Determination of the scattering anisotropy with optical coherence tomography

V. M. Kodach, D. J. Faber, J. van Marle, T. G. van Leeuwen, and J. Kalkman  »View Author Affiliations


Optics Express, Vol. 19, Issue 7, pp. 6131-6140 (2011)
http://dx.doi.org/10.1364/OE.19.006131


View Full Text Article

Acrobat PDF (1115 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

In this work we demonstrate measurements with optical coherence tomography (OCT) of the scattering phase function in the backward direction and the scattering anisotropy parameter g. Measurements of the OCT attenuation coefficient and the backscattering amplitude are performed on calibrated polystyrene microspheres with a time-domain OCT system. From these measurements the phase function in the backward direction is determined. The measurements are described by the single scattering model and match Mie calculations very well. Measurements on Intralipid demonstrate the ability to determine the g of polydisperse samples and, for Intralipid, g = 0.35 ± 0.03 is measured, which is well in agreement with g from literature. These measurements are validated using the Intralipid particle size distribution determined from TEM measurements. Measurements of g and the scattering phase function in the backward direction can be used to monitor changes in backscattering, which can indicate morphological changes of the sample or act as contrast enhancement mechanism.

© 2011 OSA

1. Introduction

Optical coherence tomography (OCT) is a non-invasive imaging modality that is used to image the morphology of turbid media. An important application of OCT is to quantitatively determine the optical properties of tissue. The attenuation coefficient of the OCT signal can be used to characterize tissue [1

1. D. J. Faber, F. J. van der Meer, M. C. G. Aalders, and T. G. van Leeuwen, “Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography,” Opt. Express 12(19), 4353–4365 (2004). [CrossRef] [PubMed]

3

3. F. J. van der Meer, D. J. Faber, J. Perree, G. Pasterkamp, D. B. Sassoon, and T. G. van Leeuwen, "Quantitative optical coherence tomography of arterial wall components," Lasers in Medical Science 20, 45-51 (2005). [CrossRef] [PubMed]

] and to determine the OCT imaging depth [4

4. V. M. Kodach, J. Kalkman, D. J. Faber, and T. G. van Leeuwen, “Quantitative comparison of the OCT imaging depth at 1300 nm and 1600 nm,” Biomed. Opt. Express 1(1), 176–185 (2010). [CrossRef]

]. In contrast to frequently reported measurements of the OCT attenuation coefficient, quantitative measurements of the OCT backscattering are reported less often. The OCT backscattering coefficient can be quantified [5

5. D. Levitz, L. Thrane, M. H. Frosz, P. E. Andersen, C. B. Andersen, S. Andersson-Engels, J. Valanciunaite, J. Swartling, and P. R. Hansen, “Determination of optical scattering properties of highly-scattering media in optical coherence tomography images,” Opt. Express 12(2), 249–259 (2004). [CrossRef] [PubMed]

] and used for tissue characterization, visualization, and/or contrast enhancement [6

6. C. Xu, J. M. Schmitt, S. G. Carlier, and R. Virmani, “Characterization of atherosclerosis plaques by measuring both backscattering and attenuation coefficients in optical coherence tomography,” J. Biomed. Opt. 13(3), 034003 (2008). [CrossRef] [PubMed]

]. In addition, the variation of the OCT magnitude in time can provide quantitative information about flow [7

7. Y. Wang and R. Wang, “Autocorrelation optical coherence tomography for mapping transverse particle-flow velocity,” Opt. Lett. 35(21), 3538–3540 (2010). [CrossRef] [PubMed]

] and diffusion [8

8. J. Kalkman, R. Sprik, and T. G. van Leeuwen, “Path-length-resolved diffusive particle dynamics in spectral-domain optical coherence tomography,” Phys. Rev. Lett. 105(19), 198302 (2010). [CrossRef]

]. Quantitative measurement of the OCT (or low coherence interferometry) magnitude as a function of angle and/or wavelength [9

9. J. Pyhtila, R. Graf, and A. Wax, “Determining nuclear morphology using an improved angle-resolved low coherence interferometry system,” Opt. Express 11(25), 3473–3484 (2003). [CrossRef] [PubMed]

, 10

10. A. E. Desjardins, B. J. Vakoc, G. J. Tearney, and B. E. Bouma, “Backscattering spectroscopic contrast with angle-resolved optical coherence tomography,” Opt. Lett. 32(21), 3158–3160 (2007). [CrossRef] [PubMed]

] can provide information on the scattering phase function and the size of the scatterer.

