Abstract

We present a continuous-wave terahertz (THz) vector spectroscopy and imaging system based on a 1.5-µm fiber optic uni-traveling-carrier photodiode and InGaAs photo-conductive receiver. Using electro-optic (EO) phase modulators for THz phase control with shortened optical paths, the system achieves fast vector measurement with effective phase stabilization. Dynamic ranges of 100 dB·Hz and 75 dB·Hz at 300 GHz and 1 THz, and phase stability of 1.5° per minute are obtained. With the simultaneous measurement of absorbance and relative permittivity, we demonstrate non-destructive analyses of pharmaceutical cocrystals inside tablets within a few minutes.

© 2014 Optical Society of America

1. Introduction

Terahertz (THz) waves have received much attention due to their interesting features such as spectral fingerprints of biological or chemical materials and a relatively long penetration depth for spectroscopy and imaging [1

1. M. Tonouchi, “Cutting-edge terahertz technology,” Nat. Photonics 1(2), 97–105 (2007). [CrossRef]

]. Concurrently, many potential applications, such as non-destructive inspection of pharmaceuticals, drug detection, material testing, and safety screening, are being investigated [2

2. K. Ajito and Y. Ueno, “THz Chemical Imaging for Biological Applications,” IEEE Trans. THz Sci. Tech. (Paris) 1, 293–300 (2011).

5

5. F. Friederich, W. V. Spiegel, M. Bauer, F. Meng, M. D. Thomson, S. Boppel, A. Lisauskas, B. Hils, V. Krozer, A. Keil, T. Löffler, R. Henneberger, A. K. Huhn, G. Spickermann, P. H. Bolívar, and H. G. Roskos, “THz Active Imaging Systems With Real-Time Capabilities,” IEEE Trans. THz Sci. Tech. (Paris) 1, 183–200 (2011).

]. Many of these practical applications need high-frame-rate imaging close to real time and adaptive spectroscopy in the test protocol. Fast measurement is also important for reducing instabilities of samples and the setup with time.

One potential advantage of a continuous-wave (CW) system is frequency selectivity with high and easily scalable spectral resolution, which has stimulated a lot of active research using electrical MMICs [5

5. F. Friederich, W. V. Spiegel, M. Bauer, F. Meng, M. D. Thomson, S. Boppel, A. Lisauskas, B. Hils, V. Krozer, A. Keil, T. Löffler, R. Henneberger, A. K. Huhn, G. Spickermann, P. H. Bolívar, and H. G. Roskos, “THz Active Imaging Systems With Real-Time Capabilities,” IEEE Trans. THz Sci. Tech. (Paris) 1, 183–200 (2011).

], quantum cascade lasers [6

6. A. W. M. Lee, T.-Y. Kao, D. Burghoff, Q. Hu, and J. L. Reno, “Terahertz tomography using quantum-cascade lasers,” Opt. Lett. 37(2), 217–219 (2012). [CrossRef] [PubMed]

], or photo-mixing [7

7. J. R. Demers, R. T. Logan Jr, N. J. Bergeron, and E. R. Brown, “A coherent frequency-domain THz spectrometer with a signal-to-noise ratio 60 dB at 1 THz,” Proc. SPIE 6949, 694909 (2008). [CrossRef]

, 8

8. A. Roggenbuck, H. Schmitz, A. Deninger, I. C. Mayorga, J. Hemberger, R. Güsten, and M. Grüninger, “Coherent broadband continuous-wave terahertz spectroscopy on solid-state samples,” New J. Phys. 12(4), 043017 (2010). [CrossRef]

]. Among them, the photo-mixing THz generation method can support the widest frequency range for phase-sensitive spectroscopy and imaging simply with free-running CW lasers by using homodyne detection [9

9. R. Wilk, F. Breitfeld, M. Mikulics, and M. Koch, “Continuous wave terahertz spectrometer as a noncontact thickness measuring device,” Appl. Opt. 47(16), 3023–3026 (2008). [CrossRef] [PubMed]

