Theoretical comparison of three X-ray phase-contrast imaging techniques: propagation-based imaging, analyzer-based imaging and grating interferometry |
Optics Express, Vol. 20, Issue 3, pp. 2789-2805 (2012)
http://dx.doi.org/10.1364/OE.20.002789
Acrobat PDF (2171 KB)
Abstract
Various X-ray phase-contrast imaging techniques have been developed and applied over the last twenty years in different domains, such as material sciences, biology and medicine. However, no comprehensive inter-comparison exists in the literature. We present here a theoretical study that compares three among the most used techniques: propagation-based imaging (PBI), analyzer-based imaging (ABI) and grating interferometry (GI). These techniques are evaluated in terms of signal-to-noise ratio, figure of merit and spatial resolution. Both area and edge signals are considered. Dependences upon the object properties (absorption, phase shift) and the experimental acquisition parameters (energy, system point-spread function etc.) are derived and discussed. The results obtained from this analysis can be used as the reference for determining the most suitable technique for a given application.
© 2012 OSA
1. Introduction
U. Bonse and M. Hart, “An X-ray interferometer,” Appl. Phys. Lett. 6(8), 155–156 (1965). [CrossRef]
A. Momose, T. Takeda, Y. Itai, and K. Hirano, “Phase-contrast X-ray computed tomography for observing biological soft tissues,” Nat. Med. 2(4), 473–475 (1996). [CrossRef] [PubMed]
A. Snigirev, I. Snigireva, V. Kohn, S. Kuznetsov, and I. Schelokov, “On the possibility of x-ray phase contrast microimaging by coherent high-energy synchrotron radiation,” Rev. Sci. Instrum. 66(12), 5486–5492 (1995). [CrossRef]
S. W. Wilkins, T. E. Gureyev, D. Gao, A. Pogany, and A. W. Stevenson, “Phase-contrast imaging using polychromatic hard X-rays,” Nature 384(6607), 335–338 (1996). [CrossRef]
E. Förster, K. Goetz, and P. Zaumseil, “Double crystal diffractometry for the characterization of targets for laser fusion experiments,” Krist. Tech. 15(8), 937–945 (1980). [CrossRef]
A. Bravin, “Exploiting the X-ray refraction contrast with an analyser: the state of the art,” J. Phys. D Appl. Phys. 36(10A), A24–A29 (2003). [CrossRef]
C. David, B. Nöhammer, H. Solak, and E. Ziegler, “Differential X-ray phase contrast imaging using a shearing interferometer,” Appl. Phys. Lett. 81(17), 3287–3289 (2002). [CrossRef]
F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources,” Nat. Phys. 2(4), 258–261 (2006). [CrossRef]
A. Olivo, K. Ignatyev, P. R. T. Munro, and R. D. Speller, “Noninterferometric phase-contrast images obtained with incoherent x-ray sources,” Appl. Opt. 50(12), 1765–1769 (2011). [CrossRef] [PubMed]
E. Castelli, M. Tonutti, F. Arfelli, R. Longo, E. Quaia, L. Rigon, D. Sanabor, F. Zanconati, D. Dreossi, A. Abrami, E. Quai, P. Bregant, K. Casarin, V. Chenda, R. H. Menk, T. Rokvic, A. Vascotto, G. Tromba, and M. A. Cova, “Mammography with synchrotron radiation: first clinical experience with phase-detection technique,” Radiology 259(3), 684–694 (2011). [CrossRef] [PubMed]
P. Coan, F. Bamberg, P. C. Diemoz, A. Bravin, K. Timpert, E. Mützel, J. G. Raya, S. Adam-Neumair, M. F. Reiser, and C. Glaser, “Characterization of osteoarthritic and normal human patella cartilage by computed tomography X-ray phase-contrast imaging: a feasibility study,” Invest. Radiol. 45(7), 437–444 (2010). [PubMed]
G. Schulz, T. Weitkamp, I. Zanette, F. Pfeiffer, F. Beckmann, C. David, S. Rutishauser, E. Reznikova, and B. Mueller, “High-resolution tomographic imaging of a human cerebellum: comparison of absorption and grating-based phase contrast,” J. R. Soc. Interface 7(53), 1665–1676 (2010). [CrossRef] [PubMed]
E. Pagot, S. Fiedler, P. Cloetens, A. Bravin, P. Coan, K. Fezzaa, J. Baruchel, J. Härtwig, K. von Smitten, M. Leidenius, M. L. Karjalainen-Lindsberg, and J. Keyriläinen, “Quantitative comparison between two phase contrast techniques: diffraction enhanced imaging and phase propagation imaging,” Phys. Med. Biol. 50(4), 709–724 (2005). [CrossRef] [PubMed]
2. Brief description of PBI, ABI and GI techniques