The scattering phase function, which describes the angular probability distribution of the scattered light, is parameterized by the scattering anisotropy parameter g, which is the average of the cosine of the scattering angle. The scattering anisotropy plays an important role in diffusive light transport, for example in diffuse reflectance spectroscopy. Also in OCT, g plays an important role as it determines the amount of backscattering and in the quantification of multiple scattering effects in (Doppler) OCT [11

11. J. Kalkman, A. V. Bykov, D. J. Faber, and T. G. van Leeuwen, “Multiple and dependent scattering effects in Doppler optical coherence tomography,” Opt. Express 18(4), 3883–3892 (2010). [CrossRef] [PubMed]

]. The possibility to extract g from OCT measurements is very attractive, because, in comparison with commonly used goniometric or integrating sphere methods, the non-invasive nature of OCT allows for in-vivo application.

Here we present a simple method based on the single scattering model to determine g from OCT measurements. We demonstrate experimentally, for the first time to our knowledge, that this model can be used for measurements of the scattering phase function in the backward direction and in specific cases for measuring g.

2. Materials and methods

2.1 Single scattering model of the OCT signal

2.2 OCT measurements

For our study we utilize a home-built time-domain OCT system, which is described in detail in Ref [4

4. V. M. Kodach, J. Kalkman, D. J. Faber, and T. G. van Leeuwen, “Quantitative comparison of the OCT imaging depth at 1300 nm and 1600 nm,” Biomed. Opt. Express 1(1), 176–185 (2010). [CrossRef]

]. In summary, light from a Fianium light source is filtered to obtain a spectrum centered at 1300 nm, which is coupled into a fixed focus OCT setup. The axial resolution of this system is 9.7 ± 0.1 µm as was determined from the full width at half maximum of the OCT magnitude point spread function. The signal to noise ratio is 90 dB. The Rayleigh length, measured with a mirror in the sample arm, is 292 ± 9 µm. The corresponding Gaussian beam waist is 9.6 ± 0.2 µm and the numerical aperture NA of the sample arm lens is 0.043 ± 0.001. Prior to the experiment, the OCT system is calibrated for quantitative measurements of the backscattered power. Due to the limited dynamic range of the OCT system, the power from the mirror is measured using different calibrated neutral density filters in the sample arm. From the dependence of the OCT signal on the optical attenuation the reflected power and the OCT magnitude can be directly compared to the signals for the scattering sample (no optical attenuation in the sample arm) in Eq. (6).

OCT measurements on suspensions of scatterers are performed in a 1 mm thick glass cuvette. The cuvette is placed in the sample arm at ~70° angle relative to the incident beam to avoid specular reflections in the OCT signal. The sample arm beam is focused at the first glass-medium interface. Measurements for every solution are performed independently for 5 times. For each measurement, the average of 100 A-scans is taken. After background subtraction the OCT signal magnitude is corrected for the confocal point spread function [16

16. T. G. van Leeuwen, D. J. Faber, and M. C. Aalders, “Measurement of the axial point spread function in scattering media using single-mode fiber-based optical coherence tomography,” IEEE J. Sel. Top. Quantum Electron. 9(2), 227–233 (2003). [CrossRef]

]. The OCT attenuation coefficient is determined with a two parameter single exponential fit of the measured OCT signal in depth (corrected for the refractive index of water). To reduce the effects of multiple scattering, the scatterers concentration is kept low to create samples with scattering coefficients below 5 mm−1 (as calculated with Mie theory). In addition, only the first 500 data points (190 µm) of the OCT signal (starting at ~60 µm depth after sample front surface) are used for fitting the single exponential decay. The scattering coefficient is calculated by subtracting the water absorption coefficient (µa= 0.2 mm−1) from the fitted attenuation coefficient. Finally, the scattering cross section σscat is calculated by dividing the scattering coefficient by the known particle concentration. The OCT signal amplitude at the front glass-sample boundary is determined from the exponential fit by extending the fitting line to zero depth (see Fig. 1
Fig. 1 OCT measurement of backscattered power from a suspension of 400 nm diameter polystyrene microspheres in water. The vertical scale on the left is converted to absolute units of square root of power using the power calibration and is indicated on the right. The attenuation coefficient is determined from the single exponential decay fit (solid red line); the amplitude of the OCT signal from the sample surface is determined by extending the exponential fit to zero depth (dashed red line). Zero depth is indicated (dashed blue line).
). Zero depth was determined from the crossing of the OCT signal with the vertical drop line at half height. With this method the OCT signal magnitude is determined on an absolute scale in [mW1/2] units. This is indicated in Fig. 1 on the right hand side scale for a measurement of a scattering medium for 400 nm diameter particles. The peak of the backscattered power in the heterodyne (OCT) signal is of the order of 90 picoWatts.