, 10

10. I. S. Gregory, W. R. Tribe, C. Baker, B. E. Cole, M. J. Evans, L. Spencer, M. Pepper, and M. Missous, “Continuous-wave terahertz system with a 60 dB dynamic range,” Appl. Phys. Lett. 86(20), 204104 (2005). [CrossRef]

]. In addition, in the 1.55-µm-wavelength range, the well-developed electro-optic (EO) devices, such as compact dual-mode tunable lasers, have merits in cost and size [11

11. N. Kim, S.-P. Han, H. Ko, Y. A. Leem, H.-C. Ryu, C. W. Lee, D. Lee, M. Y. Jeon, S. K. Noh, and K. H. Park, “Tunable continuous-wave terahertz generation/detection with compact 1.55 μm detuned dual-mode laser diode and InGaAs based photomixer,” Opt. Express 19(16), 15397–15403 (2011). [CrossRef] [PubMed]

].

A homodyne detection scheme requires phase rotation of THz waves for acquisition of a complete waveform, which has been achieved with a mechanical delay stage [9

9. R. Wilk, F. Breitfeld, M. Mikulics, and M. Koch, “Continuous wave terahertz spectrometer as a noncontact thickness measuring device,” Appl. Opt. 47(16), 3023–3026 (2008). [CrossRef] [PubMed]

, 10

10. I. S. Gregory, W. R. Tribe, C. Baker, B. E. Cole, M. J. Evans, L. Spencer, M. Pepper, and M. Missous, “Continuous-wave terahertz system with a 60 dB dynamic range,” Appl. Phys. Lett. 86(20), 204104 (2005). [CrossRef]

]. However, the low speed of mechanical stages limits the acquisition time and requires additional magnitude chopping to avoid the flicker noise of the detector. Moreover, the bulky delay stage prevents compact all-fiber implementation. An alternative approach is to utilize an EO phase modulator. Because the frequency or phase modulation of a lightwave before the photo-mixing is directly proportional to that of the resulting THz wave, EO phase modulation enables fast THz phase control [12

12. A. M. Sinyukov, Z. Liu, Y. L. Hor, K. Su, R. B. Barat, D. E. Gary, Z.-H. Michalopoulou, I. Zorych, J. F. Federici, and D. Zimdars, “Rapid-phase modulation of terahertz radiation for high-speed terahertz imaging and spectroscopy,” Opt. Lett. 33(14), 1593–1595 (2008). [CrossRef] [PubMed]

] and synchronized lock-in detection [13

13. T. Göbel, D. Schoenherr, C. Sydlo, M. Feiginov, P. Meissner, and H. L. Hartnagel, “Single-sampling-point coherent detection in continuous-wave photomixing terahertz systems,” Electron. Lett. 45(1), 65–66 (2009). [CrossRef]

, 14

14. J.-Y. Kim, H.-J. Song, K. Ajito, M. Yaita, and N. Kukutsu, “Continuous-Wave THz Homodyne Spectroscopy and Imaging System With Electro-Optical Phase Modulation for High Dynamic Range,” IEEE Trans. THz Sci. Tech. (Paris) 3, 158–164 (2013).

]. In addition, the phase-sensitive detection can enhance the contrast by using depth information [15

15. S. Hisatake, G. Kitahara, K. Ajito, Y. Fukada, N. Yoshimoto, and T. Nagatsuma, “Phase-sensitive terahertz self-heterodyne system based on photodiode and low-temperature-grown GaAs photoconductor at 1.55 μm,” IEEE Sens. J. 13(1), 31–36 (2013). [CrossRef]

, 16

16. M. Suga, Y. Sasaki, T. Sasahara, T. Yuasa, and C. Otani, “THz phase-contrast computed tomography based on Mach-Zehnder interferometer using continuous wave source: proof of the concept,” Opt. Express 21(21), 25389–25402 (2013). [CrossRef] [PubMed]

]. However, in the early versions of the system, the use of a long-fiber setup caused environmental phase instability which prevented the practical use of the phase information.

