Effects of spatial coherence in diffraction phase microscopy

Chris Edwards, Basanta Bhaduri, Tan Nguyen, Benjamin G. Griffin, Hoa Pham, Taewoo Kim, Gabriel Popescu, and Lynford L. Goddard  »View Author Affiliations


Optics Express, Vol. 22, Issue 5, pp. 5133-5146 (2014)
http://dx.doi.org/10.1364/OE.22.005133


View Full Text Article

Acrobat PDF (1731 KB)





Browse Journals / Lookup Meetings

Browse by Journal and Year


   


Lookup Conference Papers

Close Browse Journals / Lookup Meetings

Article Tools

Share
Citations

Abstract

Quantitative phase imaging systems using white light illumination can exhibit lower noise figures than laser-based systems. However, they can also suffer from object-dependent artifacts, such as halos, which prevent accurate reconstruction of the surface topography. In this work, we show that white light diffraction phase microscopy using a standard halogen lamp can produce accurate height maps of even the most challenging structures provided that there is proper spatial filtering at: 1) the condenser to ensure adequate spatial coherence and 2) the output Fourier plane to produce a uniform reference beam. We explain that these object-dependent artifacts are a high-pass filtering phenomenon, establish design guidelines to reduce the artifacts, and then apply these guidelines to eliminate the halo effect. Since a spatially incoherent source requires significant spatial filtering, the irradiance is lower and proportionally longer exposure times are needed. To circumvent this tradeoff, we demonstrate that a supercontinuum laser, due to its high radiance, can provide accurate measurements with reduced exposure times, allowing for fast dynamic measurements.

© 2014 Optical Society of America

1. Introduction

1.1. Quantitative phase imaging (QPI)

QPI methods exploit the fact that the phase of the imaging field is typically much more informative than its amplitude [1

1. G. Popescu, Quantitative Phase Imaging of Cells and Tissues, McGraw-Hill biophotonics (McGraw-Hill, 2011).

]. The relative phase shifts measured using these techniques contain topographical as well as refractive index information about the specimen under investigation. Diffraction phase microscopy (DPM) is a QPI method that was first demonstrated in 2006 using laser illumination [2

2. G. Popescu, T. Ikeda, R. R. Dasari, and M. S. Feld, “Diffraction phase microscopy for quantifying cell structure and dynamics,” Opt. Lett. 31(6), 775–777 (2006). [CrossRef] [PubMed]

]. DPM utilizes a compact Mach-Zehnder interferometer to combine many of the best attributes of current QPI methods [2

2. G. Popescu, T. Ikeda, R. R. Dasari, and M. S. Feld, “Diffraction phase microscopy for quantifying cell structure and dynamics,” Opt. Lett. 31(6), 775–777 (2006). [CrossRef] [PubMed]

4

4. C. Edwards, K. Wang, R. Zhou, B. Bhaduri, G. Popescu, and L. L. Goddard, “Digital projection photochemical etching defines gray-scale features,” Opt. Express 21(11), 13547–13554 (2013). [CrossRef] [PubMed]

]. This common-path geometry inherently cancels out most mechanisms responsible for noise and the single-shot measurement allows acquisition rates limited only by the speed of the camera employed. This technique does not require coating or staining of the specimen and is therefore non-destructive. This unique collection of features enables the DPM system to accurately monitor the dynamics of various nanoscale phenomena in their natural environments. Our laser DPM system operates in both transmission and reflection in order to accommodate both transparent and reflective samples and is currently being used to monitor wet etching [3

3. C. Edwards, A. Arbabi, G. Popescu, and L. L. Goddard, “Optically monitoring and controlling nanoscale topography during semiconductor etching,” Light Sci. Appl. 1(9), e30 (2012). [CrossRef]

], photochemical etching [4

4. C. Edwards, K. Wang, R. Zhou, B. Bhaduri, G. Popescu, and L. L. Goddard, “Digital projection photochemical etching defines gray-scale features,” Opt. Express 21(11), 13547–13554 (2013). [CrossRef] [PubMed]

], surface wetting and evaporation, expansion and deformation of materials, and to perform semiconductor wafer defect detection [5

5. R. Zhou, G. Popescu, and L. L. Goddard, “22 nm node wafer inspection using diffraction phase microscopy and image post-processing,” Proc. SPIE 8681, 8610G (2013).

]. Other research groups have exploited the multimodal imaging capabilities of DPM by combining it with other imaging modalities such as fluorescence and confocal/Raman in order to study morphology and dynamics of biological structures [6

6. Y. Park, G. Popescu, K. Badizadegan, R. R. Dasari, and M. S. Feld, “Diffraction phase and fluorescence microscopy,” Opt. Express 14(18), 8263–8268 (2006). [CrossRef] [PubMed]

13

13. S. Witte, A. Plauşka, M. C. Ridder, L. van Berge, H. D. Mansvelder, and M. L. Groot, “Short-coherence off-axis holographic phase microscopy of live cell dynamics,” Biomed. Opt. Express 3(9), 2184–2189 (2012). [CrossRef] [PubMed]

].

Recently, white light DPM (wDPM) was developed for use in biological applications [9

9. Y. Park, T. Yamauchi, W. Choi, R. Dasari, and M. S. Feld, “Spectroscopic phase microscopy for quantifying hemoglobin concentrations in intact red blood cells,” Opt. Lett. 34(23), 3668–3670 (2009). [CrossRef] [PubMed]

, 14

14. B. Bhaduri, H. Pham, M. Mir, and G. Popescu, “Diffraction phase microscopy with white light,” Opt. Lett. 37(6), 1094–1096 (2012). [CrossRef] [PubMed]

17

17. H. V. Pham, B. Bhaduri, K. Tangella, C. Best-Popescu, and G. Popescu, “Real time blood testing using quantitative phase imaging,” PLoS ONE 8(2), e55676 (2013). [CrossRef] [PubMed]

]. White light imaging systems using standard halogen lamp illumination can exhibit lower spatial phase noise than their laser counterparts [3

3. C. Edwards, A. Arbabi, G. Popescu, and L. L. Goddard, “Optically monitoring and controlling nanoscale topography during semiconductor etching,” Light Sci. Appl. 1(9), e30 (2012). [CrossRef]

, 4

4. C. Edwards, K. Wang, R. Zhou, B. Bhaduri, G. Popescu, and L. L. Goddard, “Digital projection photochemical etching defines gray-scale features,” Opt. Express 21(11), 13547–13554 (2013). [CrossRef] [PubMed]

, 14

14. B. Bhaduri, H. Pham, M. Mir, and G. Popescu, “Diffraction phase microscopy with white light,” Opt. Lett. 37(6), 1094–1096 (2012). [CrossRef] [PubMed]

, 18

18. T. Slabý, P. Kolman, Z. Dostál, M. Antoš, M. Lošťák, and R. Chmelík, “Off-axis setup taking full advantage of incoherent illumination in coherence-controlled holographic microscope,” Opt. Express 21(12), 14747–14762 (2013). [CrossRef] [PubMed]

]. This is a result of the lower coherence, both spatially and temporally, which reduces some noise mechanisms such as speckle [19

19. J. W. Goodman, “Statistical properties of laser speckle patterns,” in Laser Speckle and Related Phenomena (Springer, 1975).