- 1. Propagation-based imaging (PBI) is the method with the simplest experimental setup. It requires that the sample is irradiated with highly spatially coherent radiation and that the detector is positioned at a sufficient distance r from the sample; no optical elements are needed between the sample and the detector. Thanks to Fresnel diffraction, the differences in phase shift introduced by the object onto the beam lead to a measurable intensity modulation onto the detector [1]. This intensity can be expressed, in the near-field diffraction regime, by the transport of intensity equation (TIE) [17]. The near-field regime approximation is valid for sufficiently small propagation distances and for objects introducing a slowly varying phase shift in the plane (x,y) transversal to the optical axis z [18
M. R. Teague, “Irradiance moments - their propagation and use for unique retrieval of phase,” J. Opt. Soc. Am. 72(9), 1199–1209 (1982). [CrossRef]
]. In the case of additional slowly varying object absorption, the TIE can be written as (for sake of simplicity the spatial variables are omitted):T. E. Gureyev, Y. I. Nesterets, A. W. Stevenson, P. R. Miller, A. Pogany, and S. W. Wilkins, “Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging,” Opt. Express 16(5), 3223–3241 (2008). [CrossRef] [PubMed]
S. W. Wilkins, T. E. Gureyev, D. Gao, A. Pogany, and A. W. Stevenson, “Phase-contrast imaging using polychromatic hard X-rays,” Nature 384(6607), 335–338 (1996). [CrossRef]
- 2. Analyzer-based imaging (ABI) makes use of a quasi-monochromatic and quasi-parallel beam (typically produced by diffraction from a perfect crystal, called the monochromator) irradiating the sample and of a perfect “analyzer” crystal placed between the sample and the detector. The analyzer acts as an angular filter of the radiation refracted and scattered by the sample. In a typical ABI setup the sample-to-detector distance is small, so that the propagation phase contrast can be neglected [19]. The general expression for the intensity recorded on the detector [20
P. Coan, E. Pagot, S. Fiedler, P. Cloetens, J. Baruchel, and A. Bravin, “Phase-contrast X-ray imaging combining free space propagation and Bragg diffraction,” J. Synchrotron Radiat. 12(2), 241–245 (2005). [CrossRef] [PubMed]
] can be greatly simplified under the geometrical optics approximation [21Y. I. Nesterets, P. Coan, T. E. Gureyev, A. Bravin, P. Cloetens, and S. W. Wilkins, “On qualitative and quantitative analysis in analyser-based imaging,” Acta Crystallogr. A 62(4), 296–308 (2006). [CrossRef] [PubMed]
], which is valid if the phase of the wave incident onto the crystal is a slowly varying function over the length scales on the order of the crystal extinction length [22T. E. Gureyev and S. W. Wilkins, “Regimes of X-ray phase-contrast imaging with perfect crystals,” Nuovo Cimento D 19(2-4), 545–552 (1997). [CrossRef]
]. In this case the intensity for each detector pixel can then be expressed, if the crystal diffraction plane is assumed to be parallel to the (y,z) plane, as:K. M. Pavlov, T. E. Gureyev, D. Paganin, Y. I. Nesterets, M. J. Morgan, and R. A. Lewis, “Linear systems with slowly varying transfer functions and their application to x-ray phase-contrast imaging,” J. Phys. D Appl. Phys. 37(19), 2746–2750 (2004). [CrossRef]
R. Tanuma and M. Ohsawa, “Submicron-resolved X-ray topography using asymmetric-reflection magnifiers,” Spectrochim. Acta, Part B 59(10-11), 1549–1555 (2004). [CrossRef]
D. Chapman, W. Thomlinson, R. E. Johnston, D. Washburn, E. Pisano, N. Gmur, Z. Zhong, R. Menk, F. Arfelli, and D. Sayers, “Diffraction enhanced x-ray imaging,” Phys. Med. Biol. 42(11), 2015–2025 (1997). [CrossRef] [PubMed]
- 3. The grating interferometry (GI) technique consists in illuminating the imaged object with highly spatially coherent X-rays, and in analyzing the radiation transmitted through the object by using a pair of gratings [10]. The first is generally a phase grating, of period p1, which introduces a periodic phase shift onto the beam but negligible absorption. A second absorption grating of period p2 is then placed downstream at one of the so-called fractional Talbot distances dTalbot, where the interference fringes created by the first grating give rise to the so-called self-imaging effect.
