T. Wei and A. Chiou, “Three-dimensional tracking of Brownian motion of a particle trapped in optical tweezers with a pair of orthogonal tracking beams and the determination of the associated optical force constants,” Opt. Express 13(15), 5798–5806 (2005). [CrossRef] [PubMed]

].

Elective approaches to be adopted in microfluidic environments are optical/photonics ones that have the remarkable advantage to be non-contact, full-field, non-invasive and can be packaged thanks to the integrated optics and optofluidic modalities [2

2. J. W. Hong and S. R. Quake, “Integrated nanoliter systems,” Nat. Biotechnol. 21(10), 1179–1183 (2003). [CrossRef] [PubMed]

,29

29. S. O. Isikman, W. Bishara, H. Zhu, and A. Ozcan, “Optofluidic Tomography on a Chip,” Appl. Phys. Lett. 98(16), 161109 (2011). [CrossRef] [PubMed]

]. Various microfluidic platforms have been developed for manipulating droplets, handling micro-nano-objects, visualize and quantify processes occurring in-loco and for direct application for lab-on-a chip configurations. In fact, phase-contrast approaches, adapted to a lab-on-a-chip configurations, have given the possibility to get quantitative information with remarkable lateral and vertical resolution directly in situ. Moreover, techniques for tracking of micro-objects needs to be developed for measuring velocity fields, trajectories patterns, motility of cancer cell and so on [3

3. F. Dubois, C. Yourassowsky, O. Monnom, J. C. Legros, O. Debeir, P. Van Ham, R. Kiss, and C. Decaestecker, “Digital holographic microscopy for the three-dimensional dynamic analysis of in vitro cancer cell migration,” J. Biomed. Opt. 11(5), 054032 (2006). [CrossRef] [PubMed]

,30

30. H. Yang, N. Halliwell, and J. Coupland, “Application of the digital shearing method to extract three-component velocity in holographic particle image velocimetry,” Meas. Sci. Technol. 15(4), 694–698 (2004). [CrossRef]

]. In ref [31

31. F. C. Cheong, B. J. Krishnatreya, and D. G. Grier, “Strategies for three-dimensional particle tracking with holographic video microscopy,” Opt. Express 18(13), 13563–13573 (2010). [CrossRef] [PubMed]

]. the video stream captured by an in-line holographic microscope can be analyzed on a frame-by-frame basis to track individual colloidal particles experiencing three-dimensional motions with nanometer resolution. In [32

32. O. Otto, F. Czerwinski, J. L. Gornall, G. Stober, L. B. Oddershede, R. Seidel, and U. F. Keyser, “Real-time particle tracking at 10,000 fps using optical fiber illumination,” Opt. Express 18(22), 22722–22733 (2010). [CrossRef] [PubMed]

] is introduced optical fiber illumination for real-time tracking of optically trapped micrometer-sized particles with microsecond time resolution. Digital shearing method was adopted in [30

30. H. Yang, N. Halliwell, and J. Coupland, “Application of the digital shearing method to extract three-component velocity in holographic particle image velocimetry,” Meas. Sci. Technol. 15(4), 694–698 (2004). [CrossRef]

] to extract three-component velocity in particle image velocimetry.

Several particle-tracking and imaging methods have achieved three-dimensional sensitivity through the introduction of angled micromirrors into the observation volume of an optical microscope [18

18. M. D. McMahon, A. J. Berglund, P. Carmichael, J. J. McClelland, and J. A. Liddle, “3D particle trajectories observed by orthogonal tracking microscopy,” ACS Nano 3(3), 609–614 (2009). [CrossRef] [PubMed]

,33

33. H. Hajjoul, S. Kocanova, I. Lassadi, K. Bystricky, and A. Bancaud, “Lab-on-Chip for fast 3D particle tracking in living cells,” Lab Chip 9(21), 3054–3058 (2009). [CrossRef] [PubMed]

]. In ref [34

34. A. J. Berglund, M. D. McMahon, J. J. McClelland, and J. A. Liddle, “Theoretical model of errors in micromirror-based three-dimensional particle tracking,” Opt. Lett. 35(11), 1905–1907 (2010). [CrossRef] [PubMed]

]. the authors developed a theoretical model of the imaging response of such devices and show how the direct and reflected images of a fluorescent particle are affected. In particle-tracking applications, asymmetric image degradation manifests itself as systematic tracking errors. Recently some experiments have been performed to get simultaneous trapping and tracking, but, in order to achieve this double function, they used two coupled lasers systems [31

31. F. C. Cheong, B. J. Krishnatreya, and D. G. Grier, “Strategies for three-dimensional particle tracking with holographic video microscopy,” Opt. Express 18(13), 13563–13573 (2010). [CrossRef] [PubMed]

,35

35. D. B. Conkey, R. P. Trivedi, S. R. P. Pavani, I. I. Smalyukh, and R. Piestun, “Three-dimensional parallel particle manipulation and tracking by integrating holographic optical tweezers and engineered point spread functions,” Opt. Express 19(5), 3835–3842 (2011). [CrossRef] [PubMed]

,36

36. B. Kemper, P. Langehanenberg, A. Hoink, G. von Bally, F. Wottowah, S. Schinkinger, J. Guck, J. Kas, I. Bredebusch, J. Schnekenburger, and K. Schütze, “Monitoring of laser micromanipulated optically trapped cells by digital holographic microscopy,” J. Biophotonics 3(7), 425–431 (2010). [CrossRef] [PubMed]