]. Unfortunately, white light systems can also exhibit some object-dependent artifacts not present in laser systems. In this work, we show that both a lack of spatial coherence and inadequate spatial filtering of the reference beam can individually cause high-pass filtering of the DPM image, resulting in the halo effect and shade-off. The halo effect can result in erroneous height values, typically negative, around the perimeter of an object. Shade-off occurs in conjunction with the halo effect and results in the center of large area objects having reduced contrast in measured intensity and/or reduced phase shifts in QPI [20

20. P. Gao, B. Yao, I. Harder, N. Lindlein, and F. J. Torcal-Milla, “Phase-shifting Zernike phase contrast microscopy for quantitative phase measurement,” Opt. Lett. 36(21), 4305–4307 (2011). [CrossRef] [PubMed]

, 21

21. T. Wilson and C. J. R. Sheppard, “The halo effect of image processing by spatial frequency filtering,” Optik (Stuttg.) 59(1), 19–23 (1981).

]. By conducting a thorough theoretical and experimental investigation, we present guidelines for obtaining adequate spatial coherence and filtering and demonstrate the removal of these object-dependent artifacts.

The spatial filtering used to obtain adequate spatial coherence reduces the illuminating power. This in turn increases the necessary camera exposure time and prevents high-throughput real-time imaging [17

17. H. V. Pham, B. Bhaduri, K. Tangella, C. Best-Popescu, and G. Popescu, “Real time blood testing using quantitative phase imaging,” PLoS ONE 8(2), e55676 (2013). [CrossRef] [PubMed]

]. We solve this tradeoff by showing that the much higher radiance of a supercontinuum laser results in accurate measurements with low exposure times.

1.2 Experimental setup

Fig. 1 wDPM imaging system. A diffraction grating, 4f lens system, and a spatial light modulator (SLM) are used to achieve interference. The SLM performs spatial filtering and one order can be gray-scaled in order to match intensities and optimize the fringe visibility. The interferogram captured by the CCD is a spatially modulated signal, which allows us to extract the phase information via a Hilbert transform and reconstruct the surface topography.
Figure 1 shows a schematic of the wDPM system, which can be placed at the output port of a conventional light microscope. We create an interferometer by using a diffraction grating, a 4f lens system, and a spatial light modulator (SLM). The diffraction grating creates multiple copies of the image at various angles, some of which are captured by the 1st lens. Under a 4f configuration, the first lens takes the Fourier transform. An SLM is placed in the Fourier plane where the spatial filtering is performed. Here, all of the orders except the 0 and + 1 are blocked. The 1st order contains the full image and the 0th order is filtered down using a small circular mask which serves as a pinhole. The pinhole diameter is chosen to be small enough such that, after the 2nd lens takes another Fourier transform, the field approaches a plane wave and is uniform across the CCD to serve as our reference beam. The two beams interfere at the CCD to produce an interferogram. This interferogram is a spatially modulated signal, which allows us to extract the phase information via a Hilbert transform [3

3. C. Edwards, A. Arbabi, G. Popescu, and L. L. Goddard, “Optically monitoring and controlling nanoscale topography during semiconductor etching,” Light Sci. Appl. 1(9), e30 (2012). [CrossRef]

, 22

22. H. V. Pham, C. Edwards, L. L. Goddard, and G. Popescu, “Fast phase reconstruction in white light diffraction phase microscopy,” Appl. Opt. 52(1), A97–A101 (2013). [CrossRef] [PubMed]

].

1.3 Spatial coherence in conventional microscopes

Commercial microscopes are often incorporated into experimental setups when researching and developing new imaging techniques. Unfortunately, most microscopes are not designed for quantitative phase imaging. For example, the condenser aperture of the Zeiss Z2 Axio-Imager only closes down to a numerical aperture (NAcon) of about 0.09. This results in a coherence area much smaller than the field-of-view (FOV) for typical objectives. Note that for laser-based systems, this is not an issue since the effective NAcon is approximately zero. The normalized mutual intensity function formed by the circular aperture of the condenser (actually a heptagon shaped iris) under partially coherent illumination is [23

23. J. W. Goodman, Introduction to Fourier Optics (Roberts & Company, 2005).

, 24

24. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (Wiley, 2013).

],
|g(x1,x2,y1,y2)|=|2J1(πρθs/λ)πρθs/λ|
(1)
where J1 is a Bessel function of the first kind, ρ=(x2x1)2+(y2y1)2, θs=2NAcon is the angle subtended by the source, and λ is the mean wavelength of the illumination. The spatial coherence radius, ρc, is often defined as the distance from the peak to its first zero (radius of coherence area) [24

24. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (Wiley, 2013).

, 25

25. M. Born and E. Wolf, Principles of optics, 7th (expanded) ed. (Pergamon Press, 1999).

]. Thus, ρc=0.61λ/NAcon. For our case, with a mean wavelength of 574 nm and an NAcon = 0.09, we obtain ρc = 3.9 µm. Throughout this paper, images are captured using a 20X, 0.8 NA objective resulting in a lateral resolution of 0.87 μm and a FOV of 135 x 180 μm2 in the sample plane. Thus, the radius of a circle that circumscribes the FOV is r = 112 µm. Intuitively, it is desirable to have ρc ≥ r. As we will later show, this condition is actually too weak and it is necessary to have the mutual intensity function be nearly flat over the object being measured, which as a general rule of thumb, requires that the coherence diameter be at least ten times the lateral extent of the object.

To experimentally determine ρc, the grating was removed from the wDPM setup and the Fourier plane was imaged to measure the angular spread of the source. By taking the inverse Fourier transform, we obtain the autocorrelation function via the Wiener-Khintchine theorem [1

1. G. Popescu, Quantitative Phase Imaging of Cells and Tissues, McGraw-Hill biophotonics (McGraw-Hill, 2011).

, 23

23. J. W. Goodman, Introduction to Fourier Optics (Roberts & Company, 2005).

, 24

24. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (Wiley, 2013).

].
Fig. 2 Spatial coherence of the wDPM imaging system. (a) Angular spectrum of the source showing the full-width at half-maximum (FWHM) approximation. (b) Horizontal slice taken through the origin in (a). The radius of the unscattered light circle is approximately k0NAcon. (c) Autocorrelation function computed from (a). (d) Horizontal slice taken through origin in (c). The spatial coherence radius is ρc = 4.0 μm.
Figures 2(a) and 2(b) show a 2-dimensional (2D) image of the Fourier plane and the corresponding 1-dimensional (1D) kx slice. The maximum transverse component of the wave vector (kρ=kx2+ky2=k0NAcon) is inversely proportional to the degree of spatial coherence. As the condenser aperture approaches zero, full spatial coherence across the FOV is achieved. Figures 2(c) and 2(d) show the resulting normalized 2D autocorrelation function and 1D slice. The peak to first zero distance is measured to be ρc = 4.0 µm. The coherence area is simply Ac = πρc2 [24

24. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (Wiley, 2013).

]. The spatial coherence present in this conventional microscope, even with the condenser completely closed is not adequate for proper quantitative phase imaging of many structures.