In this paper, we present a CW-THz vector spectroscopy system using EO phase modulation which enables fast and simultaneous measurement of absorbance and relative permittivity for spectroscopy and imaging applications at up to 2 THz. To stabilize the system, the environmental phase changes have been suppressed by reducing the optical path length for THz phase control. By combining an advanced InGaAs photoconductive receiver [17

17. T. Göbel, D. Stanze, B. Globisch, R. J. B. Dietz, H. Roehle, and M. Schell, “Telecom technology based continuous wave terahertz photomixing system with 105 decibel signal-to-noise ratio and 3.5 terahertz bandwidth,” Opt. Lett. 38(20), 4197–4199 (2013). [CrossRef] [PubMed]

] and a wideband uni-traveling-carrier photodiode (UTC-PD) [18

18. H. Ito, T. Furuta, F. Nakajima, K. Yoshino, and T. Ishibashi, “Photonic Generation of Continuous THz Wave Using Uni-Traveling-Carrier Photodiode,” J. Lightwave Technol. 23(12), 4016–4021 (2005). [CrossRef]

], high dynamic ranges of 100 and 75 dB·Hz are obtained at 300 GHz and 1 THz. Thus, fast vector measurement with millisecond-order data acquisition is achieved for both spectroscopy and raster-scan imaging in a frequency range of more than 2 THz. The short measurement time also reduces errors caused by environmental instability of the system. Using the THz spectroscopy and imaging system, a complex-domain characterization of pharmaceutical cocrystal inside a tablet [19

19. D. M. Charron, K. Ajito, J.-Y. Kim, and Y. Ueno, “Chemical mapping of pharmaceutical cocrystals using terahertz spectroscopic imaging,” Anal. Chem. 85(4), 1980–1984 (2013). [CrossRef] [PubMed]

] has been demonstrated.

2. CW-THz homodyne system for fast vector measurement

Figure 1
Fig. 1 Schematic diagram of CW-THz homodyne spectroscopy and imaging system with EO phase modulation and direct lock-in detection. UTC-PD: Uni-traveling-carrier photodiode. PCA: Photo-conductive antenna.
shows the configuration of the CW-THz homodyne spectroscopy system [14

14. J.-Y. Kim, H.-J. Song, K. Ajito, M. Yaita, and N. Kukutsu, “Continuous-Wave THz Homodyne Spectroscopy and Imaging System With Electro-Optical Phase Modulation for High Dynamic Range,” IEEE Trans. THz Sci. Tech. (Paris) 3, 158–164 (2013).

]. For THz wave generation, two individual lightwaves from a fixed-wavelength DFB laser and a tunable laser (TLS) are combined and then photo-mixed in a UTC-PD. The EDFAs are used for both input lightwaves to compensate for the power loss throughout the fiber setup. In practical implementation, the two lighwave sources can be replaced with other combinations like two tunable DFB lasers or a dual-mode laser. For homodyne detection, the identical lightwaves are also delivered to the InGaAs-based photoconductive antenna (PCA). When the THz wave from the UTC-PD arrives at the PCA through free space, the PCA generates the detected signal by mixing the received THz-wave and the photonic THz wave propagating through the other path. In the free space, the THz wave emitted from the UTC-PD module integrated with a silicon lens is guided with lenses for beam focusing on the sample and the PCA. Both of the emitter and detector have a bow-tie antenna for broadband mode matching with polarization control of the radiated beam to minimize path loss.

For coherent detection, the phase of the photonic THz wave is linearly rotated by using EO phase modulators driven by saw-tooth signals. Then, a sine wave is formed in the detected signal by the continuous changing of the phase difference between the two THz waves. Both the phase and intensity responses of the THz wave passing through the sample directly appear in the detected signal. When the differential saw-tooth signals drive the two EO phase modulators, the frequency of THz phase rotation and the detected signal is an integer multiplication of the saw-tooth frequency. The balanced configuration also has a merit of balancing the path length from the source to detector, which suppresses the phase noise caused by the unlocked laser sources.