2.3 Scattering samples

Polystyrene microspheres (Thermo Scientific, USA) are used as scatterers, certified by NIST traceable procedures, with the concentration of the sample calculated based on the used dilution (1 wt.% concentration). The microspheres have mean diameters of 203 ± 5, 400 ± 9, 596 ± 6, 799 ± 9, and 994 ± 10 nm and size distribution standard deviations of 4.7, 7.3, 7.7, 4.8 and 10 nm, respectively. Mie calculations are performed based on mean diameters to calculate the scattering cross section and phase function of the polystyrene particles. The refractive index of water (nwater=1.32) and polystyrene are used (npolyst=1.57 [18

18. A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, V. Yakovlev, G. Tempea, C. Schubert, E. M. Anger, P. K. Ahnelt, M. Stur, J. E. Morgan, A. Cowey, G. Jung, T. Le, and A. Stingl, “Compact, low-cost Ti:Al2O3 laser for in vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett. 28(11), 905–907 (2003). [CrossRef] [PubMed]

]) as input. From Mie calculations, the scattering anisotropy, i.e. the g of these microspheres, is calculated to be: 0.07, 0.29, 0.62, 0.73, and 0.81 for increasing sphere diameter. Also, from the calculated phase functions, the scattering efficiency in the backscattering direction pNA is calculated by integrating the phase function over the NA of the sample arm focusing lens.

Measurements performed on Intralipid samples are used as an example of our technique to a polydisperse medium. Intralipid is an aqueous suspension of polydisperse lipid droplets, which is often used as a tissue phantom for optical measurements. Recently we showed that for high Intralipid concentrations multiple and dependent scattering effects play an important role and lead to nonlinear changes in the scattering coefficient with concentration [11

11. J. Kalkman, A. V. Bykov, D. J. Faber, and T. G. van Leeuwen, “Multiple and dependent scattering effects in Doppler optical coherence tomography,” Opt. Express 18(4), 3883–3892 (2010). [CrossRef] [PubMed]

]. To avoid these effects, measurements with low concentrations Intralipid (0.63, 1.25 and 2.5 wt.%) are performed. The samples are prepared by dilution of a single batch of 20 wt.% Intralipid (Fresenius-Kabi) with deionized water. The refractive indexes used for Mie calculations of Intralipid are: nwater=1.32 for water; nlip=1.46 for lipid droplets [19

19. H. J. van Staveren, C. J. M. Moes, J. van Marie, S. A. Prahl, and M. J. C. van Gemert, “Light scattering in Intralipid-10% in the wavelength range of 400-1100 nm,” Appl. Opt. 30(31), 4507–4514 (1991). [CrossRef] [PubMed]

].

To correlate our OCT measurements to the size of the scatterers in the polydisperse Intralipid suspension, transmission electron microscopy (TEM) measurements are performed to determine the size distribution of the scatterers in Intralipid. For TEM imaging the samples were cryoprotected with glycerin and frozen in liquid ethane at the temperature of liquid nitrogen. The samples were replicated in a vacuum better than 3×10−7 mbar at a temperature of −120°C. A platinum layer (2 nm) was evaporated at an angle of 45° and carbon (20 nm) was deposited at an angle of 90°. The replicas were cleaned with house hold bleach and collected on 300 mesh copper grids. The replicas were then imaged with a FEI T2 electron microscope at a magnification of 26500. Images were collected with a SIS Velata camera in a 2048 × 2048 format. QWin-pro (Leica) software was used to determine the diameters. In total, 2019 particle diameters were measured.

3. Results

3.1 Polystyrene microspheres

From an OCT measurement, like the measurement shown in Fig. 1, the scattering coefficient for a solution of particles is determined. Figure 2(a)
Fig. 2 Results of Mie calculations and experimental measurements of: a) scattering cross-section of polystyrene microspheres (error bars are smaller than symbols); b) scattering phase function pNA of polystyrene microspheres (error bars are standard deviations); the dashed line indicates the measured value of pNA for Intralipid (standard deviation is given in the text); c) Mie calculations of g versus particle diameter at 1300 nm for polystyrene microspheres. Arrows show the average particle diameter and the g of Intralipid. Dotted lines indicate the limits of our method for particle diameter and scattering anisotropy (grey zone – region of applicability).
shows the scattering cross section of polystyrene microspheres for the different diameters obtained from the scattering coefficient. The experimental results are compared to Mie calculations and good agreement is observed (typical error is within 10%). Consequently, it can be assumed that multiple scattering effects are negligible and that the single scattering model is valid for a description of the OCT signal. From the data in Fig. 2(a) and the measured OCT magnitude at the interface, pNA is determined using Eq. (6). Measurements of pNA for all diameters are used to calculate the average heterodyne intensity back-coupling efficiency Q = 0.26 ± 0.04, which is used to compare pNA to Mie calculations in Fig. 2(b). The experimental points match the calculated values reasonably well and the oscillations in pNA due to the diameter dependent lobe structure of the backscattering efficiency are clearly observable.