In addition to spatial coherence, wDPM also needs adequate temporal coherence to produce proper interferograms. In the Appendix, we present measurement data showing that the temporal coherence of the wDPM light source is at least an order of magnitude larger than what is needed to measure the samples discussed in this paper.

2. Theoretical analysis

2.1 DPM with partially coherent illumination

In this section, we present a theoretical analysis of the phase measured by DPM under partially coherent illumination. The full derivation will be presented in [26

26. T. Nguyen, C. Edwards, L. Goddard, and G. Popescu, “Quantitative phase imaging with partially coherent illumination,” (to be submitted).

]. To simplify the analysis, low-pass filtering from the microscope’s objective lens is ignored. This phenomenon is already well understood and moreover, it is negligible when imaging large objects with high NA objectives. Let U1 be the imaging field and U0 be the reference field. Both of these are random fields, but the CCD measures mutual intensity functions, which are deterministic [27

27. T. Kim, R. Zhu, T. H. Nguyen, R. Zhou, C. Edwards, L. L. Goddard, and G. Popescu, “Deterministic signal associated with a random field,” Opt. Express 21(18), 20806–20820 (2013). [CrossRef] [PubMed]

]. The interferogram captured by the CCD can be written as:
I(r)=|U0(r,t)+U1(r,t)|2T=I0(r)+I1(r)+2{U0*(r,t)U1(r,t)T}
(2)
where I0 and I1are the intensities of the reference and imaging fields respectively. The modulation term contains the temporal cross-correlation of the two interfering fields. In DPM, we recover the phase image from the modulation term as [26

26. T. Nguyen, C. Edwards, L. Goddard, and G. Popescu, “Quantitative phase imaging with partially coherent illumination,” (to be submitted).

]:
ϕ(r)=arg[T(r)]arg[(Γillum*(r)Π˜slm(r))T(r)]
(3)
where T is the complex-valued transmittance function of the sample, Πslmis the SLM filter aperture function, a tilde denotes a Fourier transform, is the 2D convolution operator, Γillum(rr')Uillum(r,t)Uillum*(r',t)Tis the mutual intensity function, UillumU˜s(r,t)Π˜con(r)is the illumination field that interacts with the sample, U˜s is the field emitted by the halogen source (in a conjugate Fourier plane), and Πcon is the condenser aperture function. Note that Γillum that depends only on rr' since we assume the source field U˜s is shift invariant or isoplanatic [24

24. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (Wiley, 2013).

, 27

27. T. Kim, R. Zhu, T. H. Nguyen, R. Zhou, C. Edwards, L. L. Goddard, and G. Popescu, “Deterministic signal associated with a random field,” Opt. Express 21(18), 20806–20820 (2013). [CrossRef] [PubMed]

].

The 1st term in (4), ϕ1(r)=arg[T(r)] is the phase information from the sample and is referred to as the true phase image. The 2nd term ϕ0(r)=arg[(Γi*(r)Π˜slm(r))T(r)] is a low-pass filtered (LPF) version of the true phase image and is referred to as the cancellation phase. Thus, the relative phase measured by DPM is the true phase image minus a LPF version of itself. This subtraction produces a high pass filtered (HPF) image, which makes the lower spatial frequencies of the sample unattainable. To recover the true phase image, the 2nd term needs to be spatially constant, which requires that Γillum*(r)Π˜slm(r) be uniform across the FOV. This requires that: (1) the condenser aperture be closed down far enough to ensure adequate spatial coherence and (2) the SLM filter be small enough to ensure a uniform reference beam.

2.2 The effects of spatial filtering on DPM image formation

Fig. 3 DPM image formation. (a) Simulation of the ideal height map for a 3 μm diameter polystyrene microbead in immersion oil. (b) Simulated height from the cancellation field under non-ideal conditions. (c) Simulation of recovered height, which is the difference between (a) and (b). (d) Horizontal slice taken through origin showing (a-c). Note that cancellation field in (b) contains residues of the structure which results in the observed halo effect and a reduction in the measured height. (e-h) Corresponding simulations for a red blood cell in Coulter solution showing a reduction in the measured height as well as a halo and a negative dimple.
Figure 3 contains simulation results showing the effects of spatial filtering. The computed height is proportional to the measured phase [3

3. C. Edwards, A. Arbabi, G. Popescu, and L. L. Goddard, “Optically monitoring and controlling nanoscale topography during semiconductor etching,” Light Sci. Appl. 1(9), e30 (2012). [CrossRef]

]. Note that as the spatial filtering is improved, h0 becomes flat and Δh approaches h1. Figure 3 shows simulations of the true height imageh1(r)=λ2πΔnϕ1(r), the height of the cancellation field h0(r)=λ2πΔnϕ0(r), the resulting height map, and the cross-section for polystyrene microbeads in immersion oil (Figs. 3(a-d)) and for red blood cells (RBCs) in Coulter solution (Figs. 3(e-h)) under non-ideal conditions. The halo can be seen in Fig. 3(d) and the halo and negative dimple can be seen in Fig. 3(h). The halo and shade-off are both produced simultaneously via the HPF effect. Such halos have been observed notoriously in phase contrast microscopy, as well as in high-throughput white light QPI systems [15

15. H. Pham, B. Bhaduri, H. Ding, and G. Popescu, “Spectroscopic diffraction phase microscopy,” Opt. Lett. 37(16), 3438–3440 (2012). [CrossRef] [PubMed]

, 17

17. H. V. Pham, B. Bhaduri, K. Tangella, C. Best-Popescu, and G. Popescu, “Real time blood testing using quantitative phase imaging,” PLoS ONE 8(2), e55676 (2013). [CrossRef] [PubMed]

, 21

21. T. Wilson and C. J. R. Sheppard, “The halo effect of image processing by spatial frequency filtering,” Optik (Stuttg.) 59(1), 19–23 (1981).

, 28

28. B. Bhaduri, D. Wickland, R. Wang, V. Chan, R. Bashir, and G. Popescu, “Cardiomyocyte Imaging Using Real-Time Spatial Light Interference Microscopy (SLIM),” PLoS ONE 8(2), e56930 (2013). [CrossRef] [PubMed]

31

31. Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, and G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express 19(2), 1016–1026 (2011). [CrossRef] [PubMed]

].

3. Experimental results

3.1 Spatial filtering study

Experiments were done to validate the theory presented above for DPM image formation. A quartz wafer was patterned and etched, resulting in a variety of 123 nm tall (as verified by the Alpha Step IQ Profilometer) square pillars with different lateral sizes, to serve as the calibration sample. Two key parameters: the NA of the condenser (NAcon) and the diameter of the SLM pinhole filter (D) were varied and DPM height images were collected.
Fig. 4 The effects of spatial filtering in DPM. (a) Log-log plot showing the investigated parameter space of SLM pinhole filter diameter (D) and condenser numerical aperture (NAcon). The solid line represents the matched filter condition. Labels b-k correspond to the operating point for subsequent subfigures. (b-k) 0th order reference amplitude images, recovered height images, and corresponding height cross-sections. The cross-section figures show the wDPM profile in red and the Alpha Step profile in blue. The spatial noise (σs) and exposure times (texp) are also listed for each subfigure. Both adequate spatial coherence and proper spatial filtering are required in order to properly reconstruct the height maps using wDPM. As both the condenser NA and the SLM pinhole diameter are reduced, aberrations such as halo and shade-off are reduced and the pillar heights converge to their proper value (h = 123 nm) as verified with the Alpha Step IQ Profilometer. † denotes a physical pinhole in the SLM plane, and ‡ denotes a physical pinhole with SLM attenuation to better match intensities. * indicates a gain setting of 2 on the Zeiss AxioCam MRm CCD.
Figure 4(a) shows a plot of the investigated parameter space. The labels in Fig. 4(a) correspond to the operating points for subsequent subfigures.