Because the detected signal at fixed frequency contains the phase and intensity responses, simple lock-in detection using the saw-tooth control signal as a reference clock enables fast and continuous readout of the THz responses in each period of the E/O phase control. In this system, the lock-in integration time for measuring a data point can be freely selected depending on the desired signal-to-noise ratio (S/N) or measurement speed. With the continuous data readout, spectroscopy and imaging can be done by a laser sweep or raster scan of the sample position. In both of these operations, the phase and intensity responses are always obtained individually.

The frequency response of the CW-THz homodyne system was measured from 0.2 to 2 THz with linear wavelength tuning of the TLS and synchronized lock-in detection of the output signal as shown in Figs. 2(a)
Fig. 2 Measured (a) intensity and (b) phase responses of the THz spectroscopy and imaging system from 0.2 to 2 THz through 30-cm-long air path.
and 2(b). The integration time of the lock-in detection was 3 ms. Because the lock-in detector used in the experiment needs a guide interval for each data point, the 64 data points were obtained in 1 s so that the total sweep time was about 15 s. for 900 frequency points with about 2-GHz steps. The measured signal linearly decreases as frequency increases due to the frequency response of the UTC-PD and PCA. At the same time, there is linear shift of the phase response due to the fixed offset of the travel lengths for the two THz waves.

In a spectroscopy and imaging system, measurement stability is another important figure of merit. This is especially true in a homodyne system with EO phase modulator: The separated lightwave propagates through individual paths from the power dividers after the EDFA to the power combiners before the emitter and detector. Because the lightwave has very short wavelength, environmental temperature ambiguity can induce very sensitive phase variation. To enhance the stability, one potential solution is to miniaturize the optical path by photonic integration or by using short fiber components, especially in the critical section from the optical power divider to combiner. Another is to achieve fast measurement, because the stability of the setup only needs to be maintained during the measurement time.

For the phase stability, short fiber components were used in the optical phase control and distribution section from the power divider to combiner, of which total length is about 40 cm. The fiber section was insulated in a box for thermal stability. Compared to our earlier version, in which the total fiber length was about 5 m [14

14. J.-Y. Kim, H.-J. Song, K. Ajito, M. Yaita, and N. Kukutsu, “Continuous-Wave THz Homodyne Spectroscopy and Imaging System With Electro-Optical Phase Modulation for High Dynamic Range,” IEEE Trans. THz Sci. Tech. (Paris) 3, 158–164 (2013).

], the setup size is reduced by ten times, resulting in a corresponding stabilization of the phase response. Figures 3(a)
Fig. 3 Measured long-term variation of (a) phase and (b) intensity responses for 1 hour.
and 3(b) shows the variation of the measured phase and intensity responses for 1 h in the open space of the experimental room. Although the phase response is largely changed as about ± 20° for 1 h, the expected phase variation inside 1 min is only about 1.5°. Therefore, in a high-speed measurement for less than a minute, the effect of environmental phase variation is negligible. The measured intensity stability is dominated by short-term variation of about 0.67%, which is equivalent to relative intensity noise of about −50 dB.

3. Fast vector spectroscopy and imaging demonstration for pharmaceutical tablet

The CW-THz homodyne system was used for vector spectroscopy and imaging of a pharmaceutical multi-component tablet. Figure 4(a)
Fig. 4 Measured THz imaging results of multi-component caffeine:oxalic-acid (Caf:Oca) cocrystal tablet with polyethylene (PE). (a) Photograph of tablet and composition of three sections with different Caf:Oxa cocrystal concentrations. (b) Intensity and (c) phase images of the tablet measured at different THz frequencies.
shows a photograph of the tablet. The tablet is composed of three sections, which cannot be well identified with a visual photograph. The sections are commonly based on polyethylene (PE) binder with different caffeine:oxalic-acid (Caf:Oxa) cocrystal concentrations of 0, 20 and 40%, respectively. Cocrystallization, a method of manipulating the crystal structure, can change the physicochemical properties of the active pharmaceutical ingredient, such as bioavailability. Previously, we demonstrated the advantages of THz spectroscopy for distinguishing the crystal structures of pharmaceuticals in a tablet [17

17. T. Göbel, D. Stanze, B. Globisch, R. J. B. Dietz, H. Roehle, and M. Schell, “Telecom technology based continuous wave terahertz photomixing system with 105 decibel signal-to-noise ratio and 3.5 terahertz bandwidth,” Opt. Lett. 38(20), 4197–4199 (2013). [CrossRef] [PubMed]

]. The diameter and thickness of the tablet are 10 and 1 mm, respectively.