3.2 Intralipid

The phase function in the backscattering direction pNA can be used to estimate the average size of the scatterers and, consequently, g. To determine the average diameter, the crossing point of the horizontal line through the experimental point of pNA and the calculated curve of pNA for each scatterer size has to be found. The obtained crossing point specifies the average particle diameter (Fig. 2(b)). Finally, the average particle diameter corresponds to a scattering anisotropy g (Fig. 2(c)).

To demonstrate our method on non-calibrated samples we apply it to Intralipid as a polydisperse suspension of scatterers. Following the same procedure as for polystyrene microspheres, pNA of Intralipid is determined. For low particle concentration, pNA is independent of the concentration of scatterers and the pNA values for all measured Intralipid concentrations (1.30·10−4, 1.26·10−4 and 1.11·10−4 for 0.63, 1.25 and 2.5 wt.% Intralipid correspondingly) are averaged. The resulting value (pNA = (1.22 ± 0.21)·10−4) is plotted in Fig. 2(b) (although the refractive indexes of polystyrene spheres and Intralipid droplets are different, this has negligible influence on the calculated pNA). From Fig. 2(b), the crossing point of the line of the calculated pNA of polystyrene microspheres and the measured pNA of Intralipid corresponds to a particle diameter of 438 ± 21 nm. Consequently, from Fig. 2(c) where we calculate g for varying particle diameter, this particle diameter corresponds to the scattering anisotropy g = 0.35 ± 0.03. This value agrees very well with previously reported measurements of g for Intralipid at 1300 nm g = 0.32 ± 0.07 [20

20. C. Chen, J. Q. Lu, H. F. Ding, K. M. Jacobs, Y. Du, and X. H. Hu, “A primary method for determination of optical parameters of turbid samples and application to intralipid between 550 and 1630 nm,” Opt. Express 14(16), 7420–7435 (2006). [CrossRef] [PubMed]

].

To validate our experimental results, Mie theory calculations are performed based on the Intralipid particle size distribution measured by TEM, see Fig. 3
Fig. 3 Particle size distribution of Intralipid measured by transmission electron microscopy. In total, 2019 cross-sections were measured. The mean diameter is 214 nm. Inset: TEM image of Intralipid (bar – 500 nm).
. The fraction of particles f vs. particle diameter d is given in the Table 1

Table 1. Particle size distribution of Intralipid.

table-icon
View This Table
. The mean particle diameter is calculated to be 214 nm. Based on the particle size distribution in Table 1, the average scattering phase function pNA of Intralipid in the backscattering direction is calculated using [21

21. J. R. Mourant, J. P. Freyer, A. H. Hielscher, A. A. Eick, D. Shen, and T. M. Johnson, “Mechanisms of light scattering from biological cells relevant to noninvasive optical-tissue diagnostics,” Appl. Opt. 37(16), 3586–3593 (1998). [CrossRef]

]
pNA=dpNA(d)μs(d)dμs(d)=dpNA(d)σsc(d)f(d)dσsc(d)f(d)
(7)
which is the average of pNA over all particle diameters weighted with the scattering coefficient for every diameter. The resulting average phase function of Intralipid in the backward direction is pNA= 0.94·10−4. This corresponds to an average scatterer diameter d = 465 nm, which gives a scattering anisotropy g = 0.40.

4. Discussion

4.1 Method

With the demonstrated procedure it is possible to determine g for monodisperse polystyrene particles with diameters smaller than ~500 nm (at an OCT operating wavelength of 1300 nm). With increasing particle diameter oscillations of the scattering phase function become present and cause the measured value of pNA to coincide with a multitude of diameters, which makes a unique determination of g impossible (Fig. 2(b)). However, using additional measurements (e.g. at different wavelengths [22

22. N. Bosschaart, D. J. Faber, T. G. van Leeuwen, and M. C. G. Aalders, “Measurements of wavelength dependent scattering and backscattering coefficients by low-coherence spectroscopy,” J. Biomed.Opt. 16, 030503 (2011). [PubMed]

] or different angles [9

9. J. Pyhtila, R. Graf, and A. Wax, “Determining nuclear morphology using an improved angle-resolved low coherence interferometry system,” Opt. Express 11(25), 3473–3484 (2003). [CrossRef] [PubMed]

]) or using additional information (e.g. the particle concentration) the average size of the scatterers and the scattering anisotropy g also can be determined for particles larger than ~500 nm.

However, even without the possibility to perform quantitative measurements, relative changes in the scattering phase function can be used to detect changes in tissue morphology and/or act as a contrast enhancement mechanism in OCT images. Furthermore, in contrast with the backscattering coefficient, the scattering phase function does not depend on the concentration of scatterers and therefore can distinguish between changes in scattering due to refractive index contrast changes or due to changes in the scatterer size.