To further reduce the numerical aperture of the condenser beyond the conventional limit (NAcon = 0.09), a small 400 μm pinhole (NAcon = 0.014) was placed inside the condenser turret and the system was realigned. This increases the coherence radius from ρc = 3.9 µm to ρc = 25 µm. Figure 4(d) shows the 0th order reference amplitude image along with its recovered height map. Here, the SLM pinhole is the same size as in Fig. 4(c), but NAcon is about six times smaller. Despite the reference still containing small residues of the pillar and not being perfectly uniform, the height map recovered under these conditions is beginning to show the structure of the pillar. Figure 4(e) shows a factor of two reduction in the SLM pinhole diameter for the same NAcon = 0.014 value. As predicted by Eq. (4), there is no discernible change in the height map when the SLM pinhole is reduced beyond its optimum value; however, a reduction in the pinhole diameter requires the exposure time to increase. Again, this is because once the pinhole is small enough, its diffraction pattern becomes uniform over the mutual intensity function's central lobe, which now becomes dominant, and further decrease in the pinhole diameter has no effect. To further reduce the numerical aperture of the condenser, a 200 μm pinhole was placed into the condenser turret (NAcon = 0.0072, ρc = 50 µm). The optimum SLM pinhole diameter for NAcon = 0.0072 is D = 40 μm. Under these conditions, the halo in Fig. 4(f) is slightly reduced and the measured height (h = 99 nm) is closer to the Profilometer value (h = 123 nm). The gain on the CCD was set to 2, which lowered the required exposure time from about 600 ms down to 300 ms.

Figures 4(g) and 4(h) show a comparison between a 35 μm physical pinhole and a 2x2 pixel (26 μm) SLM filter with NAcon = 0.0072. This comparison was done to test whether the measured 400:1 contrast ratio of the SLM or the fact that a 2x2 pixel filter is square and not circular has a considerable effect on the recovered height map. These effects appear to be negligible. It is important though to note that the exposure time was higher for the SLM filter than the physical pinhole due to the presence of the crossed polarizers required with the SLM, which decreases the power in the interferogram by at least a factor of two.

Figures 4(i)-4(k) compare different NAcon values with a 10 μm physical pinhole. This is the typical pinhole diameter used in laser DPM because it produces an adequately uniform reference across the CCD for laser DPM [3

3. C. Edwards, A. Arbabi, G. Popescu, and L. L. Goddard, “Optically monitoring and controlling nanoscale topography during semiconductor etching,” Light Sci. Appl. 1(9), e30 (2012). [CrossRef]

, 4

4. C. Edwards, K. Wang, R. Zhou, B. Bhaduri, G. Popescu, and L. L. Goddard, “Digital projection photochemical etching defines gray-scale features,” Opt. Express 21(11), 13547–13554 (2013). [CrossRef] [PubMed]

]. Figure 4(i) required additional attenuation from the SLM to get adequate fringe visibility, but the SLM cut the power by at least 2x, resulting in a large exposure time, texp = 1.5 s, and additional noise in the final image. Figure 4(j) looks nearly identical to 4(f), 4(g), and 4(h) showing that reducing the SLM pinhole size below the optimum value does not bring back the low frequency content to the image. Figure 4(k) shows that using a 100 μm pinhole in the condenser (NAcon = 0.0036, ρc = 100 µm) along with a 10 μm pinhole in the SLM plane allows for proper reconstruction of the 20 μm wide pillar. The minimum exposure time under proper imaging conditions was 360 ms. The exposure time increases with decreasing NAcon, decreasing D, or using an SLM that requires polarizers. It is also important to note that the exposure time can be set arbitrarily low by increasing the CCD’s gain factor, but at the expense of lower SNR since noise is amplified along with signal. Ideally, a brighter source should be used for fast, accurate, low-noise imaging.

This spatial filtering study shows the importance of maintaining proper spatial coherence and filtering in DPM.

3.2 Required spatial coherence for a given object size

Fig. 5 Required spatial coherence for a given object size. Cross-sections were taken of square micropillars of different widths for various condenser numerical apertures using the optimum SLM pinhole filter size for each case. (a-d) 5, 10, 20, and 40 μm wide square pillars respectively. As the condenser NA is reduced and the SLM filter is optimized, the recovered structure becomes more quantitatively correct. Halo and shade-off disappear simultaneously as the pillar height converges to the proper value. Larger objects require more spatial coherence (smaller NAcon) to image properly. The 5, 10, and 20 μm wide pillars image properly for NAcon ≤ 0.014, 0.0072, and 0.0036, respectively. The 40 μm wide pillar will require an even smaller NAcon to image properly.
Figure 5 contains cross-sections of square pillars of different widths taken with various NAcon using the optimum SLM pinhole filter size for each case. This study indicates that larger lateral features are more difficult to quantitatively reconstruct and require a larger degree of spatial coherence. This is due to the simple fact that larger features in the spatial domain map more information to lower frequencies in the spatial frequency domain. The low frequency information is lost by the HPF effects in DPM if proper filtering is not obtained. For NAcon = 0.09, 0.014, 0.0072, and 0.0036, the corresponding spatial coherence diameters are 2ρc = 7.8, 50, 100, and 200 μm, respectively.

Figure 5(a) shows cross-sections of a 5 μm wide square pillar with varying NAcon values. A conventional microscope with the condenser fully closed (NAcon = 0.09) is not able to properly reproduce the height map even for a 5 μm wide square. The heights converge for all values of NAcon ≤ 0.014. Note that the shape of the pillar is slightly distorted due to LPF effects from the objective (20X, NAobj = 0.8) which broadens the edges due to convolution with the point spread function (PSF) and also causes a peak at the edges (Gibbs’ phenomenon) due to missing high frequency components required to reconstruct the vertical edge. These effects are less apparent for larger structures, where the dimensions are much larger than the diffraction spot, which has a 0.87 μm radius in the sample plane. The dimensions of the squares were verified using the Alpha Step IQ Profilometer.