T. Weitkamp, A. Diaz, C. David, F. Pfeiffer, M. Stampanoni, P. Cloetens, and E. Ziegler, “X-ray phase imaging with a grating interferometer,” Opt. Express 13(16), 6296–6304 (2005). [CrossRef] [PubMed]
P. C. Diemoz, P. Coan, I. Zanette, A. Bravin, S. Lang, C. Glaser, and T. Weitkamp, “A simplified approach for computed tomography with an X-ray grating interferometer,” Opt. Express 19(3), 1691–1698 (2011). [CrossRef] [PubMed]
P. C. Diemoz, P. Coan, I. Zanette, A. Bravin, S. Lang, C. Glaser, and T. Weitkamp, “A simplified approach for computed tomography with an X-ray grating interferometer,” Opt. Express 19(3), 1691–1698 (2011). [CrossRef] [PubMed]
F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources,” Nat. Phys. 2(4), 258–261 (2006). [CrossRef]
F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources,” Nat. Phys. 2(4), 258–261 (2006). [CrossRef]
2.1. Spatial resolution
T. E. Gureyev, Y. I. Nesterets, A. W. Stevenson, P. R. Miller, A. Pogany, and S. W. Wilkins, “Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging,” Opt. Express 16(5), 3223–3241 (2008). [CrossRef] [PubMed]
T. E. Gureyev, Y. I. Nesterets, A. W. Stevenson, P. R. Miller, A. Pogany, and S. W. Wilkins, “Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging,” Opt. Express 16(5), 3223–3241 (2008). [CrossRef] [PubMed]
R. Tanuma and M. Ohsawa, “Submicron-resolved X-ray topography using asymmetric-reflection magnifiers,” Spectrochim. Acta, Part B 59(10-11), 1549–1555 (2004). [CrossRef]
3. Definition of signal-to-noise ratio and figure-of-merit
3.1 Area signal
F. Arfelli, V. Bonvicini, A. Bravin, G. Cantatore, E. Castelli, L. Dalla Palma, M. Di Michiel, R. Longo, A. Olivo, S. Pani, D. Pontoni, P. Poropat, M. Prest, A. Rashevsky, G. Tromba, and A. Vacchi, “Mammography of a phantom and breast tissue with synchrotron radiation and a linear-array silicon detector,” Radiology 208(3), 709–715 (1998). [PubMed]
3.2 Edge signal
E. Pagot, S. Fiedler, P. Cloetens, A. Bravin, P. Coan, K. Fezzaa, J. Baruchel, J. Härtwig, K. von Smitten, M. Leidenius, M. L. Karjalainen-Lindsberg, and J. Keyriläinen, “Quantitative comparison between two phase contrast techniques: diffraction enhanced imaging and phase propagation imaging,” Phys. Med. Biol. 50(4), 709–724 (2005). [CrossRef] [PubMed]
P. Coan, E. Pagot, S. Fiedler, P. Cloetens, J. Baruchel, and A. Bravin, “Phase-contrast X-ray imaging combining free space propagation and Bragg diffraction,” J. Synchrotron Radiat. 12(2), 241–245 (2005). [CrossRef] [PubMed]
T. E. Gureyev, Y. I. Nesterets, A. W. Stevenson, P. R. Miller, A. Pogany, and S. W. Wilkins, “Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging,” Opt. Express 16(5), 3223–3241 (2008). [CrossRef] [PubMed]
4. Calculation of figure of merit
4.1 Calculation of area FoM
J. M. Boone, K. K. Lindfors, V. N. Cooper 3rd, and J. A. Seibert, “Scatter/primary in mammography: comprehensive results,” Med. Phys. 27(10), 2408–2416 (2000). [CrossRef] [PubMed]