The accuracy in the determination of pNA and g depends on a large number of factors. As can be seen from Eq. (6), μs, lc, NA and Q are parameters that need to be accurately determined to calculate pNA. The errors in all these parameters are summarized in the determination of the heterodyne intensity back-coupling efficiency Q, which has an error of 15%. Any systematic error in Q and its relation to the theoretical value needs to be investigated (see section 4.3) as well as methods to reduce the error in Q.

4.2 Results

The physical size distribution for Intralipid (Fig. 3) that we obtained shows a size distribution with a decreasing number of particles with increasing size. The average physical diameter is 214 nm, which is larger than 97 nm, reported by van Staveren et al. [19

19. H. J. van Staveren, C. J. M. Moes, J. van Marie, S. A. Prahl, and M. J. C. van Gemert, “Light scattering in Intralipid-10% in the wavelength range of 400-1100 nm,” Appl. Opt. 30(31), 4507–4514 (1991). [CrossRef] [PubMed]

], which was measured for a different weight fraction solution and of Intralipid from a different manufacturer. Also our average size is larger than 123 nm, as was measured by Michels et al. [23

23. R. Michels, F. Foschum, and A. Kienle, “Optical properties of fat emulsions,” Opt. Express 16(8), 5907–5925 (2008). [CrossRef] [PubMed]

], however, that value was obtained with an indirect method using integrating spheres measurements correlated with Monte Carlo simulations.

For polydisperse particle solutions the described method results in values for the size d and scattering anisotropy g that are averaged over the size distribution. However, since these properties are measured using an optical technique instead of a direct size measurement technique, the averages tend to underestimate the contribution of small particles since the scattering strength decreases strongly with decreasing size. For Intralipid, for example, the difference between the average physical diameter of the scatterers (214 nm) and the optically measured average scatterer diameter (438 nm) demonstrates the strong impact of the size determination method being used (TEM versus OCT). This is a well known phenomenon in, for example, dynamic light scattering where the different sizes are weighted with their respective scattering strength.

Like for the average diameter, also for g the described method is more sensitive to larger diameters [21

21. J. R. Mourant, J. P. Freyer, A. H. Hielscher, A. A. Eick, D. Shen, and T. M. Johnson, “Mechanisms of light scattering from biological cells relevant to noninvasive optical-tissue diagnostics,” Appl. Opt. 37(16), 3586–3593 (1998). [CrossRef]

]. However, since g is a parameter describing diffusive light transport, its optically weighted value is the only relevant parameter. The experimental value we obtain for Intralipid (g = 0.35 ± 0.03) is in good agreement with the value obtained using diffuse reflection spectroscopy (0.32 ± 0.07). In comparison to g determined from the Intralipid particle size distribution measured with TEM (g = 0.40), the value we obtained from our OCT measurements is lower. This difference can be explained by the underestimation of the small diameter particles in the TEM measurements.

4.3 Efficiency factor Q

The heterodyne detection efficiency factor Q for small scatterers is of fundamental interest in OCT [6

6. C. Xu, J. M. Schmitt, S. G. Carlier, and R. Virmani, “Characterization of atherosclerosis plaques by measuring both backscattering and attenuation coefficients in optical coherence tomography,” J. Biomed. Opt. 13(3), 034003 (2008). [CrossRef] [PubMed]

, 13

13. J. M. Schmitt, A. Knuttel, and R. F. Bonner, “Measurement of optical properties of biological tissues by low-coherence reflectometry,” Appl. Opt. 32(30), 6032–6042 (11993). [CrossRef]

, 24

24. B. Karamata, M. Laubscher, M. Leutenegger, S. Bourquin, T. Lasser, and P. Lambelet, “Multiple scattering in optical coherence tomography. I. Investigation and modeling,” J. Opt. Soc. Am. A 22(7), 1369–1379 (2005). [CrossRef]

] and we identified two major reasons that cause a reduction of the heterodyne detection efficiency Q for a homogenous distribution of scatterers compared to a mirror to values less than 1. First, the OCT magnitude for a sum of scatterers depends on the random axial position of the scatterers in the sample (causing speckle). As a result, the envelope of the OCT signal (e.g. the mean value of the speckle pattern) is smaller than the addition of all individual particle envelopes. Second, the OCT backscatter efficiency for small particles depends on lateral offset of the scatterer in the focused beam. The spherical wave originating from a particle is less efficiently coupled back to the single mode fiber compared to the planar wavefront reflected from the mirror. A full quantification of these effects is subject to future work.

5. Conclusions

In conclusion, we have presented experimental measurements of the scattering phase function in the backward direction pNA by OCT and demonstrated the feasibility of such measurements to determine the scattering anisotropy g. The proposed method is well applicable to scatterers with diameters below ~500 nm (for measurements at 1300 nm). We applied this method to a polydisperse Intralipid solution and obtain g = 0.35 ± 0.03, which is close to that found in literature.