Figure 5(b) shows cross-sections of a 10 μm wide square pillar. All heights converge to the proper value for NAcon ≤ 0.0072. Note that the height map produced with NAcon = 0.014 contains a small halo and reduction in height which is indicative of a lack of spatial coherence. For NAcon = 0.09, only the pillar’s edges are visible and the center height of the pillar is nearly zero. Figure 5(c) shows similar cross-sections for a 20 μm wide pillar, which are the cross sections of the images presented in Fig. 6.
Fig. 6 Halo reduction in 3 μm diameter polystyrene microbeads in immersion oil. wDPM height map of microbeads for (a) closed condenser (NAcon = 0.09), (b) a 1mm iris placed in the condenser (NAcon = 0.036), and (c) a 100 μm pinhole placed in the condenser (NAcon = 0.0036). (d) Overlay showing cross-sections of a diagonal slice taken through the center of both beads. Closing the condenser beyond the conventional limit of NAcon = 0.09 to NAcon = 0.036 allows the heights to converge to their proper values, while additional spatial coherence using NAcon = 0.0036 is needed to remove the halo. Residual ringing is due to the Gibbs’ phenomenon.
The height converges to the proper value for NAcon ≤ 0.0036. Figure 5(d) shows cross-sections for a 40 μm wide square pillar. Here, even with a 100 μm pinhole in the condenser (NAcon = 0.0036, ρc = 100 µm), the height of the pillar has not converged and aberrations are still present. This study shows that for a given size structure, the condenser NA must be chosen such that the coherence area is much greater than the FOV. The condition that the coherence area circle at least circumscribes the FOV is too weak; it is actually necessary that the mutual intensity function be relatively flat over the object of interest. For a 20X objective in our setup, the radius of a circle circumscribing the rectangular FOV is 112 μm. The spatial coherence radius with an NAcon = 0.0036 is ρc = 100 μm, yet we can only properly image structures that are 20 μm in diameter or smaller. From this object size scaling study, we obtain a general rule of thumb that the coherence diameter, 2ρc, should be at least ten times the lateral size of the object being measured.

3.3 Halo reduction in microbeads and red blood cells

Figures 6(a)-6(c) show wDPM images of 3 μm polystyrene microbeads in immersion oil with an NAcon = 0.09, 0.036, and 0.0036, using a closed condenser, a 1 mm condenser iris, and a 100 μm condenser pinhole, respectively. Figure 6(d) shows an overlay of the corresponding cross-sections which were taken diagonally through the centers of both beads. With the condenser aperture closed (NAcon = 0.09), the measured heights of the beads are lower than their expected value. Care must be taken when imaging microbeads, especially with a lack of spatial coherence, since the measured heights are very sensitive to the focus. Defocusing the bead can actually result in a measured height much larger than the correct value. To avoid such issues, the most in-focus image was taken from a stack of 60 images with 200 nm z-sections. The cross-sections in Fig. 6(d) show a reduction in the halo and convergence to the correct height upon reducing the numerical aperture of the condenser. For 3 μm diameter beads, the calculated NAcon using our rule of thumb to achieve adequate spatial coherence is 0.024, which is in between our 1 mm iris and 400 μm pinhole. The 1 mm iris results in an NAcon = 0.036, which is enough to get the heights to converge, but a small halo remains. A 100 μm pinhole with an NAcon = 0.0036 was therefore used in an attempt to completely remove the halo. Here, the halo is removed and only a small amount of ringing remains, which is a result of the Gibbs’ phenomenon mentioned earlier.

3.4 DPM using a supercontinuum laser source

To get enough spatial coherence to measure an object that extends across most of the FOV for the widest FOV scenario (5X objective), the required NA is estimated to be 1.5x10−4, which is 600x smaller than what is achievable with commercial microscopes. This would require a miniscule pinhole (~4 µm) to be placed in the condenser turret, which would result in a signal that is exceedingly weak, thereby requiring long exposure times. The long integration time makes this approach impractical given camera noise considerations and the inability to observe fast dynamics. As a practical compromise in a DPM system with a conventional incoherent source, measurements can be limited to smaller objects so that the quantitative phase values are correct for each object. Using a 100 µm condenser pinhole will allow us to greatly reduce the halo effect in images from most biological samples while still maintaining a good SNR. The shortest exposure time obtained using a 100 µm pinhole in the condenser and a 10 µm physical pinhole in the SLM plane was 360 ms. The typical exposure time for wDPM with NAcon = 0.09 is about 10-20 ms.

Fig. 8 DPM using an SCL with a 750nm shortpass filter. (a) Spectrum of the SCL source with power-equivalent bandwidth (PEB) approximation. The mean wavelength is 599.6 nm. (b) 0th order reference amplitude image is uniform. (c-f) Recovered height images and corresponding cross-sections for 5, 10, 20, and 40 μm wide square pillars, respectively. The cross-section figure shows wDPM profile in red and the Alpha Step profile in blue. † denotes a physical pinhole in the SLM plane.
In order to capture faster dynamics and/or to measure larger objects without introducing artifacts, we need to increase source brightness. This can be accomplished using a supercontinuum laser (SCL). Figure 8(a) shows the spectrum of YSL's SC-1.5-HE model, 1.5W blue enhanced SCL. The infrared (IR) portion of the spectrum was blocked using a shortpass filter (Thorlabs FESH0750) with a cut-off wavelength of 750 nm. Figure 8(b) show the 0th order reference amplitude image illustrating the uniformity. Figures 8(c)-8(f) show the recovered height maps and corresponding cross-sections of square micropillars of various lateral dimensions. The measured heights are correct and the images are free of halo and shade-off artifacts. This is due to the fact that, while the SCL is temporally broadband (white light), it has adequate spatial coherence without any additional filtering by the condenser aperture. Furthermore, the SCL source can deliver a much brighter illumination field for a given spatial coherence area yielding superior SNR and allowing for significantly faster dynamic measurements, including those of structures that have arbitrary lateral size, without introducing artifacts. For example, the exposure time can be set to the camera’s minimum, which is 400 µs. Note however that in Fig. 8, the exposure time was set to 10 ms and a neutral density filter attenuated the SCL power so that a fairer comparison could be made between the SCL data and the halogen data in Figs. 4 and 5 with NAcon = 0.09.

4. Conclusions

We have demonstrated that object-dependent artifacts such as halo and shade-off in white light DPM can be caused by a lack of spatial coherence, which prevents the lower spatial frequencies of the structure from being recovered correctly. Obtaining adequate spatial coherence at the condenser aperture as well as performing proper spatial filtering in the SLM plane are both necessary to recover quantitatively correct height maps. A clear reduction in the halo and negative dimple of red blood cells was demonstrated by computing the required condenser pinhole size in order to achieve adequate spatial coherence for the given specimen. We also demonstrated that a supercontinuum laser can be used to obtain accurate topographical measurements with minimal exposure times.

Appendix

A1. Temporal coherence

Fig. 9 Temporal coherence of the wDPM imaging system. (a) Spectrum of the HAL 100 halogen lamp showing the power-equivalent bandwidth (PEB) approximation. The center wavelength at a color temperature of 3200 K is 574 nm. This wavelength is used to convert the measured phase to height. (b) Temporal autocorrelation function and its envelope obtained via a Hilbert transform. The temporal coherence length obtained from the envelope is 2.1 μm.
In this section, we assess the temporal coherence and show that it is not an issue under our typical imaging conditions. To quantify the temporal coherence, the source spectrum was measured and used to compute the temporal autocorrelation function. The temporal coherence length was then extracted. Figure 9(a) shows the source spectrum taken from the wDPM system (HAL 100 Halogen Lamp, Z2 Axio-Imager, Zeiss) at T = 3200 K, where T is the color temperature of the halogen lamp.