]. In this paper, we show a completely new concept of a compact holographic microscope that can ensure the multi-functionality, accomplishing, by the same configuration and simultaneously, accurate 3D tracking and quantitative phase-contrast analysis. Experimental results are presented and discussed for in vitro cells in microfluidic environment. The system is very simple and compact and is based on twins-laser-beams coming from a single laser source. Through this simple conceptual design we show how two different functionalities can be accomplished by the same optical setup, i.e. 3D tracking of micro-object and quantitative phase contrast imaging. It is important to note that by same system it is possible accomplish other two different functionalities, i.e. for driving particles along appropriate paths, performing simultaneously their interferometric analysis [37

37. F. Merola, L. Miccio, M. Paturzo, A. Finizio, S. Grilli, and P. Ferraro, “Driving and analysis of micro-objects by digital holographic microscope in microfluidics,” Opt. Lett. 36(16), 3079–3081 (2011). [CrossRef] [PubMed]

].

2. Working principle

2.1 Optical configuration

The optical configuration is illustrated in Fig. 1
Fig. 1 Twin-beams digital holography microscope.
. The particles/cells are loaded in a chamber of about 5x5x0.3mm, assembled by using two cover glasses (0.15 mm thick) with a double-sided tape spacer (0.3 mm thick). Two beams, coming from the same DPSS laser (532nm, 250mW), enter into the microscope objective (oil immersion Nikon 100x, 1.2 NA). The two beams are sent slightly off-axis through the same microscope objective. The angle between the beams is about 4 degrees. Since the two beams passes through the same microscope objective, in principle they are in focus in the same image plane. Nevertheless the twin beams can experience aberrations giving rise to some focus-shift too. A 20x microscope objective is used to obtain an image on a CCD plane. The tracking is performed by evaluating the double out-of-focus projections of the particles due to the twin-beams onto the array detector plane. The principle is simple and allows tracking in 3D. In fact as shown in Fig. 1 each particle forms two shadows on the CCD array, The separation between the two shadows is a function of the longitudinal position of the particle. By the very same configuration also digital holograms can be directly recorded with the aim to obtain quantitative phase contrast images of the micro-objects as depicted in Fig. 1.

2.2 Modeling for 3D tracking

In our model, we have two plane beams sent through an objective microscope. The experimental configuration is depicted in Fig. 2
Fig. 2 Simulated interference between two laser beams
. The lens makes a Fourier transformation thus focusing the two beams, in ideal condition without considering the aberrations, in its back focal plane. An additional lens is used to image the microfluidic volume. Essentially we can sketch the imaged volume as composed by two cones, transmitted into the microfluidic sample. The two cones (see Fig. 2) are partially superimposed in the image space. On the CCD digital sensor, the intersection between the beam and the sensor plane of each cone is a circle ellipse with radius r<<dc, where dc is the distance between the digital sensor and the vertex of cone. Supposing that the principal axes of cones are parallel, we consider a reference system of coordinates Oxyz where the plane xy is on the digital sensor with O in the center. In this reference system, we define the vertex of both cones C1(x1,y1,z1) and C2(x2,y2,z2) where (x1,y1) and (x2,y2) are given by the center of two circles in the sensor plane and z1=z2=dc. Then we suppose that a microscopic object, that can be considered a point with coordinates P(xp,yp,zp), is in the volume defined by the union of the two cones, that is P(C1,Ω1)(C2,Ω2), where Ω1,Ω2 are the cones solid angles. In this scenario, we have three possible situations. P(C1,Ω1), in this case we will have a projection of the point P on the sensor plane in the circle relative to the cone 1. If P(C2,Ω2) we have a projection of the point P on the sensor plane in the circle relative to the cone 2. Finally if P(C1,Ω1)(C2,Ω2), in this case we have two projections of the point P on the sensor plane (see Fig. 1), i.e. each twin laser beam makes a projection (or shadow) of the particle on the CCD array detector. This latter situation is interesting because we demonstrate that it is possible to find the path of a point P(C1,Ω1)(C2,Ω2) using the information on the projections coordinates, by the knowledge of the position of the two vertices C1, C2. Supposing that the projections of P in the two circles are the points P1(xp1,yp1,0) and P2(xp2,yp2,0) respectively, the estimation of P is given by
P^=C1P1¯C2P2¯
(1)
where AB¯ denote the segment joining the points A(xA,yA,zA) and B(xB,yB,zB):
{x'=xA+lxty'=yA+lytz'=zA+lzt
(2)
where (x',y',z')AB¯, t[0,1] and l()=()B()A.

The hypothesis P(C1,Ω1)(C2,Ω2) assure that one and only one intersection's point P^ exists.

2.2 Modeling for 3D tracking

To demonstrate how the tracking functionality works we firstly present here some numerical simulations to validate the reliability of the working principle of the proposed method. In our numerical simulations, we set C1(112,76,dc), C2(112,76,dc) with dc=44478, r=700 and the unit are in pixels. Using these parameters it is possible to find the aperture angles of cones:
α1=α2=arctan(rdc)=0.0157rad
(3)
from which we can find the solid angles

Ω1=Ω2=2π(1cos(α1/2))1.95104sr
(4)

We report here simulations of two different paths of particle P. A linear path and helix paths, respectively. The following equations give the math formulation of the two simulated trajectories. The linear path with xp0, yp0, zp[30000,30000].

Elliptical helix given by the following equation:

{xp=100sin(t/1000)yp=10cos(t/1000)zp=tt[8000π,0]
(5)

For both simulated paths, we compute the mean error and the standard deviation error on the difference between the real path and the path estimated with the proposed method.