Figures 4(b) and 4(c) show the intensity and phase images of the tablet at several different frequencies, measured with a raster scan of the sample position by continuously moving the translation stage. The imaging resolution is 64 × 40 and the measurement time for a pair of intensity and phase images at a fixed frequency point is about 40 s. For calibration, the reference intensity and phase response of the air path were measured at the frequency before each imaging was taken. The intensity images are plotted as relative transmittance, which was obtained by subtracting the reference intensity of the air path from the measured raw data. In the phase images, the relative phase differences of each pixel from the center of the tablet are plotted. In the intensity images, we can see the high absorption property in both side areas of the tablet at around 1.4 THz and the losses at the borderlines. In the phase images, the changes of the phase response between the areas more directly show that there are different materials having different refractive indexes. With the phase images, the intensity losses at the borderline can be identified as a scattering effect. From 1.2 to 1.6 THz, the phase responses do not increase linearly, which means that the refractive index is distorted in this frequency region. Therefore, from the intensity and phase images, we can identify the divided areas of the tablet and the different concentrations of the Caf:Oxa cocrytal in both side areas, which has absorption peak at 1.4 THz.

Figures 5(a) and (b)
Fig. 5 THz vector spectroscopy results of (a) absorbance and (b) relative permittivity from measured intensity and phase responses of sampled spatial points in each section of multi-component tablet.
show THz absorbance and relative permittivity calculated from measured THz intensity and phase spectra, respectively. The spectroscopic measurements were performed only at a sampled spatial point for each area of the tablet shown in Fig. 4(a). The spectroscopy results were obtained by subtracting the response of the air path as a reference from the measured sample responses. With integration time of 3 ms, measurement of single pair of intensity and phase spectra from 0.2 to 2 THz takes about 15 s for 900 frequency points with about 2-GHz steps. The absorbance is defined as the power loss of the sample on a logarithmic scale, and the relative permittivity was calculated from the measured phase spectra by using a simple wave equation.

4. Conclusion

A continuous-wave THz vector spectroscopy and imaging system was implemented using 1.55-µm fiber optics, including EO phase modulators, a UTC-PD and an InGaAs PCA. The reduced size of optical fiber setup enhances the phase stability of the system by more than ten times, resulting in phase error of 1.5° per minute. High dynamic ranges of 100 and 75 dB·Hz obtained at 300 GHz and 1 THz allow fast millisecond-order data acquisition and both the spectroscopy and imaging to be performed inside 1 min. Using the vector spectroscopy and imaging system, we demonstrated non-destructive characterization, including absorbance and relative permittivity for pharmaceutical cocrystals inside tablets at up to 2 THz, emphasizing the flexibility of the continuous-wave system for practical THz applications.

Acknowledgments

The authors thank Dr. T. Ishibashi and Dr. A. Wakatsuki of NTT, for providing the UTC-PDs, and Dr. O. Kagami of NTT for his encouragement.

References and links

1.

M. Tonouchi, “Cutting-edge terahertz technology,” Nat. Photonics 1(2), 97–105 (2007). [CrossRef]

2.

K. Ajito and Y. Ueno, “THz Chemical Imaging for Biological Applications,” IEEE Trans. THz Sci. Tech. (Paris) 1, 293–300 (2011).

3.

Z. Yan, Y. Ying, H. Zhang, and H. Yu, “Research progress of terahertz wave technology in food inspection,” Proc. SPIE 6373, 63730R (2006). [CrossRef]

4.

K. Kawase, Y. Ogawa, Y. Watanabe, and H. Inoue, “Non-destructive terahertz imaging of illicit drugs using spectral fingerprints,” Opt. Express 11(20), 2549–2554 (2003). [CrossRef] [PubMed]

5.