The spectrum was measured using a standard spectrometer (Ocean Optics Spectrometer USB 2000+). Several spectra were taken in different planes including the sample plane, camera plane, and directly off the source. All of these measurements resulted in the same center wavelength and temporal coherence value.

The autocorrelation function was then calculated using MATLAB and the temporal coherence was extracted by looking at the distance along the optical axis where the envelope was greater than half its maximum value. Figure 9(b) shows the autocorrelation function of the source along with its envelope computed via a Hilbert transform. The measured temporal coherence length using this approach was 2.1 µm. The temporal coherence length of white light systems is typically on the order of a couple of microns depending on both the source spectrum and the definition of coherence length used in the calculation [31

31. Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, and G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express 19(2), 1016–1026 (2011). [CrossRef] [PubMed]

].

The temporal coherence length indicates the maximum optical path length (OPL) difference over which we can obtain interference and thus accurately measure height. Most biological specimens in their typical imaging media exhibit low index contrast. For example, red blood cells (RBCs) in Coulter solution have a very low index contrast of Δn ≈ 0.06 [32

32. M. Mir, K. Tangella, and G. Popescu, “Blood testing at the single cell level using quantitative phase and amplitude microscopy,” Biomed. Opt. Express 2(12), 3259–3266 (2011). [CrossRef] [PubMed]

]. For a lower Δn, a larger height (i.e. h = 2.0 / 0.06 = 33 µm) can be measured for a given temporal coherence length. Thus, we can measure most biological specimens without any issues regarding the temporal coherence. Further, the maximum thickness we can measure without issues is 3.9 µm for quartz pillars in air and 27 µm for polystyrene beads in immersion oil. Thus, the source has an order of magnitude larger temporal coherence than what is needed to measure the samples in this paper.

It is also important to mention that phase unwrapping errors can occur when the measured phase jumps by more than 2π between adjacent pixels. For most biological structures, changes in height are gradual and this is not an issue. However, for semiconductor samples such as dry etched structures with vertical sidewalls, this may be a concern depending on the height. Both the temporal coherence and phase ambiguity should be considered before imaging a particular sample. Phase unwrapping errors can be eradicated by employing different wavelength lasers (i.e., 2 or more) or by using a color camera to separate the red-green-blue (RGB) planes under white light illumination [33

33. J. Gass, A. Dakoff, and M. K. Kim, “Phase imaging without 2pi ambiguity by multiwavelength digital holography,” Opt. Lett. 28(13), 1141–1143 (2003). [CrossRef] [PubMed]

35

35. M. T. Rinehart, N. T. Shaked, N. J. Jenness, R. L. Clark, and A. Wax, “Simultaneous two-wavelength transmission quantitative phase microscopy with a color camera,” Opt. Lett. 35(15), 2612–2614 (2010). [CrossRef] [PubMed]

].

Acknowledgments

This work is supported in part by NSF CBET-1040462 MRI award with matching funds from the University of Illinois. The authors thank Ware Photonics for use of the supercontinuum laser and Renjie Zhou and Steve McKeown for helpful discussions and assistance in the lab.

References and links

1.

G. Popescu, Quantitative Phase Imaging of Cells and Tissues, McGraw-Hill biophotonics (McGraw-Hill, 2011).

2.

G. Popescu, T. Ikeda, R. R. Dasari, and M. S. Feld, “Diffraction phase microscopy for quantifying cell structure and dynamics,” Opt. Lett. 31(6), 775–777 (2006). [CrossRef] [PubMed]

3.

C. Edwards, A. Arbabi, G. Popescu, and L. L. Goddard, “Optically monitoring and controlling nanoscale topography during semiconductor etching,” Light Sci. Appl. 1(9), e30 (2012). [CrossRef]

4.

C. Edwards, K. Wang, R. Zhou, B. Bhaduri, G. Popescu, and L. L. Goddard, “Digital projection photochemical etching defines gray-scale features,” Opt. Express 21(11), 13547–13554 (2013). [CrossRef] [PubMed]

5.

R. Zhou, G. Popescu, and L. L. Goddard, “22 nm node wafer inspection using diffraction phase microscopy and image post-processing,” Proc. SPIE 8681, 8610G (2013).

6.

Y. Park, G. Popescu, K. Badizadegan, R. R. Dasari, and M. S. Feld, “Diffraction phase and fluorescence microscopy,” Opt. Express 14(18), 8263–8268 (2006). [CrossRef] [PubMed]

7.

N. Lue, W. Choi, K. Badizadegan, R. R. Dasari, M. S. Feld, and G. Popescu, “Confocal diffraction phase microscopy of live cells,” Opt. Lett. 33(18), 2074–2076 (2008). [CrossRef] [PubMed]

8.

Y. K. Park, M. Diez-Silva, G. Popescu, G. Lykotrafitis, W. Choi, M. S. Feld, and S. Suresh, “Refractive index maps and membrane dynamics of human red blood cells parasitized by Plasmodium falciparum,” Proc. Natl. Acad. Sci. U.S.A. 105(37), 13730–13735 (2008). [CrossRef] [PubMed]

9.

Y. Park, T. Yamauchi, W. Choi, R. Dasari, and M. S. Feld, “Spectroscopic phase microscopy for quantifying hemoglobin concentrations in intact red blood cells,” Opt. Lett. 34(23), 3668–3670 (2009). [CrossRef] [PubMed]

10.

Y. K. Park, C. A. Best, T. Auth, N. S. Gov, S. A. Safran, G. Popescu, S. Suresh, and M. S. Feld, “Metabolic remodeling of the human red blood cell membrane,” Proc. Natl. Acad. Sci. U.S.A. 107(4), 1289–1294 (2010). [CrossRef] [PubMed]

11.

Y. K. Park, C. A. Best, K. Badizadegan, R. R. Dasari, M. S. Feld, T. Kuriabova, M. L. Henle, A. J. Levine, and G. Popescu, “Measurement of red blood cell mechanics during morphological changes,” Proc. Natl. Acad. Sci. U.S.A. 107(15), 6731–6736 (2010). [CrossRef] [PubMed]

12.

J. W. Kang, N. Lue, C.-R. Kong, I. Barman, N. C. Dingari, S. J. Goldfless, J. C. Niles, R. R. Dasari, and M. S. Feld, “Combined confocal Raman and quantitative phase microscopy system for biomedical diagnosis,” Biomed. Opt. Express 2(9), 2484–2492 (2011). [CrossRef] [PubMed]

13.

S. Witte, A. Plauşka, M. C. Ridder, L. van Berge, H. D. Mansvelder, and M. L. Groot, “Short-coherence off-axis holographic phase microscopy of live cell dynamics,” Biomed. Opt. Express 3(9), 2184–2189 (2012). [CrossRef] [PubMed]

14.

B. Bhaduri, H. Pham, M. Mir, and G. Popescu, “Diffraction phase microscopy with white light,” Opt. Lett. 37(6), 1094–1096 (2012). [CrossRef] [PubMed]

15.