Media 1 shows the results of these two simulation (frames on the left). From the simulated images we evaluate the corresponding points P1 and P2 in the sensor plane for both paths. By the knowledge of the P1 and P2 coordinates of and through Eq. (1) we reconstructed the paths. Figure 3
Fig. 3 Comparison between real and estimated path for (a) linear and (b) coil path.
shows the comparison between the simulated and reconstructed paths for the linear and the coil path, respectively. The results show very good agreement between the simulated values and the retrieved paths.

3. Experiments

Here we report two experiments to demonstrate the capability and usefulness of the proposed method. First, to show the reliability of the method, we performed the system calibration by using a spherical particle (latex) displaced along the optical axis with known steps, simulating a simple linear path. In a second experiment we show how the method can be applied in real situation by tracking 3D paths of multiple motile in-vitro cells.

3.1 System calibration displacing a particle along the optical axis with known steps

In Fig. 4
Fig. 4 Three frames from the recorded sequence of the Z-axis motion of a microscopic particle.
are shown the images recorded at different distances. The experimental path is given by xp0, yp0, zp[200,0]μm, with acquisition step equal to 20μm. The CCD pixels dimension are px = py = 6.7µm.

We reconstructed the path by estimating the centroid of two images of micro-spheres for each frame of the recorded sequence, by applying the above equations. The centroids are computed using the classical algorithm for detecting regions of interest [38

38. R. C. Gonzalez and R. E. Woods, Digital Image Processing, 2nd ed. (Addison - Wesley Longman, Boston 1992).

] and their center of mass. The result are shown in Fig. 5
Fig. 5 Comparison between real and estimated path for real sequence.
(and fully visualized in the supplementary Media 2) for this experiment. We calculate the average value and standard deviation value of difference between the estimated positions and real position having 1.07µm and 2.02µm, respectively. Since the ratio between the standard deviation computed and the acquisition step is approximately 10%, we have the similar accuracy show in the z-localization proposed in [22

22. S. Ram, P. Prabhat, E. S. Ward, and R. J. Ober, “Improved single particle localization accuracy with dual objective multifocal plane microscopy,” Opt. Express 17(8), 6881–6898 (2009). [CrossRef] [PubMed]

].

3.2 Tracking 3D paths of multiple cells

The results of the retrieved 3D paths of the cells were easily estimated by the proposed method. The results, i.e. the 3D plot of the paths, are shown in Fig. 7
Fig. 7 Estimated path for random motion of the three cells shows in Fig. 6
. The dynamic evolutions can be seen in full in the supplementary Media 3.

From the last result, it is clear that all the three cells follow paths along the same streamlines in the microfluidic flux. However, can be noted that cells experience a displacement mainly along the longitudinal axis. This is due to the twins beams that have surely a role in influencing and/or determining the paths followed by the particles, because of both scattering (radiation pressure) and the attractive force (similar to optical tweezers). Nevertheless the optical configuration can be designed to cancel the effect of the light on the paths. In fact by using a trivial stroboscopic illumination, depending also on the velocity fields of the particles, the influence of the light on the paths can be drastically reduced up to become completely negligible, making this method a full non-invasive 3D tracking approach.

4. Quantitative phase – contrast microscopy of tracked particles

In addition to the tracking functionality we show here that the same set-up can be adopted as quantitative phase contrast microscope. In fact due to the coherence properties of the adopted laser source we can see that the twin-beams interfere to produce nice interference fringe patterns at the CCD sensor plane. Indeed the interference patterns, as those in Figs. 4 and 6, are really digital holograms. Such digital holograms contain quantitative information about the cells. In fact optical path length variations due to the presence of the cells along can be retrieved easily. Digital holograms (i.e., each frames of the recorded movies) can be numerically reconstructed by well known diffraction propagation methods to get the quantitative phase contrast maps of the cells (QPMs). We adopted the holographic reconstruction by the convolution method [8

8. G. Coppola, G. Di Caprio, M. Gioffré, R. Puglisi, D. Balduzzi, A. Galli, L. Miccio, M. Paturzo, S. Grilli, A. Finizio, and P. Ferraro, “Digital self-referencing quantitative phase microscopy by wavefront folding in holographic image reconstruction,” Opt. Lett. 35(20), 3390–3392 (2010). [CrossRef] [PubMed]

,37

37. F. Merola, L. Miccio, M. Paturzo, A. Finizio, S. Grilli, and P. Ferraro, “Driving and analysis of micro-objects by digital holographic microscope in microfluidics,” Opt. Lett. 36(16), 3079–3081 (2011). [CrossRef] [PubMed]

41

41. E. Allaria, S. Brugioni, S. Denicola, P. Ferraro, S. Grilli, and R. Meucci, “Digital holography at 10.6 μm,” Opt. Commun. 215(4-6), 257–262 (2003). [CrossRef]

]. In Fig. 8
Fig. 8 (a) phase reconstruction of a single acquisition with estimated in-focus distance; (b) unwrapping of in-focus phase image.
, we show an example of the phase-contrast map of the cell A.

It is important to note that digital holography method could also give the 3D tracking of the particles in the imaged volume. In fact the numerical reconstruction of the whole complex optical wavefront would allow to have reconstructions at different depths (i.e. at various distances from the CCD array, on the hologram plane). However it is well known that have accurate information of the axial position of a particles it would be necessary to assess with high accuracy of the in-focus position [39

39. A. El Mallahi and F. Dubois, “Dependency and precision of the refocusing criterion based on amplitude analysis in digital holographic microscopy,” Opt. Express 19(7), 6684–6698 (2011). [CrossRef] [PubMed]

]. Various method have been investigated and demonstrated for this aim and the topic is still matter of intensive investigation. Different strategies have been proposed and often quite heavy computational efforts are necessary to get the good in-focus images and sometimes high uncertainties still remain. The method proposed here, on the contrary, allows quite easy calculation of the depth coordinate of the particles.