J. Persliden and G. A. Carlsson, “Scatter rejection by air gaps in diagnostic radiology. Calculations using a Monte Carlo collision density method and consideration of molecular interference in coherent scattering,” Phys. Med. Biol. 42(1), 155–175 (1997). [CrossRef] [PubMed]
4.2 Calculation of peak edge signal FoM
M. Sanchez del Rio, C. Ferrero, and V. Mocella, “Computer simulations of bent perfect crystal diffraction profiles,” Proc. SPIE 3152, 312–323 (1997), http://www.esrf.eu/UsersAndScience/Experiments/TBS/ SciSoft/xop2.3. [CrossRef]
M. Sanchez del Rio, C. Ferrero, and V. Mocella, “Computer simulations of bent perfect crystal diffraction profiles,” Proc. SPIE 3152, 312–323 (1997), http://www.esrf.eu/UsersAndScience/Experiments/TBS/ SciSoft/xop2.3. [CrossRef]
E. Pagot, S. Fiedler, P. Cloetens, A. Bravin, P. Coan, K. Fezzaa, J. Baruchel, J. Härtwig, K. von Smitten, M. Leidenius, M. L. Karjalainen-Lindsberg, and J. Keyriläinen, “Quantitative comparison between two phase contrast techniques: diffraction enhanced imaging and phase propagation imaging,” Phys. Med. Biol. 50(4), 709–724 (2005). [CrossRef] [PubMed]
4.3 Calculation of integral edge signal FoM
T. E. Gureyev, Y. I. Nesterets, A. W. Stevenson, P. R. Miller, A. Pogany, and S. W. Wilkins, “Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging,” Opt. Express 16(5), 3223–3241 (2008). [CrossRef] [PubMed]
5. Conclusions
Acknowledgments
References and links
M. Born and E. Wolf, Principles of Optics: Electromagnetic Theory of Propagation, Interference and Diffraction of Light (Cambridge University Press, Cambridge, 1999). | |
U. Bonse and M. Hart, “An X-ray interferometer,” Appl. Phys. Lett. 6(8), 155–156 (1965). [CrossRef] | |
A. Momose, T. Takeda, Y. Itai, and K. Hirano, “Phase-contrast X-ray computed tomography for observing biological soft tissues,” Nat. Med. 2(4), 473–475 (1996). [CrossRef] [PubMed] | |
A. Snigirev, I. Snigireva, V. Kohn, S. Kuznetsov, and I. Schelokov, “On the possibility of x-ray phase contrast microimaging by coherent high-energy synchrotron radiation,” Rev. Sci. Instrum. 66(12), 5486–5492 (1995). [CrossRef] | |
S. W. Wilkins, T. E. Gureyev, D. Gao, A. Pogany, and A. W. Stevenson, “Phase-contrast imaging using polychromatic hard X-rays,” Nature 384(6607), 335–338 (1996). [CrossRef] | |
E. Förster, K. Goetz, and P. Zaumseil, “Double crystal diffractometry for the characterization of targets for laser fusion experiments,” Krist. Tech. 15(8), 937–945 (1980). [CrossRef] | |
D. Chapman, W. Thomlinson, R. E. Johnston, D. Washburn, E. Pisano, N. Gmur, Z. Zhong, R. Menk, F. Arfelli, and D. Sayers, “Diffraction enhanced x-ray imaging,” Phys. Med. Biol. 42(11), 2015–2025 (1997). [CrossRef] [PubMed] | |
A. Bravin, “Exploiting the X-ray refraction contrast with an analyser: the state of the art,” J. Phys. D Appl. Phys. 36(10A), A24–A29 (2003). [CrossRef] | |
C. David, B. Nöhammer, H. Solak, and E. Ziegler, “Differential X-ray phase contrast imaging using a shearing interferometer,” Appl. Phys. Lett. 81(17), 3287–3289 (2002). [CrossRef] | |