F. Friederich, W. V. Spiegel, M. Bauer, F. Meng, M. D. Thomson, S. Boppel, A. Lisauskas, B. Hils, V. Krozer, A. Keil, T. Löffler, R. Henneberger, A. K. Huhn, G. Spickermann, P. H. Bolívar, and H. G. Roskos, “THz Active Imaging Systems With Real-Time Capabilities,” IEEE Trans. THz Sci. Tech. (Paris) 1, 183–200 (2011).

6.

A. W. M. Lee, T.-Y. Kao, D. Burghoff, Q. Hu, and J. L. Reno, “Terahertz tomography using quantum-cascade lasers,” Opt. Lett. 37(2), 217–219 (2012). [CrossRef] [PubMed]

7.

J. R. Demers, R. T. Logan Jr, N. J. Bergeron, and E. R. Brown, “A coherent frequency-domain THz spectrometer with a signal-to-noise ratio 60 dB at 1 THz,” Proc. SPIE 6949, 694909 (2008). [CrossRef]

8.

A. Roggenbuck, H. Schmitz, A. Deninger, I. C. Mayorga, J. Hemberger, R. Güsten, and M. Grüninger, “Coherent broadband continuous-wave terahertz spectroscopy on solid-state samples,” New J. Phys. 12(4), 043017 (2010). [CrossRef]

9.

R. Wilk, F. Breitfeld, M. Mikulics, and M. Koch, “Continuous wave terahertz spectrometer as a noncontact thickness measuring device,” Appl. Opt. 47(16), 3023–3026 (2008). [CrossRef] [PubMed]

10.

I. S. Gregory, W. R. Tribe, C. Baker, B. E. Cole, M. J. Evans, L. Spencer, M. Pepper, and M. Missous, “Continuous-wave terahertz system with a 60 dB dynamic range,” Appl. Phys. Lett. 86(20), 204104 (2005). [CrossRef]

11.

N. Kim, S.-P. Han, H. Ko, Y. A. Leem, H.-C. Ryu, C. W. Lee, D. Lee, M. Y. Jeon, S. K. Noh, and K. H. Park, “Tunable continuous-wave terahertz generation/detection with compact 1.55 μm detuned dual-mode laser diode and InGaAs based photomixer,” Opt. Express 19(16), 15397–15403 (2011). [CrossRef] [PubMed]

12.

A. M. Sinyukov, Z. Liu, Y. L. Hor, K. Su, R. B. Barat, D. E. Gary, Z.-H. Michalopoulou, I. Zorych, J. F. Federici, and D. Zimdars, “Rapid-phase modulation of terahertz radiation for high-speed terahertz imaging and spectroscopy,” Opt. Lett. 33(14), 1593–1595 (2008). [CrossRef] [PubMed]

13.

T. Göbel, D. Schoenherr, C. Sydlo, M. Feiginov, P. Meissner, and H. L. Hartnagel, “Single-sampling-point coherent detection in continuous-wave photomixing terahertz systems,” Electron. Lett. 45(1), 65–66 (2009). [CrossRef]

14.

J.-Y. Kim, H.-J. Song, K. Ajito, M. Yaita, and N. Kukutsu, “Continuous-Wave THz Homodyne Spectroscopy and Imaging System With Electro-Optical Phase Modulation for High Dynamic Range,” IEEE Trans. THz Sci. Tech. (Paris) 3, 158–164 (2013).

15.

S. Hisatake, G. Kitahara, K. Ajito, Y. Fukada, N. Yoshimoto, and T. Nagatsuma, “Phase-sensitive terahertz self-heterodyne system based on photodiode and low-temperature-grown GaAs photoconductor at 1.55 μm,” IEEE Sens. J. 13(1), 31–36 (2013). [CrossRef]

16.

M. Suga, Y. Sasaki, T. Sasahara, T. Yuasa, and C. Otani, “THz phase-contrast computed tomography based on Mach-Zehnder interferometer using continuous wave source: proof of the concept,” Opt. Express 21(21), 25389–25402 (2013). [CrossRef] [PubMed]

17.