5. Conclusion

We developed a simple and compact digital holographic microscope in off-axis configuration by using two twins laser beams passing through the same microscope objective. We demonstrate the effectiveness of this novel configuration that can be especially suitable for phase contrast analysis and 3D imaging of biological sample and microfluidic devices. We implemented experimental the set-up, validated the model either numerically and from the experimental point of view. Demonstration of the 3D tracking and the capability to perform Quantitative Phase-Contrast microscopic analysis, on multiple moving in-vitro cells in a microfluidic environment, was finally performed and we believe can be useful in lab-on-a-chip devices. The optical configuration is very simple as it is made by only two beams. The beams can be produced by various configurations such a diffraction grating at the input pupil of the microscope objective, or even by using a spatial light modulator [13

13. Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, and G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express 19(2), 1016–1026 (2011). [CrossRef] [PubMed]

]. Furthermore, to make the optical configuration very compact, specialized laser devices can be used for generating the twin-beams simultaneously. For example, in the visible spectral region, the second harmonic generation in PPLN crystals can be used for the simultaneous generation of twins beams at 532 nm [42

42. T. Ellenbogen, A. Ganany-Padowicz, and A. Arie, “Nonlinear photonic structures for all-optical deflection,” Opt. Express 16(5), 3077–3082 (2008). [CrossRef] [PubMed]

], while in mid-IR (10 μm) multibeams generation can be achieved by quantum cascade lasers [43

43. N. Yu, M. A. Kats, C. Pflügl, M. Geiser, Q. J. Wang, M. A. Belkin, F. Capasso, M. Fischer, A. Wittmann, J. Faist, T. Edamura, S. Furuta, M. Yamanishi, and H. Kan, “Multi-beam multi-wavelength semiconductor lasers,” Appl. Phys. Lett. 95(16), 161108 (2009). [CrossRef]

] in a region of the spectrum in which digital holography has been demonstrated too [41

41. E. Allaria, S. Brugioni, S. Denicola, P. Ferraro, S. Grilli, and R. Meucci, “Digital holography at 10.6 μm,” Opt. Commun. 215(4-6), 257–262 (2003). [CrossRef]

,44

44. M. Paturzo, A. Pelagotti, A. Finizio, L. Miccio, M. Locatelli, A. Gertrude, P. Poggi, R. Meucci, and P. Ferraro, “Optical reconstruction of digital holograms recorded at 10.6 microm: route for 3D imaging at long infrared wavelengths,” Opt. Lett. 35(12), 2112–2114 (2010). [CrossRef] [PubMed]

].

References and links

1.

D. Psaltis, S. R. Quake, and C. Yang, “Developing optofluidic technology through the fusion of microfluidics and optics,” Nature 442(7101), 381–386 (2006). [CrossRef] [PubMed]

2.

J. W. Hong and S. R. Quake, “Integrated nanoliter systems,” Nat. Biotechnol. 21(10), 1179–1183 (2003). [CrossRef] [PubMed]

3.

F. Dubois, C. Yourassowsky, O. Monnom, J. C. Legros, O. Debeir, P. Van Ham, R. Kiss, and C. Decaestecker, “Digital holographic microscopy for the three-dimensional dynamic analysis of in vitro cancer cell migration,” J. Biomed. Opt. 11(5), 054032 (2006). [CrossRef] [PubMed]

4.

B. Rappaz, F. Charrière, C. Depeursinge, P. J. Magistretti, and P. Marquet, “Simultaneous cell morphometry and refractive index measurement with dual-wavelength digital holographic microscopy and dye-enhanced dispersion of perfusion medium,” Opt. Lett. 33(7), 744–746 (2008). [CrossRef] [PubMed]

5.

W. M. Ash 3rd and M. K. Kim, “Digital holography of total internal reflection,” Opt. Express 16(13), 9811–9820 (2008). [CrossRef] [PubMed]

6.

B. Kemper, A. Vollmer, C. E. Rommel, J. Schnekenburger, and G. Bally, “Simplified approach for quantitative digital holographic phase contrast imaging of living cells,” J. Biomed. Opt. 16(2), 026014 (2011). [CrossRef] [PubMed]

7.

J. Garcia-Sucerquia, W. Xu, S. K. Jericho, M. H. Jericho, and H. J. Kreuzer, “4-D imaging of fluid flow with digital in-line holographic microscopy,” Optik 114(9), 419–423 (2007).

8.

G. Coppola, G. Di Caprio, M. Gioffré, R. Puglisi, D. Balduzzi, A. Galli, L. Miccio, M. Paturzo, S. Grilli, A. Finizio, and P. Ferraro, “Digital self-referencing quantitative phase microscopy by wavefront folding in holographic image reconstruction,” Opt. Lett. 35(20), 3390–3392 (2010). [CrossRef] [PubMed]

9.

C. Song, T.-D. Luong, T. F. Kong, N.-T. Nguyen, and A. K. Asundi, “Disposable flow cytometer with high efficiency in particle counting and sizing using an optofluidic lens,” Opt. Lett. 36(5), 657–659 (2011). [CrossRef] [PubMed]

10.

M. Antkowiak, M. L. Torres-Mapa, K. Dholakia, and F. J. Gunn-Moore, “Quantitative phase study of the dynamic cellular response in femtosecond laser photoporation,” Biomed. Opt. Express 1(2), 414–424 (2010). [CrossRef] [PubMed]

11.