T. Weitkamp, A. Diaz, C. David, F. Pfeiffer, M. Stampanoni, P. Cloetens, and E. Ziegler, “X-ray phase imaging with a grating interferometer,” Opt. Express 13(16), 6296–6304 (2005). [CrossRef] [PubMed] | |
F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources,” Nat. Phys. 2(4), 258–261 (2006). [CrossRef] | |
A. Olivo, K. Ignatyev, P. R. T. Munro, and R. D. Speller, “Noninterferometric phase-contrast images obtained with incoherent x-ray sources,” Appl. Opt. 50(12), 1765–1769 (2011). [CrossRef] [PubMed] | |
E. Castelli, M. Tonutti, F. Arfelli, R. Longo, E. Quaia, L. Rigon, D. Sanabor, F. Zanconati, D. Dreossi, A. Abrami, E. Quai, P. Bregant, K. Casarin, V. Chenda, R. H. Menk, T. Rokvic, A. Vascotto, G. Tromba, and M. A. Cova, “Mammography with synchrotron radiation: first clinical experience with phase-detection technique,” Radiology 259(3), 684–694 (2011). [CrossRef] [PubMed] | |
P. Coan, F. Bamberg, P. C. Diemoz, A. Bravin, K. Timpert, E. Mützel, J. G. Raya, S. Adam-Neumair, M. F. Reiser, and C. Glaser, “Characterization of osteoarthritic and normal human patella cartilage by computed tomography X-ray phase-contrast imaging: a feasibility study,” Invest. Radiol. 45(7), 437–444 (2010). [PubMed] | |
G. Schulz, T. Weitkamp, I. Zanette, F. Pfeiffer, F. Beckmann, C. David, S. Rutishauser, E. Reznikova, and B. Mueller, “High-resolution tomographic imaging of a human cerebellum: comparison of absorption and grating-based phase contrast,” J. R. Soc. Interface 7(53), 1665–1676 (2010). [CrossRef] [PubMed] | |
E. Pagot, S. Fiedler, P. Cloetens, A. Bravin, P. Coan, K. Fezzaa, J. Baruchel, J. Härtwig, K. von Smitten, M. Leidenius, M. L. Karjalainen-Lindsberg, and J. Keyriläinen, “Quantitative comparison between two phase contrast techniques: diffraction enhanced imaging and phase propagation imaging,” Phys. Med. Biol. 50(4), 709–724 (2005). [CrossRef] [PubMed] | |
M. R. Teague, “Irradiance moments - their propagation and use for unique retrieval of phase,” J. Opt. Soc. Am. 72(9), 1199–1209 (1982). [CrossRef] | |
T. E. Gureyev, Y. I. Nesterets, A. W. Stevenson, P. R. Miller, A. Pogany, and S. W. Wilkins, “Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging,” Opt. Express 16(5), 3223–3241 (2008). [CrossRef] [PubMed] | |
P. Coan, E. Pagot, S. Fiedler, P. Cloetens, J. Baruchel, and A. Bravin, “Phase-contrast X-ray imaging combining free space propagation and Bragg diffraction,” J. Synchrotron Radiat. 12(2), 241–245 (2005). [CrossRef] [PubMed] | |
Y. I. Nesterets, P. Coan, T. E. Gureyev, A. Bravin, P. Cloetens, and S. W. Wilkins, “On qualitative and quantitative analysis in analyser-based imaging,” Acta Crystallogr. A 62(4), 296–308 (2006). [CrossRef] [PubMed] | |
T. E. Gureyev and S. W. Wilkins, “Regimes of X-ray phase-contrast imaging with perfect crystals,” Nuovo Cimento D 19(2-4), 545–552 (1997). [CrossRef] | |