T. Göbel, D. Stanze, B. Globisch, R. J. B. Dietz, H. Roehle, and M. Schell, “Telecom technology based continuous wave terahertz photomixing system with 105 decibel signal-to-noise ratio and 3.5 terahertz bandwidth,” Opt. Lett. 38(20), 4197–4199 (2013). [CrossRef] [PubMed]

18.

H. Ito, T. Furuta, F. Nakajima, K. Yoshino, and T. Ishibashi, “Photonic Generation of Continuous THz Wave Using Uni-Traveling-Carrier Photodiode,” J. Lightwave Technol. 23(12), 4016–4021 (2005). [CrossRef]

19.

D. M. Charron, K. Ajito, J.-Y. Kim, and Y. Ueno, “Chemical mapping of pharmaceutical cocrystals using terahertz spectroscopic imaging,” Anal. Chem. 85(4), 1980–1984 (2013). [CrossRef] [PubMed]

20.

X. Xin, H. Altan, A. Saint, D. Matten, and R. R. Alfano, “Terahertz absorption spectrum of para and ortho water vapors at different humidities at room temperature,” J. Appl. Phys. 100(9), 094905 (2006). [CrossRef]

OCIS Codes
(300.6495) Spectroscopy : Spectroscopy, teraherz
(110.6795) Imaging systems : Terahertz imaging

ToC Category:
Terahertz optics

History
Original Manuscript: October 9, 2013
Revised Manuscript: November 22, 2013
Manuscript Accepted: January 9, 2014
Published: January 17, 2014

Virtual Issues
Vol. 9, Iss. 3 Virtual Journal for Biomedical Optics

Citation
Jae-Young Kim, Ho-Jin Song, Makoto Yaita, Akihiko Hirata, and Katsuhiro Ajito, "CW-THz vector spectroscopy and imaging system based on 1.55-µm fiber-optics," Opt. Express 22, 1735-1741 (2014)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-22-2-1735