W. Bishara, T.-W. Su, A. F. Coskun, and A. Ozcan, “Lensfree on-chip microscopy over a wide field-of-view using pixel super-resolution,” Opt. Express 18(11), 11181–11191 (2010). [CrossRef] [PubMed]

12.

N. T. Shaked, L. L. Satterwhite, N. Bursac, and A. Wax, “Whole-cell-analysis of live cardiomyocytes using wide-field interferometric phase microscopy,” Biomed. Opt. Express 1(2), 706–719 (2010). [CrossRef] [PubMed]

13.

Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, and G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express 19(2), 1016–1026 (2011). [CrossRef] [PubMed]

14.

L. Friedrich and A. Rohrbach, “Improved interferometric tracking of trapped particles using two frequency-detuned beams,” Opt. Lett. 35(11), 1920–1922 (2010). [CrossRef] [PubMed]

15.

M. J. Padgett and R. Di Leonardo, “Holographic optical tweezers and their relevance to lab on chip devices,” Lab Chip 11(7), 1196–1205 (2011). [CrossRef] [PubMed]

16.

P. Schiro, C. Dubois, and A. Kwok, “Large capture-range of a single-beam gradient optical trap,” Opt. Express 11(25), 3485–3489 (2003). [CrossRef] [PubMed]

17.

A. Rohrbach, H. Kress, and E. H. Stelzer, “Three-dimensional tracking of small spheres in focused laser beams: influence of the detection angular aperture,” Opt. Lett. 28(6), 411–413 (2003). [CrossRef] [PubMed]

18.

M. D. McMahon, A. J. Berglund, P. Carmichael, J. J. McClelland, and J. A. Liddle, “3D particle trajectories observed by orthogonal tracking microscopy,” ACS Nano 3(3), 609–614 (2009). [CrossRef] [PubMed]

19.

E. Toprak, H. Balci, B. H. Blehm, and P. R. Selvin, “Three-dimensional particle tracking via bifocal imaging,” Nano Lett. 7(7), 2043–2045 (2007). [CrossRef] [PubMed]

20.

H. P. Kao and A. S. Verkman, “Tracking of single fluorescent particles in three dimensions: use of cylindrical optics to encode particle position,” Biophys. J. 67(3), 1291–1300 (1994). [CrossRef] [PubMed]

21.

S. R. P. Pavani and R. Piestun, “Three dimensional tracking of fluorescent microparticles using a photon-limited double-helix response system,” Opt. Express 16(26), 22048–22057 (2008). [CrossRef] [PubMed]

22.

S. Ram, P. Prabhat, E. S. Ward, and R. J. Ober, “Improved single particle localization accuracy with dual objective multifocal plane microscopy,” Opt. Express 17(8), 6881–6898 (2009). [CrossRef] [PubMed]

23.

A. J. Berglund, M. D. McMahon, J. J. McClelland, and J. A. Liddle, “Fast, bias-free algorithm for tracking single particles with variable size and shape,” Opt. Express 16(18), 14064–14075 (2008). [CrossRef] [PubMed]

24.

Y. Park, G. Popescu, K. Badizadegan, R. R. Dasari, and M. S. Feld, “Fresnel particle tracing in three dimensions using diffraction phase microscopy,” Opt. Lett. 32(7), 811–813 (2007). [CrossRef] [PubMed]

25.

M. J. Mlodzianoski, M. F. Juette, G. L. Beane, and J. Bewersdorf, “Experimental characterization of 3D localization techniques for particle-tracking and super-resolution microscopy,” Opt. Express 17(10), 8264–8277 (2009). [CrossRef] [PubMed]

26.

M. DaneshPanah, S. Zwick, F. Schaal, M. Warber, B. Javidi, and W. Osten, “3D Holographic Imaging and Trapping for Non-Invasive Cell Identification and Tracking,” J. Disp. Technol. 6(10), 490–499 (2010). [CrossRef]

27.

J. A. Dominguez-Caballero, N. Loomis, G. Barbastathis, and J. Milgram, Techniques Based on Digital Multiplexing Holography for Three-Dimensional Object Tracking”, in Conference on Lasers and Electro-Optics/Quantum Electronics and Laser Science Conference and Photonic Applications Systems Technologies, OSA Technical Digest (CD) (Optical Society of America, 2007), paper JThD84.

28.

M.-T. Wei and A. Chiou, “Three-dimensional tracking of Brownian motion of a particle trapped in optical tweezers with a pair of orthogonal tracking beams and the determination of the associated optical force constants,” Opt. Express 13(15), 5798–5806 (2005). [CrossRef] [PubMed]

29.

S. O. Isikman, W. Bishara, H. Zhu, and A. Ozcan, “Optofluidic Tomography on a Chip,” Appl. Phys. Lett. 98(16), 161109 (2011). [CrossRef] [PubMed]

30.

H. Yang, N. Halliwell, and J. Coupland, “Application of the digital shearing method to extract three-component velocity in holographic particle image velocimetry,” Meas. Sci. Technol. 15(4), 694–698 (2004). [CrossRef]

31.

F. C. Cheong, B. J. Krishnatreya, and D. G. Grier, “Strategies for three-dimensional particle tracking with holographic video microscopy,” Opt. Express 18(13), 13563–13573 (2010). [CrossRef] [PubMed]

32.

O. Otto, F. Czerwinski, J. L. Gornall, G. Stober, L. B. Oddershede, R. Seidel, and U. F. Keyser, “Real-time particle tracking at 10,000 fps using optical fiber illumination,” Opt. Express 18(22), 22722–22733 (2010). [CrossRef] [PubMed]

33.