K. M. Pavlov, T. E. Gureyev, D. Paganin, Y. I. Nesterets, M. J. Morgan, and R. A. Lewis, “Linear systems with slowly varying transfer functions and their application to x-ray phase-contrast imaging,” J. Phys. D Appl. Phys. 37(19), 2746–2750 (2004). [CrossRef] | |
R. Tanuma and M. Ohsawa, “Submicron-resolved X-ray topography using asymmetric-reflection magnifiers,” Spectrochim. Acta, Part B 59(10-11), 1549–1555 (2004). [CrossRef] | |
P. C. Diemoz, P. Coan, I. Zanette, A. Bravin, S. Lang, C. Glaser, and T. Weitkamp, “A simplified approach for computed tomography with an X-ray grating interferometer,” Opt. Express 19(3), 1691–1698 (2011). [CrossRef] [PubMed] | |
A. L. Evans, The Evaluation of Medical Images (Adam Hilger Ltd, Bristol, 1981). | |
F. Arfelli, V. Bonvicini, A. Bravin, G. Cantatore, E. Castelli, L. Dalla Palma, M. Di Michiel, R. Longo, A. Olivo, S. Pani, D. Pontoni, P. Poropat, M. Prest, A. Rashevsky, G. Tromba, and A. Vacchi, “Mammography of a phantom and breast tissue with synchrotron radiation and a linear-array silicon detector,” Radiology 208(3), 709–715 (1998). [PubMed] | |
S. Webb, The Physics of Medical Imaging (Institute of Physics Publishing, Bristol, 1988). | |
J. M. Boone, K. K. Lindfors, V. N. Cooper 3rd, and J. A. Seibert, “Scatter/primary in mammography: comprehensive results,” Med. Phys. 27(10), 2408–2416 (2000). [CrossRef] [PubMed] | |
J. Persliden and G. A. Carlsson, “Scatter rejection by air gaps in diagnostic radiology. Calculations using a Monte Carlo collision density method and consideration of molecular interference in coherent scattering,” Phys. Med. Biol. 42(1), 155–175 (1997). [CrossRef] [PubMed] | |
M. Sanchez del Rio, C. Ferrero, and V. Mocella, “Computer simulations of bent perfect crystal diffraction profiles,” Proc. SPIE 3152, 312–323 (1997), http://www.esrf.eu/UsersAndScience/Experiments/TBS/ SciSoft/xop2.3. [CrossRef] | |
T. Matsushita and H. Hashizume, “X-Ray monochromators,” in Handbook on Synchrotron Radiation, E. Koch, ed. (North Holland Publishing Company, New York, 1983), pp. 261–314. |
OCIS Codes
(100.2960) Image processing : Image analysis
(110.2990) Imaging systems : Image formation theory
(110.4980) Imaging systems : Partial coherence in imaging
(110.7440) Imaging systems : X-ray imaging
ToC Category:
Imaging Systems
History
Original Manuscript: October 17, 2011
Manuscript Accepted: November 30, 2011
Published: January 23, 2012
Virtual Issues
Vol. 7, Iss. 3 Virtual Journal for Biomedical Optics
Citation
P. C. Diemoz, A. Bravin, and P. Coan, "Theoretical comparison of three X-ray phase-contrast imaging techniques: propagation-based imaging, analyzer-based imaging and grating interferometry," Opt. Express 20, 2789-2805 (2012)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-20-3-2789
Sort: Year | Journal | Reset
References
- M. Born and E. Wolf, Principles of Optics: Electromagnetic Theory of Propagation, Interference and Diffraction of Light (Cambridge University Press, Cambridge, 1999).