H. Pham, B. Bhaduri, H. Ding, and G. Popescu, “Spectroscopic diffraction phase microscopy,” Opt. Lett. 37(16), 3438–3440 (2012). [CrossRef] [PubMed]

16.

P. Girshovitz and N. T. Shaked, “Compact and portable low-coherence interferometer with off-axis geometry for quantitative phase microscopy and nanoscopy,” Opt. Express 21(5), 5701–5714 (2013). [CrossRef] [PubMed]

17.

H. V. Pham, B. Bhaduri, K. Tangella, C. Best-Popescu, and G. Popescu, “Real time blood testing using quantitative phase imaging,” PLoS ONE 8(2), e55676 (2013). [CrossRef] [PubMed]

18.

T. Slabý, P. Kolman, Z. Dostál, M. Antoš, M. Lošťák, and R. Chmelík, “Off-axis setup taking full advantage of incoherent illumination in coherence-controlled holographic microscope,” Opt. Express 21(12), 14747–14762 (2013). [CrossRef] [PubMed]

19.

J. W. Goodman, “Statistical properties of laser speckle patterns,” in Laser Speckle and Related Phenomena (Springer, 1975).

20.

P. Gao, B. Yao, I. Harder, N. Lindlein, and F. J. Torcal-Milla, “Phase-shifting Zernike phase contrast microscopy for quantitative phase measurement,” Opt. Lett. 36(21), 4305–4307 (2011). [CrossRef] [PubMed]

21.

T. Wilson and C. J. R. Sheppard, “The halo effect of image processing by spatial frequency filtering,” Optik (Stuttg.) 59(1), 19–23 (1981).

22.

H. V. Pham, C. Edwards, L. L. Goddard, and G. Popescu, “Fast phase reconstruction in white light diffraction phase microscopy,” Appl. Opt. 52(1), A97–A101 (2013). [CrossRef] [PubMed]

23.

J. W. Goodman, Introduction to Fourier Optics (Roberts & Company, 2005).

24.

B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (Wiley, 2013).

25.

M. Born and E. Wolf, Principles of optics, 7th (expanded) ed. (Pergamon Press, 1999).

26.

T. Nguyen, C. Edwards, L. Goddard, and G. Popescu, “Quantitative phase imaging with partially coherent illumination,” (to be submitted).

27.

T. Kim, R. Zhu, T. H. Nguyen, R. Zhou, C. Edwards, L. L. Goddard, and G. Popescu, “Deterministic signal associated with a random field,” Opt. Express 21(18), 20806–20820 (2013). [CrossRef] [PubMed]

28.

B. Bhaduri, D. Wickland, R. Wang, V. Chan, R. Bashir, and G. Popescu, “Cardiomyocyte Imaging Using Real-Time Spatial Light Interference Microscopy (SLIM),” PLoS ONE 8(2), e56930 (2013). [CrossRef] [PubMed]

29.

B. Bhaduri, K. Tangella, and G. Popescu, “Fourier phase microscopy with white light,” Biomed. Opt. Express 4(8), 1434–1441 (2013). [CrossRef] [PubMed]

30.

T. H. Nguyen and G. Popescu, “Spatial Light Interference Microscopy (SLIM) using twisted-nematic liquid-crystal modulation,” Biomed. Opt. Express 4(9), 1571–1583 (2013). [CrossRef] [PubMed]

31.

Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, and G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express 19(2), 1016–1026 (2011). [CrossRef] [PubMed]

32.

M. Mir, K. Tangella, and G. Popescu, “Blood testing at the single cell level using quantitative phase and amplitude microscopy,” Biomed. Opt. Express 2(12), 3259–3266 (2011). [CrossRef] [PubMed]

33.

J. Gass, A. Dakoff, and M. K. Kim, “Phase imaging without 2pi ambiguity by multiwavelength digital holography,” Opt. Lett. 28(13), 1141–1143 (2003). [CrossRef] [PubMed]

34.

C. J. Mann, P. R. Bingham, V. C. Paquit, and K. W. Tobin, “Quantitative phase imaging by three-wavelength digital holography,” Opt. Express 16(13), 9753–9764 (2008). [CrossRef] [PubMed]

35.

M. T. Rinehart, N. T. Shaked, N. J. Jenness, R. L. Clark, and A. Wax, “Simultaneous two-wavelength transmission quantitative phase microscopy with a color camera,” Opt. Lett. 35(15), 2612–2614 (2010). [CrossRef] [PubMed]

OCIS Codes
(030.1640) Coherence and statistical optics : Coherence
(070.6110) Fourier optics and signal processing : Spatial filtering
(120.3180) Instrumentation, measurement, and metrology : Interferometry
(120.4630) Instrumentation, measurement, and metrology : Optical inspection
(120.5050) Instrumentation, measurement, and metrology : Phase measurement
(170.0110) Medical optics and biotechnology : Imaging systems

ToC Category:
Coherence and Statistical Optics

History
Original Manuscript: December 23, 2013
Revised Manuscript: February 12, 2014
Manuscript Accepted: February 19, 2014
Published: February 26, 2014

Virtual Issues
Vol. 9, Iss. 5 Virtual Journal for Biomedical Optics

Citation
Chris Edwards, Basanta Bhaduri, Tan Nguyen, Benjamin G. Griffin, Hoa Pham, Taewoo Kim, Gabriel Popescu, and Lynford L. Goddard, "Effects of spatial coherence in diffraction phase microscopy," Opt. Express 22, 5133-5146 (2014)
http://www.opticsinfobase.org/vjbo/abstract.cfm?URI=oe-22-5-5133