H. Hajjoul, S. Kocanova, I. Lassadi, K. Bystricky, and A. Bancaud, “Lab-on-Chip for fast 3D particle tracking in living cells,” Lab Chip 9(21), 3054–3058 (2009). [CrossRef] [PubMed]

34.

A. J. Berglund, M. D. McMahon, J. J. McClelland, and J. A. Liddle, “Theoretical model of errors in micromirror-based three-dimensional particle tracking,” Opt. Lett. 35(11), 1905–1907 (2010). [CrossRef] [PubMed]

35.

D. B. Conkey, R. P. Trivedi, S. R. P. Pavani, I. I. Smalyukh, and R. Piestun, “Three-dimensional parallel particle manipulation and tracking by integrating holographic optical tweezers and engineered point spread functions,” Opt. Express 19(5), 3835–3842 (2011). [CrossRef] [PubMed]

36.

B. Kemper, P. Langehanenberg, A. Hoink, G. von Bally, F. Wottowah, S. Schinkinger, J. Guck, J. Kas, I. Bredebusch, J. Schnekenburger, and K. Schütze, “Monitoring of laser micromanipulated optically trapped cells by digital holographic microscopy,” J. Biophotonics 3(7), 425–431 (2010). [CrossRef] [PubMed]

37.

F. Merola, L. Miccio, M. Paturzo, A. Finizio, S. Grilli, and P. Ferraro, “Driving and analysis of micro-objects by digital holographic microscope in microfluidics,” Opt. Lett. 36(16), 3079–3081 (2011). [CrossRef] [PubMed]

38.

R. C. Gonzalez and R. E. Woods, Digital Image Processing, 2nd ed. (Addison - Wesley Longman, Boston 1992).

39.

A. El Mallahi and F. Dubois, “Dependency and precision of the refocusing criterion based on amplitude analysis in digital holographic microscopy,” Opt. Express 19(7), 6684–6698 (2011). [CrossRef] [PubMed]

40.

L. Miccio, D. Alfieri, S. Grilli, P. Ferraro, A. Finizio, L. De Petrocellis, and S. D. Nicola, “Direct full compensation of the aberrations in quantitative phase microscopy of thin objects by a single digital hologram,” Appl. Phys. Lett. 90(4), 041104 (2007). [CrossRef]

41.

E. Allaria, S. Brugioni, S. Denicola, P. Ferraro, S. Grilli, and R. Meucci, “Digital holography at 10.6 μm,” Opt. Commun. 215(4-6), 257–262 (2003). [CrossRef]

42.

T. Ellenbogen, A. Ganany-Padowicz, and A. Arie, “Nonlinear photonic structures for all-optical deflection,” Opt. Express 16(5), 3077–3082 (2008). [CrossRef] [PubMed]

43.

N. Yu, M. A. Kats, C. Pflügl, M. Geiser, Q. J. Wang, M. A. Belkin, F. Capasso, M. Fischer, A. Wittmann, J. Faist, T. Edamura, S. Furuta, M. Yamanishi, and H. Kan, “Multi-beam multi-wavelength semiconductor lasers,” Appl. Phys. Lett. 95(16), 161108 (2009). [CrossRef]

44.

M. Paturzo, A. Pelagotti, A. Finizio, L. Miccio, M. Locatelli, A. Gertrude, P. Poggi, R. Meucci, and P. Ferraro, “Optical reconstruction of digital holograms recorded at 10.6 microm: route for 3D imaging at long infrared wavelengths,” Opt. Lett. 35(12), 2112–2114 (2010). [CrossRef] [PubMed]

OCIS Codes
(100.0100) Image processing : Image processing
(180.3170) Microscopy : Interference microscopy
(180.6900) Microscopy : Three-dimensional microscopy
(350.4990) Other areas of optics : Particles
(090.1995) Holography : Digital holography

ToC Category:
Holography

History
Original Manuscript: September 8, 2011
Revised Manuscript: October 19, 2011
Manuscript Accepted: October 20, 2011
Published: December 2, 2011

Virtual Issues
Vol. 7, Iss. 2 Virtual Journal for Biomedical Optics

Citation
Pasquale Memmolo, Andrea Finizio, Melania Paturzo, Lisa Miccio, and Pietro Ferraro, "Twin-beams digital holography for 3D tracking and quantitative phase-contrast microscopy in microfluidics," Opt. Express 19, 25833-25842 (2011)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-19-25-25833