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. M. Tonouchi, “Cutting-edge terahertz technology,” Nat. Photonics 1(2), 97–105 (2007). [CrossRef]
  2. K. Ajito, Y. Ueno, “THz Chemical Imaging for Biological Applications,” IEEE Trans. THz Sci. Tech. (Paris) 1, 293–300 (2011).
  3. Z. Yan, Y. Ying, H. Zhang, H. Yu, “Research progress of terahertz wave technology in food inspection,” Proc. SPIE 6373, 63730R (2006). [CrossRef]
  4. K. Kawase, Y. Ogawa, Y. Watanabe, H. Inoue, “Non-destructive terahertz imaging of illicit drugs using spectral fingerprints,” Opt. Express 11(20), 2549–2554 (2003). [CrossRef] [PubMed]
  5. F. Friederich, W. V. Spiegel, M. Bauer, F. Meng, M. D. Thomson, S. Boppel, A. Lisauskas, B. Hils, V. Krozer, A. Keil, T. Löffler, R. Henneberger, A. K. Huhn, G. Spickermann, P. H. Bolívar, H. G. Roskos, “THz Active Imaging Systems With Real-Time Capabilities,” IEEE Trans. THz Sci. Tech. (Paris) 1, 183–200 (2011).
  6. A. W. M. Lee, T.-Y. Kao, D. Burghoff, Q. Hu, J. L. Reno, “Terahertz tomography using quantum-cascade lasers,” Opt. Lett. 37(2), 217–219 (2012). [CrossRef] [PubMed]
  7. J. R. Demers, R. T. Logan, N. J. Bergeron, E. R. Brown, “A coherent frequency-domain THz spectrometer with a signal-to-noise ratio 60 dB at 1 THz,” Proc. SPIE 6949, 694909 (2008). [CrossRef]
  8. A. Roggenbuck, H. Schmitz, A. Deninger, I. C. Mayorga, J. Hemberger, R. Güsten, M. Grüninger, “Coherent broadband continuous-wave terahertz spectroscopy on solid-state samples,” New J. Phys. 12(4), 043017 (2010). [CrossRef]
  9. R. Wilk, F. Breitfeld, M. Mikulics, M. Koch, “Continuous wave terahertz spectrometer as a noncontact thickness measuring device,” Appl. Opt. 47(16), 3023–3026 (2008). [CrossRef] [PubMed]
  10. I. S. Gregory, W. R. Tribe, C. Baker, B. E. Cole, M. J. Evans, L. Spencer, M. Pepper, M. Missous, “Continuous-wave terahertz system with a 60 dB dynamic range,” Appl. Phys. Lett. 86(20), 204104 (2005). [CrossRef]
  11. N. Kim, S.-P. Han, H. Ko, Y. A. Leem, H.-C. Ryu, C. W. Lee, D. Lee, M. Y. Jeon, S. K. Noh, K. H. Park, “Tunable continuous-wave terahertz generation/detection with compact 1.55 μm detuned dual-mode laser diode and InGaAs based photomixer,” Opt. Express 19(16), 15397–15403 (2011). [CrossRef] [PubMed]
  12. A. M. Sinyukov, Z. Liu, Y. L. Hor, K. Su, R. B. Barat, D. E. Gary, Z.-H. Michalopoulou, I. Zorych, J. F. Federici, D. Zimdars, “Rapid-phase modulation of terahertz radiation for high-speed terahertz imaging and spectroscopy,” Opt. Lett. 33(14), 1593–1595 (2008). [CrossRef] [PubMed]
  13. T. Göbel, D. Schoenherr, C. Sydlo, M. Feiginov, P. Meissner, H. L. Hartnagel, “Single-sampling-point coherent detection in continuous-wave photomixing terahertz systems,” Electron. Lett. 45(1), 65–66 (2009). [CrossRef]
  14. J.-Y. Kim, H.-J. Song, K. Ajito, M. Yaita, N. Kukutsu, “Continuous-Wave THz Homodyne Spectroscopy and Imaging System With Electro-Optical Phase Modulation for High Dynamic Range,” IEEE Trans. THz Sci. Tech. (Paris) 3, 158–164 (2013).
  15. S. Hisatake, G. Kitahara, K. Ajito, Y. Fukada, N. Yoshimoto, T. Nagatsuma, “Phase-sensitive terahertz self-heterodyne system based on photodiode and low-temperature-grown GaAs photoconductor at 1.55 μm,” IEEE Sens. J. 13(1), 31–36 (2013). [CrossRef]
  16. M. Suga, Y. Sasaki, T. Sasahara, T. Yuasa, C. Otani, “THz phase-contrast computed tomography based on Mach-Zehnder interferometer using continuous wave source: proof of the concept,” Opt. Express 21(21), 25389–25402 (2013). [CrossRef] [PubMed]
  17. T. Göbel, D. Stanze, B. Globisch, R. J. B. Dietz, H. Roehle, M. Schell, “Telecom technology based continuous wave terahertz photomixing system with 105 decibel signal-to-noise ratio and 3.5 terahertz bandwidth,” Opt. Lett. 38(20), 4197–4199 (2013). [CrossRef] [PubMed]
  18. H. Ito, T. Furuta, F. Nakajima, K. Yoshino, T. Ishibashi, “Photonic Generation of Continuous THz Wave Using Uni-Traveling-Carrier Photodiode,” J. Lightwave Technol. 23(12), 4016–4021 (2005). [CrossRef]
  19. D. M. Charron, K. Ajito, J.-Y. Kim, Y. Ueno, “Chemical mapping of pharmaceutical cocrystals using terahertz spectroscopic imaging,” Anal. Chem. 85(4), 1980–1984 (2013). [CrossRef] [PubMed]
  20. X. Xin, H. Altan, A. Saint, D. Matten, R. R. Alfano, “Terahertz absorption spectrum of para and ortho water vapors at different humidities at room temperature,” J. Appl. Phys. 100(9), 094905 (2006). [CrossRef]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.

Figures

Fig. 1 Fig. 2 Fig. 3
 
Fig. 4 Fig. 5
 

« Previous Article  |  Next Article »

OSA is a member of CrossRef.

CrossCheck Deposited