Sort:  Author  |  Year  |  Journal  |  Reset  

References

  1. G. Popescu, Quantitative Phase Imaging of Cells and Tissues, McGraw-Hill biophotonics (McGraw-Hill, 2011).
  2. G. Popescu, T. Ikeda, R. R. Dasari, M. S. Feld, “Diffraction phase microscopy for quantifying cell structure and dynamics,” Opt. Lett. 31(6), 775–777 (2006). [CrossRef] [PubMed]
  3. C. Edwards, A. Arbabi, G. Popescu, L. L. Goddard, “Optically monitoring and controlling nanoscale topography during semiconductor etching,” Light Sci. Appl. 1(9), e30 (2012). [CrossRef]
  4. C. Edwards, K. Wang, R. Zhou, B. Bhaduri, G. Popescu, L. L. Goddard, “Digital projection photochemical etching defines gray-scale features,” Opt. Express 21(11), 13547–13554 (2013). [CrossRef] [PubMed]
  5. R. Zhou, G. Popescu, L. L. Goddard, “22 nm node wafer inspection using diffraction phase microscopy and image post-processing,” Proc. SPIE 8681, 8610G (2013).
  6. Y. Park, G. Popescu, K. Badizadegan, R. R. Dasari, M. S. Feld, “Diffraction phase and fluorescence microscopy,” Opt. Express 14(18), 8263–8268 (2006). [CrossRef] [PubMed]
  7. N. Lue, W. Choi, K. Badizadegan, R. R. Dasari, M. S. Feld, G. Popescu, “Confocal diffraction phase microscopy of live cells,” Opt. Lett. 33(18), 2074–2076 (2008). [CrossRef] [PubMed]
  8. Y. K. Park, M. Diez-Silva, G. Popescu, G. Lykotrafitis, W. Choi, M. S. Feld, S. Suresh, “Refractive index maps and membrane dynamics of human red blood cells parasitized by Plasmodium falciparum,” Proc. Natl. Acad. Sci. U.S.A. 105(37), 13730–13735 (2008). [CrossRef] [PubMed]
  9. Y. Park, T. Yamauchi, W. Choi, R. Dasari, M. S. Feld, “Spectroscopic phase microscopy for quantifying hemoglobin concentrations in intact red blood cells,” Opt. Lett. 34(23), 3668–3670 (2009). [CrossRef] [PubMed]
  10. Y. K. Park, C. A. Best, T. Auth, N. S. Gov, S. A. Safran, G. Popescu, S. Suresh, M. S. Feld, “Metabolic remodeling of the human red blood cell membrane,” Proc. Natl. Acad. Sci. U.S.A. 107(4), 1289–1294 (2010). [CrossRef] [PubMed]
  11. Y. K. Park, C. A. Best, K. Badizadegan, R. R. Dasari, M. S. Feld, T. Kuriabova, M. L. Henle, A. J. Levine, G. Popescu, “Measurement of red blood cell mechanics during morphological changes,” Proc. Natl. Acad. Sci. U.S.A. 107(15), 6731–6736 (2010). [CrossRef] [PubMed]
  12. J. W. Kang, N. Lue, C.-R. Kong, I. Barman, N. C. Dingari, S. J. Goldfless, J. C. Niles, R. R. Dasari, M. S. Feld, “Combined confocal Raman and quantitative phase microscopy system for biomedical diagnosis,” Biomed. Opt. Express 2(9), 2484–2492 (2011). [CrossRef] [PubMed]
  13. S. Witte, A. Plauşka, M. C. Ridder, L. van Berge, H. D. Mansvelder, M. L. Groot, “Short-coherence off-axis holographic phase microscopy of live cell dynamics,” Biomed. Opt. Express 3(9), 2184–2189 (2012). [CrossRef] [PubMed]
  14. B. Bhaduri, H. Pham, M. Mir, G. Popescu, “Diffraction phase microscopy with white light,” Opt. Lett. 37(6), 1094–1096 (2012). [CrossRef] [PubMed]
  15. H. Pham, B. Bhaduri, H. Ding, G. Popescu, “Spectroscopic diffraction phase microscopy,” Opt. Lett. 37(16), 3438–3440 (2012). [CrossRef] [PubMed]
  16. P. Girshovitz, N. T. Shaked, “Compact and portable low-coherence interferometer with off-axis geometry for quantitative phase microscopy and nanoscopy,” Opt. Express 21(5), 5701–5714 (2013). [CrossRef] [PubMed]
  17. H. V. Pham, B. Bhaduri, K. Tangella, C. Best-Popescu, G. Popescu, “Real time blood testing using quantitative phase imaging,” PLoS ONE 8(2), e55676 (2013). [CrossRef] [PubMed]
  18. T. Slabý, P. Kolman, Z. Dostál, M. Antoš, M. Lošťák, R. Chmelík, “Off-axis setup taking full advantage of incoherent illumination in coherence-controlled holographic microscope,” Opt. Express 21(12), 14747–14762 (2013). [CrossRef] [PubMed]
  19. J. W. Goodman, “Statistical properties of laser speckle patterns,” in Laser Speckle and Related Phenomena (Springer, 1975).
  20. P. Gao, B. Yao, I. Harder, N. Lindlein, F. J. Torcal-Milla, “Phase-shifting Zernike phase contrast microscopy for quantitative phase measurement,” Opt. Lett. 36(21), 4305–4307 (2011). [CrossRef] [PubMed]
  21. T. Wilson, C. J. R. Sheppard, “The halo effect of image processing by spatial frequency filtering,” Optik (Stuttg.) 59(1), 19–23 (1981).
  22. H. V. Pham, C. Edwards, L. L. Goddard, G. Popescu, “Fast phase reconstruction in white light diffraction phase microscopy,” Appl. Opt. 52(1), A97–A101 (2013). [CrossRef] [PubMed]
  23. J. W. Goodman, Introduction to Fourier Optics (Roberts & Company, 2005).
  24. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (Wiley, 2013).
  25. M. Born and E. Wolf, Principles of optics, 7th (expanded) ed. (Pergamon Press, 1999).
  26. T. Nguyen, C. Edwards, L. Goddard, G. Popescu, “Quantitative phase imaging with partially coherent illumination,” (to be submitted).
  27. T. Kim, R. Zhu, T. H. Nguyen, R. Zhou, C. Edwards, L. L. Goddard, G. Popescu, “Deterministic signal associated with a random field,” Opt. Express 21(18), 20806–20820 (2013). [CrossRef] [PubMed]
  28. B. Bhaduri, D. Wickland, R. Wang, V. Chan, R. Bashir, G. Popescu, “Cardiomyocyte Imaging Using Real-Time Spatial Light Interference Microscopy (SLIM),” PLoS ONE 8(2), e56930 (2013). [CrossRef] [PubMed]
  29. B. Bhaduri, K. Tangella, G. Popescu, “Fourier phase microscopy with white light,” Biomed. Opt. Express 4(8), 1434–1441 (2013). [CrossRef] [PubMed]
  30. T. H. Nguyen, G. Popescu, “Spatial Light Interference Microscopy (SLIM) using twisted-nematic liquid-crystal modulation,” Biomed. Opt. Express 4(9), 1571–1583 (2013). [CrossRef] [PubMed]
  31. Z. Wang, L. Millet, M. Mir, H. Ding, S. Unarunotai, J. Rogers, M. U. Gillette, G. Popescu, “Spatial light interference microscopy (SLIM),” Opt. Express 19(2), 1016–1026 (2011). [CrossRef] [PubMed]
  32. M. Mir, K. Tangella, G. Popescu, “Blood testing at the single cell level using quantitative phase and amplitude microscopy,” Biomed. Opt. Express 2(12), 3259–3266 (2011). [CrossRef] [PubMed]
  33. J. Gass, A. Dakoff, M. K. Kim, “Phase imaging without 2pi ambiguity by multiwavelength digital holography,” Opt. Lett. 28(13), 1141–1143 (2003). [CrossRef] [PubMed]
  34. C. J. Mann, P. R. Bingham, V. C. Paquit, K. W. Tobin, “Quantitative phase imaging by three-wavelength digital holography,” Opt. Express 16(13), 9753–9764 (2008). [CrossRef] [PubMed]
  35. M. T. Rinehart, N. T. Shaked, N. J. Jenness, R. L. Clark, A. Wax, “Simultaneous two-wavelength transmission quantitative phase microscopy with a color camera,” Opt. Lett. 35(15), 2612–2614 (2010). [CrossRef] [PubMed]

Cited By

Alert me when this paper is cited

OSA is able to provide readers links to articles that cite this paper by participating in CrossRef's Cited-By Linking service. CrossRef includes content from more than 3000 publishers and societies. In addition to listing OSA journal articles that cite this paper, citing articles from other participating publishers will also be listed.