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. D. Psaltis, S. R. Quake, and C. Yang, “Developing optofluidic technology through the fusion of microfluidics and optics,” Nature442(7101), 381–386 (2006). [CrossRef] [PubMed]
  2. J. W. Hong and S. R. Quake, “Integrated nanoliter systems,” Nat. Biotechnol.21(10), 1179–1183 (2003). [CrossRef] [PubMed]
  3. F. Dubois, C. Yourassowsky, O. Monnom, J. C. Legros, O. Debeir, P. Van Ham, R. Kiss, and C. Decaestecker, “Digital holographic microscopy for the three-dimensional dynamic analysis of in vitro cancer cell migration,” J. Biomed. Opt.11(5), 054032 (2006). [CrossRef] [PubMed]
  4. B. Rappaz, F. Charrière, C. Depeursinge, P. J. Magistretti, and P. Marquet, “Simultaneous cell morphometry and refractive index measurement with dual-wavelength digital holographic microscopy and dye-enhanced dispersion of perfusion medium,” Opt. Lett.33(7), 744–746 (2008). [CrossRef] [PubMed]
  5. W. M. Ash and M. K. Kim, “Digital holography of total internal reflection,” Opt. Express16(13), 9811–9820 (2008). [CrossRef] [PubMed]
  6. B. Kemper, A. Vollmer, C. E. Rommel, J. Schnekenburger, and G. Bally, “Simplified approach for quantitative digital holographic phase contrast imaging of living cells,” J. Biomed. Opt.16(2), 026014 (2011). [CrossRef] [PubMed]
  7. J. Garcia-Sucerquia, W. Xu, S. K. Jericho, M. H. Jericho, and H. J. Kreuzer, “4-D imaging of fluid flow with digital in-line holographic microscopy,” Optik114(9), 419–423 (2007).
  8. G. Coppola, G. Di Caprio, M. Gioffré, R. Puglisi, D. Balduzzi, A. Galli, L. Miccio, M. Paturzo, S. Grilli, A. Finizio, and P. Ferraro, “Digital self-referencing quantitative phase microscopy by wavefront folding in holographic image reconstruction,” Opt. Lett.35(20), 3390–3392 (2010). [CrossRef] [PubMed]
  9. C. Song, T.-D. Luong, T. F. Kong, N.-T. Nguyen, and A. K. Asundi, “Disposable flow cytometer with high efficiency in particle counting and sizing using an optofluidic lens,” Opt. Lett.36(5), 657–659 (2011). [CrossRef] [PubMed]
  10. M. Antkowiak, M. L. Torres-Mapa, K. Dholakia, and F. J. Gunn-Moore, “Quantitative phase study of the dynamic cellular response in femtosecond laser photoporation,” Biomed. Opt. Express1(2), 414–424 (2010). [CrossRef] [PubMed]
  11. W. Bishara, T.-W. Su, A. F. Coskun, and A. Ozcan, “Lensfree on-chip microscopy over a wide field-of-view using pixel super-resolution,” Opt. Express18(11), 11181–11191 (2010). [CrossRef] [PubMed]
  12. N. T. Shaked, L. L. Satterwhite, N. Bursac, and A. Wax, “Whole-cell-analysis of live cardiomyocytes using wide-field interferometric phase microscopy,” Biomed. Opt. Express1(2), 706–719 (2010). [CrossRef] [PubMed]
  13. Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, and G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express19(2), 1016–1026 (2011). [CrossRef] [PubMed]
  14. L. Friedrich and A. Rohrbach, “Improved interferometric tracking of trapped particles using two frequency-detuned beams,” Opt. Lett.35(11), 1920–1922 (2010). [CrossRef] [PubMed]
  15. M. J. Padgett and R. Di Leonardo, “Holographic optical tweezers and their relevance to lab on chip devices,” Lab Chip11(7), 1196–1205 (2011). [CrossRef] [PubMed]
  16. P. Schiro, C. Dubois, and A. Kwok, “Large capture-range of a single-beam gradient optical trap,” Opt. Express11(25), 3485–3489 (2003). [CrossRef] [PubMed]
  17. A. Rohrbach, H. Kress, and E. H. Stelzer, “Three-dimensional tracking of small spheres in focused laser beams: influence of the detection angular aperture,” Opt. Lett.28(6), 411–413 (2003). [CrossRef] [PubMed]
  18. M. D. McMahon, A. J. Berglund, P. Carmichael, J. J. McClelland, and J. A. Liddle, “3D particle trajectories observed by orthogonal tracking microscopy,” ACS Nano3(3), 609–614 (2009). [CrossRef] [PubMed]
  19. E. Toprak, H. Balci, B. H. Blehm, and P. R. Selvin, “Three-dimensional particle tracking via bifocal imaging,” Nano Lett.7(7), 2043–2045 (2007). [CrossRef] [PubMed]
  20. H. P. Kao and A. S. Verkman, “Tracking of single fluorescent particles in three dimensions: use of cylindrical optics to encode particle position,” Biophys. J.67(3), 1291–1300 (1994). [CrossRef] [PubMed]
  21. S. R. P. Pavani and R. Piestun, “Three dimensional tracking of fluorescent microparticles using a photon-limited double-helix response system,” Opt. Express16(26), 22048–22057 (2008). [CrossRef] [PubMed]
  22. S. Ram, P. Prabhat, E. S. Ward, and R. J. Ober, “Improved single particle localization accuracy with dual objective multifocal plane microscopy,” Opt. Express17(8), 6881–6898 (2009). [CrossRef] [PubMed]
  23. A. J. Berglund, M. D. McMahon, J. J. McClelland, and J. A. Liddle, “Fast, bias-free algorithm for tracking single particles with variable size and shape,” Opt. Express16(18), 14064–14075 (2008). [CrossRef] [PubMed]
  24. Y. Park, G. Popescu, K. Badizadegan, R. R. Dasari, and M. S. Feld, “Fresnel particle tracing in three dimensions using diffraction phase microscopy,” Opt. Lett.32(7), 811–813 (2007). [CrossRef] [PubMed]