Acknowledgments

V. M. Kodach and J. Kalkman are supported by the IOP Photonic Devices program (IPD067774) managed by the Technology Foundation STW and SenterNovem. D. J. Faber is funded by a personal grant in the Vernieuwingsimpuls program by the Netherlands Organization of Scientific Research (NWO) and the Technology Foundation STW (AGT07544).

References and Links

1.

D. J. Faber, F. J. van der Meer, M. C. G. Aalders, and T. G. van Leeuwen, “Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography,” Opt. Express 12(19), 4353–4365 (2004). [CrossRef] [PubMed]

2.

F. J. van der Meer, D. J. Faber, D. M. Baraznji Sassoon, M. C. Aalders, G. Pasterkamp, and T. G. van Leeuwen, “Localized measurement of optical attenuation coefficients of atherosclerotic plaque constituents by quantitative optical coherence tomography,” IEEE Trans. Med. Imaging 24(10), 1369–1376 (2005). [CrossRef] [PubMed]

3.

F. J. van der Meer, D. J. Faber, J. Perree, G. Pasterkamp, D. B. Sassoon, and T. G. van Leeuwen, "Quantitative optical coherence tomography of arterial wall components," Lasers in Medical Science 20, 45-51 (2005). [CrossRef] [PubMed]

4.

V. M. Kodach, J. Kalkman, D. J. Faber, and T. G. van Leeuwen, “Quantitative comparison of the OCT imaging depth at 1300 nm and 1600 nm,” Biomed. Opt. Express 1(1), 176–185 (2010). [CrossRef]

5.

D. Levitz, L. Thrane, M. H. Frosz, P. E. Andersen, C. B. Andersen, S. Andersson-Engels, J. Valanciunaite, J. Swartling, and P. R. Hansen, “Determination of optical scattering properties of highly-scattering media in optical coherence tomography images,” Opt. Express 12(2), 249–259 (2004). [CrossRef] [PubMed]

6.

C. Xu, J. M. Schmitt, S. G. Carlier, and R. Virmani, “Characterization of atherosclerosis plaques by measuring both backscattering and attenuation coefficients in optical coherence tomography,” J. Biomed. Opt. 13(3), 034003 (2008). [CrossRef] [PubMed]

7.

Y. Wang and R. Wang, “Autocorrelation optical coherence tomography for mapping transverse particle-flow velocity,” Opt. Lett. 35(21), 3538–3540 (2010). [CrossRef] [PubMed]

8.

J. Kalkman, R. Sprik, and T. G. van Leeuwen, “Path-length-resolved diffusive particle dynamics in spectral-domain optical coherence tomography,” Phys. Rev. Lett. 105(19), 198302 (2010). [CrossRef]

9.

J. Pyhtila, R. Graf, and A. Wax, “Determining nuclear morphology using an improved angle-resolved low coherence interferometry system,” Opt. Express 11(25), 3473–3484 (2003). [CrossRef] [PubMed]

10.

A. E. Desjardins, B. J. Vakoc, G. J. Tearney, and B. E. Bouma, “Backscattering spectroscopic contrast with angle-resolved optical coherence tomography,” Opt. Lett. 32(21), 3158–3160 (2007). [CrossRef] [PubMed]

11.

J. Kalkman, A. V. Bykov, D. J. Faber, and T. G. van Leeuwen, “Multiple and dependent scattering effects in Doppler optical coherence tomography,” Opt. Express 18(4), 3883–3892 (2010). [CrossRef] [PubMed]

12.

L. Thrane, H. T. Yura, and P. E. Andersen, “Analysis of optical coherence tomography systems based on the extended Huygens-Fresnel principle,” J. Opt. Soc. Am. A 17(3), 484–490 (2000). [CrossRef]

13.

J. M. Schmitt, A. Knuttel, and R. F. Bonner, “Measurement of optical properties of biological tissues by low-coherence reflectometry,” Appl. Opt. 32(30), 6032–6042 (11993). [CrossRef]

14.

D. J. Faber and T. G. van Leeuwen, “Are quantitative attenuation measurements of blood by optical coherence tomography feasible?” Opt. Lett. 34(9), 1435–1437 (2009). [CrossRef] [PubMed]

15.

J. W. Goodman, Statistical optics (Wiley, New York, 1985).

16.

T. G. van Leeuwen, D. J. Faber, and M. C. Aalders, “Measurement of the axial point spread function in scattering media using single-mode fiber-based optical coherence tomography,” IEEE J. Sel. Top. Quantum Electron. 9(2), 227–233 (2003). [CrossRef]

17.

C. Akcay, P. Parrein, and J. P. Rolland, “Estimation of Longitudinal Resolution in Optical Coherence Imaging,” Appl. Opt. 41(25), 5256–5262 (2002). [CrossRef] [PubMed]

18.