- U. Bonse and M. Hart, “An X-ray interferometer,” Appl. Phys. Lett.6(8), 155–156 (1965). [CrossRef]
- A. Momose, T. Takeda, Y. Itai, and K. Hirano, “Phase-contrast X-ray computed tomography for observing biological soft tissues,” Nat. Med.2(4), 473–475 (1996). [CrossRef] [PubMed]
- A. Snigirev, I. Snigireva, V. Kohn, S. Kuznetsov, and I. Schelokov, “On the possibility of x-ray phase contrast microimaging by coherent high-energy synchrotron radiation,” Rev. Sci. Instrum.66(12), 5486–5492 (1995). [CrossRef]
- S. W. Wilkins, T. E. Gureyev, D. Gao, A. Pogany, and A. W. Stevenson, “Phase-contrast imaging using polychromatic hard X-rays,” Nature384(6607), 335–338 (1996). [CrossRef]
- E. Förster, K. Goetz, and P. Zaumseil, “Double crystal diffractometry for the characterization of targets for laser fusion experiments,” Krist. Tech.15(8), 937–945 (1980). [CrossRef]
- D. Chapman, W. Thomlinson, R. E. Johnston, D. Washburn, E. Pisano, N. Gmur, Z. Zhong, R. Menk, F. Arfelli, and D. Sayers, “Diffraction enhanced x-ray imaging,” Phys. Med. Biol.42(11), 2015–2025 (1997). [CrossRef] [PubMed]
- A. Bravin, “Exploiting the X-ray refraction contrast with an analyser: the state of the art,” J. Phys. D Appl. Phys.36(10A), A24–A29 (2003). [CrossRef]
- C. David, B. Nöhammer, H. Solak, and E. Ziegler, “Differential X-ray phase contrast imaging using a shearing interferometer,” Appl. Phys. Lett.81(17), 3287–3289 (2002). [CrossRef]
- T. Weitkamp, A. Diaz, C. David, F. Pfeiffer, M. Stampanoni, P. Cloetens, and E. Ziegler, “X-ray phase imaging with a grating interferometer,” Opt. Express13(16), 6296–6304 (2005). [CrossRef] [PubMed]
- F. Pfeiffer, T. Weitkamp, O. Bunk, and C. David, “Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources,” Nat. Phys.2(4), 258–261 (2006). [CrossRef]
- A. Olivo, K. Ignatyev, P. R. T. Munro, and R. D. Speller, “Noninterferometric phase-contrast images obtained with incoherent x-ray sources,” Appl. Opt.50(12), 1765–1769 (2011). [CrossRef] [PubMed]
- E. Castelli, M. Tonutti, F. Arfelli, R. Longo, E. Quaia, L. Rigon, D. Sanabor, F. Zanconati, D. Dreossi, A. Abrami, E. Quai, P. Bregant, K. Casarin, V. Chenda, R. H. Menk, T. Rokvic, A. Vascotto, G. Tromba, and M. A. Cova, “Mammography with synchrotron radiation: first clinical experience with phase-detection technique,” Radiology259(3), 684–694 (2011). [CrossRef] [PubMed]
- P. Coan, F. Bamberg, P. C. Diemoz, A. Bravin, K. Timpert, E. Mützel, J. G. Raya, S. Adam-Neumair, M. F. Reiser, and C. Glaser, “Characterization of osteoarthritic and normal human patella cartilage by computed tomography X-ray phase-contrast imaging: a feasibility study,” Invest. Radiol.45(7), 437–444 (2010). [PubMed]
- G. Schulz, T. Weitkamp, I. Zanette, F. Pfeiffer, F. Beckmann, C. David, S. Rutishauser, E. Reznikova, and B. Mueller, “High-resolution tomographic imaging of a human cerebellum: comparison of absorption and grating-based phase contrast,” J. R. Soc. Interface7(53), 1665–1676 (2010). [CrossRef] [PubMed]
- E. Pagot, S. Fiedler, P. Cloetens, A. Bravin, P. Coan, K. Fezzaa, J. Baruchel, J. Härtwig, K. von Smitten, M. Leidenius, M. L. Karjalainen-Lindsberg, and J. Keyriläinen, “Quantitative comparison between two phase contrast techniques: diffraction enhanced imaging and phase propagation imaging,” Phys. Med. Biol.50(4), 709–724 (2005). [CrossRef] [PubMed]