  25. M. J. Mlodzianoski, M. F. Juette, G. L. Beane, and J. Bewersdorf, “Experimental characterization of 3D localization techniques for particle-tracking and super-resolution microscopy,” Opt. Express17(10), 8264–8277 (2009). [CrossRef] [PubMed]
  26. M. DaneshPanah, S. Zwick, F. Schaal, M. Warber, B. Javidi, and W. Osten, “3D Holographic Imaging and Trapping for Non-Invasive Cell Identification and Tracking,” J. Disp. Technol.6(10), 490–499 (2010). [CrossRef]
  27. J. A. Dominguez-Caballero, N. Loomis, G. Barbastathis, and J. Milgram, Techniques Based on Digital Multiplexing Holography for Three-Dimensional Object Tracking”, in Conference on Lasers and Electro-Optics/Quantum Electronics and Laser Science Conference and Photonic Applications Systems Technologies, OSA Technical Digest (CD) (Optical Society of America, 2007), paper JThD84.
  28. M.-T. Wei and A. Chiou, “Three-dimensional tracking of Brownian motion of a particle trapped in optical tweezers with a pair of orthogonal tracking beams and the determination of the associated optical force constants,” Opt. Express13(15), 5798–5806 (2005). [CrossRef] [PubMed]
  29. S. O. Isikman, W. Bishara, H. Zhu, and A. Ozcan, “Optofluidic Tomography on a Chip,” Appl. Phys. Lett.98(16), 161109 (2011). [CrossRef] [PubMed]
  30. H. Yang, N. Halliwell, and J. Coupland, “Application of the digital shearing method to extract three-component velocity in holographic particle image velocimetry,” Meas. Sci. Technol.15(4), 694–698 (2004). [CrossRef]
  31. F. C. Cheong, B. J. Krishnatreya, and D. G. Grier, “Strategies for three-dimensional particle tracking with holographic video microscopy,” Opt. Express18(13), 13563–13573 (2010). [CrossRef] [PubMed]
  32. O. Otto, F. Czerwinski, J. L. Gornall, G. Stober, L. B. Oddershede, R. Seidel, and U. F. Keyser, “Real-time particle tracking at 10,000 fps using optical fiber illumination,” Opt. Express18(22), 22722–22733 (2010). [CrossRef] [PubMed]
  33. H. Hajjoul, S. Kocanova, I. Lassadi, K. Bystricky, and A. Bancaud, “Lab-on-Chip for fast 3D particle tracking in living cells,” Lab Chip9(21), 3054–3058 (2009). [CrossRef] [PubMed]
  34. A. J. Berglund, M. D. McMahon, J. J. McClelland, and J. A. Liddle, “Theoretical model of errors in micromirror-based three-dimensional particle tracking,” Opt. Lett.35(11), 1905–1907 (2010). [CrossRef] [PubMed]
  35. D. B. Conkey, R. P. Trivedi, S. R. P. Pavani, I. I. Smalyukh, and R. Piestun, “Three-dimensional parallel particle manipulation and tracking by integrating holographic optical tweezers and engineered point spread functions,” Opt. Express19(5), 3835–3842 (2011). [CrossRef] [PubMed]
  36. B. Kemper, P. Langehanenberg, A. Hoink, G. von Bally, F. Wottowah, S. Schinkinger, J. Guck, J. Kas, I. Bredebusch, J. Schnekenburger, and K. Schütze, “Monitoring of laser micromanipulated optically trapped cells by digital holographic microscopy,” J. Biophotonics3(7), 425–431 (2010). [CrossRef] [PubMed]
  37. F. Merola, L. Miccio, M. Paturzo, A. Finizio, S. Grilli, and P. Ferraro, “Driving and analysis of micro-objects by digital holographic microscope in microfluidics,” Opt. Lett.36(16), 3079–3081 (2011). [CrossRef] [PubMed]
  38. R. C. Gonzalez and R. E. Woods, Digital Image Processing, 2nd ed. (Addison - Wesley Longman, Boston 1992).
  39. A. El Mallahi and F. Dubois, “Dependency and precision of the refocusing criterion based on amplitude analysis in digital holographic microscopy,” Opt. Express19(7), 6684–6698 (2011). [CrossRef] [PubMed]
  40. L. Miccio, D. Alfieri, S. Grilli, P. Ferraro, A. Finizio, L. De Petrocellis, and S. D. Nicola, “Direct full compensation of the aberrations in quantitative phase microscopy of thin objects by a single digital hologram,” Appl. Phys. Lett.90(4), 041104 (2007). [CrossRef]
  41. E. Allaria, S. Brugioni, S. Denicola, P. Ferraro, S. Grilli, and R. Meucci, “Digital holography at 10.6 μm,” Opt. Commun.215(4-6), 257–262 (2003). [CrossRef]
  42. T. Ellenbogen, A. Ganany-Padowicz, and A. Arie, “Nonlinear photonic structures for all-optical deflection,” Opt. Express16(5), 3077–3082 (2008). [CrossRef] [PubMed]
  43. N. Yu, M. A. Kats, C. Pflügl, M. Geiser, Q. J. Wang, M. A. Belkin, F. Capasso, M. Fischer, A. Wittmann, J. Faist, T. Edamura, S. Furuta, M. Yamanishi, and H. Kan, “Multi-beam multi-wavelength semiconductor lasers,” Appl. Phys. Lett.95(16), 161108 (2009). [CrossRef]
  44. M. Paturzo, A. Pelagotti, A. Finizio, L. Miccio, M. Locatelli, A. Gertrude, P. Poggi, R. Meucci, and P. Ferraro, “Optical reconstruction of digital holograms recorded at 10.6 microm: route for 3D imaging at long infrared wavelengths,” Opt. Lett.35(12), 2112–2114 (2010). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Multimedia

Multimedia FilesRecommended Software
» Media 1: AVI (393 KB)      QuickTime
» Media 2: AVI (152 KB)      QuickTime
» Media 3: AVI (1150 KB)      QuickTime

« Previous Article

OSA is a member of CrossRef.

CrossCheck Deposited