A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, V. Yakovlev, G. Tempea, C. Schubert, E. M. Anger, P. K. Ahnelt, M. Stur, J. E. Morgan, A. Cowey, G. Jung, T. Le, and A. Stingl, “Compact, low-cost Ti:Al2O3 laser for in vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett. 28(11), 905–907 (2003). [CrossRef] [PubMed]

19.

H. J. van Staveren, C. J. M. Moes, J. van Marie, S. A. Prahl, and M. J. C. van Gemert, “Light scattering in Intralipid-10% in the wavelength range of 400-1100 nm,” Appl. Opt. 30(31), 4507–4514 (1991). [CrossRef] [PubMed]

20.

C. Chen, J. Q. Lu, H. F. Ding, K. M. Jacobs, Y. Du, and X. H. Hu, “A primary method for determination of optical parameters of turbid samples and application to intralipid between 550 and 1630 nm,” Opt. Express 14(16), 7420–7435 (2006). [CrossRef] [PubMed]

21.

J. R. Mourant, J. P. Freyer, A. H. Hielscher, A. A. Eick, D. Shen, and T. M. Johnson, “Mechanisms of light scattering from biological cells relevant to noninvasive optical-tissue diagnostics,” Appl. Opt. 37(16), 3586–3593 (1998). [CrossRef]

22.

N. Bosschaart, D. J. Faber, T. G. van Leeuwen, and M. C. G. Aalders, “Measurements of wavelength dependent scattering and backscattering coefficients by low-coherence spectroscopy,” J. Biomed.Opt. 16, 030503 (2011). [PubMed]

23.

R. Michels, F. Foschum, and A. Kienle, “Optical properties of fat emulsions,” Opt. Express 16(8), 5907–5925 (2008). [CrossRef] [PubMed]

24.

B. Karamata, M. Laubscher, M. Leutenegger, S. Bourquin, T. Lasser, and P. Lambelet, “Multiple scattering in optical coherence tomography. I. Investigation and modeling,” J. Opt. Soc. Am. A 22(7), 1369–1379 (2005). [CrossRef]

OCIS Codes
(170.3880) Medical optics and biotechnology : Medical and biological imaging
(170.4500) Medical optics and biotechnology : Optical coherence tomography
(290.7050) Scattering : Turbid media

ToC Category:
Medical Optics and Biotechnology

History
Original Manuscript: January 19, 2011
Revised Manuscript: March 1, 2011
Manuscript Accepted: March 3, 2011
Published: March 17, 2011

Virtual Issues
Vol. 6, Iss. 4 Virtual Journal for Biomedical Optics

Citation
V. M. Kodach, D. J. Faber, J. van Marle, T. G. van Leeuwen, and J. Kalkman, "Determination of the scattering anisotropy with optical coherence tomography," Opt. Express 19, 6131-6140 (2011)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-19-7-6131