- M. R. Teague, “Irradiance moments - their propagation and use for unique retrieval of phase,” J. Opt. Soc. Am.72(9), 1199–1209 (1982). [CrossRef]
- T. E. Gureyev, Y. I. Nesterets, A. W. Stevenson, P. R. Miller, A. Pogany, and S. W. Wilkins, “Some simple rules for contrast, signal-to-noise and resolution in in-line x-ray phase-contrast imaging,” Opt. Express16(5), 3223–3241 (2008). [CrossRef] [PubMed]
- P. Coan, E. Pagot, S. Fiedler, P. Cloetens, J. Baruchel, and A. Bravin, “Phase-contrast X-ray imaging combining free space propagation and Bragg diffraction,” J. Synchrotron Radiat.12(2), 241–245 (2005). [CrossRef] [PubMed]
- Y. I. Nesterets, P. Coan, T. E. Gureyev, A. Bravin, P. Cloetens, and S. W. Wilkins, “On qualitative and quantitative analysis in analyser-based imaging,” Acta Crystallogr. A62(4), 296–308 (2006). [CrossRef] [PubMed]
- T. E. Gureyev and S. W. Wilkins, “Regimes of X-ray phase-contrast imaging with perfect crystals,” Nuovo Cimento D19(2-4), 545–552 (1997). [CrossRef]
- K. M. Pavlov, T. E. Gureyev, D. Paganin, Y. I. Nesterets, M. J. Morgan, and R. A. Lewis, “Linear systems with slowly varying transfer functions and their application to x-ray phase-contrast imaging,” J. Phys. D Appl. Phys.37(19), 2746–2750 (2004). [CrossRef]
- R. Tanuma and M. Ohsawa, “Submicron-resolved X-ray topography using asymmetric-reflection magnifiers,” Spectrochim. Acta, Part B59(10-11), 1549–1555 (2004). [CrossRef]
- P. C. Diemoz, P. Coan, I. Zanette, A. Bravin, S. Lang, C. Glaser, and T. Weitkamp, “A simplified approach for computed tomography with an X-ray grating interferometer,” Opt. Express19(3), 1691–1698 (2011). [CrossRef] [PubMed]
- A. L. Evans, The Evaluation of Medical Images (Adam Hilger Ltd, Bristol, 1981).
- F. Arfelli, V. Bonvicini, A. Bravin, G. Cantatore, E. Castelli, L. Dalla Palma, M. Di Michiel, R. Longo, A. Olivo, S. Pani, D. Pontoni, P. Poropat, M. Prest, A. Rashevsky, G. Tromba, and A. Vacchi, “Mammography of a phantom and breast tissue with synchrotron radiation and a linear-array silicon detector,” Radiology208(3), 709–715 (1998). [PubMed]
- S. Webb, The Physics of Medical Imaging (Institute of Physics Publishing, Bristol, 1988).
- J. M. Boone, K. K. Lindfors, V. N. Cooper, and J. A. Seibert, “Scatter/primary in mammography: comprehensive results,” Med. Phys.27(10), 2408–2416 (2000). [CrossRef] [PubMed]
- J. Persliden and G. A. Carlsson, “Scatter rejection by air gaps in diagnostic radiology. Calculations using a Monte Carlo collision density method and consideration of molecular interference in coherent scattering,” Phys. Med. Biol.42(1), 155–175 (1997). [CrossRef] [PubMed]
- M. Sanchez del Rio, C. Ferrero, and V. Mocella, “Computer simulations of bent perfect crystal diffraction profiles,” Proc. SPIE3152, 312–323 (1997), http://www.esrf.eu/UsersAndScience/Experiments/TBS/ SciSoft/xop2.3 . [CrossRef]
- T. Matsushita and H. Hashizume, “X-Ray monochromators,” in Handbook on Synchrotron Radiation, E. Koch, ed. (North Holland Publishing Company, New York, 1983), pp. 261–314.
Cited By |
OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.





OSA is a member of 