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. D. J. Faber, F. J. van der Meer, M. C. G. Aalders, and T. G. van Leeuwen, “Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography,” Opt. Express 12(19), 4353–4365 (2004). [CrossRef] [PubMed]
  2. F. J. van der Meer, D. J. Faber, D. M. Baraznji Sassoon, M. C. Aalders, G. Pasterkamp, and T. G. van Leeuwen, “Localized measurement of optical attenuation coefficients of atherosclerotic plaque constituents by quantitative optical coherence tomography,” IEEE Trans. Med. Imaging 24(10), 1369–1376 (2005). [CrossRef] [PubMed]
  3. F. J. van der Meer, D. J. Faber, J. Perree, G. Pasterkamp, D. B. Sassoon, and T. G. van Leeuwen, "Quantitative optical coherence tomography of arterial wall components," Lasers in Medical Science 20, 45-51 (2005). [CrossRef] [PubMed]
  4. V. M. Kodach, J. Kalkman, D. J. Faber, and T. G. van Leeuwen, “Quantitative comparison of the OCT imaging depth at 1300 nm and 1600 nm,” Biomed. Opt. Express 1(1), 176–185 (2010). [CrossRef]
  5. D. Levitz, L. Thrane, M. H. Frosz, P. E. Andersen, C. B. Andersen, S. Andersson-Engels, J. Valanciunaite, J. Swartling, and P. R. Hansen, “Determination of optical scattering properties of highly-scattering media in optical coherence tomography images,” Opt. Express 12(2), 249–259 (2004). [CrossRef] [PubMed]
  6. C. Xu, J. M. Schmitt, S. G. Carlier, and R. Virmani, “Characterization of atherosclerosis plaques by measuring both backscattering and attenuation coefficients in optical coherence tomography,” J. Biomed. Opt. 13(3), 034003 (2008). [CrossRef] [PubMed]
  7. Y. Wang and R. Wang, “Autocorrelation optical coherence tomography for mapping transverse particle-flow velocity,” Opt. Lett. 35(21), 3538–3540 (2010). [CrossRef] [PubMed]
  8. J. Kalkman, R. Sprik, and T. G. van Leeuwen, “Path-length-resolved diffusive particle dynamics in spectral-domain optical coherence tomography,” Phys. Rev. Lett. 105(19), 198302 (2010). [CrossRef]
  9. J. Pyhtila, R. Graf, and A. Wax, “Determining nuclear morphology using an improved angle-resolved low coherence interferometry system,” Opt. Express 11(25), 3473–3484 (2003). [CrossRef] [PubMed]
  10. A. E. Desjardins, B. J. Vakoc, G. J. Tearney, and B. E. Bouma, “Backscattering spectroscopic contrast with angle-resolved optical coherence tomography,” Opt. Lett. 32(21), 3158–3160 (2007). [CrossRef] [PubMed]
  11. J. Kalkman, A. V. Bykov, D. J. Faber, and T. G. van Leeuwen, “Multiple and dependent scattering effects in Doppler optical coherence tomography,” Opt. Express 18(4), 3883–3892 (2010). [CrossRef] [PubMed]
  12. L. Thrane, H. T. Yura, and P. E. Andersen, “Analysis of optical coherence tomography systems based on the extended Huygens-Fresnel principle,” J. Opt. Soc. Am. A 17(3), 484–490 (2000). [CrossRef]
  13. J. M. Schmitt, A. Knuttel, and R. F. Bonner, “Measurement of optical properties of biological tissues by low-coherence reflectometry,” Appl. Opt. 32(30), 6032–6042 (11993). [CrossRef]
  14. D. J. Faber and T. G. van Leeuwen, “Are quantitative attenuation measurements of blood by optical coherence tomography feasible?” Opt. Lett. 34(9), 1435–1437 (2009). [CrossRef] [PubMed]
  15. J. W. Goodman, Statistical optics (Wiley, New York, 1985).
  16. T. G. van Leeuwen, D. J. Faber, and M. C. Aalders, “Measurement of the axial point spread function in scattering media using single-mode fiber-based optical coherence tomography,” IEEE J. Sel. Top. Quantum Electron. 9(2), 227–233 (2003). [CrossRef]
  17. C. Akcay, P. Parrein, and J. P. Rolland, “Estimation of Longitudinal Resolution in Optical Coherence Imaging,” Appl. Opt. 41(25), 5256–5262 (2002). [CrossRef] [PubMed]
  18. A. Unterhuber, B. Povazay, B. Hermann, H. Sattmann, W. Drexler, V. Yakovlev, G. Tempea, C. Schubert, E. M. Anger, P. K. Ahnelt, M. Stur, J. E. Morgan, A. Cowey, G. Jung, T. Le, and A. Stingl, “Compact, low-cost Ti:Al2O3 laser for in vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett. 28(11), 905–907 (2003). [CrossRef] [PubMed]
  19. H. J. van Staveren, C. J. M. Moes, J. van Marie, S. A. Prahl, and M. J. C. van Gemert, “Light scattering in Intralipid-10% in the wavelength range of 400-1100 nm,” Appl. Opt. 30(31), 4507–4514 (1991). [CrossRef] [PubMed]
  20. C. Chen, J. Q. Lu, H. F. Ding, K. M. Jacobs, Y. Du, and X. H. Hu, “A primary method for determination of optical parameters of turbid samples and application to intralipid between 550 and 1630 nm,” Opt. Express 14(16), 7420–7435 (2006). [CrossRef] [PubMed]
  21. J. R. Mourant, J. P. Freyer, A. H. Hielscher, A. A. Eick, D. Shen, and T. M. Johnson, “Mechanisms of light scattering from biological cells relevant to noninvasive optical-tissue diagnostics,” Appl. Opt. 37(16), 3586–3593 (1998). [CrossRef]
  22. N. Bosschaart, D. J. Faber, T. G. van Leeuwen, and M. C. G. Aalders, “Measurements of wavelength dependent scattering and backscattering coefficients by low-coherence spectroscopy,” J. Biomed.Opt. 16, 030503 (2011). [PubMed]
  23. R. Michels, F. Foschum, and A. Kienle, “Optical properties of fat emulsions,” Opt. Express 16(8), 5907–5925 (2008). [CrossRef] [PubMed]
  24. B. Karamata, M. Laubscher, M. Leutenegger, S. Bourquin, T. Lasser, and P. Lambelet, “Multiple scattering in optical coherence tomography. I. Investigation and modeling,” J. Opt. Soc. Am. A 22(7), 1369–1379 (2005). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1 Fig. 2 Fig. 3
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited